Presentation is loading. Please wait.

Presentation is loading. Please wait.

Calibration of Balance

Similar presentations


Presentation on theme: "Calibration of Balance"— Presentation transcript:

1 Calibration of Balance
Bangladesh BEST Programme Calibration of Balance Sri Lanka Nihal Gunasekara

2 Types of Balances Electromagnetic force balance (Top loading)
Bangladesh BEST Programme Types of Balances Electromagnetic force balance (Top loading) Single-pan, two-knife-edge balance Two-pan, three-knife-edge balance Triple beam balance Weighing scale

3 Bangladesh BEST Programme
Analytical Balance Balance with an enclosed weighing chamber and a resolution 0.01 or 0.1 mg Micro Balance Balance with a resolution of 1 microgram

4 Bangladesh BEST Programme
Micro Balance

5 Ultra (Super) Micro Balance
Bangladesh BEST Programme Semi Micro Balance Balance with a resolution of 10 microgram Ultra (Super) Micro Balance Balance with a resolution of 0.1 microgram

6 Resolution or Discrimination
Bangladesh BEST Programme Resolution or Discrimination The smallest change in mass which can be detect by the balance Sensitivity This is a measure of the ability of the balance to detect changes in the load applied to it

7 Response time Stability
Bangladesh BEST Programme Response time Time taken for the weighing instrument to indicate the value of a change in load Stability A measure of the time for which the reading on the balance remains unchanged

8 Correction = True Value - Reading Error = Reading - True Value
Bangladesh BEST Programme Correction Correction = True Value - Reading Error Error = Reading - True Value

9 Departure From Nominal Value
Bangladesh BEST Programme Departure From Nominal Value The amount by which the reading on an instrument depart from its nominal value Hysteresis The indications of a measuring instrument when the same value of the quantity measured is reached by increasing or decreasing that quantity

10 Standby Accuracy class. (mass)
Bangladesh BEST Programme Standby Operating state of an electronic instrument where one part is switched off for energy saving reasons Accuracy class. (mass) Legal metrology term. Classification of weights for calibration purposes by OIML R111-1 weights of Accuracy class E1, E2, F1, F2, M1, M2, M3 from 1 mg to 50 kg

11 Different accuracy classes of mass
Bangladesh BEST Programme Different accuracy classes of mass

12 Repeatability Bangladesh BEST Programme
Closeness of the agreement between the results of successive measurements of the same measurement carried out under the same conditions of measurement. Repeatability condition include : the same observer the same measuring instrument, used under the same conditions the same location repetition over a short period of time

13 Reproducibility Bangladesh BEST Programme
Closeness of the agreement between the results of measurements of the same measurement carried out under changed conditions of measurement. A valid statement of reproducibility requires specification of the conditions change The changed conditions may include principle of measurement method of measurement observer measuring instrument reference standard location conditions of use Time

14 Influence Factors Air movements and heat sources
Bangladesh BEST Programme Influence Factors Air movements and heat sources Support surfaces and vibration Electrical and electromagnetic interference Magnetic effects Buoyancy effects

15 Influence factor Bangladesh BEST Programme Influence factor Parameter
Source Effect Test for effect Precautions Environmental Vibration Various, Eg. Machinery, roads, railways, aircraft operator Vibration causes damages Changes forces on the weighing machine due to acceleration Primary Stability of reading under constant load Secondary Accelerometer measurements Use of a high-mass support, set on a solid surface, for the weighing machine, independent of the adjacent flooring, Eg fixed to bedrock Note : low frequency, vibration is likely to be more serious in its effects than high (acoustic) frequencies Rubber pads used to damp high frequencies may pass low frequency vibration, and may cause tilting of the weighing machine

