Presentation is loading. Please wait.

Presentation is loading. Please wait.

A Patient Safety Initiative For Insulin Pumps Manufacturing Standards to improve insulin pump use and medical outcomes These proposals are near final,

Similar presentations


Presentation on theme: "A Patient Safety Initiative For Insulin Pumps Manufacturing Standards to improve insulin pump use and medical outcomes These proposals are near final,"— Presentation transcript:

1 A Patient Safety Initiative For Insulin Pumps Manufacturing Standards to improve insulin pump use and medical outcomes These proposals are near final, although suggestions for editorial changes are still welcome. Send your approval or comments to: John Walsh, PA, CDE at jwalsh@diabetesnet.com or by calling (619) 497-0900 jwalsh@diabetesnet.com

2 Introduction Over 500,000 insulin pumps are in use around the world, yet no formal guidelines regarding manufacturing standards and medical practice have been adopted by the diabetes clinical community and pump manufacturing industry. These are suggestions for such standards and once approved will be incorporated into future insulin pumps or into current pumps where software changes allow. These manufacturing standards are designed to Provide safer dosing increments to pump users Allow clinicians to make consistent dosing decisions when managing a variety of pumps Allow adjunctive medical and accessory personnel (ER, surgical, school nurses, etc.) to more easily be trained and interact with insulin pumps

3 Background Older insulin pumps were primarily designed to improve insulin delivery. Technological advances have transformed todays pumps and controllers into data collection centers. As additional data from continuous glucose monitoring devices becomes more widely used, the value of this data becomes even greater. Data needed for clinical monitoring and decisions is available in todays pump and can be accessed through screen displays, alerts, or recommendations. Using routine monitoring techniques and data analyses, a pump can inform the wearer regarding their current control and changes in control.

4 These Mechanical Standards Are Supported By: John Walsh, PA, CDE Ruth Roberts, MA Gary Scheiner, MS, CDE Timothy Bailey, MD, FACE Steve Edelman, MD Carol Wysham, MD Joseph Largay, PA, CDE David Horwitz, MD Etc * Reservations by a signatory about a standard will be noted

5 Definitions TDD – total daily dose of insulin (all basals and boluses) Basal –background insulin pumped slowly through the day to keep BG flat Bolus – a quick surge of insulin as Carb boluses to cover carbs Correction boluses to lower high readings that arise from too little basal insulin delivery or insufficient carb boluses Bolus On Board (BOB) – the units of bolus insulin with glucose-lowering activity still working from recent boluses Duration of Insulin Action (DIA) – time that a bolus will lower the BG. This is used to calculate BOB.

6 Why Insulin Pump Guidelines Are Needed These mechanical standards are designed to improve: 1.Consistency of pump settings between pump manufacturers 2.Accuracy and safety of carb and correction factor increments 3.Safety and consistency of DIA defaults and increments 4.Consistency in the handling of BOB and insulin stacking 5.Improved monitoring for hypoglycemia & hyperglycemia 6.Improved entry of glucose values into bolus calculations 7.Faster notification of excessive use of correction boluses 8.Faster identification of control problems related to infusion sets

7 Overview #Topic 1Carb Factor Increments 2Correction Factor Increments 3Carb Factor Accuracy 4Correction Factor Accuracy 5DIA Default Times 6DIA Time Increments 7Handling Of BOB #Topic 8.Multi-Linear And Curvilinear DIA 9.Hypoglycemia Alert 10.Hyperglycemia Alert 11.Correction Bolus Alert 12.Insulin Stacking Alert 13.Automatic Entry Of BG Values 14.Infusion Set Monitoring Slides are numbered by topic for easy reference.

8 Review What % of The TDD Changes The BG? To understand some slides that follow, it helps to know the significance of the effect that a change in the TDD has on the glucose level. Using a 450 Rule to set the carb factor and a 2000 Rule to set the correction factor: 1.25% of an appropriate TDD for an individual is sufficient to change the glucose about 25 mg/dl when given as a single dose A 5% change in the TDD is equivalent to about a 25 mg/dl increase or decrease in the glucose through the day A 5 to 6% change in the carb factor (about 2.5% to 3% of the TDD) is sufficient to change the glucose about 20 mg/dl per meal.

