Optimisation of Computed Radiography chest imaging utilising a digitally reconstructed radiograph simulation technique Craig Moore Radiation Physics Service.

Slides:



Advertisements
Similar presentations
DIGITAL RADIOGRAPHY.
Advertisements

Advanced Biomedical Imaging Lecture 3
Technique Guidance Systems
X Rays Medical Physics Notes.
Quantitative Quality Assurance in Hull & East Yorkshire Dr. Craig Moore & Dr. Tim Wood Radiation Physics Department.
Computers and Computed Tomography
Image Reconstruction.
Formation Characteristics
Improved Conspicuity of Abdominal Lesions with Single-Source Dual-Energy MDCT Hadassah Hebrew University Medical Center Jerusalem, Israel Ruth Eliahou.
Advanced Biomedical Imaging Lecture 4 Dr. Azza Helal A. Prof. of Medical Physics Faculty of Medicine Alexandria University.
Image Quality Bushong Chapter 7.
1 Low-Dose Dual-Energy CT for PET Attenuation Correction with Statistical Sinogram Restoration Joonki Noh, Jeffrey A. Fessler EECS Department, The University.
IMAGE QUALITY.
Computed Tomography Physics, Instrumentation, and Imaging
Conventional Radiography
Primary Exposure Factors IV
Topic 3 Selection of kV High kV scattered radiation is energetic. Angle of rays is same as those of primary beam and is not deflected very much. Follows.
RADIOLOGY. NEXT GENERATION SCIENCE / COMMON CORE STANDARDS ADDRESSED! CCSS.ELA-Literacy.RST Determine the central ideas or conclusions of a text;
Lower Dosage CT-guided Lung Biopsy Protocol Maintains Quality, Minimizes Exposure Jeremy Collins, MD Pegah E, Lewandowski RJ, Yaghmai V, Nemcek jr AA,
MAMMO QC – covered in week 8
5 th September 2005 Paul Collins Computed Tomography Dosimetry Assessment of Effective Dose in Computed Tomography using an Anthropomorphic Phantom Paul.
Advanced Biomedical Imaging
Dr. Mustafa Zuhair Mahmoud Mr.Ali B Alhailiy بسم الله الرحمن الرحيم 1.
Digital Image Manipulation I By Professor Stelmark.
IMAGE QUALITY NOISE LINEARITY CROSS-FIELD UNIFORMITY IMAGE ARTIFACTS.
Optical Density and Brightness
tomos = slice, graphein = to write
8.1 PRODUCTION AND CHARACTERISTICS OF X-RAYS
TERPSSC 2001Robert M. Gagne OVERVIEW OF DIGITAL IMAGING AND RADIATION PROTECTION ISSUES Robert M. Gagne MICAB/DECS/OST [From: Handbook.
Conventional and Computed Tomography
Seeram Chapter 11: Image Quality
Quality Control Rad T 110.
Digital Image Characteristic
Automated Image Analysis Software for Quality Assurance of a Radiotherapy CT Simulator Andrew J Reilly Imaging Physicist Oncology Physics Edinburgh Cancer.
Resident Physics Lectures Christensen, Chapter 6Filters George David Associate Professor Department of Radiology Medical College of Georgia.
Computers in Imaging Q & A Robert Metzger, Ph.D..
Quality Control.
Resident Physics Lectures
Image Quality Radiographic Resolution.
Experiences with radiographic simulation software, student evaluation and an overview of simulation strategies for learning: A work in progress Rob Stewart:
Factors affecting CT image RAD
Exposure Factors or Prime Factors
IAEA International Atomic Energy Agency Optimization of Protection in Computed Tomography (CT)-What can radiographers do? IAEA Regional Training Course.
Preparing Variable kVp Technique Charts By Prof. Stelmark.
PRINCIPLES OF TECHNIQUE AND EXPOSURE
Radiographic Quality Visibility and Sharpness
Image Quality Q&A. RAPHEX Diagnostic Question 2001 D9: The imaging system which is best for visualizing small high contrast objects is: A. Computed tomography.
Optimisation in digital radiography 9 October 2015 Radiation Dose Metrics Dr Tim Wood Hull and East Yorkshire Hospitals NHS Trust
The production of X-Rays X-Rays are produced whenever charged particles are slowed down or stopped. X-rays are generated via interactions of the accelerated.
Image Display. But first a review Remember the 3 main steps 1. Data Acquisition 2. Image Reconstruction 3. Image Display.
Part No...., Module No....Lesson No
X-ray SNR in 3 steps. I ∆I. X-ray transmission SNR Review Let N = average number of transmitted x-rays N = N 0 exp [ - ∫  dz ] Emission and transmission.
Stacy Kopso, M.Ed., R.T.(R)(M).  Xray photon loses energy and changes direction  Responsible for most of the scatter that fogs the image  Related to.
IAEA International Atomic Energy Agency General Radiography Radiation Sources in medicine diagnostic Radiology Day 7 – Lecture 1(1)
Image Quality and Patient Dose. X-ray tube design - basic principles Electrons generated by thermionic emission from a heated filament (cathode) Accelerating.
Tube Exposure Factors Math Technique Contrast and Density
Analog X-ray Imaging Recommended Book: Walter Huda, REVIEW OF RADIOLOGIC PHYSICS By: Maisa Alhassoun
Basis of Mammography F. Milano Dept. Clinical Physiopathology University of Florence
Computed Tomography Computed Tomography is the most significant development in radiology in the past 40 years. MRI and Ultrasound are also significant.
Physics Case of the Day History : Patient presented for a follow-up mammogram. The cranio-caudal view was suboptimal due to very low contrast and was repeated.
Imaging: Images were obtained over a range of tube potentials (81 to 125 kV) without scatter rejection, with a stationary anti- scatter grid and using.
Technique Guidance Systems By Prof. Stelmark. Anatomic Programming Anatomic programming, or anatomically programmed radiography (APR), refers to a radiographic.
Impact of Obesity on Medical Imaging and Image-Guided Intervention
Optimisation of Patient Protection for Radiography
Optimisation of Chest Computed Tomography Using a Phantom: Impact of mAs and Reconstruction Techniques on Image Quality C.S. Reis1; T. Faqir2; V. Harsaker3;
CT dose optimization software synergy – A clinical perspective
Hayley Milne & Andrew England
Resident Physics Lectures (Year 1)
Test tool for assessing lead equivalence in protective lead apparels
Roshan S Livingstone and Benedicta R Department of Radiology
Presentation transcript:

