Ultrasonic imaging parameters ~Attenuation coefficient Advisor: Pai-Chi Li Student: Mei-Ru Yang Wei-Ning Lee.

Slides:



Advertisements
Similar presentations
Current techniques for measuring
Advertisements

Foundations of Medical Ultrasonic Imaging
Ultrasound Microscopy and High Frequency Coded Signals Antti Meriläinen, Edward Hæggström.
11.7 Properties of Waves – see also lecture notes 11.7
Pegasus Lectures, Inc. COPYRIGHT 2006 Volume I Companion Presentation Frank R. Miele Pegasus Lectures, Inc. Ultrasound Physics & Instrumentation 4 th Edition.
Ultrasonic Sound Properties. Examples of oscillation ball on a spring pendulum rotating earth.
Hospital Physics Group
Millimeter Wave Sensor: An Overview
EE 5340/7340, SMU Electrical Engineering Department, © Carlos E. Davila, Electrical Engineering Dept. Southern Methodist University slides can be.
ECE 501 Introduction to BME
Chapter 6: Data Transmission Business Data Communications, 4e.
Echoes. Reflection  The reflection of sound waves applies to frequencies above human hearing. 20 kHz is human maximum20 kHz is human maximum Bats can.
Ultrasound Medical Imaging Imaging Science Fundamentals.
Multipath fading and reflections The signal takes many paths to the destination. The propagation delay along each path is different. How many meters difference.
Chapter 15: Data Transmission Business Data Communications, 5e.
Wavelet Spectral Finite Elements for Wave Propagation in Composite Plates with Damages Ratneshwar Jha, Clarkson University S. Gopalakrishnan, Indian Institute.
High Frequency Ultrasonic Characterization of Carrot Tissue Christopher Vick Advisor: Dr. Navalgund Rao Center for Imaging Science Rochester Institute.
Sound Transmission and Echolocation Sound transmission –Sound properties –Attenuation Echolocation –Decoding information from echos.
EELE 5490, Fall, 2009 Wireless Communications
Module 3.0: Data Transmission
Ultrasound Imaging Atam Dhawan.
Ultrasound Physics Sound is a mechanical, longitudinal wave that travels in a straight line Sound requires a medium through which to travel Ultrasound.
Scattering and Attenuation Seismology and the Earth’s Deep Interior Scattering and Attenuation Propagating seismic waves loose energy due to geometrical.
Parameters, pulse parameters, attenuation, and resolution.
Review Doppler Radar (Fig. 3.1) A simplified block diagram 10/29-11/11/2013METR
Objective To study the effect of sub surface defects in surface roughness monitoring through ultrasonic flaw detector. To study the sizing of defects.
Ultrasound measurements on tissue Penny Probert Smith Institute of Biomedical Engineering Department of Engineering Science University of Oxford (also.
Review: The application layer. –Network Applications see the network as the abstract provided by the transport layer: Logical full mesh among network end-points.
Doppler echocardiography & Magnetic Resonance Imaging.
1 Business Telecommunications Data and Computer Communications Chapter 3 Data Transmission.
