MSE Radial Resolution with ‘Dirty’ Lens

Similar presentations


Presentation on theme: "MSE Radial Resolution with ‘Dirty’ Lens"— Presentation transcript:

1 MSE Radial Resolution with ‘Dirty’ Lens

2 MSE Radial Resolution with ‘Clean’ Lens
FWHM about a factor 2-3 smaller. ‘Tails’ are reduced. Tried various fitting functions … settled on a spline

3 Actual magnetic Pitch-angle Degrees geometry Angle that MSE sees radius

4 The Poor Spatial Resolution of the Dirty Lens Changes the
Measured Field-Line Pitch Angle by up to ~0.5 degrees.

5 There is Little Difference in Inferred Pitch-Angle between a Local
And a Radially-averaged MSE Measurement with a Clean Lens

6 Measured MSE Light is the Convolution of the Diagnostic Response (spatial resolution) and MSE Emission Intensity Profile At edge: emission intensity weighted to inside of viewing region. At center: emission intensity weighted to outside of viewing region. Beam-into-gas: emission intensity weighted to outside of viewing region. MSE Edge Channel Density profile MSE Intensity Profile radius

7 Edge Density Gradients in L- and H-modes
L-mode H-mode

8

9

10