9 Carb Factor Increments 1

10 Carb factor increments shall be less than or equal to 5% of the next larger whole number so that each single step adjustment causes subsequent carb boluses to change by no more than 5% from previous doses. We recommend minimum carb factor increments of: 5 1.0 g/u above 20 g/u 0.5 g/u for 10 to 20 g/u 0.2 g/u for 5 to 9.8 g/u 0.1 g/u for 3 to 4.9 g/u 0.05 g/u for 0.1 to 2.95 g/u Standard For: Carb Factor Increments 1 5 Improved carb factor increments recommended by Gary Scheiner, MS, CDE

11 Issue: Current carb factor increments are too large to provide accurate carb boluses, especially for those who use smaller carb factors. This can represent a safety issue in situations where current carb factors lack the precision required to avoid excessive hyperglycemia and hypoglycemia. 1 Carb Factor (CarbF) Increments

12 Most pumps offer 1 gram per unit as their smallest CarbF increment. This increment becomes relatively large for CarbFs below 15 or 20 g/u. For instance, when the carb factor is reduced from 10 to 9 g/u, all subsequent carb boluses are increased by 11.1%. A shift in the carb factor from 1u/5g to 1u/4g causes each subsequent carb bolus to increase by 25%. For most pump users, a change in the carb factor larger than 5 or 6% would be expected to create more than a 20 mg/dl shift in the glucose following each meal. 1 Example Carb Factor Increments

13 Example Impact On BG From CarbF Adjustments This table shows the average additional fall in glucose after each meal of the day when a carb factor is reduced from 10 grams per unit to 9 grams per unit (for appropriate weight & TDD), and from 5 gr per unit to 4 gr per unit. How A 1-Step Reduction In CarbF Impacts Avg. Postprandial BG Change in CarbF Weight (~TDD) Avg Carb/day Avg carb gms/meal Change per meal Change in BG per meal* 1/10 to 1/9 160 lb (~40 u) 220 gr73 gr + 1.47 u (CorrF = 50) - 73.3 mg/dl** 1/5 to 1/4 240 lb (~100 u) 330 gr110 gr + 5.5 u (CorrF = 20) - 110 mg/dl** 1 * Calculated as avg. carbs/day – avg. carbs/day X 1 X 2000 new carb factor old carb factor 3 TDD ** Meals with higher carb intake would magnify these sample glucose changes

14 Review Median Carb Factor In unpublished data from the Cozmo Data Analysis Study: The median (middle) carb factor was 11.2 g/u Almost all pumpers used carb factors below 20 g/u 40% or more use carb factors of 10 g/u. 1

15 What Current Changes In CarbFs Do Table shows how subsequent carb boluses are affected by a one-step reduction in the CarbF using different CarbF increments. Yellow area shows values for most current pumps. Green areas show safer increments that impact subsequent boluses less than 5%. 1

16 Correction Factor Increments 2

17 For similar reasons, correction factor increments shall be less than or equal to 5% of the next larger whole number so that each single step adjustment causes subsequent correction boluses to change by no more than 5% from previous doses. We recommend minimum correction factor increments of: 5.0 mg/dl per uabove 80 mg/dl per u 2.0 mg/dl per ufor 40 to 78 mg/dl per u 1.0 mg/dl per u for 20 to 39 mg/dl per u 0.5 mg/dl per u for 10 to 19.5 mg/dl per u 0.2 mg/dl per u for 5 to 9.8 mg/dl per u 0.1 mg/dl per u for 3 to 4.9 mg/dl per u 0.05 mg/dl per u for 0.1 to 2.95 mg/dl per u Standard For: Correction Factor Increments 2

18 Carb Factor Accuracy 3

19 Standard For: Verification Of Carb Factor Accuracy 3 1.Insulin pump companies shall record and publish each year the carb factors used in insulin pumps returned for upgrade or repair. This report will include sufficient numbers of pumps to ensure statistical significance for commonly used carb factors between 5 and 20 grams per unit to ensure that pump training and clinical followup are assisting in the selection of accurate carb factors. 2.To improve accurate selection of carb factors, efforts shall be undertaken to automate carb factor testing.