16 Influence factor Bangladesh BEST Programme Influence factor Parameter
Source Effect Test for effect Precautions Environmental Temperature: air, weight, weighing machine Sunlight, heating systems, operator’s body, greenhouse heating in the weighing chamber Temperature changes and differentials can change the weighing machine response, cause convection effects and change buoyancy Temperature monitoring instruments Avoid direct sunlight and other connections to the external environment; avoid changes in heat and lighting input. Leave balance cases ajar when not in use. Avoid excessive handling of equipment during weighing. Allow weighted object to acclimatize

17 Influence factor Bangladesh BEST Programme Influence factor Parameter
Source Effect Test for effect Precautions Environmental Humidity Air supply ,air conditioning, changes in activity levels Buoyancy, absorbed moisture, condensation Rh monitoring instruments Avoid liquid water in the working area unless rh is too low . Air conditioning, humidity controller

18 Influence factor Bangladesh BEST Programme Influence factor Parameter
Source Effect Test for effect Precautions Environmental Airflow Air conditioning, wind, convection, operator movement Varies force on weighing machine Primary Stability of reading under constant load Secondary Enclose weighing machine and observe change of reading stability Avoid temperature gradients. draught free working space, enclose weighing machines

19 Influence factor Bangladesh BEST Programme Influence factor Parameter
Source Effect Test for effect Precautions Environmental Magnetic Constructional steelwork, ferromagnetic weight (s) electrical induction earth’s magnetic field, moving-coil meters Change of current in force compensation coil, direct attraction to/from surrounding structure Motion-dependent restoring forces Primary Use a magnet to check for induced reading changes Secondary Monitoring readings when weights are rotated and at varying heights above the weight receptor (place on/under non-magnetic mount) Avoid weights with residual magnetism (de-gauss) Test with compass or gaussmeter Limit magnetic permeability of weights weighed check for magnetic fields with gaussmeter Screen with mumetal

20 Type of error Source of error Bangladesh BEST Programme Repeatability
Load cell hysteresis Electrical Interference Zero Mains fluctuation Sensitivity, Zero, Repeatability Incorrect horizontal support Mechanical errors Type of error Source of error

21 ACCURACY CLASS (OIML R111-1 :2004)
Bangladesh BEST Programme ACCURACY CLASS (OIML R111-1 :2004) 0.20 0.06 0.020 0.006 1 mg 0.003 5 mg 0.25 0.08 0.025 0.008 10 mg 0.4 0.12 0.04 0.012 0.004 50 mg 1.6 0.5 0.16 0.05 0.016 0.005 100 mg 10 3.0 1.0 0.3 0.10 0.03 0.010 1 g 12 4.0 1.2 2 g 16 5.0 5 g 20 6.0 2.0 0.6 10 g 30 3 0.030 50 g 50 5 100 g 100 1. 0 200 g 500 160 1 kg 1000 300 2 kg 2500 800 250 80 25 8.0 2.5 5 kg 5000 1600 10 kg Class M3 Class M2 Class M1 Class F2 Class F1 Class E2 Class E1 Tolerance in mg Nominal value 0.003

22 Bangladesh BEST Programme
Lowest class of weights required for different types of weighing machine Capacity Resolution 100g 10g 1g 100mg 10mg 1mg 0.1mg 0.01mg or less Up to 200g M1 M1 F2 F1 E2 200g to 1 kg M1 M1 F2 F1/E2 E2 E2 1 kg to 30kg M2 M2 M1 F2 E2 E2 E2 30kg to 100kg M2 M1 F2 F1 E2 Above 100 kg M2 M1/F2 F1 E2

23 Calibration of Top Loading and Analytical Balances
Bangladesh BEST Programme Calibration of Top Loading and Analytical Balances

24 Before use a Balance…………
Bangladesh BEST Pregramme Before use a Balance………… Keep the Balance power on appropriate time to warm up. Clean the balance as best as no contamination on the balance pan. If it is applicable, level the balance before use. Don’t move the balance after the calibration. Lint or any other soft gloves should be worn by the user. A pair of forceps or a pair tweezers with thermal insulating ends should be used to lift and place weights.

