Presentation is loading. Please wait.

Presentation is loading. Please wait.

Quasar 950 & 960 Series Toxic Gas Open Path Detectors.

Similar presentations


Presentation on theme: "Quasar 950 & 960 Series Toxic Gas Open Path Detectors."— Presentation transcript:

1 Quasar 950 & 960 Series Toxic Gas Open Path Detectors

2 Contents 2 | 87 1Principle of Gas Detection 2SafEye 950 & 960 Models and Types 3SafEye 950 & 960 Benefits 4SafEye 950 & 960 Technical Specifications 5SafEye 950 & 960 Features & Functions 6Accessories 7Part Numbers 8Installation & Commissioning 9Testing & Maintenance 10Applications

3 Agenda 3 | 87 1Principles of Gas Detection 2SafEye 950 & 960 Models and Types 3SafEye 950 & 960 Benefits 4SafEye 950 & 960 Technical Specifications 5SafEye 950 & 960 Features & Functions 6Accessories 7Part Numbers 8Installation & Commissioning 9Testing & Maintenance 10 Applications

4 4 | 87 Toxic Open Path Gas Detector SafEye 950 & 960 Principles of Gas Detection GAS Current Driver Flashlamp Source CPU Outputs Detector Signal 1 Reference Signal 2 The open path gas detector system detects gases through a dual spectral range monitoring, analyzing the absorption of radiation caused by gases in the atmosphere and comparing it (ratio) to background atmospheric absorption.

5 5 | 87 20 ppm 20 m Separation of the measuring length in e.g. 20 single measuring point 5 m “Cloud“ Length5 m Average gas concentration of the “cloud“20 ppm Sum100 ppm.m Receiver RX Transmitter TX Detector Head Reading (ppm) 10 ppm 2 3 4 5 6 7 820 ppm 9 1020 ppm 1120 ppm 1220 ppm 130 ppm 140 ppm 150 ppm 160 ppm 170 ppm 180 ppm 190 ppm 200 ppm Sum100 ppm Toxic Open Path Gas Detector SafEye 950 & 960 Principles of Gas Detection

6 6 | 87 High gas dilution on complete measuring length and or low local / partial gas concentration Gas Leakage Minor gas dilution on complete measuring length and or high local / partial gas concentration The gas concentration is measured as integral /added sum over the complete measuring length: ∑ % ppm x m Both system are measuring the same total gas concentration. Simplified view: Receiver RX Transmitter TX Receiver RX Transmitter TX Principles of Gas Detection

7 Contents 7 | 87 1Principle of Gas Detection 2SafEye 950 & 960 Models and Types 3SafEye 950 & 960 Benefits 4SafEye 950 & 960 Technical Specifications 5SafEye 950 & 960 Features & Functions 6Accessories 7Part Numbers 8Installation & Commissioning 9Testing & Maintenance 10Applications

8 8 | 87 Product and Variant concept SafEye 950 & 960: Flash Source unit (Tx) – 3 for different distances:  Tx for 5 up to 20m  Tx for 15 up to 40m  Tx for 35 up to 80m The UV Source unit emits UV radiation pulses at the rate of pulse per second. The front of the source has a internal reflector that collimates the UV beam for maximum intensity. The sources models have the same electronics and optical device. The only difference is in the apparatus. Toxic Open Path Gas Detector SafEye 950 & 960 SafEye 950 & 960 Models and Types

9 9 | 87 Product and Variant concept SafEye 950 & 960: Detector unit (Rx) – 2 for different gases:  SafEye 950 for Hydrogen sulfide (H 2 S)  SafEye 960 Ammonia (NH 3 ) The Detector receives the transmitted pulsed radiation signals from the UV Source. The signals are then amplified and fed into an analog to digital signal converter to be processed by the internal microprocessor. When the signals drop below a prescribed level, the internal microprocessor will compensate for them. This will allow the signal to be maintained even in severe weather conditions. The data is sent to the output interface section. In the UV band, H 2 S has characteristically strong absorption in the solar blind range of 189–270 nm and NH 3 within the range of 189 - 210 nm. Aromatic Hydrocarbons such as Benzene, Toluene, Xylene, Styrene, etc., have the unique “conjugated double bonding” chemical configuration responsible for their spectral absorption in the UV band. In the UV band, the aromatic compounds have a typically strong absorption in ranges of 200 –230 nm and 250-270 nm. Toxic Open Path Gas Detector SafEye 950 & 960 SafEye 950 & 960 Models and Types

10 10 | 87 The Quasar 950 (H 2 S) & 960 (NH 3 ) are represented by 2 different Detectors, for H 2 S and NH 3, combined with 3 different Sources. This offers the possibility to cover detection ranges from 5 up to 80m. Toxic Open Path Gas Detector SafEye 950 & 960 SafEye 950 & 960 Models and Types

11 Contents 11 | 87 1Principle of Gas Detection 2SafEye 950 & 960 Models and Types 3SafEye 950 & 960 Benefits 4SafEye 950 & 960 Technical Specifications 5SafEye 950 & 960 Features & Functions 6Accessories 7Part Numbers 8Installation & Commissioning 9Testing & Maintenance 10Applications

12 12 | 87  Fence Line monitoring of Toxic Gases  H 2 S, NH 3 and Aromatic Hydrocarbons  Coverage of low toxic concentration 0 - 500 ppm.m  Coverage of long distances up to 80m  Reliable and fast detection principle  strong and long life time of UV source  Detection within solar blind UV range  Response time < 10 sec  Designed to meet SIL2  Can withstand harsh and extreme environments  Heated optic to eliminate effects of icing, snow, condensation  Continuous operation in extreme environments -55°C to +65°C  Easy alignment & commissioning  One-man set-up  Communication Interfaces with HART 7 and RS485 capability Toxic Open Path Gas Detector SafEye 950 & 960 SafEye 950 & 960 Benefits

13 Contents 13 | 87 1Principle of Gas Detection 2SafEye 950 & 960 Models and Types 3SafEye 950 & 960 Benefits 4SafEye 950 & 960 Technical Specifications 5SafEye 950 & 960 Features & Functions 6Accessories 7Part Numbers 8Installation & Commissioning 9Testing & Maintenance 10Applications

14 14 | 87 Spectral Response:190 -300nm Response Time:< 10 sec Sensitivity Range:0-500 ppm.m Minimum Detectable Level:50 ppm.m Drifter:±10 % of reading or ± 5% of full scale (whichever is greater) Displacement/Misalignment:> ±1° tolerance Heated Optics:To eliminate condensation and icing on the window Warm up Time: 30 sec for transmitter and receiver Temperature Range:-55°C to +65°C (-67°F to +149°F) Immunity to False Alarms:Not influenced by solar radiation, hydrocarbon flames Toxic Open Path Gas Detector SafEye 950 & 960 SafEye 950 & 960 Technical Specifications

