ESD Protection 02/12/2018 Radiometer Medical ApS 2 December, 2018.

Similar presentations


Presentation on theme: "ESD Protection 02/12/2018 Radiometer Medical ApS 2 December, 2018."— Presentation transcript:

1 ESD Protection 02/12/2018 Radiometer Medical ApS 2 December, 2018

2 Electro Static Discharge (ESD)
02/12/2018 RADIOMETER PRESENTATION Electro Static Discharge (ESD)

3 Electro Static Charge and Discharge Fundamentals
02/12/2018 RADIOMETER PRESENTATION Electro Static Charge and Discharge Fundamentals Whenever there is movement, static charge is produced Static charge is caused by an excess or deficiency of electrons collected on a surface Electro Static Discharge is the rapid flow of charge between two objects

4 When Is a Static Charge Present?
02/12/2018 RADIOMETER PRESENTATION When Is a Static Charge Present? A person walking on a carpet up to volts A person walking on a vinyl floor volts A person on a padded chair up to volts Styrofoam coffee cup up to 5000 volts Plastic or scotch tape up to 5000 volts Vinyl covered notebook (Service Manual) up to 8000 volts

5 Electro Static Discharge
02/12/2018 RADIOMETER PRESENTATION Electro Static Discharge TO FEEL IT volts or more TO HEAR IT volts or more TO SEE THE SPARK volts or more

6 Component damaging Voltage Levels
02/12/2018 RADIOMETER PRESENTATION Component damaging Voltage Levels Device type Range of ESD Vulnerability (volts) VMOS MOSFET EPROM 100 JFET SAW OP AMP CMOS Diodes The range where a injury can occur. Bipolar Transistors TTL Reference: Radiometer ESD book, code number Radiometer Medical ApS 2 December, 2018

7 Electro Static Discharge
02/12/2018 RADIOMETER PRESENTATION Electro Static Discharge 90 % That are damaged by ESD are Zinged 10 % That are damaged by ESD are Zapped. Zinged: Components’ life span is reduced Operating Characteristic changed (intermittent failures) Temperature sensitivity Zapped: - Component totally destroyed

8 ESD Protected Area Static Dissipative Bench Mat ESD protected floor
02/12/2018 RADIOMETER PRESENTATION ESD Protected Area Static Dissipative Bench Mat ESD protected floor Wrist Strap (1MΩ) Look for the sign:

9 General Requirements Instrument for repair must be grounded
02/12/2018 RADIOMETER PRESENTATION General Requirements Instrument for repair must be grounded Service person must have wrist strap connected Mains supplied test instruments must be grounded Solder iron must be grounded Discharge tools on bench mat before use Clean up Make ESD protected area visible Train the personnel Check facilities regularly

10 02/12/2018 RADIOMETER PRESENTATION Grounding Example 1

11 02/12/2018 RADIOMETER PRESENTATION Grounding Example 2

12 02/12/2018 RADIOMETER PRESENTATION Grounding Example 3

13 Working Instructions Perform grounding (instrument, person, bench mat)
02/12/2018 RADIOMETER PRESENTATION Working Instructions Perform grounding (instrument, person, bench mat) Dismantle instrument Unpack replacement board and mount in instrument Place faulty/suspect board in ESD protected bag Remount instrument case

14 02/12/2018 RADIOMETER PRESENTATION Inventory / Transport ESD sensitive devices will be delivered in protected bags or boxes ESD sensitive devices must be kept in protected bags or boxes outside protected area Unprotected ESD sensitive devices must be rejected

15 02/12/2018 RADIOMETER PRESENTATION Cleaning ESD Vacuum cleaner

16 RADIOMETER ESD stock items
02/12/2018 RADIOMETER PRESENTATION RADIOMETER ESD stock items 3M bench mat type R- part no 3M field kit type R- part no 3M wrist band type R- part no 3M cord for 2204, 1.5 m -R- part no 3M cord for 2204, 3 m -R- part no 3M bag 150 x 250 mm -R- part no 3M bag 200 x 300 mm -R- part no 3M bag 250 x 350 mm -R- part no 3M bag 350 x 450 mm -R- part no

17 02/12/2018 RADIOMETER PRESENTATION Implementation Manufacturing and stocking facilities in Copenhagen, January 15, 1993 Distributors service and stocking facilities, May 15, 1993

18 Questionnaire Have you implemented ESD protection in your company ?
02/12/2018 RADIOMETER PRESENTATION Questionnaire Have you implemented ESD protection in your company ? Yes No Have the person responsible for your stock received training in ESD ? Yes No Have you got ESD protected areas in your workshop ? Yes No Have you got portable ESD protection equipment ? Yes No Do you feel that you now have enough information to implement ESD protection in your company ? Yes No

19 End of Session 02/12/2018 Radiometer Medical ApS 2 December, 2018