Download presentation
Presentation is loading. Please wait.
1
Generalized instrumentation system
Perceptible output Output display Signal processing Variable Conversion element Sensor Primary Sensing Measurand
2
Generalized instrumentation system
Perceptible output Output display Control And feedback Signal processing Data transmission storage Variable Conversion element Sensor Primary Sensing Measurand Calibration signal Power source
3
Generalized instrumentation system
Perceptible output Output display Control And feedback Signal processing Data transmission storage Variable Conversion element Sensor Primary Sensing Measurand Calibration signal Radiation, electric current, or other applied energy Power source Eg. Heart beats Figure 1.1 The sensor converts energy or information from the measurand to another form (usually electric). This signal is the processed and displayed so that humans can perceive the information. Elements and connections shown by dashed lines are optional for some applications.
4
Measurand: Physical quantity
Biopotential Pressure Flow Dimensions (imaging) Displacement (velocity, acceleration, force) Impedance Temperature Chemical Concentration
5
Sensor and Transducer Transducer Sensor
Converts one form of energy to another Sensor Converts a physical measurand to an electrical output Interface with living system Minimize the energy extracted Minimally invasive displacement electric voltage pressure diaphragm Strain gage
6
Signal Conditioning Amplification Filtering Impedance matching
Analog/Digital for signal processing Signal form (time and frequency domains)
7
Output Display Numerical Graphical Discrete or continuous Visual
Hearing
8
Auxiliary Element Calibration Signal
Control and Feedback (auto or manual) Adjust sensor and signal conditioning
9
1.3 Alternative Operational Modes
Direct Mode: Measurand is readily accessible Temperature Heart Beat Indirect Mode: desired measurand is measured by measuring accessible measurand. Morphology of internal organ: X-ray shadows Volume of blood pumped per minute by the heart: respiration and blood gas concentration Pulmonary volumes: variation in thoracic impedance
10
Characteristics of Instrument Performance
Two classes of characteristics are used to evaluated and compare new instrument Static Characteristics: describe the performance for dc or very low frequency input. Dynamic Characteristics: describe the performance for ac and high frequency input.
11
1.9 Generalized Static Characteristics
Parameters used to evaluate medical instrument: Accuracy: The difference between the true value and the measured value divided by the true value Precision: The number of distinguishable alternatives from which a given results is selected {2.434 or 2.43} Resolution: The smallest increment quantity that can be measured with certainty Reproducibility: The ability to give the same output for equal inputs applied over some period of time.
12
1.9 Generalized Static Characteristics
Parameters used to evaluate medical instrument: Statistical Control: Systematic errors or bias are tolerable or can be removed by calibration. Statistical Sensitivity: the ratio of the incremental output quantity to the incremental input quantity, Gd.
13
Zero Drift: all output values increase or decrease by the same amount due to manufacturing misalignment, variation in ambient temperature, vibration,…. Sensitivity Drift: Output change in proportion to the magnitude of the input. Change in the slope of the calibration curve. Figure 1.3 (b) Static sensitivity: zero drift and sensitivity drift. Dotted lines indicate that zero drift and sensitivity drift can be negative.
14
Independent nonlinearity
- A% deviation of the reading - B% deviation of the full scale x1 y1 (x1 + x2) Linear Linear (y1 + y2) system system and and x2 y2 Kx1 Ky1 Linear Linear system system (a) Least-squares straight line y (Output) B% of full scale A% of reading Figure 1.4 (a) Basic definition of linearity for a system or element. The same linear system or element is shown four times for different inputs. (b) A graphical illustration of independent nonlinearity equals A% of the reading, or B% of full scale, whichever is greater (that is, whichever permits the larger error). Overall tolerance band xd (Input) Point at which Input Ranges (I): Minimum resolvable input < I < normal linear operating range A% of reading = B% of full scale (b)
15
Example A linear system described by the following equation y=2x+3. Find the overall tolerance band for the system if the input range is 0 to 10 and its independent nonlinearity is 0.5% deviation of the full scale and 1.5% deviation of the reading.
16
Input Impedance: disturb the quantity being measured.