15 15 | 87 Electrical Specifications Multiple Output Options: 0-20 mA current output; Sink (source option) HART 7.0 protocol RS485, Modbus compatible Power Supply:24VDC nominal (18-32 VDC) Power Consumption:Detector: 150mA (200mA with heated window) Source: 200mA (260mA with heated window) Electrical Input Protection:per EN/IEC 50270 Electromagnetic Compatibility:EMI/RFI protected per EN/IEC 50270 Toxic Open Path Gas Detector SafEye 950 & 960 SafEye 950 & 960 Technical Specifications

16 16 | 87 Mechanical Specifications Enclosure:Stainless steel 316L Cable entries:2 x 3/4" – 14NPT conduits or 2 x M25 x 1.5mm ISO Dimensions: Detector/Source: 267 x 130 x 130mm (10.5" x 5.1" x 5.1") Tilt Mount: 120 x 120 x 158mm (4.7" x 4.7" x 5.5") Weight:Detector/Source: 5.0 kg (11.0lb) Tilt Mount: 1.9 kg (4.2lb) Ingress Protection: Water and IP66 and IP68 Dust Tight NEMA 250 6P Toxic Open Path Gas Detector SafEye 950 & 960 SafEye 950 & 960 Technical Specifications

17 17 | 87 Approvals Ex Approvals: ATEX, IECEx (pending): Ex II2 (2) GD Ex d e ib [ib Gb] IIB + H2 T4 Gb Ex tb IIIC T135°C Db IP66 Ta= (-55°C to + 65°C) TR CU (GOST R & K) Reliability Approval:TÜV SIL 2 approved per IEC 61508 (pending) Warranty:SafEye system – 3 years Flash source bulb – 3 years Toxic Open Path Gas Detector SafEye 950 & 960 SafEye 950 & 960 Technical Specifications

18 18 | 87 UV Flash Source (Tx) Unit No. Component No. Component 1Front Window Section6 Rear Cover 2Label7Earth Terminal 3 Main Housing / Enclosure 8Front Window 4 Mounting Plate10Cable Inlet / Entry 5 Junction Box11Indication LED The front window is heated to improve performance in icing, condensation and snow conditions. Toxic Open Path Gas Detector SafEye 950 & 960 SafEye 950 & 960 Technical Specifications

19 19 | 87 Detector (Rx) Units No.ComponentNo.Component 1Front Window Section 7Earth Terminal 2Label8Front Window 3 Main Housing / Enclosure 9Handheld Fast Connection 4 Mounting Plate10Cable Inlet / Entry 5 Junction Box11Indication LED 6 Back Cover The front window is heated to improve performance in icing, condensation and snow conditions. Toxic Open Path Gas Detector SafEye 950 & 960 SafEye 950 & 960 Technical Specifications

20 Contents 20 | 87 1Principle of Gas Detection 2SafEye 950 & 960 Models and Types 3SafEye 950 & 960 Benefits 4SafEye 950 & 960 Technical Specifications 5SafEye 950 & 960 Features & Functions 6Accessories 7Part Numbers 8Installation & Commissioning 9Testing & Maintenance 10Applications

21 21 | 87 SafEye 950 & 960 – Operational Modes : The Quasar 950/960 has four (4) operational modes:  Normal Mode  Maintenance Call Mode (3 mA Output)  Fault Mode  Zero Calibration Mode (1mA Output) Normal Mode The Normal Mode is used for gas detection. In Normal Mode, the following statuses are possible:  Normal (N): Gases have been detected at safe levels  Warning (W)*: Gases have been detected at warning levels  Alarm (A)*: Gases have been detected at alarm levels Note: The user chooses these alarm levels at the controller. The Detector output is 4 mA at zero reading and 20 mA for full-scale reading. For the standard 0-20 mA output, the Warning and Alarm levels are not relevant. If the RS-485 output is used, the Detector changes its status from N to W at Warning level and to A at Alarm. Toxic Open Path Gas Detector SafEye 950 & 960 SafEye 950 & 960 Features & Functions

22 22 | 87 Visual Indicators : One 3-color LED indicator is located at the rear. Indicator LED Detector StatusLED colorLED mode FaultYellow4 Hz – flashing Alignment / StandbyYellow1 Hz - flashing Zero CalibrationYellowConstant NormalGreen1 Hz - flashing WarningRed2 Hz – flashing AlarmRedConstant Source StatusLED colorLED mode FaultYellow4 Hz – flashing NormalGreen1 Hz - flashing Toxic Open Path Gas Detector SafEye 950 & 960 SafEye 950 & 960 Features & Functions

23 23 | 87 Current Reading Status and Description 0mA + 0.2mAFault 2 or Low Voltage (Critical) 1mA ± 0.2mAZero Calibration (In Progress) 2mA ± 0.2mAFault 1 (non-critical) 3mA ± 0.2mA “Maintenance Call” 4mA ± 0.2mAZero reading – No gas present 4-20mAContinuous measuring of gas concentration between 0 and full scale 21mAConcentration is over the range limit (more than full-scale deflection) Standard 0-20 mA Fully linearized 4-20 mA gas signal in accordance with gas concentration Toxic Open Path Gas Detector SafEye 950 & 960 SafEye 950 & 960 Features & Functions

24 Operational Modes: Maintenance Call Mode (3 mA Output) The Maintenance Call Mode indicates a low signal or low signal ratio that may be caused by a dirty window, misalignment, poor source, or one of the Detector’s parameters is at the “limit” value. The Detector continues to function and reads any gas present, but provides a (3 mA) pre-warning signal that a maintenance procedure is required. Zero Calibration Mode (1mA Output) This mode calibrates the base level from which the Detector read gas to zero. It should only be performed in the following situations:  There are no combustible gases present  A clear path exists between the UV Source and Detector  Clear weather conditions are present Zero Calibration must be performed after installation, re-alignment and window cleaning, by using the handheld unit, Host software on PC or the magnet. Toxic Open Path Gas Detector SafEye 950 & 960 SafEye 950 & 960 Features & Functions

25 Fault Modes In Fault Mode, there are three fault types:  Fault 1 (2 mA Output): Non critical This occurs due to a blockage, very low signal or partial obscuration, or full beam block. In this situation, detection is no longer possible. The Detector’s proper operation can be restored (Auto reset) during operation if the condition causing the problem is removed or resolved. This mode occurs after a delay of 60 sec. from the moment of the fault. This delay is important to eliminate momentary obscuration due to an object/person passing through the beam.  Low Voltage Fault (1 mA Output) In this situation, detection is disabled due to low voltage being supplied to the Detector. The Detector returns to proper operation only when proper voltage level restored.  Fault 2 (1 mA Output): Critical In this situation, the detection is disabled due to an electrical/software operational failure or a central device (memory, processor) fault. Such a fault causes the Detector to cease operation. If there is a fault in the 0-20 mA loop, the output is 0 mA. Toxic Open Path Gas Detector SafEye 950 & 960 SafEye 950 & 960 Features & Functions

