Mechanical Design of the Fabry-Perot Spectrometer for Balloon Flight

Similar presentations


Presentation on theme: "Mechanical Design of the Fabry-Perot Spectrometer for Balloon Flight"— Presentation transcript:

1 Mechanical Design of the Fabry-Perot Spectrometer for Balloon Flight
By: Mohammed Kagalwala Faculty: Lassonde School of Engineering Dept. : ESSE Supervisor: Dr. Jinjun Shan

2 “We do not inherit the Earth from our fathers; we are borrowing it from our children” ~ Lester Brown, environmentalist

3 Contents Background My Role Design Criteria Challenges Final Design
Design to Construction

4 Mission Objectives Outcomes
To obtain accurate, detailed, and collocated measurements of A-band oxygen at km To retrieve simultaneously surface pressure, aerosol information, and surface albedo from the measurements Outcomes The O2 measurement provides information on the surface pressure, which is needed to develop accurate column measurements of gases like CO2 and CH4

5 Mission Briefing Spectrometer will be flying on a high altitude balloon from Kiruna, Sweden at the end of August Fig.1 – Atmospheric division,UCAR1

6 Mission Briefing Fig.2 – Fabry- Perot (F-P) optical spectrometer made by MPB2

7 Mission Briefing Fig.3 – PI Amplifier Fig.4 – PI Amplifier network
Fig. 5- PC104

8 My Role Primary Duties Design and assembly of mechanical structure for optical and electrical system Perform structural simulations in NX Secondary Duties Design electrical system to power electronics and instruments Assist in thermal management and Arduino health system

9 CSA Requirements Table 1 – List of CSA requirements based on gondola structure Criteria Type Criteria Dimensions Optical System: 300mm x 300mm x 750mm Electrical System: 150mm x 200mm x 200mm Gondola Mass <200kg Deadline July

10 Payload Requirements Table 2 – List of payload requirements
Criteria Type Criteria Operational Pressure 1 atm Spectrometer Temperature 25°C - 27°C CCD Cooling Ensure CCD temperature remains cool

11 Pressure Problem Due to capacitors on the F-P and PI Amplifiers, the operability of system is under question in vacuum conditions Solution Build entire system inside a sealed container

12 Pressure Interior Pressure: 1 atm Exterior Pressure: 0.08 atm
Stress Range: 1.42 MPa – 1.88 MPa Interior Pressure: 1 atm Exterior Pressure: 0.08 atm Stress Range: 4.45 MPa – MPa Fig.6 – Structural simulation results

13 CCD Cooling Problem Inside a sealed container, small fans will not be able to create enough circulation according to simulation, thus will not be able to cool CCD properly Fig.7 – CCD fans Solution Replace fans with copper blocks and to use conduction as primary heat transfer mechanism

14 CCD Cooling Fig.8 – CCD copper blocks for conduction

15 Temperature Problem F-P spectrometer operation temperature should be 25-27°C, due calibration environment At 30 km the avg. temperature ranges from -20 °C to -40 °C. According to CSA thermal simulations, inside our optical tube the temperature will range from 0 °C to 16 °C.

16 Temperature Fig.9 – Thermal simulation with copper blocks, CSA

17 Temperature Solution Localized heating and insulation around the spectrometer From thermal simulations we found three 10” heater strips evenly spread around the spectrometer gave the best heating results. Using FR4 to isolate the spectrometer from rest of the system Paint exterior surfaces white to prevent overheating

18 Temperature Fig.11 – Thermal insulation rings Fig.10 – Thermal results
Fig.12 – F-P with thermal insulation rings

19 Weight Problem After implementing our solutions we were overweight
Cut section from structure components to reduce weight

20 Weight Fig.13 – Weight reducing implementations

21 Final Design Electrical Box
Fig.15 – Interior views Fig.14 – Exterior views

22 Final Design Optical System
Fig. 16 – Side-side comparison of real setup to model

23 Final Design Skeleton Structure
Fig. 17 – Component layout on skeletal structure

24 Final Design Optical Tube
Fig. 18 – Tube interior and exterior

25 Constructed Electrical Box
Fig. 19 – Electrical box exterior Fig. 20 – Electrical interior

26 Heating and Health Fig. 21 – F-P heaters
Fig. 22 – Thermal control and health system

27 Constructed Skeleton Structure
Fig.23 – Components mounted to structure

28 Constructed Optical Tube
Fig.25 – Top face Fig.24 – Side face Fig.26 – Nadir face

29 Gondola Fitting Fig.27 – Gondola fitting, CSA

30 Special Thanks To the CSA, Lassonde School of Engineering, and SDCN Lab for this opportunity