Comparison between pulse waveform analysis and thermodilution cardiac output determination in patients with severe pre-eclampsia  R.A. Dyer, J.L. Piercy,

Similar presentations


Presentation on theme: "Comparison between pulse waveform analysis and thermodilution cardiac output determination in patients with severe pre-eclampsia  R.A. Dyer, J.L. Piercy,"— Presentation transcript:

1 Comparison between pulse waveform analysis and thermodilution cardiac output determination in patients with severe pre-eclampsia  R.A. Dyer, J.L. Piercy, A.R. Reed, G.W. Strathie, C.J. Lombard, J.A. Anthony, M.F. James  British Journal of Anaesthesia  Volume 106, Issue 1, Pages (January 2011) DOI: /bja/aeq292 Copyright © 2011 The Author(s) Terms and Conditions

2 Fig 1 Bland and Altman comparison between TD and LiDCOplus after central venous calibration. Bias: −0.58 litre min−1, 95% confidence interval: −0.77 to −0.39 litre min−1. Dotted lines show limits of agreement. British Journal of Anaesthesia  , 77-81DOI: ( /bja/aeq292) Copyright © 2011 The Author(s) Terms and Conditions

3 Fig 2 Bland and Altman comparison between TD and LiDCOplus after peripheral venous calibration. Bias: −0.16 litre min−1, 95% confidence interval: −0.37 to 0.06 litre min−1. Dotted lines show limits of agreement. British Journal of Anaesthesia  , 77-81DOI: ( /bja/aeq292) Copyright © 2011 The Author(s) Terms and Conditions