26 26 | 87 SafEye 950 & 960 – Function Set-Up: The functions to set:  Full Scale Deflection / Measuring Range  Analog Output mode  Background Zero Calibration Mode  Address Setup (RS485)  Mode of Operation / Function Set-Up  Alarm Latch  Heated Optics Operation  Heat Mode  Heat on Temperature Toxic Open Path Gas Detector SafEye 950 & 960 SafEye 950 & 960 Features & Functions

27 27 | 88 SafEye 950 & 960 – Zero Calibration: There are several ways to perform and initiate a Zero calibration :  Magnet (Enclosure)  HART / RS485 Handheld terminal  HOST Software RS485 PC via  Handheld Fast Connection Port  Segregated Terminals Magnet Toxic Open Path Gas Detector SafEye 950 & 960 SafEye 950 & 960 Features & Functions

28 28 | 87 Function Set-Up: Heated Optics This feature relates to the Detector only:  The heated optics can be defined as one of the following modes:  OFF: Not operated  ON: Continuously  AUTO: Per temperature change  In AUTO mode, the HEAT ON setting is 41°F (5°C). default Note: When operated per temperature change, the user can define the start temperature below which the window will be heated between 0 to 50°C. The Detector will stop heating the window when the temperature is 15°C above the defined temperature. The Source heated optic must be defined with the order in two options:  Heated continuously  Start heating below 41°F (5°C - Default ). Toxic Open Path Gas Detector SafEye 950 & 960 SafEye 950 & 960 Features & Functions

29 HART Interface “HART” - Highway Addressable Remote Transducer. The HART Protocol provides two simultaneous communication channels: the 4-20mA analog signal and a digital signal. The 4-20mA signal communicates the primary measured value (in the case of a field instrument) using the 4-20mA current loop. Additional device information is communicated using the digital signal (Frequency Shift Keying - FSK) that is superimposed on the analog signal. SafEye 950 & 960 functionality via HART communication:  Detector  Display set-up  Reconfigure set-up  Detector health and status  Display detector status and definition  Perform detector diagnostic  Detector troubleshooting SafEye 950 & 960 fully compatible with HART 7.0 protocol and certified by HART. Toxic Open Path Gas Detector SafEye 950 & 960 SafEye 950 & 960 Features & Functions

30 30 | 87 HART Interface Spectrex device descriptions are also compatible with the following Process Management Systems:  Meriam MFC-5150X  Emerson Hand terminal 375 & 475  SDC 625  Simatic PDM, Siemens  ABB800xA  Smar 401  Honeywell MC Toolkit Configurator Toxic Open Path Gas Detector SafEye 950 & 960 SafEye 950 & 960 Features & Functions

31 31 | 87 RS 485 Interface The RS-485 input/output sends complete data information to a PC and receives data or control commands from the PC. The protocol is Modbus compatible. The communication with the PC that is operated through the interface is executed only when used with appropriate Spectrex host software. Toxic Open Path Gas Detector SafEye 950 & 960 SafEye 950 & 960 Features & Functions

32 Contents 32 | 87 1Principle of Gas Detection 2SafEye 950 & 960 Models and Types 3SafEye 950 & 960 Benefits 4SafEye 950 & 960 Technical Specifications 5SafEye 950 & 960 Features & Functions 6Accessories 7Part Numbers 8Installation & Commissioning 9Testing & Maintenance 10Applications

33 33 | 87 Accessories – Tilt Mounting / Alignment bracket  All SafEye open path gas detector systems need to be supplied with 2 stainless steel, Tilt Mounting / Alignment brackets to allow easy installation and direction of the detector to the required line of sight. Tilt mount SafEye 9xx Series (SS), p/n 888270 Toxic Open Path Gas Detector SafEye 950 & 960 Accessories

34 34 | 87 Accessories:  Weather shield / cover: Protects the detector from extreme weather conditions, such as heavy snow and heavy rain. Material: Stainless Steel. SafEye 9xx Sunshade (SS), p/n 888263  U-Bolt / Pole Mount: The U-bolt mount is available to facilitate 5” pipe mounting. Material: Stainless steel Pole Mount (U-Bolt) 5" - (SS), p/n 799225 Toxic Open Path Gas Detector SafEye 950 & 960 Accessories

35 35 | 87  Commissioning Kit: The Commissioning/Alignment Kit is required for commissioning and maintenance checks. Only one kit is required per site. The kit includes: 1 pc - Alignment Telescope 1 pc - Magnetic Mode Selector 1 pc - Function Check Filter 1 set - Socket keys for access to units 1 pc – USB / RS485 Harness converter kit (non-Ex approved) for PC / laptop USB port, incl software. Connects via RS485 terminals in safe area Commissioning Kit SafEye 950 (H2S), p/n 888847-1 Commissioning Kit SafEye 960 (NH3), p/n 888847-2 Toxic Open Path Gas Detector SafEye 950 & 960 Accessories

36 36 | 87 Commissioning, maintenance and diagnostics tools for the Quasar 9xx Series, which provides verification, status and instructions for changing detector parameters. Includes cable harness with RS485 / USB converter.  RS485 Hand-Held Diagnostic Kit (Ex): Certified I.S. (Ex ia) for use in the hazardous area and connects to I.S. port on 950/960. RS485 Hand-Held Diagnostic Kit (Ex), p/n 799810  HART Hand-Held Diagnostic Kit (Ex): Certified I.S. (Ex ia) for use in the hazardous area and connects to I.S. port on 950/960 HART Hand-Held Diagnostic Kit (Ex), p/n 888810  HART Hand-Held Harness Kit: For standard HART Hand-Held (I.S.) to connect between the Hand-Held and the I.S. Port on 900, including a harness HART Hand-Held Harness Kit, p/n??? Toxic Open Path Gas Detector SafEye 950 & 960 Accessories

37 37 | 87  Mini Laptop Diagnostic Kit (Non-Ex): preloaded with Spectrex software. For use in Safe area only. Connects, for convenience, to I.S port on 950 or RS 485 terminals. Mini Laptop Diagnostic Kit (Non-Ex), p/n 777820  USB / RS 485 Converter Kit: This kit with RS485/USB converter, is used with Spectrex Host software, enables the user to connect to any available PC or laptop to re-configure settings or perform diagnostics on Quasar 900 series. For use in Safe area only. Connects, for convenience, to I.S port on 950 or RS 485 terminals USB / RS 485 Converter Kit, incl. SW, p/n 794079-6 (included in Commissioning kit Toxic Open Path Gas Detector SafEye 950 & 960 Accessories Accessories: Commissioning, maintenance and diagnostics tools for the Quasar 9xx Series, which provides verification, status and instructions for changing detector parameters. Includes cable harness with RS485 / USB converter.

