The pointing calibration of the WFCTA tracking system

Similar presentations


Presentation on theme: "The pointing calibration of the WFCTA tracking system"— Presentation transcript:

1 The pointing calibration of the WFCTA tracking system
Lingling Ma IHEP

2 Outline Introduction of the WFCTA telescope
The method of the pointing direction calibration Night sky background Method ---- the guide of bright stars The results of the calibration Accuracy The pointing error of the tracking system Conclusions

3 Wild Field Cherenkov Telescope Array.
Field of view 256 PMTs (1°×1°) 14°×16° It takes a star 40~50minites to pass through the FOV Step-tracking system How to estimate the pointing accuracy How to estimate the pointing error of the tracking system Trace of crab

4 Method ----Night sky background
The diffuse night sky background The star light A bright star pass through the FOV of a PMT

5 Method ----The guide of bright stars
TD1 catalog Search the trace of the brightest star on the camera by the PMTs with the highest signal Fit the trace to get the pointing direction of telescope Tune the pointing direction by other bright stars Compare the pointing direction with the one given by the tracking system to check the accuracy and the systematics

6 Results ----accuracy The differences between the positions of the bright stars and the one observed by the telescope in local coordinates are defined as the accuracy of the calibration.

7 Accuracy Azimuth Elevation The accuracy is better than 0.1°

8 The pointing error Delta-degree Delta-degree

9 Conclusions The accuracy of the calibration is better than 0.1°
The pointing error of the tracking system Azimuth: 1.5° Elevation: 1.2° More data is expected