Optimisation of Computed Radiography chest imaging utilising a digitally reconstructed radiograph simulation technique Craig Moore Radiation Physics Service Hull & East Yorkshire Hospitals NHS Trust

Introduction - Literature Lots of publications have shown that patient anatomy is the limiting factor in the reading normal structures, and detection of lesions (lung nodules) in chest images –Bochud et al 1999 –Samei et al 1999, 2000 –Burgess et al 2001 –Huda et al 2004 –Keelan et al 2004 –Sund et al 2004 –Tingberg et al 2004 European wide RADIUS chest trial (2005)

Introduction - Literature Chest radiography is now generally considered to be limited by the projected anatomy Patient anatomy = anatomical noise So if we want to optimize digital system for chest imaging, vital that anatomical noise is present in the images!!!

Introduction However, the radiation dose/image quality relationship must not be ignored Doses must be kelp ALARP –ICRP 2007 –IR(ME)R2000 – (required legally in the UK) We would therefore want system (quantum) noise present in an image for dose reduction studies

Digitally Reconstructed Radiograph (DRR) Hypothesis: –Can use CT data of humans to provide realistic anatomy (anatomical noise) Clinically realistic computerized ‘phantom’ –Simulate the transport of x-rays through the ‘phantom’ and produce a digitally reconstructed radiograph (DRR – a simulation of a conventional 2D x-ray image created from CT data) –Add frequency dependent system noise post DRR calculation –Add radiation scatter post DRR calculation –Validate –Use for optimization studies