38 Contents 38 | 87 1Principle of Gas Detection 2SafEye 950 & 960 Models and Types 3SafEye 950 & 960 Benefits 4SafEye 950 & 960 Technical Specifications 5SafEye 950 & 960 Features & Functions 6Accessories 7Part Numbers 8Installation & Commissioning 9Testing & Maintenance 10Applications

39 39 | 87 SafEye 950 & 960 Series Part No. definition : Part code example: QRU-C-112  QRUReceiver  1H2S gas detector  1Standard  2¾ NPT cable entry Part code example: QTU-C-211  QTUTransmitter / Source  2Medium Range 15-40m  1Standard  1M25 cable entry Toxic Open Path Gas Detector SafEye 950 & 960 Part Numbers Receiver Transmitter

40 Contents 40 | 87 1Principle of Gas Detection 2SafEye 950 & 960 Models and Types 3SafEye 950 & 960 Benefits 4SafEye 950 & 960 Technical Specifications 5SafEye 950 & 960 Features & Functions 6Accessories 7Part Numbers 8Installation & Commissioning 9Testing & Maintenance 10Applications

41 Installation Guidance: The Source (Tx) and Detector (Rx) The appropriate Detector should be selected for the length of open path to be monitored. To allow for ageing of the Source and a reduction of the UV signal due to adverse weather, we recommend using a Detector that is not at the limit of its operating range. The open path between the Source and Detector, and the immediate surroundings, should be kept clear of obscuration that might hinder the free movement of air in the protected area or block the infrared beam. Guidance Tips for Gas Detector Locations The following are guidance tips for gas detector locations, in order to provide the best detection coverage:  For heavier than air gases: below potential leak source  For lighter than air gases: above potential leak sources  Along the expected leak trajectory: near to leak sources, taking into account prevailing wind directions  Between leak source and potential ignition sources  In areas with expected heavy fog, rain or snow, consider the effect of long range installation and install the Detector at a shorter range, with the maximum intensity model available. Toxic Open Path Gas Detector SafEye 950 & 960 Installation & Commissioning

42 Gas heavier than air: dense gas Gas lighter than air: buoyant gas Hydrogen Methane Ammonia Acetylene Carbon dioxide Propane Chlorine Benzene Phosgene liquefied gas Gas jet Elevated / pressurised release Gas behaving indifferent Gas behaving like dense gas Toxic Open Path Gas Detector SafEye 950 & 960 Installation & Commissioning

43 SafEye 950 & 960 – Installation Guidance: Guidance tips for gas detector locations in order to provide the best detection coverage:  The appropriate detector should be selected for the length of open path to be monitored.  To allow for ageing of the source and a reduction of the UV signal due to adverse weather, we do not recommend using a detector at the limit of its operating range.  As a guide, select a detector that is installed at a distance from the source of not more than 75% of the specified operating distance.  In severe weather conditions, such as offshore oil production and exploration, this should be reduced to 50%.  The open path between the source and detector and the immediate surroundings should be kept clear of obscuration that might hinder the free movement of air in the protected area or block the infrared beam. Note: Installation shall be carried out by suitably trained personnel in accordance with the applicable code of practice such as. EN 60079-14 and EN 60079-29-2. Toxic Open Path Gas Detector SafEye 950 & 960 Installation & Commissioning

44 44 | 87 SafEye 950 & 960 – Installation Guidance: Certification Instructions Warning: Do not open the Detector, even when isolated, in a flammable atmosphere. Use the following certification instructions:  The cable entry point may exceed 182°F (83°C). Suitable precautions should be taken when selecting the cable.  The equipment may be used with flammable gases and vapors with apparatus groups IIA and IIB +H2 T4 in the ambient temperature range -67°F (-55°C) to +149°F (+65°C).  Installation should be carried out by suitably trained personnel, in accordance with the applicable code of practice e.g. EN 60079-14.  Inspection and maintenance of this equipment should be carried out by suitably trained personnel, in accordance with the applicable code of practice e.g. EN 60079-17. Toxic Open Path Gas Detector SafEye 950 & 960 Installation & Commissioning

45 Wiring: For wiring, use color-coded conductors or suitable wire markings or labels. The wire cross-section must be between 1 mm² to 2.5 mm² (18- 14 AWG). The selected wire gauge should be based on the number of Detectors used on the same loop, and the distance from the control unit. The maximum number of wire connections in one terminal is two (2) wire cross-sections, each of 1 mm². Conduit/Cable Installation The conduit and cable installation must comply with the following guidelines:  To avoid water condensation in the Detector, install it with the conduits/cable entries facing downwards.  Use flexible conduits/cables for the last portion that connects to the Detector.  When pulling the cables through the conduits, ensure that they are not tangled or stressed. Extend the cables about 12 inches (30 cm) beyond the Detector location to accommodate wiring after installation.  After the conductor cables have been pulled through the conduits, perform a continuity test. To fully comply with EMC directive and protect against interference caused by RFI and EMI, the cable to the detector must be shielded and the detector must be grounded. The shield should be grounded at the detector end. Toxic Open Path Gas Detector SafEye 950 & 960 Installation & Commissioning

46 Detector & Tilt Mount Assembly : Toxic Open Path Gas Detector SafEye 950 & 960 Installation & Commissioning No. Description 1Tilt Mount Holding Plate 2Detector/Source Holding Plate 3Horizontal Crude Alignment Tightening Screw 4Horizontal Fine Alignment Tightening Screw 5Vertical Fine Alignment Tightening Screw 6Vertical Crude Alignment Tightening Screw 7Vertical Find Alignment Screw 8Horizontal Fine Alignment Screw 9Detector Tightening Screw 10Detector Tightening Washer 11Detector 12Telescope 13Telescope Tightening Bolt 14Detector Back Cover 15Detector Back Cover Secure Bolt

47 47 | 87 SafEye 950 & 960 – Tilt Mount Assembly : Toxic Open Path Gas Detector SafEye 950 & 960 Installation & Commissioning No. Description 1Tilt Mount Holding Plate 2Detector/Source Holding Plate 3Horizontal Crude Alignment Tightening Screw 4Horizontal Fine Alignment Tightening Screw 5Vertical Fine Alignment Tightening Screw 6Vertical Crude Alignment Tightening Screw 7Vertical Find Alignment Screw 8Horizontal Fine Alignment Screw

