Presentation is loading. Please wait.

Presentation is loading. Please wait.

BOLD functional MRI Magnetic properties of oxyhemoglobin and deoxyhemoglobin L. Pauling and C. Coryell, PNAS USA 22:210-216 (1936) BOLD effects in vivo.

Similar presentations


Presentation on theme: "BOLD functional MRI Magnetic properties of oxyhemoglobin and deoxyhemoglobin L. Pauling and C. Coryell, PNAS USA 22:210-216 (1936) BOLD effects in vivo."— Presentation transcript:

1 BOLD functional MRI Magnetic properties of oxyhemoglobin and deoxyhemoglobin L. Pauling and C. Coryell, PNAS USA 22: (1936) BOLD effects in vivo S. Ogawa, et al., MRM, 14:68-78 (1990) BOLD activation experiments K. K. Kwong, et al., PNAS USA, 89: (1992) S. Ogawa, et al., PNAS USA, 89: (1992) P. A. Bandettini, et al., MRM, 25: (1992) J. Frahm, et al., JMRI, 2: (1992)

2 Mechanism of BOLD Functional MRI
Brain activity Oxygen consumption Cerebral blood flow Oxyhemoglobin Deoxyhemoglobin Magnetic susceptibility T2* MRI signal intesity

3 Magnetic Properties of Oxyhemoglobin and Deoxyhemoglobin
Deoxyhemoglobin: paramagnetic (c > 0) paramagnetic with respect to the surrounding tissue Oxyhemoglobin: diamagnetic (c < 0) isomagnetic with respect to the surrounding tissue

4 Magnetic Susceptibility

5 Oxyhemoglobin and Deoxyhemoglobin in Veins during Brain Activation
Rest Activation Normal blood flow High blood flow Oxyhemoglobin Deoxyhemoglobin

6 T2* Effect in fMRI action MR signal (S) rest TE t excitation reception

7 Time Series and Activation Maps
Signal Intensity Off On Off On Off On Off On Scan Number

8 Challenges in Functional FMRI
Sensitivity (Contrast-to-noise ratio) BOLD signal change is ~1-2% at 1.5 T; signal-to-noise ratio in single-shot EPI images is ~100. Physiological pulsations (cardiac and respiratory); Head motion; instrumental instability Specificity Location of activation – neurons or veins Susceptibility artifacts

9 Challenges in Functional FMRI
Temporal resolution Limited by BOLD impulse-response function, image sampling rate, and spin relaxation times Spatial resolution Limited by BOLD point-spread function, signal-to-noise ratio, and image sampling rate Non-linearity Neurological and hemodynamic Acoustic noise

10 Contrast-to-Noise Ratio
Brain Activation-related signal change Sensitivity = Temporal fluctuation of image intensity

11 Enhancement of BOLD Contrast
Higher magnetic fields BOLD signal change DS ~ Ba (1 < a < 2) Standard clinical MRI scanner at 1.5 T Research scanner up to 8 T currently Optimization of image acquisition parameters Optimal echo time (TE) to maximize BOLD signal Optimal repetition time (TR) to increase number of images acquired per unit time, and to decrease motion artifacts

12 TE Dependence of Signal Change

13 Suppression of Temporal Fluctuations
Head motion reduction Head holder modified from a football helmet Image realignment in data processing Physiological pulsations Correction using simultaneously recorded cardiac and respiratory signals Ultra-fast imaging techniques Single-shot echo-planner imaging (EPI) Single-shot Spiral imaging Post-processing Denoising

14 Ultra-Fast Spiral Scanning
An image (64x64) can be acquired in ~ 20 ms Reduce head motion Increase number of images collected per unit time Stable Spiral trajectory is insensitive to motion and flow artifacts Zero gradient moments at the center of k-space (self-navigated) First-order gradient moments vary smoothly over k-space

15 Multi-Slice Spiral Images

16 Multi-Slice EPI Images

17 Activation Maps on Anatomical Images
MS Spiral MS EPI 3D Spiral

18 Histograms of Temporal Standard Deviations
80 60 40 20 MS-Spiral MS-EPI 3D-Spiral Number of Activated Voxels SD/Mean

19 Comparison of Activation Studies Using MS-spiral, MS-EPI, and 3D-spiral

20 Specificity in fMRI Inflow Effects
Generated by fresh (fully recovered) spins moving into the region excited under saturating conditions Flow change in large vessels can lead to substantial signal increases (20-30%), and compromise spatial accuracy in activation studies

21 Inflow Effects in fMRI Suppression of inflowing spins a b
d b Suppression of inflowing spins No suppression of inflowing spins

22 Gradient Echo vs. Spin Echo in fMRI
Diameter (mm) DR2, DR2 * (1/sec) 20 10 DR2 DR2* Contribution of large vessels and capillary beds to the BOLD signals; Separation/suppression of signals from large vessels.

23 Temporal resolution Impulse-response function 12s 5s 2s t

24 Temporal resolution Sampling rate (single-shot EPI ~10-15 slices/sec)
Whole brain (~30 4mm-slices): 2-3 sec T1 relaxation times Grey matter: 1 sec White matter: 0.8 sec CSF: 2-3 sec

25 Spatial resolution BOLD point-spread function Image spatial resolution
Spatial extent of neuronal activity, CBF, and BOLD Image spatial resolution 64x64 with FOV 240 mm: 3.75mm 128x128 with FOV 240 mm: mm Signal-to-noise ration Single-shot EPI with voxel size 4x4x4 mm3: ~100

26 Non-linearity of BOLD Response
BOLD response vs. length of stimulation t 2t BOLD response during rapidly-repeated stimulation ts

27 Experimental Designs in fMRI
Block-Design fMRI Task Rest 20-60s Event-Related fMRI 8-12s

28 Hemodynamic Response vs. ISI
5 10 15 Time (sec) Signal Changes (%) 1.00 1.01 ISI=8s ISI=12s ISI=16s

29 Visual Activation Maps (ISI=12s)

30 Data Analysis Methods For fMRI
Hypothesis-driven approaches t-test, cross-correlation, GLM, etc. Data-driven approaches Principal component analysis (PCA), independent component analysis (ICA), and clustering analysis.


Download ppt "BOLD functional MRI Magnetic properties of oxyhemoglobin and deoxyhemoglobin L. Pauling and C. Coryell, PNAS USA 22:210-216 (1936) BOLD effects in vivo."

Similar presentations


Ads by Google