Download presentation
Presentation is loading. Please wait.
1
Diagnostic Equipment Quality Control
George David, MS, FAAPM, FACR Associate Professor of Radiology Augusta University Augusta, Georgia
2
Why do Quality Control? Legal Requirements
Manufacturers regulated by federal law Use of imaging equipment Regulated by states --- Except --- Mammography Regulated by federal law
3
Why do Quality Control? Accreditation Because we make a difference
JCAHO ACR Because we make a difference Equipment performance Image quality Patient & staff dose
4
Why is Q.C. Important? Without a QC program the only way to identify problems is on patient images. And some problems aren’t visible on images. See? Oversized collimation Insufficient beam filtration
5
Our Goal Best Possible Image Oh yeah, dose
6
Our Goal Provide DIAGNOSTIC IMAGE at lowest dose
If both images are diagnostic, did zee patient receive benefit from additional 1.5 mSv? Diagnostic but more noisy Image #1 (2 mSv ESE) Diagnostic but less noisy (prettier) Image #2 (3.5 mSv ESE) Dose
7
Q.C. Goals Obtain diagnostic images while minimizing dose to
patients staff Monitor image quality Ensure predictable equipment operation Establish baselines data on properly operating equipment assist in problem diagnosis by identifying performance changes
8
A QC program can detect Malfunctions (some) Unpredictability
may be hard to isolate clinically Inefficient use of Radiation high fluoroscopic outputs Radiation not reaching receptor inadequate filtration oversized collimation
9
X-Ray Quality Control Filtration Focal Spot Size / System Resolution
Collimation Radiographic Tube Leakage Fluoroscopy Scatter Outputs Calibration Verification Phototiming
10
Filtration: Equivalent Absorber Present in Beam
A filtration of 4 mm Al means the beam spectrum is the same as if 4 mm Al were in the beam
11
Filtration Locations inherent filtration x-ray tube and housing
12
Filtration not in Tube Metal sheets (usually Al) between tube & collimator or in collimator Mirror May be removable If removable, must pass minimum HVL with plate removed Added Filter Filter
13
Why is Filtration Important?
Tube emits spectrum of x-ray energies Filtration preferentially attenuates low energy photons If not removed, low energy photons expose patients do not contribute to image low penetration
14
Filtration: What do we measure?
Half Value Layer (HVL) absorber (aluminum) thickness that reduces beam intensity by 50%
15
Check at single kVp value
kVp Minimum HVL (mm Al) Half Value Layer HVL depends upon Filtration Waveform kVp (1 or 3 phase, medium frequency) Minimum HVL regulated by law Check at single kVp value Georgia State Rules & Regulations for X-Ray
16
Checking HVL Compliance (Radiography)
Fancy 1-shot meters can measure HVL
17
Checking HVL Compliance (Radiography)
Find aluminum thickness that reduces beam intensity 50%
18
Checking HVL Compliance (Radiographic)
What Al thickness reduces beam intensity 50%? 90 kVp, 20 mAs 2.5 mm Al HVL minimum filter mR (mm Al) Room #2 HVL > 2.5 mm Room #1 filter mR (mm Al) Room #3 HVL < 2.5 mm filter mR (mm Al) HVL = 2.5 mm Marginal HVL exactly 2.5 mmAL Must add Al to reduce beam to 50% (125) Must remove Al to reduce beam to 50% (110)
19
Added Filtration Some collimators have selectable filtration
Check for all settings Minimum HVL should be at minimum filtration setting All settings should comply with state regs Filter Selection Wheel
20
Fluoroscopic HVL Setup / ABS on (can’t turn it off)
Tabletop Image Receptor Filter Absorber to protect image receptor
21
Fluoroscopic HVL Setup (ABS on)
Place all filters above chamber Obtain desired kVp by manipulating Absorber SID mag mode Measure exposure rate move filters from top to bottom technique remains constant total absorber in beam remains constant
22
Apparent Focal Spot Size (system resolution)
Trade-off smaller spot reduces geometric unsharpness larger spot provides improved heat ratings Actual Focal Spot Apparent Focal Spot
23
Bar Phantom Setup Focal spot 18”
24
Bar Phantom Setup Evaluate digital image on workstation.