48 48 | 87 Detector / Source Installation The Detector and the Source can be installed in two (2) ways with the same Tilt Mount. To install the Detector / Source: 1.Place the Tilt Mount holding plate in its designated location and secure it with four (4) fasteners through four (4) holes of diameter 8.5 mm. Notes:  Skip this step if the Tilt Mount is already installed.  Detector removal for maintenance purposes does not require Tilt Mount removal. 2.Place the Detector, with its conduit/cable inlets pointing downwards on the Detector holding plate of the Tilt Mount. Secure the Detector with a M10 x 1.5 screw with No. M10 spring washers (9, 10). Secure the Detector to the Tilt Mount using Hex Key No. 7 for M10 x 1.5 screws (item 9). 3.Repeat steps 1 – 2 to install the Source. Toxic Open Path Gas Detector SafEye 950 & 960 Installation & Commissioning

49 49 | 87 SafEye 950 & 960 – Installation : Connecting the Detector 1.Release the rear cover securing bolt, and open the Detector rear cover. The terminal chamber is now exposed. 2.Remove the protective plug mounted on the Detector conduit/cable entry inlet, and pull the wires through the Detector inlet. Use a 3/4” – 14 NPT or M25x1.5 explosion-proof conduit connection/cable gland to assemble the cable/explosion-proof conduit to the Detector. 3.Connect the wires to the required terminals according to the wiring diagram. See Detector Terminal Wiring. 4.Connect the grounding wire to the ground screw outside the Detector. The Detector must be well grounded to Earth Ground. 5.Place and secure the Detector rear cover by screwing the cover and securing it using the secure bolt. Toxic Open Path Gas Detector SafEye 950 & 960 Installation & Commissioning

50 50 | 87 SafEye 950 & 960 – Wiring Options: UV Source (Tx) Wiring Toxic Open Path Gas Detector SafEye 950 & 960 Installation & Commissioning Terminal No.Function 1Power +24VDC 2Return -24VDC 3 Spare 4 5 RS-485 (+) 6RS-485 (-) No.ComponentNo.Component 1Housing / Enclosure5 Internal Earth Connection 2Terminal Board6N/A 3 Earth Terminal 7Source Holding Plate 4 Inlet Conduit

51 51 | 87 SafEye 950 & 960 – Wiring Options : Detector (Rx) Wiring Toxic Open Path Gas Detector SafEye 950 & 960 Installation & Commissioning Terminal No. Function 1Power +24VDC 2Return -24VDC 30-20 mA (Input) 4 0-20 mA (Output) 5 RS-485 (+) 6RS-485 (-) No.ComponentNo.Component 1Housing / Enclosure5 Internal Earth Connection 2Terminal Board6 Connection to Handheld Unit 3 Earth Terminal 7Source Holding Plate 4 Inlet Conduit

52 52 | 87 SafEye 950 & 960 – Wiring Options: Toxic Open Path Gas Detector SafEye 950 & 960 Installation & Commissioning Detector terminals / wiringSource terminals / wiring

53 53 | 87 SafEye 950 & 960 – Wiring Options: Toxic Open Path Gas Detector SafEye 950 & 960 Installation & Commissioning 0-20 mA Sink 4-Wire0-20 mA non-isolated Sink 3-Wire

54 54 | 87 SafEye 950 & 960 – Wiring Options: Toxic Open Path Gas Detector SafEye 950 & 960 Installation & Commissioning 0-20 mA Sink 3-Wire

55 55 | 87 SafEye 950 & 960 – Wiring Options: RS-485 Networking Toxic Open Path Gas Detector SafEye 950 & 960 Installation & Commissioning By using the RS-485 network capability of the SafEye Quasar 950 & 960 detector and additional software, it is possible to connect up to 32 Detectors in an addressable system with four (4) wires only (2 for power and 2 for communication). Using repeaters, the number of Detectors can be much larger (32 Detectors for each repeater) up to 247 on the same four (4) wires. When using the RS- 485 network, it is possible to read each Detector status (FAULT, WARNING, and ALARM).

56 56 | 87 SafEye 950 & 960 – Detector Alignment:  The telescope includes a periscope that consists of a prism and an ocular located vertical to the Telescope assembly.  Prior to telescope installation, the telescope and its mounting site need to be free from any dirt, to ensure proper alignment according to factory calibration. Toxic Open Path Gas Detector SafEye 950 & 960 Installation & Commissioning

57 SafEye 950 & 960 – Powering up the System: Important! Prior to any operation or maintenance, follow the Safety Precautions. ➣ To power up the system: 1.Make sure that the Source and Detector are connected to power. 2.Make sure that the 4-20 mA meter is connected to the Detector. 3.Power up the system using 18-32 VDC. 4.After 30 seconds, the current meter indicates 4 mA. Note: Perform Zero Calibration after powering up the system Safety Precautions: After power up, the Detector requires minimal attention for proper functioning, but the following should be noted: 1.Follow the instructions in the manual, and refer to the drawings and specifications issued by the manufacturer. 2.Do not open the Detector/Source housing while power is supplied. 3.Disconnect external devices, such as automatic extinguishing systems, before carrying out any maintenance task required by warranty.

58 58 | 87 SafEye 950 & 960 – Function Set-Up: The functions to set:  Full Scale Deflection / Measuring Range  Analog Output mode  Background Zero Calibration Mode  Address Setup (RS485)  Mode of Operation / Function Set-Up  Alarm Latch  Beam Block During Alarm Indication  Heated Optics Operation  Heat Mode  Heat on Temperature Toxic Open Path Gas Detector SafEye 950 & 960 Installation & Commissioning Factory default settings:  Full Scale Deflection – 500 ppm m  Analog Output mode - continuous  Background Zero Calibration Mode - Enabled  Address Setup (RS485) -1  Mode of Operation / Function Set-Up  Alarm Latch – Non Latching  Beam Block During Alarm Indication – No Latching  Heated Optics Operation  Heat Mode - Auto  Heat on Temperature - +5°C

59 59 | 87 SafEye 950 & 960 – Zero Calibration : Precise alignment must be performed prior to the zero calibration procedure. Zero calibration should be performed in good weather conditions with insignificant gas concentrations in the surrounding environment or indoors. Toxic Open Path Gas Detector SafEye 950 & 960 Installation & Commissioning 1.Switch from ‘Normal’ to ‘Alignment’ mode indication. 2.Switch from ‘Alignment’ to ‘Standby’ mode. 3.Switch from ‘Standby’ to ‘Zero Calibration’ mode.  The 0-20mA output should now be at 1mA.  Wait up to 60 seconds until it switches to ‘Normal’ mode.  The detector reading is now set to normal.  The 0-20mA output should now indicate 4mA. Magnet Switching from each position (first three instructions below) can be done by HOST HART or RS485, or by moving the magnetic mode selector above the magnetic switch (see sketch).

