1 Calibration of CO2 cuvette Reusable or disposable for V-family
Before or during the patient ventilation Remove the CO2 sensor (D) from the cuvette (E) and place it to the ambiant air (without the cuvette) 2. Touch the button Sensors/Parameters 3. Touch the tab CO2 sensor Check that the selected cuvette corresponds to cuvette used reusable or disposable and that CO2 monitoring is on 5. Touch the button Start and confirm with the rotary know. « Calibration in progress » is displayed Once the calibration is done and succesful, fit the CO2 sensor back on the cuvette already placed into the breathing circuit The calibration is identical for a reusable and a disposable CO2 cuvette. * This document doesn’t replace the Instructions for use. For more information, please, refer to the Instructions for use


Download ppt "Calibration of CO2 cuvette Reusable or disposable for V-family"