© 2013 Fluke Corporation 1 1586A Super-DAQ Precision Temperature Scanner KRISS MEASUREMENT CLUB November 14, 2013 FLUKE CALIBRATION JAEHEE JANG Public.

Similar presentations


Presentation on theme: "© 2013 Fluke Corporation 1 1586A Super-DAQ Precision Temperature Scanner KRISS MEASUREMENT CLUB November 14, 2013 FLUKE CALIBRATION JAEHEE JANG Public."— Presentation transcript:

1 © 2013 Fluke Corporation 1 1586A Super-DAQ Precision Temperature Scanner KRISS MEASUREMENT CLUB November 14, 2013 FLUKE CALIBRATION JAEHEE JANG Public Launch: November 18

2 © 2013 Fluke Corporation 2 1586A Super-DAQ Precision Temperature Scanner The most accurate and flexible temperature data acquisition system  Scans and records temperature, dc voltage, dc current, and resistance  Measures up to 40 channels  Maximum scan rate of 10 channels per second Designed for both the factory and the lab:  Factory: Multi-channel temperature data acquisition applications  Laboratory: Benchtop temperature sensor calibration

3 © 2013 Fluke Corporation 3 1586A Super-DAQ Industries:  Pharmaceutical  Biotechnology  Food processing  Aerospace  Automotive Applications:  Thermal mapping  Temperature validation  Process sensor calibration  Process monitoring  Quality control testing  Life cycle testing

4 © 2013 Fluke Corporation 4 1586A Key Features Six key features set the 1586A Super-DAQ apart from other products in its class:  Best temperature measurement accuracy  Flexible configuration for the factory or benchtop  Multiple modes of operation  Real-time graphing in color  Data portability and security  Automated sensor calibration

5 © 2013 Fluke Corporation 5  Best temperature measurement accuracy The 1586A Super-DAQ reads PRTs, thermocouples, and thermistors with best-in-class accuracy: PRTs ± 0.005 °C (using external DAQ-STAQ Multiplexer) Thermocouples: ± 0.5 °C (using High-Capacity Module and internal CJC) Thermistors: ± 0.002 °C

6 © 2013 Fluke Corporation 6  Flexible configuration For the factory… 1586A Super-DAQ + internal High-Capacity Module 1586-2586 High-Capacity Module Universal input channels designed to measure temperature, resistance, dc voltage, and dc current For data acquisition applications on the factory floor where channel count and scan speed are important Save wiring time and maintenance. Wire a module once for a thermal test. Disconnect it from the 1586A and reconnect another one for a different test.

7 © 2013 Fluke Corporation 7  Flexible configuration continued For the calibration lab… 1586A Super-DAQ + external DAQ-STAQ Multiplexer 1586-2588 DAQ-STAQ Multiplexer Designed for high-accuracy measurements in secondary temperature calibration labs Easily connect/disconnect thermocouples, PRTs, and thermistors Stack it on the 1586A Super-DAQ in busy labs to reduce footprint

8 © 2013 Fluke Corporation 8  Multiple modes of operation The 1586A Super-DAQ can operate in four modes from a single instrument… Scan – Monitor – Measure – Digital Multimeter (DMM) Sequentially scan through channels based on a user defined test Monitor any single channel during a scan Measure and record data on a single channel without the need of a pre- configured test file Or use the 1586A like a familiar bench digital multimeter

9 © 2013 Fluke Corporation 9  Real-time graphing in color View real time data for all channels in table format Chart up to four channels in color at the same time Zoom in or out to see data of interest and monitor trends Switch between chart view and table view of various measurement data and statistics History mode lets you scroll through collected data in a scan file (without a PC and expensive charting programs)

10 © 2013 Fluke Corporation 10  Data portability and security 20MB internal memory Transfer data and set-up files using USB flash drive or over network using LAN interface Two-levels of data security (admin and user profiles) to prevent unauthorized users from tampering with or forging test data or setup files

11 © 2013 Fluke Corporation 11  Automated sensor calibration Use the 1586A Super-DAQ to control a Fluke Calibration drywell or bath for an automated calibration routine The Super-DAQ ramps the drywell or bath through a series of user defined setpoint temperatures At each setpoint, the 1586A automatically records readings for the reference thermometer and one or more sensors under test Sensor calibration can be greatly automated using this feature… improving your lab efficiency

12 © 2013 Fluke Corporation 12 1586A Super-DAQ at-a-glance Five-way binding post for reference thermometer input. Also measures dc voltage, dc current, and resistance. USB support for easy transfer of data and setup files Navigation keys for fast selection of information Color display with channel status indicators. Chart up to 4 channels simultaneously. Backlit main function keys – always know where you are and the recording status Memory management – for downloading and moving data and setup folders Measure/DMM function allows quick measurements and recording with minimal setup Function soft keys support fast setup and operation Standby key turns the display and keypad off—eliminating warm-up time Support for menu and displayed text in nine languages – English, Chinese, French, German, Portuguese, Spanish, Russian, Japanese, and Korean. Front Panel