Introduction to Adaptive Digital Filters Algorithms
CE 4228 Data Communications and Networking
Characteristics Radio Frequency signals consist of the following: Polarity Wavelength Frequency Amplitude Phase These characteristics are defined by the.
Resident Categorical Course
1 Ultrasonic Elasticity Imaging. 2 Elasticity Imaging Image contrast is based on tissue elasticity (typically Young’s modulus or shear modulus).
Ultrasonic imaging parameters ~Attenuation coefficient Advisor: Pai-Chi Li Student: Mei-Ru Yang Wei-Ning Lee.
Trasmission lines.
Acoustic Emission Wave Propagation and Source Location Dr. Boris Muravin More at More about AE at
BMI I FS05 – Class 9 “Ultrasound Imaging” Slide 1 Biomedical Imaging I Class 9 – Ultrasound Imaging Doppler Ultrasonography; Image Reconstruction 11/09/05.
10/20/2015Copyright © 2008 Ballios, Dow, Vogtmann, Zofchak.
Ultrasonic Attenuation Tomography Based on Log-Spectrum Analysis Radovan Jiřík, Rainer Stotzka, Torfinn Taxt Brno University of Technology Department of.
Display of Motion & Doppler Ultrasound
Review of Ultrasonic Imaging
Adaphed from Rappaport’s Chapter 5
Communications Systems. 1Analogue modulation: time domain (waveforms), frequency domain (spectra), amplitude modulation (am), frequency modulation (fm),
Saudi Board of Radiology: Physics Refresher Course Kostas Chantziantoniou, MSc 2, DABR Head, Imaging Physics Section King Faisal Specialist Hospital &
Ultrasound Physics Reflections & Attenuation ‘97.
Ultrasonic Imaging Parameters Student: Mei-Ru Yang Wei-Ning Lee Advisor: Pai-Chi Li.
Review Doppler Radar (Fig. 3.1) A simplified block diagram 10/29-11/11/2013METR
University of Kurdistan Food Quality Evaluation Methods (FQEM) Lecturer: Kaveh Mollazade, Ph.D. Department of Biosystems Engineering, Faculty of Agriculture,
A 5-Pulse Sequence for Harmonic and Sub-Harmonic Imaging
Ultrasound Computed Tomography 何祚明 陳彥甫 2002/06/12.
Artifacts Ultrasound Physics George David, M.S.
Ultrasound Computed Tomography
 Ultrasound waves are longitudinal with high frequencies ( ≈ > 20,000 Hz, though medical Ultrasound is between 1 to 15 MHz.)  When an ultrasound reaches.
Signal Propagation Basics
© Fraunhofer HIGH RESOLUTION ULTRASOUND GONIOMETER Enabling versatile acoustic angular measurements.
Data and Computer Communications Chapter 3 – Data Transmission.
1 Opto-Acoustic Imaging 台大電機系李百祺. 2 Conventional Ultrasonic Imaging Spatial resolution is mainly determined by frequency. Fabrication of high frequency.
Imaging of propagating wave fronts resulting from ultrasonic pulses incident on heel bones using refracto-vibrometry Acoustical Society of America Meeting.
Lecture 2 Transmission Line Characteristics
Reflections & Attenuation
Radio Coverage Prediction in Picocell Indoor Networks
Ultrasound.
Ultrasound.
Resident Physics Lectures
Spatially Varying Frequency Compounding of Ultrasound Images
Date of download: 3/3/2018 Copyright © ASME. All rights reserved.
Sound Waves and Ultrasound
Ultrasonic Imaging Using Contrast Agents
Presentation transcript:

