APPLICATION NOTES THE NQAD IS HIGHLY SENSITIVE,

Similar presentations


Presentation on theme: "APPLICATION NOTES THE NQAD IS HIGHLY SENSITIVE,"— Presentation transcript:

1 APPLICATION NOTES THE NQAD IS HIGHLY SENSITIVE,
ENSURING ALL SAMPLE COMPONENTS CAN BE DETECTED Application note # QAN-0083 CHROMATOGRAMS 0.00 0.25 0.50 0.75 1.00 1.25 1.50 1.75 2.00 2.25 2.50 2.75 3.00 3.25 3.50 3.75 4.00 4.25 4.50 4.4400 4.5200 mV min PARAMETERS NQAD S/N = 12.2 FLOW RATE: 1.0 mL/min COLUMN: Thermo BetaBasic C18, 150 x 4.6 mm, 5u COLUMN TEMP: Ambient INJECTION VOLUME :10 µL CONCENTRATION: 16 ng on column ANALYTE: Aspartame NQAD GAS: Air FILTER: 1.3 sec GAIN: 1X EVAPORATION TEMP: 35C ELSD GAS: Nitrogen, 1.5 L/min GAIN: 1 EVAPORATION TEMP: 40C HPLC MOBILE PHASE: 20:80 ACN:Water APPLICATION NOTES 0.00 0.25 0.50 0.75 1.00 1.25 1.50 1.75 2.00 2.25 2.50 2.75 3.00 3.25 3.50 3.75 4.00 4.25 4.50 44.680 44.800 mV min ELSD Demonstrates better sensitivity than ELSD QUANT TECHNOLOGIES, LLC TH Lane NE Blaine, MN, USA Note: This application was run at Quant Technologies on a single ELSD and may not be representative of all ELSD’s. 1