DRR Algorithm: Virtual Patient Virtual patient derived from chest portion of real CT datasets Voxel resolution = 0.34 x 0.34 x 0.8 mm CT number converted to linear attenuation coefficient (LAC) using tissue equivalent inserts –Measure mean CT No. in each insert –We know elemental composition of each so can derive LAC –Can derive relationship between CT No. and LAC

PA slice 1 PA slice N CT axial batch 1 (i.e. CT slices 1 to 20) CT axial batch 2 (i.e. CT slices 21 to 40) Final CT axial batch (i.e. slices 681 to 700) CT dataset re- orientated in the ‘PA’ direction X-ray spectra derived from IPEM 78 X-ray attenuated exponentially through CT dataset using a ray casting method of DRR calculation Energy absorbed in CR phosphor Intensity of X-rays exiting is calculated

Scatter Addition Must add to DRR generated image –DRR algorithm does not calculate scatter Measured scatter in CR chest images using lead pellet array Use chest portion of RANDO phantom

Noise Addition Uniform noise image Corrected noise image lung spine diaphragm Based on slightly modified work by Bath et al, 2005

Lung Nodule Simulation Added artificial nodules to the CT data prior to X- ray projection Baed on work by Li et al 2009

Human DRR v Human CR DRR CR

DRRs: 50 kVp v 150 kVp

Scatter Rejection The DRR algorithm can produce images with scatter rejection (b) (c) (a) DRR CR DRR No RejectionDRR GridCR Grid

Obese Patients DRR algorithm can also produce images of large/obese patients (a) (b) DRRCR

Validation Decided to validate with RANDO and real patient images: –Histogram of pixel values –Signal to noise ratios (SNR) Important because signal and noise affects the visualisation of pathology –Tissue to rib ratios (TRR) Pixel value ratio of soft tissue to that of rib Important as rib can distract the Radiologist from detecting pathology

Phantom Histograms a b c d

PATIENT - Histograms Typical histogram of average patient DRR Typical histogram of average patient CR image

SNRs Good agreement in lung, spine and diaphragm areas of chest Maximum deviation approx 15% –Mean deviation = 7% Addition of frequency dependent noise not perfect: –CR system noise is ergodic (changes with time) –Noise added here is a snapshot (and so not ergodic) –However, quantum noise dominates over ‘ergodic noise’ so not such an issue –DQE is NOT constant with dose variation in image

Validation - TRRs Good agreement –Within 2% As tube potential increases TRR decreases –Due to rib attenuating higher percentage in incident photons at lower potentials than soft tissue, thus forcing up TRR

Validation - Radiologists Have told me DRR images contain sufficient clinical data to allow diagnosis and subsequent optimisation They have scored the images out of 10 –‘are the images sufficiently like real CR images? –Average score of 7.8

Conclusions – DRR Algorithm DRR computer program has been produced that adequately simulates chest radiographs of average and obese patients –Anatomical noise simulated by real human CT data –System noise and scatter successfully added post DRR generation that provides: SNRs TRRs Histograms –in good agreement with those measured in real CR images Provides us with a tool that can be used by Radiologists to grade image quality with images derived with different x-ray system parameters WITHOUT THE NEED TO PERFORM REPEAT EXPOSURE ON PATIENTS ACCEPTED FOR PUBLICATION IN THE BJR –AUGUST 2011

Optimisation of CR Chest Radiography using DRR Generated Images In Hull chest exposure factors were not standardised (historical reasons!!!!) Three main hospital sites: –60 kVp & 10 mAs –70 kVp & 5 mAs –80 kVp & 5 mAs In the last 6 months, four expert image evaluators have scored DRR reconstructed images –Two Consultant Radiologists –Two Reporting Radiographers Scoring criteria based on European guidelines (CEC)

What to Optimise? Optimum tube potential for Average Patients (70 kg ± 10 kg) –Without scatter rejection (as per Hull protocol) –With scatter rejection (grid and air gap) Optimum tube potential for Obese patients –Without scatter rejection –With scatter rejection Is Scatter rejection indicated? Dose reduction?