Xd1 : desired input (voltage, force, pressure) Xd2 : implicit input (current, velocity, flow) P = Xd1.Xd2 :Power transferred across the tissue-sensor interface Generalized input impedance Zx Goal: Minimize P, when measuring effort variable Xd1, by maximizing Zx which in return will minimize the flow variable Xd2. Loading effect is minimized when source impedance Zs is much smaller then the Zx
17
1.10 Generalized Dynamic Characteristics
Most medical instrument process signals that are functions of time. The input x(t) is related to the output y(t) by ai and bi depend on the physical and electrical parameters of the system. Transfer Functions The output can be predicted for any input (transient, periodic, or random)
18
Frequency Transfer Function
Can be found by replacing D by j Example: If x(t) = Ax sin ( t) then y(t) = |H()| Ax sin ( t + /_H())
19
Zero-Order Instrument a0 y(t) = b0 x(t)
K: static sensitivity Figure 1.5 (a) A linear potentiometer, an example of a zero-order system. (b) Linear static characteristic for this system. (c) Step response is proportional to input. (d) Sinusoidal frequency response is constant with zero phase shift.
20
First-Order Instrument
Where is the time constant
21
First-Order Instrument
Output y(t) R + + Slope = K = 1 x(t) C y(t) - - Input x(t) (a) (b) x(t) Log Y (jw) scale X (jw) 1 1.0 0.707 S L wL wS Log scale w t (c) (d) f y(t) 1 0° S Log scale w L 0.63 - 45° -90° S L t Example 1.1: High-pass filter
22
Second-Order Instrument
Many medical instrument are 2nd order or higher Operational Transfer Function Frequency Transfer Function
23
2nd order mechanical force-measuring Instrument
Output y ( t ) (b) (d) (c) 1 Ks x(t) y(t) yn yn + 1 Resonance 2 Log scale -90° 0.5 -180° Log scale w K Input x(t) Slope K = 0° wn Y (jw) X (jw) f displacement (a) Input Force x(t) B = viscosity constant Ks = spring constant Natural freq. Damping ratio Figure 1.7 (a) Force-measuring spring scale, an example of a second-order instrument. (b) Static sensitivity. (c) Step response for overdamped case = 2, critically damped case = 1, underdamped case = (d) Sinusoidal steady-state frequency response, = 2, = 1, = 0.5.
24
Overdamped Critically damped Underdamped Damped natural freq. y(t) 1
Ks 0.5 Damped natural freq. t
25
Example 1.2: for underdamped second-order instruments, find the damping ratio from the step response
and Logarithmic decrement
26
Time Delay System Y (jw) X (jw) Log scale K Log scale w f Log scale w
0°
27
Design Criteria Figure 1.8 Design process for medical instruments Choice and design of instruments are affected by signal factors, and also by environmental, medical, and economic factors.
28
Commercial Medical Instrumentation Development Process
Ideas: come from people working in the health care Detailed evaluation and signed disclosure Feasibility analysis and product description Medical need Technical feasibility Brief business plan (financial, sales, patents, standards, competition) Product Specification (interface, size, weight, color) “What” is required but nothing about “how” Design and development (software and hardware)
29
Commercial Medical Instrumentation Development Process
Prototype development Testing on animals or human subjects Final design review (test results for, specifications, subject feedback, cost) Production (packaging, manual and documents) Technical support
30
Regulation of Medical Devices
Medical devices is “any item promoted for a medical purpose that does not rely on chemical action to achieve its intended effect” 2 Ways for Medical Devices Classification First Way: (based on potential hazards) Class I: general controls Class II: performance standards Class III: premarketing approval Second Method: (see Table 1.2 in textbook) preamendment, postamendment, substantially equivalent, implant, custom, investigational, transitional
31
Regulation of Medical Devices
Second Way of classifications: (see Table 1.2 in textbook) Preamendment: Devices on the market before 5/28/1976 Postamendment: Devices on the market after 5/28/1976 Substantially equivalent: Equivalent to preamendment devices Implant: devices inserted in human body and intended to remain there for >30 days. Custom: Devices not available to other licensed and not in finished form Investigational: Unapproved devices undergoing clinical investigation Transitional: devices that were regulated as drugs and now defined as medical devices
Similar presentations
© 2025 SlidePlayer.com Inc.
All rights reserved.