20 Personal Carb Factors Issue: Many carb factors used in insulin pumps today are poorly tuned to the users need. When a carb factor does not match an individuals need, other sources of error in carb bolus calculations are significantly magnified. 3

21 Review Carb Factors In Use 1 Avg. carb factors* for 468 consecutive Cozmo insulin pump downloads (>126,000 boluses) are shown in blue Note that they are NOT bell-shaped or physiologic People prefer magic numbers – 7, 10, 15, and 20 g/unit – for their carb factors * Determined directly from grams of carb divided by carb bolus units for each carb bolus 7 10 115 20 1 3

22 Review Carb Factors In Use 1 MANY magic carb factors, shown in blue, are inaccurate. A more normal or physiologic distribution is shown in green Use of magic numbers creates major, consistent bolus errors that magnify other sources for error 7 10 115 20 1 3

23 Correction Factor Accuracy 4

24 Standard For: Verification Of Corr Factor Accuracy 4 1.Insulin pump companies shall record and publish each year the correction factors used in insulin pumps returned for upgrade or repair. This report will include sufficient numbers of pumps to ensure statistical significance for commonly used correction factors between 20 and 80 mg/dl per unit to ensure that pump training and clinical followup are assisting in the selection of accurate correction factors. 2.To improve accurate selection of correction factors, efforts shall be undertaken to automate correction factor testing.

25 Personal Correction Factors Issue: Many correction factors used in insulin pumps today are poorly tuned to the users need. This inaccuracy significantly magnifies other sources of error in correction bolus calculations. 4

26 Review Correction Factors In Use 1 Avg. correction factors in use for 452 consecutive Cozmo insulin pump downloads Like carb factors, correction factors in use are NOT bell-shaped or physiologic. A more accurate choice of correction factors would create a bell-shaped curve. Users or clinicians appear to frequently select magic numbers for correction factors. 7 10 115 20 1 4

27 DIA Default Times 5

28 Standard For: DIA Default Times 1.Default duration of insulin action (DIA) times in current pumps vary widely between 3 and 6 hours. For safety in bolus calculations that depend on DIA, the DIA default shall be set no shorter than 4.5 hours in pumps that determine DIA in a linear fashion and no shorter than 5 hours in pumps that determine DIA in a curvilinear or multi-linear fashion. These default times apply for the rapid insulins (lispro, aspart, and glulisine) in use at this time. 5

29 DIA Default Time Settings Issue: DIA measures the glucose-lowering activity of a carb or correction bolus over time. Current default times for DIA range from 3 to 6 hours in different pumps. The DIA is often considered another tool to improve control rather than being set at an appropriate value and focusing on more appropriate changes in basal rates or carb and correction factors to improve control. A DIA that is too short allows excess unrecognized bolus insulin to accumulate, usually in the afternoon and evening hours. Example: a bolus given at 7 am appears to have no activity after 10 am. If a high BG occurs 10 am, more bolus than needed will be given. At lunch, the bolus will be excessive, regardless of the BG at that time, creating a high likelihood of hypoglycemia. 5

