1 Planmeca ProMax tomography

2 Planmeca ProMax tomography
Four tomography modes Wide beam linear tomography (film)‏ Wide beam true linear tomography (film)‏ Digital linear tomography Digital Transtomography™

3 Film tomography Wide beam linear tomography
Cassette is perpendicular to the X-ray beam Wide beam true linear tomography Cassette is kept parallel to the image layer Constant resolution over the whole exposed area

4 Linear vs true linear tomoraphy
pin phantom Linear: The thickness of the layer is not regular True linear: The thickness of the layer is regular

5 Linear vs true linear tomoraphy
The resolution is optimum only in the middle of the image True linear: The resolution is constant all over the image Tomography

6 Digital tomography Digital linear tomography
One sweep - the rotating point moves along the layer during imaging Straight 6mm layer Fast Digital Transtomography™ Multiple sweep - the beam makes oscillating and transitional movement Straight layers, adjustable thickness Better layer separation

7 Tomography features Straight cuttings
Layer thickness selectable (1, 3, 6, 9, 18, 36 mm)‏ Adjustable step between layers Constant 1.50x magnification Pre-programmed cutting sites (based on jaw size)‏ Pre-programmed cutting angles (based on jaw shape)‏ Automatic or manual operations Specialised imaging at the dental practice High value for implantology

8 Tomography programs Cross-sectional tomograms 1-4 cross-sectional images (manual)‏ 4 cross sectional images (automatic)‏ 3 cross sectional images with different angles (automatic)‏ 4 cross-sectional with anatomical angles 3 cross sectional images from sinus and nasal cavity (automatic)‏ The interval between cuttings is selectable by the user (1, 3, 6, 9 or 12 mm)‏ Angle between cuttings in multi- angle tomograms is ±6°, the image no 2 corresponds to the selected target position Anatomical angles are according to the curvature of the jaw, the image no 2 corresponds to the selected target position

9 Tomography programs Longitudinal tomograms 1-4 longitudinal images (manual)‏ 3 longitudinal images (automatic)‏ 3 longitudinal images with different angles (automatic)‏ 3 longitudinal images from the sinus and nasal cavity angle between cuttings in multi- angle tomograms is transtomogram

10 Tomography programs Mixed tomograms 1 cross-sectional and 1 longitudinal images (automatic)‏ 3 cross-sectional images and 1 longitudinal image (automatic)‏ 3 cross-sectional images with different angles and 1 longitudinal image (automatic)‏ Mixed tomograms give 3D- information from the target site Necessary for exact measurements the interval between cuttings is selectable by the user angle between cuttings in multi- angle tomograms is transtomogram

11 Taking tomogram Step #1 Take impression Step #2 Mark the target
Step #3 Impression in machine Step #4 Make program selections Step #5 Set layer to mark Step #6 Patient in machine Step #7 Touch Ready Step #8 Exposure

12 Taking tomogram Step #1 Take Impression

13 Taking tomogram Step #1 Take impression Step #2 Mark the target

14 Taking tomogram Step #3 Impression in machine

15 Taking tomogram Step #4 Make program selections

16 Taking tomogram ProMax's concept positions the patient automatically so the area to be exposured is precisely within the focal layer. The operator simply selects the proper coordinates of the object, and ProMax goes to the right position.

17 Taking tomogram Step #5 Set layer to mark with joystick
Adjust angle with the angle buttons

18 Taking tomogram Step # 6 Patient in machine
Patient bites the impression When the target is upper jaw, adjust the maxillary alveolar ridge horizontal When the target is lower jaw adjust the lower part of the mandible horizontal Fine adjust the target with the positioning lights

19 Taking tomogram Step # 7 touch Ready Step # 8 Press exposure button

20 Taking tomogram 3 x CRS 1xLNG Auto Angle
The diagnostic set must include the longitudinal and the one best cross-sectional images.

21 The End More information: Erkki Hiltunen Product Manager, X-rays
tel: Mark Niemi tel: 4/2011