25 Before use a Balance Bangladesh BEST Programme
Make sure with the soft placement of weights on the balance pan as level best of possible. Readings should be taken after the balance is stabilized well. The place should be free from vibration. Any other relevant precaution practiced by the user.

26 Test 2 - Repeatability of Reading
Bangladesh BEST Programme Balance Calibration The calibration of the electromagnetic force compensation balance consists of the following tests: Test 1 - Scale Value Test 2 - Repeatability of Reading Test 3 - Departure from Nominal Value Test 4 - Off centre Loading Test 5 – Hysteresis

27 Procedure – Balance Operating Manual
Bangladesh BEST Programme Scale Value Internal Scale Adjustment –Internal Mass External Scale Adjustment – External Mass Procedure – Balance Operating Manual

28 Repeatability of Reading
Bangladesh BEST Programme Repeatability of Reading Objective of Measurement : The repeatability test is carried out to determine the ability of the balance to give under defined conditions of use closely similar responses for repeated measurements The balance Standard Deviation is the “repeatability of measurement”

29 Repeatability of Reading
Bangladesh BEST Programme Repeatability of Reading : M 1 -Set the reading to zero without any mass on the pan M 2 - Record the zero reading of the balance (z) M 3 - Place the mass on the pan and record the reading (r) M 4 - Remove the mass from the pan M 5 – Repeat M2 to M4 ten times during the process the balance should not be zeroed

30 Repeatability of Reading:
Bangladesh BEST Programme Repeatability of Reading: Performed at three points : Near zero Near half load Near full load

31 DATA RECORDING Mass 1 : Mass 2: Mass 3: Bangladesh BEST Programme No 1
Zero reading (z) Mass reading (r) Corr. reading (r-z) Corr. reading (r ) 1 2 3 4 5 6 7 8 9 10

32 Repeatability of Reading
Bangladesh BEST Programme Repeatability of Reading Mass 1: Mass 2: Mass 3 : Mean value : Std. Dev. ( σ ) : Max. Diff : Uncer. Type A :

33 Repeatability of Reading
Bangladesh BEST Programme Repeatability of Reading Repeat three different values of M ; near zero, half maximum and maximum load. Load on Pan (g) Standard Deviation of Reading (mg) Maximum Difference Between Successive Readings (mg) 0.05 100 200 0.067 0.052 0.2 0.1

34 Departure from Nominal Value:
Bangladesh BEST Programme Departure from Nominal Value: Objective of Measurement : “The departure from nominal value measurements are carried out to determine the deviation of the balance reading from the expected value” This is the Correction to Balance Reading

35 Departure from Nominal Value:
Bangladesh BEST Programme Departure from Nominal Value: 1 Divide the range into at least 10 equally spaced steps 2 The masses corresponding to M, 2M, 3M, 4M, M, 6M etc. until the capacity of balance is reached 3 In addition choose four more test points corresponding to 20 %, 40 %,60%and 80 % of the first step for single range.

36 Departure from Nominal Value:
Bangladesh BEST Programme Departure from Nominal Value: 1 - Read the zero z1 2 - Place the weight on the pan &record reading r1 3 – Remove the weight momentarily and place it back on the pan and note the new reading r2 4 - Remove the weight and read the zero z2 5 – Repeat 1 to 4 for weights of masses 2M,3M,4M,5M etc until the capacity of balance is reached

37 Departure from Nominal Value:
Bangladesh BEST Programme Departure from Nominal Value: Average zero reading zi = (z1 + z2 )/ 2 Average mass reading ri = (r1 + r2 )/ 2 Difference di = ri – zi Correction ci = Mi – di Mi is the conventional value of the mass

38 Correction to Balance Reading
Bangladesh BEST Programme Correction to Balance Reading Pan Load Conv. Value Mi Reading Mean Difference ri - zi Correction Mi - di M- 100 100.02 0.00 100.00 100.01 0.01 2M 3M 4M 5M 6M 7M 8M 9M 10M

