CMS Galaxy varian CP-4900 GC Gas Analysis

Similar presentations


Presentation on theme: "CMS Galaxy varian CP-4900 GC Gas Analysis"— Presentation transcript:

1 CMS Galaxy varian CP-4900 GC Gas Analysis
Status of the setup

2 Introduction The Varian CP-4900 is a gas chromatography system Connected to 46 streams of gas from CMS: CSC, RPC and DT Setup with three types of detectors: OV1 (separate hydrocarbons from C4 to C12) PPU (separate hydrocarbons from C1 to C4) MS (separate permanent and noble gas)

3 System summarized A gas chromatographic system consists of:
• A regulated and purified carrier gas source, which moves the sample through the GC • An inlet, which also acts as a vaporizer for liquid samples • A column, in which the time separation occurs • A detector, which responds to the components as they occur by changing its electrical output • Data interpretation of some sort

4 Schematic view Each valves are connected to 16 stream + 1 output.
46 different stream can be analysed: 11 from CSC 6 from DT 27 from RPC

5 SEtup

6 Gc reparation First problem: Impossibility to modify the position of the valve - Solved by resetting the position using the manual controller and the multiposition actuator control module Second problem: Valve automatisation through software not working - Solved by reconstruction of the system inside the configuration software of the Galaxy GC

7 CHROMATOGRAM

8 Carrier Gas Use of gas bottles for now because of the possibility of contamination of helium supply lines. Helium bottle: 40 bar. Argon bottle: 200 bar.

9 NEXT STEP Check if the helium supply line is contaminated:
Need of high purity heluim to use as carrier gas (Risk of damage of the µGc if not). Open the gas lines to begin the gas analysis of CMS.