Presentation is loading. Please wait.

Presentation is loading. Please wait.

Product specification

Similar presentations


Presentation on theme: "Product specification"— Presentation transcript:

1

2 Product specification
Software specification Main - exclusive features Overview Parameters Reagents & Controls Rack loader & sample principles CDR Measurement principles Customer maintenance Environmental protection

3 Compact bench top system
Overview Compact bench top system No external computer is requested Touch screen Reagent Compartment Processed Tubes Rack Loader 10 racks of 10 tubes Stat Sample Bay

4 WBC, RBC, HGB, HCT, MCV, MCH, MCHC, RDW, PLT, MPV, PCT, PDW
Parameters CBC Mode (12 Parameters) WBC, RBC, HGB, HCT, MCV, MCH, MCHC, RDW, PLT, MPV, PCT, PDW 5 DIFF Mode ( 26 Parameters) LYM, MON, NEU, EOS, BASO (% and #) ALY (Atypical Lymphocytes) (% and #) LIC (Large Immature Cells) (% and #) 80 samples per hour Manual or automatic CBC or DIF (random access)

5 5 reagents Control bloods : ABX Cleaner ABX Biolyse ABX Basolyse II
Reagents & Controls 5 reagents ABX Cleaner ABX Biolyse ABX Basolyse II ABX Eosinofix + ABX Diluent Onboard reagents Bar-coded Waste level sensor Control bloods : Bar-coded Target values uploaded by floppy disk

6 Sampler capacity : 10 racks of 10 samples
Rack loader Sampler capacity : 10 racks of 10 samples XL walk-away autonomy : 1hour 15mn

7 Automatic mode with end over end mixing
Sampling Automatic mode with end over end mixing 30 µl (CBC) 53 µl (CBC+DIFF) Manual (open or close) mode Internal and external barcode reader

8 CDR = Customized Dilution Ratio
CDR - Purpose CDR = Customized Dilution Ratio Purpose : Pushing the limits of linearity without blood manipulation and in a very user friendly way The CDR eliminates manual dilution error and secure the result !

9 1- Manual selection of the dilution ratio
CDR - Operation The end-user can use the CDR in 2 modes : 1- Manual selection of the dilution ratio The end-user can program 3 dilution ratios:

10 2- Automatic selection of dilution ratio
CDR - Operations 2- Automatic selection of dilution ratio CDR are programmed to rerun samples for every “DIL” flag with predefined dilution ratio

11 Measurement principles
RBC/PLT - Impedance HGB - Cyanmethemoglobin or cyanide free WBC - Impedance & optical BASO - Impedance & specific lyse HGB measurement Method Photometry Wavelenght nm Dilution ratio /250 Reaction temperature °C HCT measurement Method Numeric integration WBC & BASO count Method Impedance Ruby diameter µm Depression of count mb Duration of count  6 seconds Dilution ratio /200 Reaction temperature °C

12 Measurement principles
1. Cytochemistry Lyses red blood cells Stains nuclei and cytoplasmic granules of leukocytes Fixes the leukocytes in their original state (48 hours post-draw stability) 2. Double hydrodynamic focusing (DHSS) Flow-cytometer with hydrodynamic focusing before and after the aperture Differentiation Ruby diameter 60µm Diameter of the flow 42µm Injection duration 12 seconds Dilution ratio 1/80 Incubation duration 12 seconds Reaction duration 35°C 3. Impedance method Cell volume (“X” axis) 4. Light absorbance Analysis of the internal structure (“Y” axis”)

13 Measurement principles
The determination of the 5 per differential uses : X -> Cellular volume Y -> Intensity of staining Eosinophils Large Immature Cells Neutrophils Monocytes Lymphocytes Atypical Lymphocytes

14 General flags RBC and PLT curves “l” and “h” (normal limit)
Flags and alarms General flags “l” and “h” (normal limit) “L” and “H” (panic limit) “x” (reject) “!” (suspicion) Pathologic messages RBC and PLT curves

15 5 per differential matrix
Flags and alarms 5 per differential matrix 1- Priority alarms ALY: (Atypical Lymphocytes) Atypical Lymphocytes Plasmocytes, hairy Cells LIC: (Large Immature Cells) Blasts, myelocytes Promyelocytes, Metamyelocytes

16 5 per differential matrix
Flags and alarms 5 per differential matrix 2- Separation alarms Nl: (Neutro-lympho) Hypo-segmented neutrophils Sezary cells, hairy cells Ne: (Neutro-eosino) Hyper-segmented neutrophils Hypo-granular eosinophils Mn: (Mono-neutro) Immature granulocytes, band cells

17 5 per differential matrix
Flags and alarms 5 per differential matrix 3- Zone alarms Rm: (Right-mono) Large monocytes, blasts No: (Noise) Platelet aggregates, NRBCs Lysis resistant RBCs Ln: (Left-neutro) Cellular debris Rn: (Right-neutro) Immature granulocytes Ll, Ll1: (Left-lympho) Platelet aggregates, NRBCs

18 Only one automatic shut-down per day
Customer maintenance Only one automatic shut-down per day Stepper motors replace the old compressor concept -> No noise, no maintenance A double needle replaces the sampling valve concept -> No clogging, no maintenance Liquid valves replace the old pinch valve concept

19 Environmental protection
Liquid waste 32 ml / sample Solid waste 1,83 g / sample

20 Environmental protection
Power Requirements Power supply From 100 V to 240 V (50 HZ to 60 Hz) Power consumption Maximum 230 VA

21 Product specification
Software specification Main - exclusive features System specification Worklist Data processing Validation station

22 Colour LCD touch screen Windows XP Query mode
System Colour LCD touch screen Windows XP Query mode Random access (DIFF or CBC mode) Onboard validation-station Onboard user manuel Real-time management of patient’s report

23 9 logs for : Maintenance Calibration Controls Errors Data Handling
System 9 logs for : Maintenance Calibration Controls Errors Data Handling Patient record Settings Blank cycles Connection

24 Manual entry or host connection
Worklist Manual entry or host connection Easy to set up Real time visualisation Maximum security Demographic data, department... Comment

25

26 Automatic validation Automatic validation based on: Flags, Alarms, Limits, Delta check

27 Building up a patient report (1)
Patient report built by selective choice from: Run Re-run Manual entry Pathology level

28

29 Product specification
Software specification Main - exclusive features

30 Qualification & quantification of ALY and LIC
Main features Qualification & quantification of ALY and LIC Automatic and immediate rerun Real manual open tube and manual close tube Real capillary mode (30 µl for CBC, 53µl for CBC+Diff) Integrated validation station Compact, simple, reliable (no shear valve, no compressor, only one PCB) Automatic shutdown cycle - No customer maintenance Small reagent consumption Low power consumption

31


Download ppt "Product specification"

Similar presentations


Ads by Google