39 Departure from Nominal Value:
Bangladesh BEST Programme Departure from Nominal Value: Reading (g) Correction(mg) Expanded Uncertainty k=2 +(mg) Before the Calibration After the Calibration 4 8 12 16 20 40 60 80 100 120 140 160 180 200 220 -- 0.2 0.4 0.6 0.5 0.0 -0.1 0.1 -0.0 -0.2 -0.3 0.096 0.097 0.098 0.099 0.10 0.11

40 Objective of Measurement :
Bangladesh Best Programme Off Centre Loading: Objective of Measurement : “The off center loading is carried out to find the variation in the displayed weighted as the object is moved to various positions on the weighing pan”

41 Off Centre Loading Bangladesh BEST Programme
This test to find the variation of the reading for several place of the balance pan. Test should be done for rectangular type balance pans and as well as for circular type balance pans. A single weight should be used 4 3 1 5 2

42 Off Centre Loading: Bangladesh BEST Programme
Place a mass on the centre of the pan and move it successively to the front rear, left and right positions ( position being 2/3 of the distance between the centre and the lip of the pan) on the pan recording the reading each time Back (g) =0.0000 Center (g) =0.0000 Right (g) = Left (g) =0.0001 Max Difference = g Front (g) =0.0001

43 Maximum Difference (g)
Bangladesh BEST Programme Off Centre Loading: Mass of 100g was moved to various positions on the pan. The balance readings obtained relative to the centre reading 0.0002 0.0001 0.0000 Maximum Difference (g) Right(g) Left(g) Back(g) Front(g) Centre(g) Place a mass on the centre of the pan and move it successively to the front rear, left and right positions ( position being 2/3 of the distance between the centre and the lip of the pan) on the pan recording the reading each time

44 Hysteresis: Objective of Measurement :
Bangladesh BEST Programme Hysteresis: Objective of Measurement : “The hysteresis measurements are performed to determine the difference in the indication at the same load , depending upon the direction of approach to that load ie increasing or decreasing the load”

45 Hysteresis Zero the balance and record the reading z1
Bangladesh BEST Programme Hysteresis Perform the test at one point about mid range Zero the balance and record the reading z1 Place a mass M1 on the plan and record the reading r1 Add extra mass M2 to bring the balance close to full range Remove the extra mass M2 and record the reading with the mass M1 still on the pan Remove the mass M1 and record the reading r2 Repeat above steps three times Take the average of three readings

46 Hysteresis Mass M = g Bangladesh BEST Programme Pan Load Reading 1
Pan Load Reading 1 Reading 2 Reading 3 Zero ( z1 ) M ( m1 ) M + M1 M ( m2 ) Zero ( z2 ) m1 - m2 z1 - z2

47 Hysteresis: 0.1 100 0.0 Hysteresis Error(mg) Pan Load (g)
Bangladesh BEST Programme Hysteresis: 0.1 100 0.0 Hysteresis Error(mg) Pan Load (g)

48 Bangladesh BEST Programme
Hysteresis: Hysteresis error

49 Calibration of Single Pan two Knife Edge Balances
Bangladesh BEST Programme Calibration of Single Pan two Knife Edge Balances

50 Test 1 - Repeatability of Reading
Bangladesh BEST Programme Balance Calibration The calibration of the single pan two knife edge balance consists of the following tests: Test 1 - Repeatability of Reading Test 2 - Departure from Nominal Value Test 3 – Built-in-Mass Test 4 - Off centre Loading Test 5 – Hysterisis

51 Bangladesh BEST Programme
Balance Calibration The calibration of the single pan two knife edge balance consists of the following tests: Test 1 - Repeatability of Reading- same as electronic balance Test 2 - Departure from Nominal Value- same as electronic balance Test 3 – Built-in-Mass Test 4 - Off centre Loading- same as electronic balance Test 5 – Hysteresis- same as electronic balance