30 Review How Long Do Boluses Lower The BG? Numerous GIR studies show rapid insulins lower the glucose for 5 hours or more. With Novolog (aspart) at 0.2 u/kg (0.091 u/lb), 23% of glucose lowering activity remained after 4 hours. 12 Another study found Novolog (0.2 u/kg) lowered the glucose for 5 hours and 43 min. +/- 1 hour. 13 After 0.3 u/kg or 0.136 u/lb of Humalog (lispro), peak glucose- lowering activity was seen at 2.4 hours and 30% of activity remained after 4 hours. 11 These times would be longer if the unmeasured basal suppression in pharmacodynamic studies were accounted for. 11 From Table 1 in Humalog Mix50/50 product information, PA 6872AMP, Eli Lilly and Company, issued January 15, 2007. 12 Mudaliar S, et al: Insulin aspart (B28 Asp-insulin): a fast-acting analog of human insulin. Diabetes Care 1999; 22:1501-1506. 13 L Heinemann, et al: Time-action profile of the insulin analogue B28Asp. Diabetic Med 1996;13:683-684. 5

31 A short DIA time hides true BOB level and its glucose- lowering activity. This can be a safety issue in that it: Leads to unexplained lows Leads to incorrect adjustments in basal rates, carb factors, and correction factors Causes users to start ignoring their smart pumps advice An inappropriately long DIA time overestimates bolus insulin activity – this leads to underdosing rather than overdosing on subsequent boluses. DIA should be based on an insulins real action time. Do NOT modify the DIA time to fix a control problem Review Short DIAs Hide Bolus Insulin Activity 5

32 Review Duration Of Insulin Action (DIA) 4 hrs 6 hrs 2 hrs 0 Glucose-lowering Activity Accurate bolus estimates require an accurate DIA. DIA times shorter than 4.5 to 7 hrs may hide BOB and its glucose lowering activity 5

33 Review DIA Large doses (0.3 u/kg = 30 u for 220 lb. person) of rapid insulin in 18 non- diabetic, obese people Med. doses (0.2 u/kg = 20 u for 220 lb. person) This study suggests that residual insulin activity can lower glucose levels for 7-8 hours Apidra product handout, Rev. April 2004a Regular 5

34 Review Does Dose Size Affect DIA? This graphic suggests that smaller boluses do not lower the BG as long as larger boluses. However, this may not be true – see next 2 slides. Size of the injected Humalog dose for a 154 lb or 70 kg person: 0.05 u/kg= 3.5 u 0.1 u/kg = 7 u 0.2 u/kg = 14 u 0.3 u/kg = 21 u Woodworth et al. Diabetes. 1993;42(Suppl. 1):54A 5

35 Review Pharmacodynamics Is Not DIA The DIA time entered into an insulin pump is based on studies of insulin pharmacodynamics. However, the traditional method used to determine the pharmacodynamics of insulin may underestimate insulins true duration of action, as shown in the next two slides. 5

36 Review Pharmacodynamics Underestimates DIA And Overestimates Impact Of Bolus Size To measure pharmaco- dynamics, glucose clamp studies are done in healthy individuals (0.05 to 0.3 u/kg) Because there is no basal suppression, this injected insulin ALSO SUPPRESSES normal basal release from the pancreas (grey area in figure) 5

37 Review Pharmacodynamic Time Does Not Equal DIA After accounting for the lack of basal suppression, True DIA times become longer than the PD times derived in traditional research If basal suppression activity is accounted for, small boluses may be found to have a longer DIA than it currently appears, erasing some of the apparent variation in DIA related to bolus size Some of the apparent inter- individual variation in pharmacodynamics may also disappear 5

38 DIA Time Increments 6

39 For safe and accurate estimates of residual BOB, DIA time increments shall be no greater than 15 minutes. Standard For: DIA Time Increments 6

40 Issue: Current DIA time increments vary from 15 minutes to 1 hour in different pumps When a pumps DIA time is adjusted, large time increments, such as 1 hr, can introduce large changes in subsequent estimates of BOB. For example, when the DIA is reduced from 5 hours to 4 hours, subsequent BOB estimates are decreased and recommendations for carb boluses are increased by about 25%. 6 DIA Time Increments

41 Most GIR studies suggest that pharmacodynamic action of insulin varies only about 25% to 40% between individuals. For a DIA time of 5 hr and 15 min, a 25% range is equivalent to 1 hr and 20 min, such as from 4 hrs and 30 min to 5hr and 50 min. A pump that has 1 hr DIA increments would enable the user to select only 1 or 2 settings within this physiologic range, while a 30 min increment would allow only 2 or 3 choices that are close to a physiologic range. 6 Review Glucose Infusion Rate (GIR) Studies

