Presentation is loading. Please wait.

Presentation is loading. Please wait.

PMT test setup in Erlangen

Similar presentations


Presentation on theme: "PMT test setup in Erlangen"— Presentation transcript:

1 PMT test setup in Erlangen
Oleg Kalekin University of Erlangen Meeting Pylos,

2 PMT Test Setup Charge response
Pulser (custom made): Rate 10 kHz FWHM 10ns Variable amplitude Dark Box: Copper shielding Various diameter PMT holders ADC LeCroy 2249A No. of channels 12 No. of Bits 10 Full scale 256 pC Oleg Kalekin University Erlangen KM3NeT Meeting, Pylos

3 PMT Test Setup Charge response
Pulser (custom made): Rate 10 kHz FWHM 10ns Variable amplitude Dark Box: Copper shielding Various diameter PMT holders Diaphragms ADC LeCroy 2249A No. of channels 12 No. of Bits 10 Full scale 256 pC Oleg Kalekin University Erlangen KM3NeT Meeting, Pylos

4 PMT Test Setup Charge response
Pulser (custom made): Rate 10 kHz FWHM 10ns Variable amplitude Dark Box: Copper shielding Various diameter PMT holders Diaphragms ADC LeCroy 2249A No. of channels 12 No. of Bits 10 Full scale 256 pC Oleg Kalekin University Erlangen KM3NeT Meeting, Pylos

5 PMT Test Setup Charge response
Reconstruction of single-/multiple- photoelectron response Model(*): 1st dynode: if  mean number of ph.e., probability to detect n ph.e. = P(n, ) - Poisson function Anode: single ph.e. charge = G(Q1,1) – Gaussian n ph.e. charge = G(Qn,n), Qn=n•Q1 , 2n= n•2n PMT charge response: P(i,)•C(i)•exp(-(x-Qi)2/(22i)), i=1,2,… Qi=i•Q1, 2i= i•21, C()=1/ 2 0 ph.e. (pedestal): P(0, )=exp(- ), if N – total number of event, number of pedestal events N0=N•exp(- ), and then =-ln(N0/N) can be used as first estimation of mean number of ph.e. Fit method: 2-minimizing with , Q1 , 1 as fit parameters Software: ROOT Version 5.08/00 (*) E.H. Bellamy, et al., NIM A 339 (1994) Oleg Kalekin University Erlangen KM3NeT Meeting, Pylos

6 PMT Test Setup Charge response
Reconstruction of single-/ multiple- photoelectron response: simulations Oleg Kalekin University Erlangen KM3NeT Meeting, Pylos

7 PMT Test Setup Charge response
Reconstruction of single-/ multiple- photoelectron response Oleg Kalekin University Erlangen KM3NeT Meeting, Pylos

8 PMT Test Setup Absolute and Spectral Calibration
Monochromator LOT-Oriel MS301 fully automated nm Tungsten lamp PMT bases for QE measurements Oleg Kalekin University Erlangen KM3NeT Meeting, Pylos

9 PMT Test Setup Absolute and Spectral Calibration
Monochromator LOT-Oriel MS301 fully automated nm Tungsten lamp PMT bases for QE measurements Oleg Kalekin University Erlangen KM3NeT Meeting, Pylos

10 PMT Test Setup Absolute and Spectral Calibration
Monochromator LOT-Oriel MS301 fully automated nm Tungsten lamp PMT bases for QE measurements Oleg Kalekin University Erlangen KM3NeT Meeting, Pylos

11 Photonis PMTs 10" PMT ordered, expected to be delivered end of April
3" PMT will be provided by NIKHEF Diam/Area inch/cm2 9/380 5/120 10/600 8/400 12/670 Euro/cm2 3.72 7.48 2.72 3.54 3.26 3.20 Oleg Kalekin University Erlangen KM3NeT Meeting, Pylos

12 Summary PMT test setup in Erlangen works
Further development is in progress to cover all important PMT characteristics Tests of Photonis PMTs will start soon Hamamatsu PMTs will be evaluated to order Methods, codes, results will be posted on the KM3NeT Portal Oleg Kalekin University Erlangen KM3NeT Meeting, Pylos


Download ppt "PMT test setup in Erlangen"

Similar presentations


Ads by Google