1 ARCHITECT c 16000 / c 8000 Differences
4/12/2019 ARCHITECT c / c Differences Document Control Number ©2006, Abbott Laboratories, Abbott Park, Illinois 1 of 18

2 ARCHITECT c 16000 / c 8000 Differences
4/12/2019 ARCHITECT c / c 8000 Differences The information, documents and related graphics published herein (the "Information") are the sole property of Abbott Laboratories. Permission to use the Information is granted, provided that the copyright notice appears on all copies; use of the information is for operation of ABBOTT products by Abbott trained personnel or information use only; the information is not modified in any way; and no graphics are used separate from accompanying text. All samples (printouts, graphics, displays, screens, etc.) are for information and illustration purposes only and shall not be used for clinical or maintenance evaluations. Data shown in sample printouts and screens do not reflect actual patient names or test results. The information was developed to be used by Abbott Laboratories-trained personnel, by other persons knowledgeable or experienced with the operation and service of the product identified, under the supervision and with cooperation from Abbott Laboratories technical support or service representatives. In no event shall Abbott Laboratories or its affiliates be liable for any damages or losses incurred in connection with or arising from the use of the information by persons not fully trained by Abbott Laboratories. This limitation shall not apply to those persons knowledgeable or experienced with the operation and service of the product identified, under the supervision and with cooperation from Abbott Laboratories technical sales or service representatives. No part of this media may be reproduced, stored, retrieved, or transmitted in any form or by any means without the prior written permission of Abbott Laboratories. Updates to the Information may be provided in either paper or electronic format. Always refer to the latest documents for the most current information. ABBOTT LABORATORIES MAKES NO REPRESENTATIONS OR WARRANTIES OF ANY KIND OR NATURE WITH RESPECT TO THE INFORMATION. ABBOTT LABORATORIES HEREBY DISCLAIMS ALL REPRESENTATIONS AND WARRANTIES, WHETHER EXPRESS OR IMPLIED, CREATED BY LAW, CONTRACT OR OTHERWISE, INCLUDING, WITHOUT LIMITATION, ANY WARRANTIES OF MERCHANTABILITY, FITNESS FOR A PARTICULAR PURPOSE, TITLE OR NON-INFRINGEMENT. IN NO EVENT SHALL ABBOTT LABORATORIES BE LIABLE FOR ANY DAMAGES OF ANY KIND OR NATURE, INCLUDING, WITHOUT LIMITATION, DIRECT, INDIRECT, SPECIAL (INCLUDING LOSS OF PROFIT) CONSEQUENTIAL OR INCIDENTAL DAMAGES ARISING FROM OR IN CONNECTION WITH THE EXISTENCE OR USE OF THE INFORMATION, REGARDLESS OF WHETHER ABBOTT LABORATORIES HAS BEEN ADVISED AS TO THE POSSIBILITY OF SUCH DAMAGES Each person assumes full responsibility and all risks arising from use of the Information. The Information is presented "AS IS" and may include technical inaccuracies or typographical errors. Abbott Laboratories reserves the right to make additions, deletions, or modifications to the Information at any time without any prior notification. ARCHITECT and c 8000 are registered trademarks and c is a trademark of Abbott Laboratories. All Abbott Laboratories product names and trademarks are owned by or licensed to Abbott Laboratories, its subsidiaries or affiliates. No use of any Abbott trademark, trade name, trade dress, or product name may be made without the prior written authorization of Abbott Laboratories, except to identify the product or services of Abbott Laboratories. All other trademarks, brands, product names, and trade names are the property of their respective companies. All rights reserved. Except as permitted above, no license or right, express or implied, is granted to any person under any patent, trademark, or other proprietary right of Abbott Laboratories. Document Control Number Revision or Release Date Content or Pages Revised, Added, or Deleted 12/05/2006 New Release Document Control Number ©2006, Abbott Laboratories, Abbott Park, Illinois

3 4/12/2019 Purpose This presentation highlights the significant differences between the ARCHITECT c and c 8000 systems and should be used in conjunction with the ARCHITECT c Online Help. Document Control Number ©2006, Abbott Laboratories, Abbott Park, Illinois

4 Location of components Dual line feature Loading onboard solutions
Topics Location of components Dual line feature Loading onboard solutions Loading reagents Assay Files Maintenance Changes Document Control Number ©2006, Abbott Laboratories, Abbott Park, Illinois

5 Location of Components
4/12/2019 Location of Components The location of the following c components has changed when compared to the c 8000 system. Front: Rear: - Sample and Reagent Syringes - ICT Aspiration Pump - Probe Wash Pumps - ICT Reference Pump - Bulk Solution Supply - High Concentration Waste Pump - Wash Solution Pumps - Cuvette Wash Pump Document Control Number ©2006, Abbott Laboratories, Abbott Park, Illinois

6 Location of Components (cont’d)
4/12/2019 Location of Components (cont’d) ARCHITECT c supply and pump center components Front View (1) Sample, reagent 1 and reagent 2 syringe drives (2) Probe wash pumps (3) Bulk solutions (4) Cuvette wash pump (5) Wash solution pump Rear View (1) ICT aspiration pump (2) ICT reference solution pump (3) High concentration waste pump Document Control Number ©2006, Abbott Laboratories, Abbott Park, Illinois

7 Dual Line Feature The c uses a dual line feature to pair assays for each run improving the throughput of the c The system aspirates reagent for two assays, one assay on a designated “A-line” and the other assay on a designated “B-line”, in the same cycle. Document Control Number ©2006, Abbott Laboratories, Abbott Park, Illinois

