Matrix-Assisted Laser Desorption Ionization Time-of-Flight (MALDI-TOF) Mass spectrometry for protein identification 2-Dimensional Gel Electrophoresis MALDI-TOF.

Similar presentations


Presentation on theme: "Matrix-Assisted Laser Desorption Ionization Time-of-Flight (MALDI-TOF) Mass spectrometry for protein identification 2-Dimensional Gel Electrophoresis MALDI-TOF."— Presentation transcript:

1 Matrix-Assisted Laser Desorption Ionization Time-of-Flight (MALDI-TOF) Mass spectrometry for protein identification 2-Dimensional Gel Electrophoresis MALDI-TOF Mass Spectrometry Speaker: Dr. J. S. Yu Date: 3/21/2002

2 The age of X-omics and biotechnology: Genomics: Human genome project
Transcriptomics: cDNA microarray Proteomics: Development and involvement of mass spectrometry Celera Genomics Inc. cDNA microarray Tandem mass spectrometer (MS/MS) MALDI-TOF MS

3 Proteomics solution IEF SDS-PAGE

4 2-Dimension Electrophoresis (2-DE) for Protein Separation
The core technology of proteomics is 2-DE: At present, there is no other technique which is capable of resolving thousands of proteins in one separation procedure. Speaker: C. C. Wu Date: 31/10/2001

5 Isoelectric point (pI):
Isoelectric point is the pH of a solution at which the net charge of protein is zero. In electrophoresis there is no motion of the particles in an electric field at the isoelectric point. NH3+ COOH pH < pI Positive charge NH3+ COO- pH = pI NH2 COO- pH > pI Negative charge Net charge pH Isoelectric point

6 General principle and protocol of 2-Dimension Electrophoresis
Ampholytes sample Isoelectric focusing (1st dimension) pH pH polyacrylamide 2nd dimension SDS-PAGE MW pH gradient

7 Traditional Equipment for Isoelectric focusing (IEF):
Ampholytes polyacrylamide Cathode (-) electrode solution Anode (+) electrode solution

8 Traditional 2-Dimensional Electrophoresis
Disadvantage: cathodic drift Cathode (-) electrode solution pH pH pH 5 polyacrylamide Ampholyte pH pH pH 3 Time Anode (+) electrode solution

9 Immobilized pH Gradient (IPG)
Acrylamide monomer Acidic buffering group: COO- CH2 = CH-C-NH-R O Basic buffering group: NH3+ Polyacrylamide gel

10 Production of Immobilized pH Gradient (IPG) strip
Gradient maker acidic basic A D plastic support film B E C F pH 3 pH 10

11 Equipment for Isoelectric focusing (IEF):
IPGphor (IEF System) Amersham Pharmacia Biotech Inc. Protein IEF Cell Bio-Rad Laboratories

12 Sample preparation Lysis solution: 8M Urea 4% NP-40 or CHAPS
40mM Tris base Cell line Lysis solution Sonication vacuum Lysis solution Centrifugation Measurement of [protein] 2-DE sample

13 IPG strip rehydration and sample loading
Rehydration solution 2-DE sample Rehydration solution: 8M Urea 2% NP-40 or CHAPS 2% IPG buffer (Ampholyte) 0.28% DTT Trace Bromophenol blue IPG strip holder Position the IPG strip

14 IPG strip rehydration and sample loading
Strip holder Anode (+) electrode Cathode (-) electrode 30 voltage 12hr

15 First dimension: Isoelectric focusing
1. Place electrode pads (?) V step-n-hold hr V step-n-hold hr V gradient vhr V gradient (?) vhr Holder cover IPG strip Electrode pads Time Voltage

16 Second dimension: SDS-PAGE
SDS equilibration SDS-PAGE SDS SDS equilibration buffer 50 mM Tris-HCl 6 M Urea 30% Glycerol 2% SDS Trace Bromophenol IPG strip SDS-PAGE 0.5% agarose in running buffer Marker in paper

17 Protocol of silver stain:
50% methanol 25% acetic acid 4hr ddH2O 30 sec ddH2O x 3 times 30min/time 3% Na2CO3 0.0185% formaldehyde 0.004% DTT solution 30min 2.3M citric acid 0.1% AgNO3 30min 5% acetic acid 25% methanol

18 2-DE separation of soluble E. coli proteins