42 Handling Of BOB Bolus On Board (BOB) 7

43 Standard For: Handling Of BOB For safe and accurate BOB measurement: BOB measurements shall include all carb and correction boluses given within the selected DIA When residual BOB is present at the time of a bolus, the BOB shall be subtracted from both carb and correction bolus recommendations. When BOB exceeds the current correction bolus need or the current carb plus correction need, the user will be alerted to how many grams of carb they need to eat. [(BOB – correction & carb bolus need) X carb factor] 7

44 Issue: Current pumps differ significantly in what is counted as BOB and in whether or not BOB is subtracted from subsequent carb boluses. Most insulin pumps assume that excess BOB does not need to be taken into account when determining the next carb bolus. Though commonly determined in this way, the resulting bolus dose recommendations can cause unexplained and unnecessary insulin stacking and hypoglycemia. Handling Of BOB 7

45 Example Insulin Stacking With a bedtime BG of 173 mg/dl, is there an insulin deficit or a carb deficit? 6 pm8 pm10 pm12 am Dinner Dessert Correction Bedtime BG = 173 mg/dl 7

46 Review Frequency Of Insulin Stacking CDA1 Study Results Of 201,538 boluses, 64.8% were given within 4.5 hrs of a previous bolus Although 4.5 hours may underestimate true DIA, use of this minimal DIA time shows that some BOB is present for MOST boluses 4.5 hrs 7

47 Review Bolus On Board (BOB) An accurate measurement of the glucose-lowering activity that remains from recent boluses: Prevents insulin stacking Improves bolus accuracy Allows the current carb or insulin deficit to be determined aka: insulin on board, active insulin, unused insulin* * Introduced as Unused Insulin in 1st ed of Pumping Insulin (1989) 7

48 Review How Current Pumps Handle BOB Whats In the BOB & What Is It Applied Against? BOB Includes This Type Of Bolus BOB Is Subtracted From This Type Of Bolus CarbCorrectionCarbCorrection Animas 2020Yes No*Yes Deltec CozmoYes Insulet OmnipodNoYesNoYes Medtronic ParadigmYes No Yes * Except when BG is below target BG * Yes is generally safer 7

49 Example Unsafe BOB 1 Handling If a pump user gets frustrated with a high BG and they overdose to speed its fall, or they exercise longer or more intensely than anticipated, they can acquire a significant excess in BOB. In this situation, most current pumps recommend that a bolus be given for all carb intake regardless of how much BOB is actually present. If BOB is greater than the correction bolus requirement at the time, the pumps bolus recommendation may introduce a risk for hypoglycemia. 1 Pumping Insulin, 1st ed, 1989, Chap 12, pgs 70-73: The Unused Insulin Rule 7

50 Example Differences In Bolus Recommendations Situation: BOB = 3.0 u and 30 gr. of carb will be eaten at these glucose levels Carb factor = 1u / 10 gr Corr. Factor = 1 u / 40 mg/dl over 100 Target BG = 100 TDD = ~50 u units mg/dl Omnipod bolus cannot be determined - it counts only correction bolus insulin as BOB The graphic shows how widely bolus recommendations vary from one pump to another for the same situation. 7

51 Review Track BOB Or Carb Digestion? For safety after meal and correction boluses, tracking the glucose- lowering action of BOB is more important than accounting for the glucose-raising action of digesting meal carbs: When a BG is taken after a meal, the BOB times the correction factor ideally represents the maximum fall in glucose expected. Accounting for the impact of the BOB on the current glucose provides the safest approach in the determination of bolus recommendations. Low glycemic index meals, gastroparesis, Symlin, and other issues may counteract a predicted fall in glucose based on BOB, but the user can more easily judge and remedy this situation than dealing with an unknown excess of insulin. 7

