1 MIPA status report (Miniature Ion Precipitation Analyzer)
David McCann, Swedish Institute of Space Physics, Kiruna, Sweden Prepared for GSST Workshop, Gothenburg

2 MIPA basic features Three major sub-units: 1. Electrostatic deflector
4. deflector high voltage plates 2. Electrostatic analyzer 5. analyzer high voltage plate 11. UV attenuation structure 3. time-of-flight unit 6. time-of-flight cell 7. start surface 8. stop surface 9. start ccem* 10. stop ccem* * Ceramic Channel Electron Multiplier Parameter Value Energy range 10eV – 15keV Energy resolution ∆E/E 7% Viewing angle 9° x 180° Angular resolution 9° x 22.5° Mass resolution 6 mass groups Energy elevation cycle 8s Geometric factor 0.14 cm2 sr eV

3 Status A TM (Technological Model) of MIPA is fully integrated and set up for tests. HV stability tests has been performed without encountering any problems. A fully functional prototype of the HVPS has been integrated and will be used for calibration tests of the TM. Layout of the time-of-flight electronics is on-going and will be concluded within the next few weeks. A fully functional rough prototype has been integrated for use on the TM. Thermal tests of the CCEM’s have been concluded. Noise levels do not exceed acceptable levels.

4 MIPA TM Left: i. View of TM set-up in vacuum chamber at IRF.
ii. The box under the sensor will be used for storing the hvps and time-of-flight electronics boards. Right: i. View of TM without sensor cover.

5 HVPS View of the HVPS. Dimensions: ~75g 120x60x30 mm

6 TOF electronics Picture of the time-of-flight electronics board (rough prototype) set up in the vacuum chamber at IRF. Dimensions (of flight board): ~80g 120x60x15 mm

7 UV trap Left: View of the base structure with deflector grid and UV rejection unit integrated. Right: Focused view of the UV rejection plates. The unit comprise 14 blackened aluminum plates. Interplate spacing is ~0.2mm with plates roughly 0.1mm thick

8 Status comments MIPA is well on schedule.
We will be able to conclude tests and calibration of the TM during oct/nov 2005. As of now, no critical action items are identified.