1
Influence of airway-occluding instruments on airway pressure during jet ventilation for rigid bronchoscopy P. Biro, M. Layer, H.D. Becker, F. Herth, K. Wiedemann, B. Seifert, D.R. Spahn British Journal of Anaesthesia Volume 85, Issue 3, Pages (September 2000) DOI: /bja/ Copyright © 2000 British Journal of Anaesthesia Terms and Conditions
2
Fig 1 Proximal (dotted line) and distal (solid line) airway pressure (Paw) in patient no. 6 during increasing obstruction of the airway: Period A: apnoea during zeroing of both pressure sensors; period B: beginning of expansion of endosonography probe balloon; period C: complete occlusion of the airway. British Journal of Anaesthesia , DOI: ( /bja/ ) Copyright © 2000 British Journal of Anaesthesia Terms and Conditions
3
Fig 2 Proximal and distal Paw during 0%, 50% and 90% obstruction of the airway and a driving pressure of 2.0 atm (upper segment) and 3.0 atm (lower segment). Values are expressed as mean±sd; *Significantly different from Paw without obstruction (P<0.025), †Pressure amplitude significantly different from that without obstruction (P<0.025). British Journal of Anaesthesia , DOI: ( /bja/ ) Copyright © 2000 British Journal of Anaesthesia Terms and Conditions