Activities at SMI/Vienna in testing the performance of SiPMs

Similar presentations


Presentation on theme: "Activities at SMI/Vienna in testing the performance of SiPMs"— Presentation transcript:

1 Activities at SMI/Vienna in testing the performance of SiPMs
J. Marton Stefan Meyer Institute for Subatomic Physics (SMI), Vienna J. Marton AMPDs for Frontier Detector Systems, GSI, February 10, 2009

2 Outline SMI Research Program SiPM test equipment at SMI
SiPM studies performed Example for application: Beam profile monitor Next steps J. Marton AMPDs for Frontier Detector Systems, GSI, February 10, 2009

3 SMI Research Program Kaon-nucleon interaction: SIDDHARTA at LNF,
E17 at J-PARC, FOPI at GSI, AMADEUS at LNF Matter-Antimatter (A)symmetry at AD/CERN Antiprotons at FAIR (PANDA, FLAIR, AIC) Tests of fundamental principles (VIP at LNGS) J. Marton AMPDs for Frontier Detector Systems, GSI, February 10, 2009

4 Motivations Photon detectors for our physics program :
Study of the performance of SiPMs for scintillating fiber detectors: AMADEUS, FOPI Cherenkov detectors for PID: PANDA J. Marton AMPDs for Frontier Detector Systems, GSI, February 10, 2009

5 Our interest in SiPM Applications
Low-level light detection and single photon read- out with SiPMs  Cherenkov detectors Medium-level light detection in SiPM-coupled scintillating fiber detectors  position sensitive charged particle detection SiPMs for fast calorimetry and large dynamic range Ultra-fast timing for TOF-applications J. Marton AMPDs for Frontier Detector Systems, GSI, February 10, 2009

6 Geiger-Mode-Matrix APDs: The trouble with the name … (AMPD, MAPD, SiPM, MPPC,…)
Geiger-mode APD Advantage: High Gain Disadvantage: On/Off characteristics („digital device“) Solution: Arrays of G-APDs 1 G-APD = „Cell“, „Pixel“ Output of all Cells added Photon counting Loss of active area Single GM-APD is just a single switch. A single noise event triggers the whole detector area.  Detector always busy with noise. Solution: Matrix design, noise triggers only a single cell. Hamamatsu 100U MPPC 1x1 mm2, 100 APD cells each 100 μm x 100 µm J. Marton AMPDs for Frontier Detector Systems, GSI, February 10, 2009

7 SiPM Test Systems at SMI
Black box with optical board („bread board“) and picosecond-laser Laser light can be attenuated and fed into optical fiber J. Marton AMPDs for Frontier Detector Systems, GSI, February 10, 2009

8 Picosecond – Laser for timing studies
Specifications: Power (pulse) 1000 mW Power (Ø max. f) 30 µW Frequency (max) 1 Mhz Wavelength λ 407 nm Pulse duration 32 ps J. Marton AMPDs for Frontier Detector Systems, GSI, February 10, 2009

9 Setup at SMI for cooling tests
Vacuum chamber for cooling tests, Peltier cooling, chiller with temperature control Light tight, prevents condensation of water vapor SiPM mounting in the vacuum chamber: Coupling into Laser beam via optical fiber Feedthroughs for supply voltages and signal Temperature control J. Marton AMPDs for Frontier Detector Systems, GSI, February 10, 2009

10 Low-level light detection (e.g. Cherenkov)
Requirements: High photon detection (PDE) efficiency at short wavelengths Moderate number of cells, high fill factor, good QE in the blue light range Low noise  cooling can help - but how the performance changes at low T J. Marton AMPDs for Frontier Detector Systems, GSI, February 10, 2009

11 Leakage Current Photonique Hamamatsu (HS 50)
Problem: How can one set a GM-APD at a certain operating point / gain? Suggestion: Use the leakage current (current without light) which is made up by current through the diode + dark counts. We investigated this leakage curret with respect to different ambiant conditions and found completely different dependencies for different types from different manufacturers. Left: lower temperature -> lower leakage current Right: lower temperature -> exp. increase of lc. J. Marton AMPDs for Frontier Detector Systems, GSI, February 10, 2009

12 Cooling Gain changes as function of T Dark count rate decreases
Afterpulses increase J. Marton AMPDs for Frontier Detector Systems, GSI, February 10, 2009

13 SiPMs tested Recently tests were also done with Hamamatsu 3x3 mm
New MAPDs from Zecotec J. Marton AMPDs for Frontier Detector Systems, GSI, February 10, 2009

14 Tests at SMI Current/voltage characteristics as function of T
Pulse-height spectra (CAMAC and VME based DAQ system), gain studies Timing studies with ps laser Study the effect of cooling on the performance Measurements at different temperatures, Peltier cooling below 0oC J. Marton AMPDs for Frontier Detector Systems, GSI, February 10, 2009

15 Recorded PH Spectra J. Marton AMPDs for Frontier Detector Systems, GSI, February 10, 2009

16 First tests of timing resolution
Different light intensity, threshold 0.5p.e.: Mean: 10 photon 240ps FWHM Mean: ~15 photon 180ps FWHM Stefan Hamamatsu 100U AMPD (HS 100) J. Marton AMPDs for Frontier Detector Systems, GSI, February 10, 2009

17 Comparison FWHM ~ 250 ps From Hamamatsu MPPC booklet
J. Marton AMPDs for Frontier Detector Systems, GSI, February 10, 2009

18 An Example for the Application of SiPMs : Beam-Profile Monitor for FOPI
K. Suzuki, P. Bühler, S. Fossati, J. Marton, M. Schafhauser and J. Zmeskal, "Development of SciFi/CheFi detectors with SiPM readout", Proc. New Developments In Photodetection 2008", Aix-les-Bains, France, Nuclear Instruments and Methods in Physic Research A, in print. Monitor of proton beam Good position resolution ~1mm 1 grid consisting of 2 layers of scintillating fibers 1mm 9x9 cm active area Resistant to magnetic field of 0.6 Tesla High rate capability (>1 MHz) J. Marton AMPDs for Frontier Detector Systems, GSI, February 10, 2009

19 Beam Profile Monitor - Design
2 Layers of scint. fibers X-Y Readout with SiPMs J. Marton AMPDs for Frontier Detector Systems, GSI, February 10, 2009

20 During Assembling J. Marton AMPDs for Frontier Detector Systems, GSI, February 10, 2009

21 Beam Profile Monitor J. Marton AMPDs for Frontier Detector Systems, GSI, February 10, 2009

22 Installation at GSI-FOPI
J. Marton AMPDs for Frontier Detector Systems, GSI, February 10, 2009

23 Installation at GSI-FOPI
J. Marton AMPDs for Frontier Detector Systems, GSI, February 10, 2009

24 Beam profile J. Marton AMPDs for Frontier Detector Systems, GSI, February 10, 2009

25 Next steps Improving cooling system (hopefully down to -30ºC)
Automatic basic characteristics evaluation system to test new products and protypes Timing performance of SiPMs with laser system J. Marton AMPDs for Frontier Detector Systems, GSI, February 10, 2009

26 Team @ SMI Ahmed Gamal Paul Bühler, Matthias Schafhauser, Ken Suzuki,
Johann Marton technicians (3) and students J. Marton AMPDs for Frontier Detector Systems, GSI, February 10, 2009

27 Thank you for your attention
J. Marton AMPDs for Frontier Detector Systems, GSI, February 10, 2009

28 Dark count rate with Peltier cooling
Gain kept constant by adjusting the bias voltage J. Marton AMPDs for Frontier Detector Systems, GSI, February 10, 2009