1 Lag Correlator Testing
Aria Meyhoefer Carnegie Mellon University Summer 2001

2 The Lag Correlator What it is How it works What I did with it
Reflections Transmissions Signal Response Conclusions

3 17 Unit Lag Correlator Inductor Diode Array Capacitor Balun

4 How It Works Antennas read signals off the sky with path difference depending on source angle Signals travel along paths of equal length to Correlator Inside correlator delay lines remove signal path delays Each element of mixer array in the correlator measures the power from a particular direction The correlator also allows measurement of the spectrum of any signal detected

5 Testing the Correlator
Signal Response Test Send same signal* into both inputs of correlator Use different path lengths to create interference Read response from oscilloscope Reflections Test Send signal* into correlator through directional coupler Terminate other input and output Power reflects back though directional coupler to oscilloscope Transmissions Test Send a signal* into one input of correlator Read signal from other input of correlator Find attenuation of correlator * Same microwave signal of 9dBm from 0.01 to 26.5 GHz was always used

6 Compared to mixer results
Expected Response: One part in 158 Measured Response: One part in 6000 Correlator is only 2.6% effective

7 Reflections Results Left input reflects 21% of the power
Right input reflects 16% of the power

8 Transmission Results 0.03 mV from right to left
0.02 mV from left to right Sweep generator outputting 0.4 V Average attenuation of about 42 dB

9 Summary Correlator Response only 2.6%
Correlator Reflects about 18.5% power Attenuation of 42 dB

10 Thanks Jeff Peterson – Carnegie Mellon