One image for each focal spot
25
Collimation Radiography Fluoroscopy
26
Radiographic Collimation
X-Ray / Light Field Alignment Bucky Alignment X-ray field should be centered on bucky Beam Central Axis should be in center of x-ray field Collimator field size indicators PBL (automatic collimation) field limited to size of receptor
27
Radiography X-Ray/Light Field Alignment
Alignment requires correct: collimator position on tube mirror tilt
28
Central Beam Alignment Check
Do 2 holes in centering tool line up? Central beam misalignment can cause distortion Centering Tool (small hole in bottom & top)
29
Bucky Alignment (use one image for both x-ray/light & bucky alignment)
Center tube to bucky using longitudinal bucky light transverse & SID detents Collimate smaller than receptor
30
Field Size Indicator Accuracy
Use correct SID scale
31
Positive Beam Limitation aka PBL & Automatic Collimation
Goal Constrain x-ray field to size of receptor Unit senses receptor size Limits beam size to receptor Receptor Receptor
32
Positive Beam Limitation aka PBL & Automatic Collimation
Receptor in bucky Check Multiple cassette sizes Cassette orientation Fixed digital receptor Beam no larger than receptor
33
Fluoroscopic Collimation
X-ray field should not exceed imaged field by more than 3% of SID (sum 1 direction) 4% of SID (sum both directions) Check all magnification modes Bad Good Image Receptor Image Receptor
34
Why is Fluoroscopic Collimation Important?
Additional scatter Degrades image Substantially increases radiation to operator Bad Good Image Receptor Image Receptor
35
Fluoroscopic Collimation
36
Fluoroscopic Collimation
SID ~ 20” 3% SID = 0.6” 4% SID = 0.8” Field Edges Compare with numbers seen on monitor
37
Radiographic Tube Leakage
Regulatory limit 100 1 meter at maximum continuous kVp & mA
38
Fluoroscopic Scatter Use "patient equivalent" phantom
Water jug Typical clinical protocol Measure at appropriate points around room Post diagram Image Receptor Phantom Tabletop
39
Fluoroscopic Scatter
40
Fluoroscopic Scatter
41
Maximum Fluoro Output 10 R/min limit for ABS fluoro center chamber
lead between chamber & image receptor Drives output to maximum measurement point conventional fluoro Table top C-arms / specials 30 cm from image receptor Image Receptor Lead Lead R
42
Image Receptor Entrance Exposure Rate
Check all mag modes
43
Image Receptor Entrance Exposure Rate
Entire probe must be in x-ray field to get valid reading
44
Fluoro Outputs Verification of proper auto brightness operation
Baseline techniques / outputs Check at several absorbor thicknesses Check various mag modes Pulse rates Fluoro “flavors” Low dose High dose
45
Calibration Performance Parameters
Timer Accuracy Repeatability Linearity/Reciprocity Kilovoltage accuracy
46
Performance Parameters
Repeatability within 5% Linearity constant mR/mAs for all mA Set kV Set distance
47
Timer Operation Accuracy Deadman function with auto reset
exposure ends immediately when switch released full exposure must follow prematurely aborted exposure
48
Kilovoltage Accuracy kV Measurement +/- 2 kV constancy desirable
Waveform represented by single number Non-invasive & invasive kV may not match +/- 2 kV constancy desirable 87
49
Phototiming check with output or exposure index
Reproducibility Density Control ESE Field Placement Field Balance Phototiming Operation should be Predictable
50
Phototiming Density Steps should be predictable & approximately even
Phototimer Density Control Settings Phototiming Density Steps should be predictable & approximately even
51
Optical Density or Exposure Index
b l e t o p Typical Phototimer Doses Optical Density or Exposure Index Phantom kVp Time (ms) Meas. FS-probe FS-"Skin" ESE mR (in) O.D. Chest 117 7 14.22 56 63 11 1.56 Abd 77 16.5 141.1 26 31 99 1.14 Spine 81 21.5 184.6 130 1.17 Skull 13.4 112.7 32 74 1.2 Ext. 60 5.2 19.8 36 10 1.15
52
Phototimer Field Placement / Balance
Fields should be approximately the same Some manufacturers purposely set center field differently than outer fields
53
Phototiming check with exposure index
a b l e t o p M s u r m n f P h i k V / R Phototiming check with exposure index kV response Thickness / Rate Response Vary phantom thickness Receptor
Similar presentations
© 2024 SlidePlayer.com Inc.
All rights reserved.