52 Exceptions To Usual Handling Of BOB When a second bolus is taken for an unplanned carb intake or a desert that is consumed within 60 minutes* or so of a meal bolus, BOB should not be taken into account for the second bolus because the impact of the first bolus cannot be accurately determined. Given that, it is wise to account for BOB as soon after a meal as possible, such as within 60 to 90 minutes,* to provide early warning if the bolus given was excessive or inadequate. Accounting for all BOB and applying it to subsequent boluses is generally safer, although not always more accurate. 7 * Adjustable setting in pump/controller

53 Multi-Linear And Curvilinear DIA 8

54 Standard For: Multi-Linear And Curvilinear DIA Insulin pumps shall use either a 100% curvilinear or a multi-linear method to improve accuracy and consistency of BOB estimates. 8

55 Linear And Curvilinear DIA Issue: Pump manufacturers use at least 3 different methods (100% curvilinear, 95% of curvilinear, and straight linear) to measure DIA and BOB. When a realistic DIA time is selected, a linear determination of residual BOB will not be as accurate as a curvilinear method that incorporates the slow onset of insulin action and its longer tailing off in activity. In most situations, an accurate determination of insulins tailing activity will be most important to the pump user. 8

56 Linear And Curvilinear DIA Compared Note how values for the 5 hr linear line in red and the thinner 5 hr curvilinear line diverge in value at several points along the graph. 5 hr Linear 5 hr 95% Curvilinear From Pumping Insulin, 4th ed., adapted fom Mudaliar et al: Diabetes Care, 22: 1501, 1999 8

57 Example A Multi-Linear DIA Use of a multi-linear method to measure DIA improves accuracy. The next page shows a triple-linear example for measurement of BOB. 8

58 Example A Triple-Linear Approximation Of DIA A triple-linear line in red can more closely approximate a curvilinear DIA. For a 5 hr DIA*: 1st 10%** – no change Mid 65%** – fall 75%** Last 25%** – fall 25%** (** adjustable as needed in device) 5 hr Triple Linear 8 * % modification suggested by Gary Scheiner, MS, CDE

59 Hypoglycemia Alert 9

60 Standards For: Hypoglycemia Alert 1.Insulin pumps that store glucose and insulin dosing data shall present this glucose control data in a readily accessible form on the pump or controller. 2.The pump shall alert the user when the glucose data from their glucose monitor or continuous monitor suggests they are experiencing frequent* or severe* patterns of hypoglycemia. 9 * Adjustable settings in pump/controller

61 Frequent/Severe Hypoglycemia Alert Issue: Although most current insulin pumps contain sufficient data to do so, pumps give no warning to a user when they are experiencing patterns of frequent or severe hypoglycemia. 9

62 Example Pump Screen Hypoglycemia Display 1 9 * Adjustable settings in pump/controller Weekly History – Low BGs # of weeks1248 # BGs/week23252832 % BGs/week < 50 mg/dl* 22%18%11%8% % BGs/week < 70 mg/dl* 35%27%19%16% Modified to % display per Gary Scheiner, MS, CDE

63 Example Pump Screen Hypoglycemia Display 2 9 * Adjustable settings in pump/controller Low BGs By Time Of Day For 1 Week* Avg BGs for 1 week23 (3.3/day) Time Period 4a-10a10a-4p4a-10p10p-4a % BGs < 50 mg/dl* 57%14%0% % BGs/week < 70 mg/dl* 71%28%0%14% Modified to % display per Gary Scheiner, MS, CDE

64 Hyperglycemia Alert 10

65 Standards For: Hyperglycemia Alert 1.Insulin pumps that store glucose and insulin dosing data shall present this glucose control data in a readily accessible form on the pump or controller. 2.The pump shall alert the user when the glucose data from their glucose monitor or continuous monitor suggests they are experiencing patterns of frequent* or severe* hyperglycemia. 10 * Adjustable settings in pump/controller