Ultrasonic imaging parameters ~Attenuation coefficient Advisor: Pai-Chi Li Student: Mei-Ru Yang Wei-Ning Lee

****Attenuation predicted by Biot’s theory (bone) ****Ultrasound attenuation in other tissues(breast dermis…) ***Ultrasound attenuation in bones (one transducers) ********** **** Ultrasound attenuation in bones (two transducers) Attenuation coefficient TotalProcessSystemApparatusMethodCircuitDevice Ultrasound attenuation in signal processing *********

引證族譜圖 Year EnterpriseVammenJorgensenLaugierBerger Paper 1 Paper 2 Paper 3 Wear Paper 4 Paper WSUMSSM Paper 6 Paper 7 Paper Yale U.Irvine Paper

Methods and Apparatus One transducer (pulse-echo mode) –Spectral - Shift –Log Spectral – Difference –Backscattering Two transducers(one transmitter and one receiver)

One Transducer Simplify the handling and avoid potential errors to misalignment or mismatching Differentiate the attenuation coefficients of adjacent tissues Problem : renewed reflection –The delay should preferably by adapted to space any mirror images of pronounced spikes away from the region to be analysis

Cont’d Log amplitude(dB) distance 1 MHz 5 MHz T S C T C B C

T Soft tissue Hard tissue Interpose a delay line (50mm) T delay line

Two Transducers Through-transmission method A pair of transducers coaxially in a water tank Submerge a sample between the transducers Signals were recorded both with and without the specimen in the acoustic path Can’t separate the contribution caused by effects in the soft tissue from the contributions caused by effects in the bone tissue

Apparatus t 1, t 2 : pulse-echo mode TR Water tank specimen

Spectral Shift Technique : transmitting pulse power spectrum : receiving pulse power spectrum T: time constant

Log Spectral Difference Technique

Cont’d Attenuation obeys Use amplitude and phase information of the pulse Ultrasound reflection at the water-specimen interface  produce error A w (f) : amplitude spectrum with water path only A s (f) : amplitude spectrum with the specimen inserted L : specimen thickness

Cont’d Given “n”, calculate attenuation coefficient β Minimize

Diffraction Correction Diffraction effect on attenuation estimate due to the two media with different ultrasound velocities Experimental Diffraction Correction technique –Using the spectrum of the reference media, i.e. water Ad(f, z) : diffraction magnitude transfer function

Cont’d Fresnel parameter : S=Sa=zλ a /a 2 ensures that the water-specimen-water and water- only paths undergo equivalent diffraction effects Then, : wavelength in water : wavelength in specimen

Difficulties Theoretical value –Biot’s theory (for bone) –Ultrasound properties of mammalian tissues –Numerous parameters are required for the computation (bulk modulus, shear modulus, transmission coefficient, etc.) Three phantoms –Actual attenuation coefficients are unknown

Strategies for Attenuation Estimate Qualitative analysis –Use various specimens with largely different hardness –Observe the attenuation trend –Relationship between attenuation and frequency Quantitative analysis –Dispersion & reflection –Diffraction correction

Experiment Transducers: 3.5 MHz 5 MHz 7.5 MHz Pulse receiver Oscilloscope A/D: GaGe fs=100 MHz LabView Phantom

Preliminary Results Transmitting mode –Signal type: Gaussian –4 cycles, continuous –PRI: 300μs –Output voltage: ± 5 V Receiving mode –Capture length: 5 μs –No. of PRI: 1 –Sampling rate: 50 MHz

Cont’d

References Paper1 : “Attenuation coefficient and speed of sound in immature and mature rat cartilage: a study in the MHz frequency range,” IEEE Ultra. Sym., Paper 2: “Prediction of ultrasound attenuation in cancellous bones using poroelasticity and scattering theories,” IEEE Ultra. Sym., Paper 3: “Ultrasonic wave propagation in cancellous and cortical bone: prediction of some experimental results by Biot’s theory,”J. Acous. Soc. Am.,Vol. 91, pp , Paper 4: “Frequency dependence of ultrasonic backscatter from human trabecular bone: Theory and experiment,” J. Acous. Soc. Am. Vol.106, pp , 1999 Paper 5: “Anisotropy of attenuation and backscatter in cancellous bone” IEEE Ultra. Sym., pp , 1999.

Cont’d Paper6: Ping He, “Acoustic parameter estimation based on attenuation and dispersion measurements,” Proc. IEEE/EMBS, Oct. 29, Nov. 1, 1998 Paper7: Wei Xu, Jonathan J. Kaufman, “Diffraction correction methods for insertion ultrasound attenuation estimation,” IEEE Trans. On Biomedical Engineering, vol. 40, No. 6, June 1993 Paper8: Roman Kuc, “Estimating acoustic attenuation from reflected ultrasound signals: comparison of spectral-shift and spectral-difference approaches,” IEEE Trans. On Acoustics, Speech, and Signal Processing, vol. ASSP-32, No. 1, Feb Paper9: Leeman, Ferrari, Jones, and Fink, “Perspectives on attenuation estimation from pulse-echo signals,” IEEE Trans. On Sonics and Ultrasonics, Vol. SU-31, No. 4, July 1984 Paper10: Parker, Lerner, and Waag, “Comparison of techniques for in vivo attenuation measurements,” IEEE Trans. On Biomedical Engineering,, vol. 36, No. 12, Dec 1988