60 Functional Check : The Functional Check Filter is a convenient operational check used to confirm that the response has not changed from previous readings. The filter is not used for calibration, since it is not necessary in the procedure, nor does it equate to a particular quantity of gas. Toxic Open Path Gas Detector SafEye 950 & 960 Installation & Commissioning Notes: 1.This functional verification procedure is for a standard 0-20 mA output. 2.Prior to starting the Functional check, verify that the power to the units is on, and that the current of the 0-20 mA channel is stable. Record the reading. To perform the functional check: 1.Position the Functional Check Filter in front of the SafEye Detector. 2.Center the Functional Check Filter’s window over the viewing window of the Detector. 3.Wait 20 seconds. 4.Read the 0-20 mA current. Determine the difference between the reading taken with and without the Functional Check Filter. This difference is the 0-20 mA current variance. 5.Record the 0-20 mA current variance in the Maintenance Logbook. If the variance is more than a 30% change when compared to the previous check (see delivery form), repeat the alignment. Caution: Do not allow automatic activation, and disconnect any external device that should not be activated during the Calibration check.

61 Contents 61 | 87 1Principle of Gas Detection 2SafEye 950 & 960 Models and Types 3SafEye 950 & 960 Benefits 4SafEye 950 & 960 Technical Specifications 5SafEye 950 & 960 Features & Functions 6Accessories 7Part Numbers 8Installation & Commissioning 9Testing & Maintenance 10Applications

62 Maintenance & Troubleshooting General Maintenance SafEye Quasar 950/960 requires only simple periodic maintenance to provide satisfactory service and achieve maximum performance. The Detector and Source units can be maintained with the use of common tools and equipment. Record the periodic test results in a Maintenance Logbook with a copy of the delivery form inside. Note: Inspection and maintenance of this equipment shall be carried out by suitably trained personnel in accordance with the applicable code of practice such as EN 60079-17. Periodic Maintenance It is recommended that the Source and Detector viewing windows be kept as clean as possible, as it is an active device. Perform proper maintenance periodically to allow the SafEye system to retain maximum performance and reliability. To perform periodic maintenance:: 1.Perform alignment procedures each time that the Source or the Detector unit has been opened or moved for any reason. 2.The Signal Verification Check corroborates the current signals from the UV Source compared to that of previous alignments. It is recommended to perform this check every 6-12 months. The signal should be checked according to threshold levels. 3.Perform the Functional Check every 6 months 4.Perform the alignment procedure only if the signals are below threshold value 5.Perform Zero Calibration every time the Detector or Source is realigned or the windows are cleaned. Note: The frequency of cleaning operations is ultimately dependent upon the existing environmental conditions and the application. Toxic Open Path Gas Detector SafEye 950 & 960 Testing & Maintenance

63 63 | 87 Routine Optical Surface Cleaning To clean the optical window, proceed as follows 1. Turn off the power to the SafEye detector and source. 2. Where dust or dirt has accumulated, clean the optical surface with a small, soft-bristle brush. 3. The surfaces must then be washed thoroughly with water and a mild non-abrasive detergent. 4. Thoroughly rinse the glass surface with clean water; ensuring no residue is left behind. 5. Dry the glass with a clean dry soft cloth. 6. Enter in the maintenance logbook: date, name of person and company who performed the maintenance service. 7. Turn on power to the SafEye detector and source. 8. Perform signal verification. 9. Perform zero calibration. 10. Perform functional check. Note: Maintenance should be performed by suitably qualified personnel, who are familiar with local codes and practice. Maintenance requires ordinary tools. Toxic Open Path Gas Detector SafEye 950 & 960 Testing & Maintenance

64 64 | 87 Maintenance & Troubleshooting Signal Verification The Signal Verification check determines the proper operation of the open path. It checks the alignment and cleanliness of the window or any problem in Source or Detector. Use the PC Host software to measure the signal verification. Functional Check of Unit The SafEye Quasar has been calibrated at the factory per the user’s specific gas or vapor detection requirements. This procedure validates the functional operation. The functional check should be performed periodically. To perform the functional check: 1.Position the Function Check Filter in front of the SafEye Detector. 2.Center the Function Check Filter’s window over the viewing window of the Detector. 3.Wait 20 seconds. 4.Read the 0-20 mA current. Determine the difference between the reading taken with and without the Function Check Filter. This difference is the 0-20 mA current variance. 5.Record the 0-20 mA current variance in a Maintenance Logbook. If the variance is more than a 30% change when compared to the previous check (see delivery form), repeat the alignment. Caution: Do not allow automatic activation, and disconnect any external device that should not be activated during the Calibration check. Toxic Open Path Gas Detector SafEye 950 & 960 Testing & Maintenance

65 ProblemCauseCorrective Action “ Maintenance call” status or R, S1 and S2 are below 2.5V DC at Gain 4 Poor alignmentPerform alignment Dirt on the windowClean the window Poor light sourceReplace the light source Detector faultReplace/repair Detector NQRat1 and NQRat2 below the permitted limit Gas in the pathMake sure that the path is clean and the weather conditions are good NQRat1 and NQRat2 above the permitted limit Poor alignmentPerform alignment R/N, S1/N, S2/N below 50 Poor alignmentPerform alignment Dirt on the windowClean the window Poor light sourceReplace the light source Detector faultReplace/repair Detector Temperature higher than 25° C beyond the room temperature Electronic problem Replace/repair Detector Electronic problemTemperature higher than 25° C beyond the room temperature Ratio1 and Ratio2 out of the limit Poor alignmentPerform alignment Dirt on the windowClean the window Detector faultReplace/repair detecor Check the power supply and installation Low input voltageVoltage less than 16V DC. The Detector at “V” fault Toxic Open Path Gas Detector SafEye 950 & 960 Testing & Maintenance

66 66 | 87 Toxic Open Path Gas Detector SafEye 950 & 960 Testing & Maintenance Maintenance Host through RS485 - General Description The Main window is divided into main areas: Device / Model Information Detector Measurement Data Indication of Device and Communication Status Toolbar: Enables access to various control and diagnostic functions.

