1 FOR DISTRIBUTION BY MEDTRONIC OFFICE OF MEDICAL AFFAIRS ONLY.
The information in this document is provided by Medtronic in response to an unsolicited request. It contains information on indications for which safety and effectiveness have not been established to the level required by the FDA, and it should not be further distributed in the United States. Medtronic provided financial support related to the information included in this document. Insurers may or may not cover all uses described, it is advisable to confirm coverage with carriers before filing claims.

2 Primary Results Following Percutaneous Coronary Intervention with the 38mm Resolute Zotarolimus-eluting Stent Shirish Hiremath, Robaayah Zambahari, Martin Leon, Laura Mauri, Alan Yeung and Michael Lee On behalf of the RESOLUTE US and RESOLUTE Asia Investigators TCT 2012

3 Background Implantation of drug-eluting stents in long coronary artery lesions is associated with a higher risk for restenosis and stent thrombosis related to the need for multiple and overlapping stents. Historical data indicate that the XIENCE PRIME LL stent is associated with a 1yr TLF rate of 7.7% in long lesions1. Furthermore a recent publication showed no statistical differences at 1 year between R-ZES and SES for the treatment of long coronary artery lesions2. The Resolute zotarolimus-eluting stent (R-ZES) is manufactured in a 38mm length to accommodate longer lesions but clinical data from the RESOLUTE Global Clinical Program to demonstrate efficacy and safety of the longer stent have not yet been reported. 1 Costa, M. et al. presented at TCT2011 2 Ahn, J-M, et al. Circ Cardiovasc Interv Oct 9. [Epub ahead of print]

4 RESOLUTE Global Clinical Program
2 Trials with Data for 38mm Resolute Stents Enrollment Complete - In Follow Up RESOLUTE1 Non-RCT First-in-Human (R=139) 5 yr RESOLUTE AC2,3 1:1 RCT vs. Xience V (R=1140; X=1152) 3 yr RESOLUTE Int4 Non-RCT Observational (R=2349) 3 yr RESOLUTE US5 2.25 – 4.0 mm Non-RCT vs. Hx Control (R=1402) 2 yr RESOLUTE Japan 2.5 – 3.5 mm Non-RCT (R=100) vs. Hx Control 2 yr R Japan SVS 2.25 Non-RCT vs. PG (R=65) < 1yr RESOLUTE US 38 mm sub-study Non-RCT vs. PG (R=114) < 1yr R-China RCT 1:1 RCT vs. Taxus (R=200; T=200) < 1yr RESOLUTE Asia Non-RCT Observational (R=312) 1yr R-China Registry Non-RCT Observational (R=1800) < 1yr Enrolling / Planning RI-US Registry Post-approval study (R=230) enrolling 1 Meredith IT, et al. EuroIntervention. 2010;5: Serruys PW, et al. N Engl J Med. 2010;363: Silber S, et al. Lancet. 2011;377: Neumann FJ, et al. EuroIntervention. 2012;7(10): Yeung AC, et al. JACC. 2011;57: 4

5 RESOLUTE 38mm Methods A sub-study of 2 prospective, multicenter clinical trials, RESOLUTE US and RESOLUTE Asia, enrolled patients with de novo coronary artery lesions amenable to treatment with the 38mm length R-ZES in target lesions of at least 35 mm. The primary endpoint was target lesion failure (TLF*) at 1 year. The rate of TLF at 1 year was compared to a performance goal of 19.0% derived from a logistic regression model. All patients were prescribed dual antiplatelet therapy for a minimum of 6 months. TLF is defined as cardiac death, target vessel myocardial infarction (TVMI), or clinically driven target lesion revascularization (TLR).