13 © 2013 Fluke Corporation 13 1586A Super-DAQ at-a-glance Back Panel Line voltage and fuse selector Power switch and line cord connector Dedicated connector for control of Fluke Calibration drywells or temperature baths for automated tests Totalizer and Digital I/O input ports Digital external alarm trigger outputs Two rear panel slots can be configured for the internal High-Capacity Module or the external DAQ-STAQ Multiplexer—or one of each USB and Ethernet ports used for remote operation

14 © 2013 Fluke Corporation 14 1586A Configuration Input Modules The 1586A has two slots on the rear panel Either slot can be configured for the internal or external module:  Internal High-Capacity Module (1586-2586) Intended for data acquisition applications on the factory floor where thermocouples will most frequently be used 20 two-wire terminals (can be wired for 3-, or 4-wire) Two dedicated current channels Serial number and cal date stored in on-board memory IC Module indicator lights up when module installed 1586A display indicates that a module is installed in slot one 1586A configured for both internal and external modules High-Capacity Module DAQ-STAQ Multiplexer 1586A Super-DAQ

15 © 2013 Fluke Corporation 15 Input Modules (cont.)  External DAQ-STAQ Multiplexer (1586-2588) Intended for benchtop calibration in secondary cal labs where accuracy and quick UUT connections are required 10 four-wire PRT inputs (mini-DWF terminal) 20 mini-jack thermocouple inputs Two 3-meter interface cables connect the DAQ-STAQ to the 1586A (cables stored in a compartment inside the module) DAQ-STAQ connector cables stored inside module 1586A Configuration continued 1586A Super-DAQ and DAQ-STAQ Multiplexer DAQ-STAQ Multiplexer inputs Four-wire PRT inputs (mini-DWF terminal) 10 total Mini-jack thermocouple inputs 20 total

16 © 2013 Fluke Corporation 16 Ordering Information 1586A/1HC 1586A/1DS 1586A/DS-HC *What’s in the box?  -1586A Super-DAQ (configured as ordered with modules & relay boards)  -USB cable  -Line cord  -Traceable certificate of calibration  -Screwdriver (included with HC module)  -User Manual on CD (English, Chinese, Japanese, Korean, French, German, Spanish, Portuguese, and Russian)  -Printed Safety Sheet  -USB serial driver on CD  -Note: RS-232 cable is NOT included. If required, order item# 2673499 (DE-9 Null Modem, FF).

17 © 2013 Fluke Corporation 17 Appendix A – Summary Specs Note: Specifications subject to change General Specifications Channel Capacity Total analog channels45 Voltage/resistance channels41 Current channels5 Digital I/O1 (8-bit TTL) Totalizer1 Alarm outputs6 Trigger input1 Sample Rate Fast: 10 channels per second (0.1 seconds per channel) Medium: 1 channel per second (1 second per channel) Slow: 4 seconds per channel Display Resolution6 1/2 digits Input Module Options 2 rear panel slots ('x' designates slot 1 or slot 2 below). Can be configured for 2 High-Capacity Modules, 2 DAQ-STAQ Multiplexers, or one of each. Internal High-Capacity Module Configuration 20 channels per module (x01 - x20) 2 dedicated current channels (x21 - x22) (3- and 4-wire connections require 2 channels) External DAQ-STAQ Multiplexer Configuration 10 PRT/thermistor channels (max). 20 mini-jack thermocouple channels (max). Math Channels 20 channels Operations: sum, difference, multiply, divide, polynomial, power, square root, reciprocal, exponential, logarithm, absolute value, average, maximum, minimum TriggersInterval, external (trigger input), alarm, remote (bus), manual, automated test Memory Scan data RAM75,000 readings with timestamp Internal memory20 MB Standard Interfaces USB host, USB device, Ethernet 10/100 LAN, RS-232 Temperature source control for automated temperature data collection from Fluke Calibration drywells and baths Mains Voltage 100 V setting: 90 to 100 V; 120 V setting: 108 to 132 V; 220 V setting: 198 to 242 V; 240 V setting: 216 to 264 V Frequency47 to 440 Hz Input Protection50 V all functions, terminals and ranges Dimensions Height x Width X Depth 150 mm x 245 mm x 385 mm (5.9 in x 9.6 in x 15.2 in) Weight6 kg (13.2 lb) -- typical configuration Shipping Weight9.5 kg (20.9 lb) -- typical configuration ConformityCE, CSA, IEC 61010 3 rd ed.

