GERB Ground Calibration Overview and Status J. Rufus

Similar presentations


Presentation on theme: "GERB Ground Calibration Overview and Status J. Rufus"— Presentation transcript:

1 GERB Ground Calibration Overview and Status J. Rufus
CERES-GIST 21 Meeting Boulder, March 29th – April 2rd 2004 © Imperial College London

2 Vacuum Calibration Chamber
GERB Ground Calibration Overview and Status Vacuum Calibration Chamber Cryogenically pumped high vacuum chamber capable of 10-9 Torr. Earth Orbital Calibration Facility Cleanroom (~60m3) maintained to better than class G (equivalent - class ISO6, 1000) and controlled to 2% relative humidity and 1C. © Imperial College London

3 GERB Ground Calibration Overview and Status
© Imperial College London

4 GERB Ground Calibration Overview and Status
© Imperial College London

5 How is the instrument modeled?
GERB Ground Calibration Overview and Status How is the instrument modeled? Designed to be linear with respect to incoming power. where and This is tested during calibration. © Imperial College London

6 What are the instrument gains?
GERB Ground Calibration Overview and Status What are the instrument gains? We need to measure gains in both channels over all pixels, And since LF(off) ~ 0, © Imperial College London

7 How is the instrument used?
GERB Ground Calibration Overview and Status How is the instrument used? Differences measured to avoid external offset and increase instrument flexibility. A full characterization of the IBB source is important for GERB’s accuracy. Where is GERB ‘looking’? Need and accurate picture of the influence of GERBs optics on the field of view. Crucial to Geolocation of observations. © Imperial College London

8 Summary of Calibration Aims Check Instrument Linearity.
GERB Ground Calibration Overview and Status Summary of Calibration Aims Check Instrument Linearity. Produce Gains for TOT and SW channels. Also measure T, filter response and use these three measurements to determine B, Gain Ratio = A, Long Wave Correction Factor = Characterize IBB Source for all pixels. Measure Instrument FOV Response (PSF, cross talk.) Perform Unit Level Spectral Response Spot Checks. Linearity Gsw GTOT IBB FOV R s (%) 0.05 0.53 0.38 0.2 0.1 PP © Imperial College London

9 GERB Ground Calibration Overview and Status
Linearity Assumption of linear response tested for all GERBs. Results shown for 2 & 3. For total channel WBB-IBB output measured for a range of WBB temperatures, 280 – 345K © Imperial College London

10 GERB Ground Calibration Overview and Status
© Imperial College London

11 GERB Ground Calibration Overview and Status
© Imperial College London

12 GERB Ground Calibration Overview and Status
Measurement of Gains Ground calibration sources have limited spatial extent, 6o E-W across the EV (24o) ~ 60 pixels N-S Measurement at 4 angles and 5+ heights. GTOT Measured using the WBB and CBB with QF out GSW Measured using the VISCs source with QF in. © Imperial College London

13 Gerb-3 Total Gains GERB Ground Calibration Overview and Status
© Imperial College London

14 Gerb-3 Short Wave Gains GERB Ground Calibration Overview and Status
© Imperial College London

15 IBB Characterization - Method
GERB Ground Calibration Overview and Status IBB Characterization - Method © Imperial College London

16 Produce IBB filtered radiance for all pixels at TIBB.
GERB Ground Calibration Overview and Status Produce IBB filtered radiance for all pixels at TIBB. Use linear fit to produce lookup table of data for all pixels and at all temperatures. Over Gerb-3 pixels, © Imperial College London

17 Setup for FOVR Measurement
GERB Ground Calibration Overview and Status Setup for FOVR Measurement © Imperial College London

18 Setup for FOVR Measurement
GERB Ground Calibration Overview and Status Setup for FOVR Measurement © Imperial College London

19 Normalised response profile of a center Pixel
GERB Ground Calibration Overview and Status Normalised response profile of a center Pixel © Imperial College London

20 Contour profile of a typical center Pixel
GERB Ground Calibration Overview and Status Contour profile of a typical center Pixel © Imperial College London

21 Contour Response of an end Pixel
GERB Ground Calibration Overview and Status Contour Response of an end Pixel © Imperial College London

22 Pixel Viewing Angle GERB Ground Calibration Overview and Status GERB2
Pixel Pointing E-W Pixel Pointing E-W Pixel Pointing N-S Pixel Pointing N-S © Imperial College London

23 Awaiting return of GERB-3 for pre-launch recalibration.
GERB Ground Calibration Overview and Status Calibration Status: Awaiting return of GERB-3 for pre-launch recalibration. Expected after Easter. Investigating correction to LIBB processing Conducting refurbishment VISCS Calibration Source. New lamps and stabilization detectors. Employ detectors to ensure greater IR stability. Planning Refit of the EOCF. © Imperial College London

24 GERB Ground Calibration Overview and Status
Calibration Status: Resolved change in detector alpha’s effect on FOVR measurement – negligible. © Imperial College London


Download ppt "GERB Ground Calibration Overview and Status J. Rufus"