52 Balance Calibration The procedure to determine the Built-in- Mass
Bangladesh BEST Programme Balance Calibration The procedure to determine the Built-in- Mass With all dials set to the zero and no load on the pan, release the balance and record the reading z1   Set the "1" on the dial to be tested, place a calibrated mass, release the pan and record the reading m1a  Remove the mass , replace it and record the reading m1b  Return the dial to zero, remove the mass from the pan and record the zero z2  Repeat the above steps for all dial settings of the balance (i.e. 2 , 3, 10 , 20 ,…………..90 , 100 , 200 ) Follow the same procedure used for linearity test to determine the correction

53 Calibration of Triple Beam Balances
Bangladesh BEST Programme Calibration of Triple Beam Balances

54 Test 1 - Repeatability of Reading - same as electronic balance
Bangladesh BEST Programme Balance Calibration The calibration of triple beam balances consists of the following tests: Test 1 - Repeatability of Reading - same as electronic balance Test 2 - Departure from Nominal Value (Linearity Test) - same as electronic balance

55 Triple Beam Balance Before the calibration
Bangladesh Best Programme Triple Beam Balance Before the calibration Clean the weighing pan and other parts of the balance taking due care Level the balance Exercise the balance by making several weighing

56 Triple Beam Balance Repeatability of readings:
Bangladesh BEST Programme Triple Beam Balance Repeatability of readings: Set the reading to zero without any mass on the pan Record the zero reading zi of the balance without any mass on the pan  Place the mass M on the pan and record the mass reading mi Remove the mass M from the pan Repeat the above steps ten times (The balance should not be zeroed during the series of readings) Repeat for this for near zero, half load and near maximum

57 Triple Beam Balance Bangladesh Best Programme
Select a minimum of 10 equally spaced points ( M , 2M ,…..10M ) on the measuring range in such a way they cover the entire range or the part of the measuring range to be calibrated. In addition four test points corresponding to 20%,40%,60% and 80 % of the first step Record the zero reading z1 Set the slider scale to the first step on the measuring range (M) , place the calibrating mass(es) on the pan until the balance comes to equilibrium and record the mass reading mi  Remove the mass(es), place it back on the pan and record the mass reading mi2 Remove the mass(es) and record the zero reading zi+1  Repeat the above steps for the remaining points ( 2M , 3M ,….)

58 Weighing Scale (digital or analogue)
Bangladesh Best Programme Weighing Scale (digital or analogue) The calibration of weighing scales consists of the following tests: Test 1 - Repeatability of Reading - same as electronic balance Test 2 - Departure from Nominal Value (Linearity Test) - same as electronic balance

59 Calibration Results of a Set of Weights –Class F1
Bangladesh BEST Programme Calibration Results of a Set of Weights –Class F1 Nominal Value Conventional Value Uncertainty k=2 Max. Perm. Error 1 mg 1 mg – mg 0.006 mg ±0.02 mg 2 mg 2 mg mg 5 mg 5 mg mg 20 mg 20 mg – mg 0.01 mg ±0.03 mg 50 mg 50 mg mg 0.012 mg ±0.04 mg 100 mg 100 mg mg 0.015 mg ±0.05 mg 500 mg 500 mg mg 0.025 mg ±0.08 mg 1 g 1 g – mg 0.03 mg ±0.1 mg 2 g 2 g mg 0.04 mg ±0.12 mg 20 g 20 g mg 0.08 mg ±0.25 mg 50 g 50 g – 0.12 mg 0.1 mg ±0.3 mg 100 g 100 g – 0.22 mg 0.15 mg ±0.5 mg 200 g 200 g mg 0.3 mg ±1 mg Issued by an accredited laboratory (DKD)

60 Bangladesh BEST Programme
Thank you


Download ppt "Calibration of Balance"

Similar presentations


Ads by Google