67 67 | 87 Toxic Open Path Gas Detector SafEye 950 & 960 Testing & Maintenance Signal Verification ChannelShort RangeMedium RangeLong RangeMaintenance 5 m20 m15 m40 m35 m80 m Reference2V Gain 0 1.5 V Gain 2 2V Gain 0 1V Gain 1 2V Gain 1 1V Gain 2 The minimum signal allowed is 2V at Gain 4 Signal 1&22V Gain 0 1.5 V Gain 2 2V Gain 0 1V Gain 1 2V Gain 1 1V Gain 2 The minimum signal allowed is 2V at Gain 4 Ratio 1&20.6 – 1.40.5 - 3 NQRat 1&20.97 – 1.03Must be 0.97 – 1.03 ppm x m0 ppm x m TempUp to 25°C beyond room temp. Voltage32 VDC > V > 18VDC

68 Toxic Open Path Gas Detector SafEye 950 & 960 Testing & Maintenance Detector Status StatusDescription0-20mA Output SSU90Startup 0 mA SSU91Return from parameter restoration SSU92Return from low voltage VVV81Wrong 5V 0 mA VVV82Wrong Vdd VVV83Wrong Vin PPP71Program memory CRC failure 0 mA PPP72Parameters memory CRC failure PPP73Setup memory CRC failure PPP74RAM CRC failure FFF34Noise or wrong offset in Sensor 1 0 mA FFF35Noise or wrong offset in Ref FFF36Noise or wrong offset in Sensor 2 FFF40Wrong 12V FFF41Shortcut between channels NNN34-20mA circuit failure0 mA XXX0Alignment 1 mA SSU0Start-up / stand by GGG0Zero Calibration OON42Wrong Flash Rate2 mA OON0Obscuration2 mA RON0Signal Obscuration2 mA KON0Absorption (low signal)2 mA

69 Toxic Open Path Gas Detector SafEye 950 & 960 Testing & Maintenance Detector Status StatusDescription0-20mA Output PON0Q0 obscuration2 mA ION0Saturation2 mA YON0Obscuration delay2 mA CON0Obscure C (lot of calibration) – no gas detection2 mA MBN0Misalignment2 mA OMN0Maintenance call for signal3 mA RMN0Maintenance call for ratio3 mA NNN0Normal4 mA NNN7Heater Fail4 mA BBN0BLOCK4 mA ZBN0R-LEL (high rise of REF)4 mA EBN0Diff LEL4 mA HBN0Fast Obscuration4 mA -bN0Lot of wetting (block Att)4 mA TNN0Pre Alarm4 mA WWN0Warning14 mA AAN0Alarm20 mA JAN0Disturbance while alarm20 mA LAN0Alarm latching20 mA

70 70 | 87 Toxic Open Path Gas Detector SafEye 950 & 960 Testing & Maintenance Setup Table Window

71 71 | 87 Toxic Open Path Gas Detector SafEye 950 & 960 Testing & Maintenance Maintenance Window ParameterDescription Gain LevelElectronic amplification. Signal 1Voltage signal of sensor 1 after electronic amplification (depends on gain). ReferenceVoltage signal of reference after electronic amplification. Signal 2Voltage signal of sensor 2 after electronic amplification (depends on gain). Ratio 1Ratio between sensor 1 and reference. Ratio 2Ratio between sensor 2 and reference. NQRatio 1Ratio1 / QO1 (Normalized ratio1) Equals 1 when no gas is present and goes down when gas is introduced. NQRatio 2Ratio1 / QO2 (Normalized ratio2) Equals 1 when no gas is present and goes down when gas is introduced. PPM x MPart per million x meter TemperatureThe temperature inside the detector in °C. VoltageDetector voltage.

72 72 | 87 Toxic Open Path Gas Detector SafEye 950 & 960 Testing & Maintenance Maintenance Window ParameterDescription Ref/NoiseReference / Noise. Sig1/NoiseSignal1 / Noise. Sig2/NoiseSignal2 / Noise. StatusCurrent status Total Operating Time Detectors operating hours. F9 - Checks signals to noise. F10 - Cancels the F9 check and returns to normal operation Align – switch from 1.Normal to Alignment mode 2.Alignment to Standby 3.Standby to Zero Calibration

73 73 | 87 Toxic Open Path Gas Detector SafEye 950 & 960 Testing & Maintenance Logging Detector Events  You can use the computer with the WinHost software to log the events of the detector, for diagnostic and other purposes.  When you start logging, you set the log file period in minutes.  A line is subsequently written to the log whenever that number of minutes passes (for example, every two minutes) and whenever there is a change in the status of the detector.  Each line in the log notes the following information: - Detector serial number - Detector gain - Detector status - Date and time - Detector temperature - Detector data channels, Rat1, Rat2, NQR1, NQR2, LEL.

74 74 | 87 SafEye Root Menu When HART communication is established, the first menu displayed is the SafEye Root Menu: SafEye Root Menu 1. Manufacturer 2. Model 3. Dev. ID 4. Dev. Model 5. PV Snsr s/n 6. Tag 7. Long Tag 8. Descriptor 9. Message 10. Final asmbly num 11. Date 12. Status 13. Status Menu 14. Device Setup Menu 15. HART Output Menu 1ManufacturerSpectrex Inc.Read 2ModelSafEye 950 or 960Read 3Dev. ID Device Unique identification numberRead 4Dev. Model Device sensor type (see Annex B, Table B1: Gas Detection Sensor Type Codes). Read 5PV Snsr s/n Serial Number of this device.Read 6 Tag Text that is associated with the field device installation (16 chars). This text can be modified and used by the operator any way. R/W 7 Long Tag Text that is associated with the field device installation (32 chars). This text can be modified and used by the operator any way. R/W 8 Descriptor Text associated with the field device that can be used by the operator in any way. R/W 9 Message Text associated with the field device that can be used by the operator in any way. R/W 10 Final asmbly num A number that is used for identification purposes, and is associated with the overall field device. R/W 11 Date Any Date chosen by the operator to be used for any purpose. R/W 12 Status Indicates a field device condition see Annex ARead 13 Status Menu Current operating status and diagnostic information.Menu 14 Device Setup Menu Provides Access to various setup and configurationsMenu 15 HART Output Menu HART Specific VariablesMenu Toxic Open Path Gas Detector SafEye 950 & 960 Testing & Maintenance

75 SafEye Status Menu : 10-20 ValueThe actual analog output of the Fire Detection Channel.Read 2Gas Detection ValueThe Gas reading in PPM*MRead 3 Internal Temperature Actual integral temperature of the detector (in deg C).Read 4 Input VoltageActual detector's supply voltage.Read 5 Heater ControlIndicates if the heater control is manual or automatic.Read 6Heater StateIndicates the current state of the heater (ON or OFF).Read 7Operating TimeDisplays the total detector's operating time from last power up.Read 8StatusIndicates a field device condition see Annex ARead 9Detector's GainThe current amplifying level of the detectorRead 10Signal 1The reading of signal 1Read 11Signal 2The reading of signal 2Read 12Reference SignalThe reading of reference signalRead 13NQRatio 1Normalized Ratio 1.Read 14NQRatio 2Normalized Ratio 2.Read 15 Reference to Noise Ratio Calculation of the SNR on Reference signal. The Detector must be at SNR test mode. Read 16 Signal 1 to Noise Ratio Calculation of the SNR on signal 1. The Detector must be at SNR test mode.Read 17 Signal 2 to Noise Ratio Calculation of the SNR on signal 2. The Detector must be at SNR test mode.Read 18 Change detector's working mode. The working modes are-Normal mode, Alignment, Stand-By, Calibration. This button does the operation of the magnet. Button 19Enter/Exit SNR testEntering and exiting to/from SNR test mode.Button 20SNR Test Indicates if the detector currently at SNR test mode. The detector is not operational at SNR test mode. Read Toxic Open Path Gas Detector SafEye 950 & 960 Testing & Maintenance

