An Unexpected Hysteresis Effect Seen in Hawaii-2RG Detectors

Similar presentations


Presentation on theme: "An Unexpected Hysteresis Effect Seen in Hawaii-2RG Detectors"— Presentation transcript:

1 An Unexpected Hysteresis Effect Seen in Hawaii-2RG Detectors
Background Distortion calibration important for JWST Coronagraphy, MSA, slits Self-calibration will be difficult ACS field observed in 2006 In LMC, in JWST’s CVZ 55: big enough for all detectors Good density of stars Many well-measured stars More than 100,000 good to 1 mas Good, but… Not big enough for cross-calibration No IR color information 17" Jay Anderson --- TIPS Feb 21, 2008 J1

2 Ground-based Observations
HAWK-I: new IR camera Being commissioned on VLT at Paranal 7.57.5 FOV 22 mosaic of 20482048 NIR detectors “Active optics” Luigi Bedin (STScI) observed the JWST field For them: exquisite distortion solution For us: IR photometry & larger field We got an additional benefit… Jay Anderson --- TIPS Feb 21, 2008 J2

3 HAWK-I distortion: After global correction: Raw residuals One chip J3
x x One chip x x x x x x Jay Anderson --- TIPS Feb 21, 2008 J3

4 Periodic residuals Same throughout detector Periodogram Square wave
Phase Diagram Phase Diagram Same throughout detector Periodogram Square wave removes periodic signature Jay Anderson --- TIPS Feb 21, 2008 J4

5 Why do we care? Rockwell Scientific HAWAII-2RG Very important to
Prototype for JWST! Very important to understand the source Geometric, like HST? WFPC2’s 34-row effect ACS/WFC’s 68-column effect Check flat fields No evidence WFC FLAT FIELDS HAWK-I Jay Anderson --- TIPS Feb 21, 2008 J5

6 Why do we care? Rockwell Scientific HAWAII 2RG
32 amplifiers  AMP#0 AMP#1 AMP#2 AMP#3 AMP#4 Rockwell Scientific HAWAII 2RG Prototype for JWST! Very important to understand the source Geometric? No! Readout artifact? HAWK-I Readout CTE-like hysteresis PSF asymmetry? Look at CRs… Likely cause pixel dwell time All CRs Even Amps Odd Amps inter-pixel capacitance Jay Anderson --- TIPS Feb 21, 2008 J6

7 Possible Implications for JWST
JWST will have four amplifiers four 512-pixel zones Not clear if effect comes before or after the amplifiers… Really a problem? How to deal with hysteresis? Mild deconvolution? (will amplify noise) Leave in, but model Similar to the inter-pixel capacitance Effective PSF will treat it implicitly as (x,y) Need to study impact on: subarray mode, guider, spectra, dithering, noise model Early NIRCam/U.AZ tests encouraging! Jay Anderson --- TIPS Feb 21, 2008 J7