Midterm results of the multicenter trial of the Powerlink bifurcated system for endovascular aortic aneurysm repair  Jeffrey P. Carpenter, MD  Journal.

Similar presentations


Presentation on theme: "Midterm results of the multicenter trial of the Powerlink bifurcated system for endovascular aortic aneurysm repair  Jeffrey P. Carpenter, MD  Journal."— Presentation transcript:

1 Midterm results of the multicenter trial of the Powerlink bifurcated system for endovascular aortic aneurysm repair  Jeffrey P. Carpenter, MD  Journal of Vascular Surgery  Volume 40, Issue 5, Pages (November 2004) DOI: /j.jvs Copyright © 2004 The Society for Vascular Surgery Terms and Conditions

2 Fig 1 Powerlink bifurcated system. Device is a unibody bifurcated self-expanding fully stented endovascular graft. Stented endoskeleton is constructed as a single-wire body with a double spine, without sutures or welds. Stented portion is constructed of cobalt-chromium alloy. Endoskeleton is covered with graft material made of expandable polytetrafluoroethylene. Graft material is sutured to the endoskeleton only at the ends of the device. Device is supplied in 2 neck diameters (25 mm, 28 mm) and 2 fabric lengths (135 mm, 155 mm). Limbs of the bifurcated stent graft are 16 mm in diameter. Journal of Vascular Surgery  , DOI: ( /j.jvs ) Copyright © 2004 The Society for Vascular Surgery Terms and Conditions

3 Fig 2 Kaplan-Meier comparison of patient survival through 24 months. There was no significant difference in survival between Powerlink (test) group and control group (P = .1507, log-rank test). Journal of Vascular Surgery  , DOI: ( /j.jvs ) Copyright © 2004 The Society for Vascular Surgery Terms and Conditions

4 Fig 3 Kaplan-Meier comparison of major adverse event (MAE)–free survival through 24 months. While there were significantly fewer perioperative adverse events in the Powerlink cohort compared with the control group, over time there was no significant difference in overall adverse event–free survival (P = .456, log-rank). Journal of Vascular Surgery  , DOI: ( /j.jvs ) Copyright © 2004 The Society for Vascular Surgery Terms and Conditions

5 Fig 4 Survival free of endoleak.
Journal of Vascular Surgery  , DOI: ( /j.jvs ) Copyright © 2004 The Society for Vascular Surgery Terms and Conditions

6 Fig 5 Aneurysm diameter and volume response after endovascular aneurysm repair with Powerlink. Diameter and volume reduction were significant at each follow-up interval (see text; Table VIII; and Table IX, online only). Journal of Vascular Surgery  , DOI: ( /j.jvs ) Copyright © 2004 The Society for Vascular Surgery Terms and Conditions