Direct Flow Medical Experience with a Conformable, Repositionable Retrievable Percutaneous Aortic Valve Reginald Low MD University of California, Davis.

Similar presentations


Presentation on theme: "Direct Flow Medical Experience with a Conformable, Repositionable Retrievable Percutaneous Aortic Valve Reginald Low MD University of California, Davis."— Presentation transcript:

1 Direct Flow Medical Experience with a Conformable, Repositionable Retrievable Percutaneous Aortic Valve Reginald Low MD University of California, Davis Sacramento, California

2 Disclosures Abbott Vascular - Consultant
Boston Scientific - Consultant Direct Flow Medical - Consultant UC Davis

3 Introduction Feasibility, safety, and efficacy of TAVR with Edwards-Sapien and Medtronic CoreValve have been well demonstrated Large Effective Orifice Area Durability Securement No coronary obstruction Multiple sizes Delivery profile Opportunities for improvement Reduction of paravalvular insufficiency Improved access Profile, flexibilty and trackability Assess function before final deployment Precise positioning Repositioning Retrievable UC Davis

4 Direct Flow Medical Percutaneous Aortic Valve
Design Features Bovine Pericardium Leaflets (anti-calcification treatment) Balloon Frame covered with Polyester Fabric Initial deployment - inflate frame with Saline/Contrast When acceptable final position and function are confirmed, Polymer media is infused to replace saline/contrast 18 French deliviery system Valve sizes - 25 mm, 27 mm treats annulus diameter mm 3 check valves Valve is immediately functional when unsheathed and distal ring is inflated UC Davis

5 Direct Flow Medical Percutaneous Aortic Valve
Design Features Delivery System Polymer nylon sleeve Multi-lumen catheter 3 position and fill wires 1 guide wire lumen Flexible nose cone Non-metallic delivery system and valve Flexible and trackable Precise positioning Retrievable with deflation and withdrawl into retrieval basket Position & Fill Lumens Aortic Ring Ventricular Ring Guidewire Lumen UC Davis

6 Direct Flow Medical Percutaneous Aortic Valve
Valve loaded in Delivery System Valve Unsheathed Valve Inflated & Steering System Valve in Retrieval Basket UC Davis

7 Inflation, Deflation & Media Exchange
UC Davis

8 Positioning & Securement
UC Davis

9 Deliverability & Sealing
UC Davis

10 Complex Valve and Vascular Anatomy
87 year old male Euroscore = 36.2 Moderately Calcified Leaflets &Native Annulus Extensive Calcification Marked Tortuosity 6 Month CT Current Status: 52 Months Post-Procedure Final Position UC Davis

11 Tracking - Aortic Arch Deliverable and precisely positionable in any arch configuration including Horizontal Arch UC Davis

12 Case Review: 009-01-003 (Cali) Access Summary No Complications
UC Davis

13 Balloon Aortic Valvuloplasty
UC Davis

14 Annulus and Procedure Results
Systole Symmetric Calcium Score 1051 Diastole UC Davis

15 Case Summary UC Davis

16 Echo Pre Procedure Post TAVR Procedure UC Davis

17 Important Feature No significant paravalvular leak - aortic insufficiency UC Davis

18 Important Feature No significant paravalvular leak - aortic insufficiency UC Davis

19 Important Feature Important Feature
No significant paravalvular leak - aortic insufficiency UC Davis

20 Important Feature No significant paravalvular leak - aortic insufficiency UC Davis

21 Important Feature No significant paravalvular leak - aortic insufficiency UC Davis

22 22 French Early Experience
University Medical Center, Hamburg-Eppendorf (n=25) Helios Heart Center - Siegburg (n=6) High risk patients for cardiac surgery Patients Enrollment Number / 22 Implants / 2 Surgical Conversions Age ± 4 years Female/Male /15 NYHA III – IV % Logistic EuroSCORE ± 7% Mean transvalvular gradient ± 14 mmHg Aortic valve area ± 0.13 cm2 LVEF ± 15% UC Davis

23 22 French Early Experience 2 Year Data
Effective orifice area cm2 Mean Echo gradient Functional Class I (75%) or II Aortic insufficiency - 73% none & 27% minimal 2 year survival % CT - No change in position, diameter and orifice area over time Echo - Gradient 50.1±11.3 to 19.6±5.7 at 30 days - No change over time AVA - Baseline 0.57±0.15 cm2 to 1.47±0.35 cm2 at 30 days - No change over time UC Davis Bujuklic K et al. Circ Cardiovasc Interv Dec 1;4(6):

24 Present Studies Latest iteration (18 Fr) improved profile, positioning, valve loading into delivery sheath, radial force, valve competence, & simplified media exchange IP South America - Colombia Study IP European CE March Trial France, Germany, Italy UC Davis

25 Summary Direct Flow Medical Percutaneous Aortic Valve
Non-metallic fabric covered balloon supported bovine pericardial valve with excellent trackability and deliverability Precise positioning, repositionable and retrievable Mechanical securement with conformability and sealing to minimize paravalvular leak Formed in-situ with exchange polymer media Design permits functional and hemodynamic assessment before final deployment Feasibility, safety and efficacy demonstrated with 22 F system European study with longest implant of 4.3 years CE Mark and South American (Colombia) trial enrolling No or trace paravalvular leak & no screening failures for peripheral access UC Davis

26 University of California Davis Medical Center
Thank You University of California Davis Medical Center