1 From: Ocular Tissue Imaging Using Ultrahigh-Resolution, Full-Field Optical Coherence Tomography
Invest. Ophthalmol. Vis. Sci ;45(11): doi: /iovs Figure Legend: Schematic representation of the full-field OCT. The light source is a tungsten halogen lamp in a Köhler illuminator setup. BS, beam splitter; MO, vertically positioned microscope objectives (water-immersion, 10×, 0.3 NA, or 20×, 0.5 NA); M, mirror (silver) for upright reference arm; GP, glass plates for dispersion compensation (antireflection coated); L, aplanatic doublet achromat lens (300-mm focal length); Ref, reference mirror (2% reflectivity); PZT, piezoelectric stage actuator; TS, translation stage; FG1 and FG2, function generators. Date of download: 11/12/2017 The Association for Research in Vision and Ophthalmology Copyright © All rights reserved.