Download presentation
Presentation is loading. Please wait.
Published byMariusz Markowski Modified over 5 years ago
1
A case series of anomalous high pacing lead impedances in normally functioning leads
Abdul R. Mourad, MD, Michael H. Kim, MD, FHRS, Malcolm M. Kirk, MD, FHRS HeartRhythm Case Reports Volume 1, Issue 6, Pages (November 2015) DOI: /j.hrcr Copyright © 2015 Heart Rhythm Society Terms and Conditions
2
Figure 1 Impedance trends from St Jude Medical Merlin remote monitoring for the 3 patients show variation in the pattern of sporadic high impedances. A: Composite impedance trend over the first 2 years for patient 1. Multiple high impedances are seen. B: A single high impedance value was seen in patient 2, a few months after he had been programmed to a mode that avoided ventricular pacing. A few moderately high impedances were also seen in this patient. C: Patient 3 had a single high impedance value occurring 27 months after implant. HeartRhythm Case Reports 2015 1, DOI: ( /j.hrcr ) Copyright © 2015 Heart Rhythm Society Terms and Conditions
Similar presentations
© 2024 SlidePlayer.com Inc.
All rights reserved.