Presentation is loading. Please wait.

Presentation is loading. Please wait.

Ch 4. Clinical Radiation Generators

Similar presentations


Presentation on theme: "Ch 4. Clinical Radiation Generators"— Presentation transcript:

1 Ch 4. Clinical Radiation Generators

2 Kilovoltage Units Up to above 1950
X-rays generated at voltages up to 300 kVps Still some use in the present era, esp. treatment of superficial skin lesions Kilovoltage Therapys Grenz-Ray Therapy Contact Therapy Superficial Therapy Orthovoltage Therapy or Deep Therapy Supervoltage Therapy

3 Grenz-Ray Therapy Contact Therapy Energy : < 20 kV
Very low depth of penetration No longer used in R/T Contact Therapy Energy: 40 – 50 kV Short SSD (< 2 cm) Produces a very rapidly decreasing depth dose Max. irradiated tissue : skin surface Application: Tumor not deeper than 1 – 2 mm

4 Superficial Therapy Energy: 50 – 150 kV HVLs: 1.0- – 8.0-mm Al
Applicator or cone attached to the diaphragm SSD: 15 – 20 cm Tube current: 5 – 8 mA Application: tumors confined to about 5-mm depth

5 Orthovoltage Therapy or Deep Therapy
Energy: 200 – 300 kV Tube current: 10 – 20 mA HVLs: 1 – 4 mm Cu Cones or movable diaphragm (continuous adjustable field size) SSD: 50 cm Application: tumor located < 2 –3 cm in depth Limitations of the treatment: skin dose Depth dose distribution Increase absorbed dose in bone Increase scattering

6 Supervoltage Therapy Energy: 500 – 1000 kV Technical problems
Insulating the high-voltage transformer Conventional transformer systems were not suitable for producing potential > 300 kVp The problem solved by invention of resonant transformer

7 Resonant transformer units
Used to generate x-rays from 300 to 2000 kV At resonant frequency Oscillating potential attains very high amplitude Peak voltage across the x-ray tube becomes very large

8 Megavoltage Therapy X-ray beams of energy > 1 MV
Accelerators or γray produced by radionuclides Examples of clinical megavoltage machines Van de Graaff generator Linear accelerator Betatron Microtron Teletherapy γray units (e.g. cobalt-60)

9 Van de Graaff Generator
Electrostatic accelerator Energy of x-rays: 2 MV (typical), up to 10 MV Limitations: size high-voltage insulation No longer produced commercially Technically better machine (e.g. Co-60 units & linear accelerators)

10 Linear Accelerator Use high frequency electromagnetic waves to accelerate charged particles (e.g. electrons) to high energies through a linear tube High-energy electron beam – treating superficial tumors X-rays – treating deep-seated tumors

11 Linear Accelerator Types of EM wave 1. Traveling EM wave
Required a terminating (“dummy”) load to absorb the residual power at the end of the structure Prevent backward reflection wave 2. Standing EM wave Combination of forward and reverse traveling waves More efficiency Axial beam transport cavities and the side cavities can be independently optimized More expensive Requires installation of a circulator (or insulator) between the power source the structure prevent reflections from reaching the power source

12 行波式與駐波式加速管的比較

13 Fig 4.5. A block diagram of typical medical linear accelerator

14 The Magnetron A device that produces microwaves
Functions as a high-power oscillator Generating microwave pulses of several microseconds with repetition rate of several hundred pulses per second Frequency of microwave within each pulse is about 3000 MHz Peak power output: 2 MW (for low-energy linacs, 6MV or less) 5 MW (for higher-energy linacs, mostly use klystrons)

15 The cathode is heated by an inner filament
Electrons are generated by thermionic emission Pulse E-field between cathode & anode Electron accelerated toward the anode Static B-field perpendicular to the plane of cavities Electrons move in complex spirals toward the resonant cavities Radiating energy in form of microwave

16 The Klystron Not a generator of microwaves Microwave amplifier
Needs to be driven by a low-power microwave oscillator

17 The Klystron Passed in the drift tube (field-free space)
Electrons produced by the cathode Electrons are accelerated by –ve pulse into buncher cavity Electrons arrive catcher cavity Generate a retarding E-field Electrons suffer deceleration KE of electrons converted into high-power microwaves Lower level microwave set up an alternating E field across the buncher cavity Velocity of e- is altered by the action of E-field (velocity modulation) Some e- are speed up Other are slowed down