8 Dual Line Feature (cont’d)
Both Reagent Supply Centers (R1 and R2) consist of an inner and outer carousel Outer carousel = A-line (shaded) Inner carousel = B-line (unshaded) Document Control Number ©2006, Abbott Laboratories, Abbott Park, Illinois

9 Dual Line Feature (cont’d)
4/12/2019 Dual Line Feature (cont’d) The following components exist in pairs (A and B) to accommodate dual line technology : Reagent probes, R1 & R2 Mixers, R1 & R2 Reagent syringes, R1 & R2 Cuvette washer nozzles 1-8 (total nozzles = 16) Probe wash pumps Wash solution pump (alkaline & acid) NOTE: A single sample probe dispenses sample into two separate cuvettes (A and B) during a single cycle. Document Control Number ©2006, Abbott Laboratories, Abbott Park, Illinois

10 Loading onboard solutions
4/12/2019 Loading onboard solutions Onboard solutions consist of Wash Solutions, ICT Diluent, Saline, and Water Bath Additive. Differences in loading onboard solutions are displayed in the following table: Onboard solution c loading c 8000 loading Detergent A 10% Detergent B 0.5% Acid wash R1 & R2 (C1 & D1) R1 & R2 (C2 & D2) R1 & R2 (C3 & D3) R1 & R2 (D1, E1 or E2) Water bath additive R1, A1 R2, A1 ICT Sample Diluent R1, outer carousel only (any position) R1 (any position) Saline (for sample dilution) R1, outer and inner carousel (as configured on Reagent Status screen) R1 (as configured on Reagent Status screen) Document Control Number ©2006, Abbott Laboratories, Abbott Park, Illinois

11 Loading onboard solutions (cont’d)
4/12/2019 Loading onboard solutions (cont’d) c Supply Status screen c 8000 Supply Status screen Document Control Number ©2006, Abbott Laboratories, Abbott Park, Illinois

12 Loading Reagents Reagent cartridge Reagent supply center placement
4/12/2019 Loading Reagents Reagents Both cartridges of a two reagent assay must be placed on the same line (A or B). Refer to Iron and Alkaline Phosphatase examples in table below: ICT sample diluent must be placed in the outer carousel (A-line) of the R1 reagent supply center. Reagent cartridge Reagent supply center placement Iron R1 cartridge Iron R1 - outer carousel (A-line) Iron R2 cartridge Iron R2 – outer carousel (A-line) Alk phos R1 cartridge Alk phos R1 – inner carousel (B-line) Alk phos R2 cartridge Alk phos R2 – inner carousel (B-line) Document Control Number ©2006, Abbott Laboratories, Abbott Park, Illinois

13 Loading Reagents (cont’d)
4/12/2019 Loading Reagents (cont’d) c Reagent Status screen c 8000 Reagent Status screen Document Control Number ©2006, Abbott Laboratories, Abbott Park, Illinois

14 Loading Reagents (cont’d)
4/12/2019 Loading Reagents (cont’d) Replenishing reagents with an active calibration curve To avoid recalibration when replacing an empty reagent cartridge, load the new cartridge on the same line as the reagent with the Active calibration status. NOTE: If a reagent with an Active calibration status is moved from one line to the other (A/B) and scanned, the calibration status for that reagent changes to No Cal. Document Control Number ©2006, Abbott Laboratories, Abbott Park, Illinois

15 Assay Files Assay Import Message Description
4/12/2019 Assay Files Assay Import When you copy an assay file from an ARCHITECT c 8000 System to an ARCHITECT c System (or vice versa), the sample probe wash protocol and/or wavelength assay parameters may change. A status message displays to indicate this change during the import process. Message Description OK - Multiple parameters converted, review Message History. The following assay parameters were changed: • Sample wash protocol for sample probe changed from Optimized throughput (c 8000 only) to Maximum wash (c 8000 and c 16000). See Add / edit SmartWash window – Sample probe view (c System) field descriptions. • Primary wavelength and/or Secondary wavelength. See Configure assay parameters window – General – Reaction definition view (photometric – c System) field Descriptions. Document Control Number ©2006, Abbott Laboratories, Abbott Park, Illinois

16 Assay Files (cont’d) Assay Import (cont’d) Message Description
OK - Sample wash protocol for sample probe converted to Maximum Wash Sample wash protocol for sample probe changed from Optimized throughput (c 8000 only) to Maximum wash (c 8000 and c 16000). See Add / edit SmartWash window – Sample probe view (c System) field descriptions. OK - One or more wavelengths were changed. Primary wavelength and/or Secondary wavelength. See Configure assay parameters window – General – Reaction definition view (photometric – c System) field Document Control Number ©2006, Abbott Laboratories, Abbott Park, Illinois

17 Assay Files (cont’d) c 16000 c 8000 Assay Wavelength Changes 416 412
The following wavelengths are different on the c16000 and c8000 systems: c 16000 c 8000 416 412 450 444 Document Control Number ©2006, Abbott Laboratories, Abbott Park, Illinois

18 4/12/2019 Maintenance Changes The following daily and weekly maintenance procedures have been modified. These changes apply to both the c and c 8000 systems. Current c 8000 procedure New c system procedure Daily Maintenance Check 1mL Syringes Procedure name did not change Moved the following maintenance step to Weekly maintenance: - Check ICT aspiration and reference pump 1 mL syringes. Weekly Maintenance 6019 – Check ICT Probe and Tubing Procedure name changed to 6019 – Check ICT components. Added the following maintenance step: - Check ICT aspiration and reference reference pump 1 mL syringes. Document Control Number ©2006, Abbott Laboratories, Abbott Park, Illinois