76 SafEye Device Setup Menu: 1ManufacturerSpectrex Inc.Read 2ModelSafEye 950 or 960Read 3 Device Model Device sensor type (see Annex B, Table B1: Gas Detection Sensor Type Codes). Read 4 Dev IDDevice unique identification number.Read 5 Config. Changed Counter The number of Device Configuration changes since the manufacture.Read 6Enable Alarm LatchEnable alarm relay to stay energized after exiting the Alarm Mode.R/W 70-20 Output Mode0-20mA continuous or discrete values for Normal, Warning, Alarm R/W 8 Beam Block during Alarm Latching at Alarm mode in case of Beam Block during Alarm R/W 9Flash RateThe actual Flash RateRead 10 Background Zero Calibration Enable the detector to perform constant zero calibration to adapt to external conditions. R/W 11Full Scale Sensitivity2 modes of full scale available – 500ppm.m or 200ppm.m R/W 12Gas TypeGas type that the detector should detect.R/W 13Heater Mode3 Heater Modes available- Auto, On, Off.R/W 14Heater On Temp. This option is active in Heater Auto mode – it allows the operator to set the temperature at which the optics heater begins operating. R/W 15Poll AddressAddress used by the host to identify a field device.R/W 16Loop Current ModeEnables choice between fixed or not fixed current mode.R/W Toxic Open Path Gas Detector SafEye 950 & 960 Testing & Maintenance

77 Contents 77 | 87 1Principle of Gas Detection 2SafEye 950 & 960 Models and Types 3SafEye 950 & 960 Benefits 4SafEye 950 & 960 Technical Specifications 5SafEye 950 & 960 Features & Functions 6Accessories 7Part Numbers 8Installation & Commissioning 9Testing & Maintenance 10Applications

78 78 | 87 Oil & Gas – H 2 S:  Exploration & Extraction: Off-/ onshore oil & gas platforms and rigs, FPSO’s & FLNG*  Gas Processing: LNG/LPG facilities, Gas sweetening systems,…  Production & Refining: chemical and petro-chemical plants,… Toxic Open Path Gas Detector SafEye 950 & 960 Applications *Floating Production Storage and Offloading Unit, Floating Liquefied Natural Gas

79 79 | 87 Natural-gas processing plants purify raw natural gas extracted from underground gas fields or at the surface from fluids produced by oil wells. Raw natural gas comes primarily from any one of three types of wells: crude oil wells, gas wells, and condensate wells. Natural gas is termed sweet gas when relatively free of hydrogen sulfide. however, Gas contain hydrogen sulfide is called sour gas. Natural gas, or any other gas mixture, containing significant quantities of hydrogen sulfide, carbon dioxide, or similar acidic gases is termed acid gas. Oil & Gas Processing: Toxic Open Path Gas Detector SafEye 950 & 960 Applications

80 80 | 87 HydrocarbonsCxHyCxHy 70-90% Carbon DioxideCO 2 0-8% NitrogenN2N2 0-5% Hydrogen SulfideH2SH2S0-5% Water, Oxygen, etc.1% HydrocarbonsCxHyCxHy 74% Carbon DioxideCO 2 9% Hydrogen SulfideH2SH2S15% Water, Oxygen, Nitrogen etc.1% Typical Composition of Sour Gas: H 2 S – TWA*: AGS 5 ppm = 0,0005 Vol% OSHA 4 ppm = 0,0004 Vol% *8 hour time-weighted average (TWA) Concentrations over 1000 ppm cause immediate collapse with loss of breathing, even after inhalation of a single breath. Hydrogen Sulfide: CAS: 7783-06-4 Typical Composition of Natural Gas: Toxic Open Path Gas Detector SafEye 950 & 960 Applications

81 81 | 87 Germany, Großenkneten Oman Kazakhstan, Tengiz H 2 S Detection within the Oil & Gas Segment: *natural gas production in cubic meters per year Toxic Open Path Gas Detector SafEye 950 & 960 Applications

82 82 | 87 App. 700 m distance Germany, Großenkneten Natural Gas Processing Plant Amount of : Natural Gas 6 Mrd. m 3 / a Sweet Gas 4,8 Mrd. m 3 / a Sulfur 800.000 t / a Fence line monitoring & point gas detectors Toxic Open Path Gas Detector SafEye 950 & 960 Applications

83 83 | 87 App. 15 km distance Drilling area Processing Plant & Refinery Sulfur storage Kazakhstan, Tengiz field Toxic Open Path Gas Detector SafEye 950 & 960 Applications

84 84 | 87 Flares 2 nd Processing Plant Fence Line for H2S Detectors Kazakhstan, Tengiz field Toxic Open Path Gas Detector SafEye 950 & 960 Applications

85 85 | 87 Flammability: PTB: LEL 15 vol% UEL 34 vol% NIOSH:LEL 15 vol% UEL 28 vol% Threshold limit values – TLV: EU:LTV 20 ppm (0.002 vol%) STEL: 50 ppm (0.005 vol%) NIOSH: LTV 25 ppm (0.0025 vol%) STEL: 35 ppm (0.0035 vol%) LTV / TWA – Long time value – 8 h time weighted average STEL - Short time exposure limit – 15 min average Ammonia: CAS: 7664-41-7 Ammonia NH 3 : Ammonia is a colorless gas with a characteristic pungent smell. It is lighter than air, its density being 0.589 times that of air. It is easily liquefied due to the strong hydrogen bonding between molecules; the liquid boils at −33.3 °C (−27.94 °F), and freezes at −77.7 °C (−107.86 °F) to white crystals. Combustion - when mixed with oxygen, it burns with a pale yellowish-green flame. At high temperature and in the presence of a suitable catalyst, ammonia is decomposed into its constituent elements. Ignition occurs when chlorine is passed into ammonia, forming nitrogen and hydrogen chloride; if chlorine is present in excess, then the highly explosive nitrogen trichloride (NCl 3 ) is also formed. Toxic Open Path Gas Detector SafEye 950 & 960 Applications

86 86 | 87 Chemical Industry and Food & Beverage – NH 3 :  Production: Fertilizer Production, Pulp & Paper, Process Areas,…  Storage & Handling: Refrigerant Gas Storage, Room monitoring,… Toxic Open Path Gas Detector SafEye 950 & 960 Applications

87 Thank you for your attention.


Download ppt "Quasar 950 & 960 Series Toxic Gas Open Path Detectors."

Similar presentations


Ads by Google