18 磁控管與調速管的比較

19 The Linac X-Ray Beam Production of x-rays
Electrons are incident on a target of a high-Z material (e.g. tungsten) Target – need water cooled & thick enough to absorb most of the incident electrons Bremsstrahlung interactions Electrons energy is converted into a spectrum of x-rays energies Max. energy of x-rays = energy of incident energy of electrons Average photon energy = 1/3 of max. energy of x-rays Designation of energy of electron beam and x-rays Electron beam - MeV (million electron volts, monoenergetic) X-ray beam – MV (megavolts, voltage across an x-ray tube, hetergeneous in energy)

20 Lead or tungsten Opening from 0 x 0 to 40 x 40 cm at SSD 100 cm

21 Narrow pencil about 3 mm in diameter
Uniform electron fluence across the treatment field e.g. lead Electron scatter readily in air Beam collimator must be achieved close to the skin surface

22 Flattening(X-ray):intensity uniform
Scattering(electron):add scattering

23 Ion chamber monitor 1.dose rate 2.integrated dose 3.filed symmetry

24 Light field and Radiation field

25 Betatron Energy of x-rays: 6 – 40 MV Disadvantages: low dose rate
Small field size

26 Microtron Electron accelerator which combines the principles of both linear accelerator and the cyclotron Advantage: 1.Easy energy selection, small beam energy spread and small size. 2.Single Microtron to supply a beam to several treatment room.

27 Cyclotron Charged particle accelerator
Mainly used for nuclear physics research As a source of high-energy protons for proton beam therapy Have been adopted for generating neutron beams recently

28 Cyclotron Structures Short metallic cylinder divided into two section (Ds) Highly evacuated Placed between the poles of a direct current magnet Alternating potential is applied between two Ds

29 Cyclotron Positive charged particles (e.g. protons or deuterons) are injected at the center of the two Ds Under B-field, the particles travel in a circular orbit Accelerated by E-field while passing from one D to the other Received an increment of energy Radius of its orbit increases

30 Machines Using Radionuclides
Radionuclides have been used as source of γrays for teletherapy Radium-226, Cesium-137, Cobalt-60 60Co has proved to be most suitable for external beam R/T Higher possible specific activity (Ci/g) Greater radiation output per curie Higher average photon energy Radionuclide Half-Life (Years) γRay Energy MeV I- Value ( Rm2_) Ci – h Specific Activity Achieved in Practice (Ci/g) Radium-226 (filtered by 0.5 mm Pt) 1622 0.83 (avg.) 0.825 ~ 0.98 Cesium-137 30.0 0.66 0.326 ~ 50 Cobalt-60 5.26 1.17, 1.33 1.30 ~ 200

31 Cobalt-60 Unit Source Treatment beam Heterogeneity of the beam
From 59Co(n, γ) nuclear reactor Stable 59Co → radioactive 60Co In form of solid cylinder, discs, or pallets Treatment beam 60Co →60Ni + 0β(0.32 MeV) + γ(1.17 & 1.33 MeV) Heterogeneity of the beam Secondary interactions βabsorbed by capsule → bremsstrahlung x-rays (0.1MeV) scattering from the surrounding capsule, the source housing and the collimation system (eletron contamination)

32 鈷六十治療機

33 鈷六十治療機

34 直線加速器與鈷六十的比較

35 直線加速器與鈷六十的比較

36 直線加速器與鈷六十的比較

37 Penumbra The region, at the edge of a radiation beam, over which the dose rate changes rapidly as function of distance from the beam axis 1. Transmission penumbra 2. Geometric penumbra

38 Transmission penumbra
Source Transmission penumbra The region irradiated by photons which are transmitted through the edge of the collimator block The inner surface of the blocks is made parallel to the central axis of the beam The extent of this penumbra will be more pronounced for larger collimator opening Minimizing the effect The inner surface of the blocks remains always parallel to the edge of the beam Collimator SDD SSD

39 Radiation source: not a point source
Geometric penumbra Radiation source: not a point source e.g. 60 Co teletherapy → cylinder of diameter ranging from 1.0 to 2.0 cm From considering similar triangles ABC and DEC DE = CE = CD = MN = OF + FN – OM AB CA CB OM OM AB = s (source diameter) OF = SSD DE = Pd ( penumbra) Pd = s (SSD + d – SDD) SDD Parameters determine the width of penumbra

40 Geometric penumbra (con’t)
Solutions Extendable penumbra trimmer Heavy metal bars to attenuate the beam in the penumbra region Secondary blocks Placed closed to the patient for redefining the field Should not be placed < 15 – 20 cm, excessive electron contaminants produced by the block carrying tray Definition of physical penumbra in dosimetry Lateral distance between two specified isodose curves at a specified depth At a depth in the patient, dose variation at the field border Geometric, transmission penumbras + scattered radiation produced in the patient

41 THANKS


Download ppt "Ch 4. Clinical Radiation Generators"

Similar presentations


Ads by Google