6 RESOLUTE 38mm Patient Flowchart RESOLUTE US RESOLUTE Asia
38mm Substudy N = 114 38mm Substudy N = 109 38 mm Substudy N = 223 1 Year Follow-up n = 222 (99.6%)

7 RESOLUTE 38mm Baseline Characteristics % N = 223 pts Age (yr)
60.9 ± 10.6 Male 78.9 Diabetes mellitus 37.7 IDDM 10.3 Hypertension 74.9 Hyperlipidemia 58.7 Current smoker 18.8 Family history of CAD 37.2 Prior MI 32.4 Prior PCI 27.4 Prior CABG 7.2 Cardiac status: Stable angina 39.2 Unstable angina 47.4 Myocardial infarction 13.4

8 RESOLUTE 38mm Lesion & Procedure Characteristics N = 223 pts,
269 lesions Number of diseased vessels (>50%) Single 46.2 Double 36.3 Triple 17.5 Lesion location LAD 52.0 LCx 20.2 RCA 44.4 Lesion length Discrete (<10mm) 4.6 Tubular ( mm) 24.5 Diffuse (> 20mm) 70.9 Branch vessel disease 47.9 B2/C lesion 91.2 Lesion length (mm) 25.22 ± 8.83 RVD (mm) 2.78 ± 0.42 MLD (mm) 0.80 ± 0.36 Pre-procedure % diameter stenosis 71.33 ± 11.61 QCA measurements may have been made by careful visual estimate, on-line QCA or IVUS.

9 RESOLUTE 38mm Acute Success Rates % N = 223 patients, 269 lesions
Lesion success (the attainment of <50% residual stenosis of the target lesion using any percutaneous method) 100% Device success (the attainment of <50% residual stenosis of the target lesion using only the assigned device) 97.0% Procedure success (the attainment of <50% residual stenosis of the target lesion and no in-hospital MACE) 96.3%

10 Time After Initial Procedure (months)
RESOLUTE 38mm Dual Antiplatelet Therapy (DAPT) Usage 92.3 91.4 90.0 89.5 Patients on DAPT (%) Time After Initial Procedure (months)

11 RESOLUTE 38mm Clinical Outcomes at 1 Year Primary endpoint met
Resolute 38mm (n = 222) Events [%] TLF Performance goal Target lesion failure (TLF) is defined as cardiac death, target vessel MI and clinically driven TLR. Primary endpoint of TLF at 12 months was 5.4% with a one-sided 95% upper confidence bound of 8.6%, which was significantly less (P < 0.01) than the performance goal of 19.0%.

12 RESOLUTE 38mm Clinical Outcomes at 1 Year 30 Days % n = 223 6 Months
Death (all) 0.4 0.9 Cardiac MI (target vessel) 3.6 Q Wave Non Q wave 2.7 Cardiac death + target vessel MI 4.0 4.5 ST Def/Prob (all) Early (< 30 days) Late ( days) -- TLR 1.4 TVR 1.3 TLF (cardiac death, TV-MI, TLR) 5.4 TVF (cardiac death, TV-MI, TVR) 4.9 6.8 ST occurred in one patient on day 1 (probable ST) and in a second patient on day 4 (definite ST). TLR and TVR are clinically driven. 12

13 RESOLUTE 38mm – Diabetic Patients
Clinical Outcomes at 1 Year Diabetic, 38mm patients (n = 84) Non-Diabetic, 38mm patients (n = 138) P = 0.79 P = 0.64 P = 0.52 P = 0.49 P = 0.20 Events [%] Cardiac Death ST (ARC Def/Prob) TLF TV-MI TLR All P-values are adjusted by propensity score for differences in baseline characteristics. TLF (Target Lesion Failure) is defined as cardiac death, TVMI, or clinically driven TLR.

14 RESOLUTE 38mm Conclusions At 12 months, the rate of TLF was 5.4%, thereby meeting the primary endpoint for the 38mm Resolute stent. Other clinical events rates were low and notably there were no late stent thrombosis events. Furthermore, there were no statistical significant differences observed between patients with and without diabetes mellitus. The 38mm length R-ZES was safe and effective in this selected population of long lesion patients.