18 © 2013 Fluke Corporation 18 Appendix A – Summary Specs 1586A with High-Capacity Module Accuracy Specifications * PRTs and Thermistors (4-wire) Temperature Range Temperature Resolution Temperature AccuracyResistance AccuracyExcitation Current PRT/RTD (100 Ω nominal) −200 °C to 1200 °C (depending on the sensor) 0.001 °C 0.008 °C at -200 °C 0.008 °C at 0 °C 0.018 °C at 300 °C 0.003% or 0.003 Ω (0 Ω to 400 Ω) 1 mA with current reversal Thermistor (10 KΩ nominal) −200 °C to 400 °C (depending on the sensor) 0.0001 °C 0.002 °C at 0 °C 0.009 °C at 50 °C 0.024 °C at 100 °C 0.004% + 0.5 Ω (0 Ω to 98 kΩ) 10 μA Thermocouples Temperature Range Temperature Resolution Temperature Accuracy (int CJC) Temperature Accuracy (ext CJC) Voltage Accuracy Thermocouple Type K−270 °C to 1372 °C0.01 °C 0.62 °C at 0 °C 0.64 °C at 1000 °C 0.15 °C at 0 °C 0.20 °C at 1000 °C 0.004% + 4 μV ch. 1 0.004% + 6 μV ch. x01-x20 Thermocouple Type T−270 °C to 400 °C0.01 °C 0.65 °C at 0 °C 0.47 °C at 200 °C 0.15 °C at 0 °C 0.12 °C at 200 °C 0.004% + 4 μV ch. 1 0.004% + 6 μV ch. x01-x20 Thermocouple Type S−50 °C to 1768 °C0.01 °C 0.76°C at 300 °C 0.60 °C at 1200 °C 0.67°C at 300 °C 0.54 °C at 1200 °C 0.004% + 4 μV ch. 1 0.004% + 6 μV ch. x01-x20 1586A with DAQ-STAQ Accuracy Specifications * PRTs and Thermistors (4-wire) Temperature Range Temperature Resolution Temperature AccuracyResistance AccuracyExcitation Current PRT/RTD (100 Ω) −200 °C to 1200 °C (depending on the sensor) 0.001 °C 0.005 °C at -200 °C 0.005 °C at 0 °C 0.012 °C at 300 °C 0.002% or 0.002 Ω (0 Ω to 400 Ω) 1 mA with current reversal Thermistor (10 KΩ) −200 °C to 400 °C (depending on the sensor) 0.0001 °C 0.002 °C at 0 °C 0.009 °C at 50 °C 0.024 °C at 100 °C 0.004% + 0.5 Ω (0 Ω to 98 kΩ) 10 μA Thermocouples Temperature Range Temperature Resolution Temperature Accuracy (int CJC) Temperature Accuracy (ext CJC) Voltage Accuracy Thermocouple Type K−270 °C to 1372 °C0.01 °C 0.29 °C at 0 °C 0.32 °C at 1000 °C 0.15 °C at 0 °C 0.20 °C at 1000 °C 0.004% + 4 μV ch. 1 0.004% + 6 μV ch. x01-x20 Thermocouple Type T−270 °C to 400 °C0.01 °C 0.30 °C at 0 °C 0.23 °C at 200 °C 0.15 °C at 0 °C 0.12 °C at 200 °C 0.004% + 4 μV ch. 1 0.004% + 6 μV ch. x01-x20 Thermocouple Type S−50 °C to 1768 °C0.01 °C 0.68°C at 300 °C 0.55 °C at 1200 °C 0.67°C at 300 °C 0.54 °C at 1200 °C 0.004% + 4 μV ch. 1 0.004% + 6 μV ch. x01-x20 * Accuracy specifications apply using medium sample rate (1 second per channel). Specifications are also available for type R, J, N, E, B, C, D, G, L M, U, and W thermocouples. Please consult the 1586A Super-DAQ Extended Specifications for additional information.

19 © 2013 Fluke Corporation 19 Appendix A – Summary Specs DC Voltage (medium or slow sample rate) RangeAccuracy Channel 1Ch. X01 - x20Fast Sample RateResolutionInput Impedance ± 100 mV0.0037 % + 0.0035 %add 2 μVadd 0.0008 % of range0.1 μV10 GΩ [1] ± 50 V0.0038 % + 0.0012 %-add 0.0008 % of range1 mV10 MΩ ± 1 % [1] Input beyond ± 12 V is clamped. The clamp current is up to 3 mA. DC Current (medium or slow sample rate) RangeAccuracyFast Sample RateResolutionBurden Voltage ± 100 μA0.015 % + 0.0035 %add 0.0008 % of range0.1 nA< 1 mV ± 10 mA0.015 % + 0.0035 %add 0.0008 % of range10 nA< 1 mV ± 100 mA0.015 % + 0.0035 %add 0.0008 % of range100 nA< 1 mV Resistance (medium or slow sample rate) RangeAccuracyFast Sample RateResolutionSource Current 100 Ω0.004 % + 0.0035 %add 0.001 % of range0.1 mΩ1 mA 10 kΩ0.004 % + 0.001%add 0.001 % of range10 mΩ100 μA 10 MΩ0.015 % + 0.001 %add 0.01% of range1 kΩ0.1 μA

20 © 2013 Fluke Corporation 20 1586A Super-DAQ Q&A


Download ppt "© 2013 Fluke Corporation 1 1586A Super-DAQ Precision Temperature Scanner KRISS MEASUREMENT CLUB November 14, 2013 FLUKE CALIBRATION JAEHEE JANG Public."

Similar presentations