1 Control circuit for an adaptive pressure targeting scheme (eg, Pressure Regulated Volume Control).
Control circuit for an adaptive pressure targeting scheme (eg, Pressure Regulated Volume Control). The controller is designed so that inspiratory pressure relative to PEEP (IP) is adjusted to meet the target tidal volume, given changes in resistance (R), compliance (C), and volume (Vmus) generated by patient effort (Pmus). IP will increase if R increases or C decreases. IP will decrease if Vmus increases. TI = inspiratory time. Robert L Chatburn, and Eduardo Mireles-Cabodevila Respir Care 2011;56:85-102 (c) 2012 by Daedalus Enterprises, Inc.