Presentation is loading. Please wait.

Presentation is loading. Please wait.

Low Cost Fundus Camera Sub-systems Review.

Similar presentations


Presentation on theme: "Low Cost Fundus Camera Sub-systems Review."— Presentation transcript:

1 Low Cost Fundus Camera Sub-systems Review

2 Project Background A fundus camera is used to photograph the retina to provide information about a patient's eye. Diabetic retinopathy is a complication associated with diabetes, eventually leading to blindness. Most people will only see a specialist after damage to the retina has occurred Used exclusively in an Opthamologist's office With proper care, patients with retinopathy can reduce their risk of blindness by 95%. Typically very expensive and heavy [Ask audience about having any corrective lenses] We are project 16590, the Low Cost Fundus Camera and I am Casey Petzel, the EDGE lead. A fundus camera is typically a medical imaging device to used diagnose diabetic retinopathy, a disease that can cause blindness if not caught early enough. This poses a problem because symptoms are typically noticed after the retina is damaged then where a patient would see an Ophthalmologist. These devices are normally stationary, heavy and mydriadic, a chemical dilation of the eye; something that many people like to stray away from. After looking at current commercial products and taking aim at the goals of our customer, we have developed what we would like to correct for our fundus camera.

3 Problem Statement Due to the need for pupil dilation, specialized camera equipment, staff, and facilities, diabetic patients often do not receive regular screenings. By creating a low-cost, portable, and easy to use fundus camera, screening for retinopathy in diabetic patients can be performed by a general practitioner. Our goal is to create a light, portable and affordable Fundus camera to be used by a General Practitioner, or trained individual. It is important that this does not use chemicals to dilate the retina. This camera would screen the patient for retinopathy, and should anything look peculiar, the screener would recommend going to an ophthalmologist for an official diagnosis.

4 Functional Decomposition
Image Eye Save Picture Stabilize Attach Camera Alignment IR Set Up Mount Capture Image Rest Face

5 Concept Generation - Selection Criteria
Low Cost Safe Portable Easy to Use Secure Quality Achievable within our MSD time frame

6 Concept Selection Process - Puegh Table

7 What is the best option for a camera?
Feasibility Analysis Our question: What is the best option for a camera? Considerations: Cost Built-in display for GUI Availability of IR in camera Quality of image Need for/compatible illumination solutions Lens attachments Ability to mount camera solution (fatigue, wear, machining necessary) Ease of acquisition

8 ODROID-XU4

9 ODROID-XU4

10 ODROID-XU4

11 ODROID-XU4 Processor Memory Audio USB3.0 Host USB2.0 Host Display
Samsung Exynos5422 ARM® Cortex™-A15 Quad 2.0GHz/Cortex™-A7 Quad 1.4GHz Memory 2Gbyte LPDDR3 RAM PoP (750Mhz, 12GB/s memory bandwidth, 2x32bit bus) Audio HDMI Digital audio output. Optional SPDIF optical output (USB module) USB3.0 Host SuperSpeed USB standard A type connector x 2 port USB2.0 Host HighSpeed USB standard A type connector x 1 port Display HDMI 1.4a with a Type-A connector Storage (Optional) eMMC module socket : eMMC 5.0 Flash Storage (up to 64GByte) MicroSD Card Slot (up to 64GByte) Power 5V 4A Power Size 82 x 58 x 22 mm approx. (weight: 60gram including cooling fan approx. 38gram without cooler)

12 ODROID-XU4 Benchmarks (CPU/ RAM)
Benchmarks Score Raspberry Pi 2 ODROID-C1 ODROID-U3 ODROID-XU4 Dhrystone-2 1006.6 1262.8 2300.7 3994.1 Double-Precision Whetstone (x3) 1076.1 1318.8 2217.9 3074.7 Nbench Integer (x40) 840.0 1208.0 1648.0 2451.1 Nbench FP (x100) 809.0 1050.0 1860.0 2536.9 SysBench CPU (1/T * 50000) 669.3 902.2 1497.0 2891.8 Memory BandWidth (mbw 100: MiB/s) 557.1 1069.6 736.6 3243.8

13 ODROID-XU4 Benchmarks (CPU/ RAM) Cont.

14 See3cam_CU40 Camera Takes images and video in both IR and visible light. Sensor details: OV4682 RGB IR 4MP USB 3.0 10 pin GPIO header Interchangeable/Removal Lens 5V Vin

15 See3cam_CU40 Camera

16 OV4682 RGB IR 4MP – 2688 x 1520 Contains RGB and IR pixels
Pixel Size: 2µm x 2µm 5440 x 3072 µm 10-bit RAW data output

17 OV4682 RGB IR

18 OV4682 RGB IR

19 HDMI Screen Previous screen’s ribbon ripped Cheap (<$50)
Sufficient resolution

20 HDMI Screen Active drivers Simple solution

21 Resistive Touch Screen
Cheaper Pressure based Allows covers Allows gloves Decreased Contrast Low Sensitivity

22 Illumination Fiber optic lighting? Very expensive

23 Illumination Currently have board containing LED and IRED lighting
Infrared lighting required to align camera with pupil avoiding dilation Axial lighting

24 Illumination FDA regulations
Beam standards developed by the International Electrotechnical Commission Maximum exposure must be below the accepted threshold limits set out in ISO :2007 ISO = International Organization for Standardization

25 Optical System Magnification to focus on the iris for alignment
Additional magnification to capture RGB image of the retina 20- diopter lens to achieve required field of view of retina Correct level of magnification to reach retina Located axially with eye, along with IR/RGB camera and equivalent illumination source

26 The App Android - Lollipop 6.0 or KitKat 5.3?
Programming Language - Java! Software Design Process - Next Phase Software Requirements - Per-Class, Per-Core Function How it will work

27 The App Alignment Image Capture IR Control RGB Control
Alignment Assist Illumination Control GUI File Manipulation Workflow File Control Feedback Settings

28 Mechanical Systems Face Mount Assembly 2D position adjustment system
Vertical position adjustment system Component Housing

29 Face Mount Assembly

30 Face Mount Assembly Similar design but alternative adjustment method
Complexity Ease of use Cost Material? Consider… Durability Potential Problems?

31 Compression Clamp or Screw Assem.

32 2D Position Adjustment Potential Problems?

33 Vertical Position Adjustment

34 Component Housing

35 3 Week Schedule Task Assigned To Due Date Go over design review notes, create new action items if necessary Team 10/22/2015 FDA about specific requirements Kelly 10/23/2015 Discuss purchase orders with Susan 10/27/2015 Design concept mechanical systems in AutoCAD Ryan/Matt 10/30/2015 Prototype optical solutions 11/3/2015 Begin working on software Cecelia 11/10/2015 Begin assembling with new components TBD Update EDGE with all new materials Casey 11/18/2015 Detailed design review presentation completed 12/7/2015


Download ppt "Low Cost Fundus Camera Sub-systems Review."

Similar presentations