66 Frequent/Severe Hyperglycemia Issue: Although most current insulin pumps contain sufficient data to do so, pumps give no warning to a user when they are experiencing patterns of frequent or severe hyperglycemia. 10

67 Example Pump Screen Hyperglycemia Display 1 10 * Adjustable settings in pump/controller Weekly History – High BGs # of weeks1248 Avg BGs/week23252832 % BGs/week > 180 mg/dl* 26%30%29%26% % BGs/week >220 mg/dl* 17%18%22%23% Modified to % display per Gary Scheiner, MS, CDE

68 Example Pump Screen Hypoglycemia Display 2 10 * Adjustable settings in pump/controller Low BGs By Time Of Day For 4 Weeks* # BGs for 4 weeks112 (4.0/day) Time Period 4a-10a10a-4p4a-10p10p-4a % BGs > 180 mg/dl* 1%3%4%21% % BGs/week > 220 mg/dl* 0%2%3%18% Modified to % display per Gary Scheiner, MS, CDE

69 Correction Bolus Alert 11

70 Standard For: Correction Bolus Alert 1.Insulin pumps shall show in a readily accessible history screen the percentage of the TDD that is used for correction boluses over time. 2.The insulin pump shall alert the wearer when they are using more than 8%* of their TDD for correction bolus doses in the most recent 4 day* period. 11 * Adjustable settings in pump/controller

71 Correction Bolus Alert Issue: Hyperglycemia is more common than hypoglycemia for most people on insulin pumps. When glucose levels consistently run high, many pump users address the problem by giving frequent correction boluses rather than correcting the core problem through an increase in their basal rates or carb boluses. If the correction bolus % becomes excessive relative to the TDD, this information is often not shown and no alert is given regarding the possible excessive use of correction boluses. 11

72 Insulin Stacking Alert Accurate accounting of BOB becomes more important in those who experience frequent or severe hypoglycemia, as well as those whose average glucose levels are closer to normal values. 12

73 Standard For: Insulin Stacking Alert Insulin pumps shall alert the wearer when they are giving a bolus and no glucose value has been entered in the pump. This is especially necessary when the user has sufficient insulin stacking* to significantly alter the bolus they would otherwise give. The alert is on by default once a DIA time is selected to measure BOB, but may be turned off if the user desires. * Such as when the BOB is greater than 1.25%** of the avg. TDD, sufficient to change the glucose about 25 mg/dl. (** Adjustable setting in pump/controller for a certain fall in glucose selected by the user or clinician) 12

74 Insulin Stacking Alert Issue: Pump users often bolus for carbs without checking their glucose first. With no glucose reading, the pump cannot account for BOB, nor appropriately adjust a bolus for the BOB or the current BG. Even without a glucose test, data available in the pump at the time of a bolus can determine whether enough BOB is present to substantially change a bolus dose. The pump can alert the user to this unseen, substantial insulin stacking. 12

75 When a carb bolus is planned without a recent BG check, but BOB is more than 1.25%* of the average TDD (enough to cause about a 25 mg/dl drop in the glucose), the pump will recommend that the wearer do a BG check due to the substantial presence of BOB. For instance, for someone with: Avg TDD1.25%* of TDD 40 units 0.5 units 50 mg/dl per u (corr factor) X 0.5 u = 25 mg/dl This individual would be alerted when they do not check their glucose and want to give a bolus but have 0.5 u or more of BOB present. Example Insulin Stacking or BOB Alert * Adjustable in pump/controller for a reasonable degree of safety 12

76 Automatic Entry Of BG Values 13

77 Standard For: Automatic Entry Of BG Values For completion of the glucose history, improved handling of BOB, and more accurate bolus recommendations, insulin pumps shall be enabled to have wireless or direct entry of BG test results from a glucose meter. Automatic glucose entry from two or more major brands of meters is recommended to increase the likelihood of insurance coverage for test strips. 13

78 Automatic Entry Of BG Values Issue: Pump users do not enter as many BG values into their pump when readings must be entered manually rather than having a glucose value automatically entered from a meter. When a glucose is checked but not entered, the lost data cannot be used to account for BOB, warn of insulin stacking (see #12), nor be used to analyze glucose patterns and the frequency of hypoglycemia and hyperglycemia. Relatively normal and hypoglycemia values are less likely to be entered manually, but are more likely to be influenced by BOB. 13

79 Review Automatic Entry Of BG Values In a study of over 500 insulin pumps where BG values could be entered either manually or automatically, users entered 2.6 BG values per day manually, compared to 4.1 values per day for pumps in which glucose values wre automatically entered. * BOB could typically be taken into account for 1.5 additional boluses per day with automatic entry of BG values. When glucose values are not automatically entered, BOB cannot be determined and bolus recommendations will not be as accurate. 13 * J. Walsh, D. Wroblewski, T.S. Bailey: unpublished data

80 Review Automatic Entry Of BG Values Automatic entry of glucose values into pumps offers a significant clinical advantage to users because more boluses will be adjusted for high and low BGs, and residual BOB is more likely to taken into accout in bolus calculations. Automatic entry of glucose values ensures a greater degree of safety for those who experience frequent or sever hypoglycemia, and those whose glucose values are closer to normal. 13

81 Infusion Set Monitoring 14

82 Standards For: Infusion Sets 1.Insulin pumps shall monitor and record in easily accessible history the duration of infusion set usage recorded as mean, median, and SD of time of use. 2.Insulin pumps shall monitor and report average glucose values in full and partial 24 hour* time intervals between set changes with the ability to change the observation interval, such as 1 to 30 set changes.* 3.These steps allow the HCP and user to identify infusion set problem from loss of glucose control and variations in patterns of use. 14 * Adjustable setting in pump/controller

83 Infusion Sets Issue: A significant number of non-patch pump wearers encounter infusion set problems. These problems may arise from poor infusion set design or inadequate site preparation. Often the source for the randomly erratic glucose readings that follow are difficult to identify by a user or clinician. 14

84 Review Infusion Set Failure One common problem source for infusion set failure arises when a Teflon infusion set comes loose beneath the skin from movement or tugging. Some of the infused insulin then leaks back to the skin surface resulting in unexplained high readings. A complete loss of glucose control is typically seen when an infusion set is pulled out entirely. Selecting the right infusion set plus good site technique, especially taping the infusion line to the skin, can significantly minimize the number of unexplained high readings for many pump wearers. 14

85 Why The Tubing Needs To Be Taped Most problems with infusion sets come from loosening of the Teflon under the skin, not from a complete pullout. A 1 tape placed on the infusion line: Stops tugging on the Teflon catheter under the skin Prevents loosening of the Teflon catheter under the skin Avoids many unexplained highs caused when insulin leaks back to the skin surface Reduces skin irritation And prevents many pull outs 14

86 Tape The Tubing This helps prevent Tugging Irritation Bleeding 14

87 No tape! 14 Examples Lack Of Anchoring Of Sets A review of dozens of pictures of infusion sets online and pump manuals finds that anchoring of the infusion line with tape is rarely recommended or practiced.

88 Review Infusion Set Monitor Many pump wearers experience random erratic readings until they change to a different infusion set or start to anchor their infusion lines with tape to stop line tugging. However, insulin pumps offer no mechanism for clinicians or pump users to detect who may be having problems with their infusion sets. 14

89 Tool Infusion Set Monitor Insulin pumps with direct BG entry can identify those who may be having intermittent loss of glucose control secondary to infusion set failure. The pump: Shows the average time and variation in time of use between reservoir loads or use of the priming function. Shows average BGs for each full or partial 24* hour time interval following set changes (indicated by the prime function) over a various number of set changes* or as soon as statistical significance is reached. 14 * Adjustable setting in pump/controller


Download ppt "A Patient Safety Initiative For Insulin Pumps Manufacturing Standards to improve insulin pump use and medical outcomes These proposals are near final,"

Similar presentations


Ads by Google