Scoring Criteria Images scored on a dual PACS monitor system Image on right hand screen held at a constant kVp Images on left hand screen displayed from 50 to 150 kVp in steps of 10kVp (approx) –Image 1 = 50 kV –Image 2 = 60 kV –Image 10 = 150 kV Test images scored against reference image All images matched effective dose

GRADINGVISIBILITY OF STRUCTURE -3Definitely inferior to the reference image -2Reasonably inferior to the reference image Slightly inferior to the reference image 0Equal to the reference image +1Slightly better than the reference image +2Reasonably better than the reference image +3Definitely better than the reference image [1][1] European guidelines on quality criteria for diagnostic radiographic images. CEC European Commission EUR EN (Luxembourg 1996) STUCTURE – NORMAL ANATOMYGRADING IMAGE 1IMAGE 2 IMAGE 3IMAGE 4 Vessels seen approx approx 3 cm from the pleural margin Thoracic vertebrae behind the heart Retrocardiac vessels Pleural margin Vessels seen en face in the central area Hilar region STUCTURE – LUNG NODULES GRADING Nodule in lateral pulmonary region Nodule in hilar region Are the ribs a distraction? Y/N

Tube Potential Optimisation

Results – Average Sized Patients, no scatter rejection VGAS = average of Radiologists results

Results – Average Patients with scatter rejection - grid

Results – Average Patients with scatter rejection – air gap

Results – Obese Patients without scatter rejection

Very weak trend for better image quality with higher kVp Lower kVps probably ‘worse’ due to combination of: –Lack of penetration through obese patient –Increased scatter from obese patient (scatter to cassette changes very little with kVp) This is not so with average patients –Less tissue (fat) so more radiation penetration –Less scatter from fat It is likely that poorer radiation penetration and increased scatter from obese patient outweighs the inherent benefit of photoelectric contrast obtained from lower kVps

Results – Obese Patients with scatter rejection - grids

Results – Obese Patients with scatter rejection – air gap

Scatter Rejection

Scatter Rejection vs no scatter rejection – average patients

Superior image quality with scatter rejection technique Grids performed much better than air gap Statistically significant differences BUT: –Image evaluators were asked if increase in dose due to use of grids was justified, even with better image quality –Answer in 100% of cases was NO –SCATTER REJECTION FOR AVERAGE PATIENTS IS NOT INDICATED

Exposure time for average patients without scatter rejection and low tube potentials? Scatter rejection not indicated Therefore low kVps should be used (remember the graph) European guidance recommends exposure times < 20 ms Can we achieve this with low kVps?? Modern Philips X-ray generator: –For lgM = 2 (Agfa CR specific Dose Indicator) –With 630 mA, all kVps are possible –Max exp time = 16 ms At the expense of increased tube loading

Scatter Rejection vs no scatter rejection – obese patients

Superior image quality with scatter rejection technique Grids performed much better than air gap Statistically significant differences BUT: –Image evaluators were asked if increase in dose due to use of grids was justified, even with better image quality –Answer in 100% of cases was YES –ANTI SCATTER GRID USE FOR OBESE PATIENTS IS INDICATED

Exposure times for obese patients with an anti-scatter grid Remember that low tube potentials were superior with an anti-scatter grid!!!! Need exposure times < 20ms Is this possible with low kVps and scatter grid for lgM = 2?? Scatter grid is a focused grid so have to use it in a fixed range of FDD –Nominal distance = 140 cm FDD –Range allowed = 115cm – 180cm At 140 cm FDD lowest exp time = 109 kVp At 115cm FDD lowest exp time = kVp So are limited to 90 kVp

Dose Reduction

Dose Reduction? Images were also presented at different doses

Dose reduction? Results suggest doses can be reduced by around 50% before image quality begins to suffer and become unacceptable Therefore could half exposure mAs

Conclusions – Use of DRR algorithm to optimise CR chest imaging Average patients: –No scatter rejection is indicated –Therefore, low kVps (< 102 kVp) are indicated –Can have exposure times < 20 ms for all kVps Obese patients –Anti-scatter grid is indicated –So low kVps (<102kVp) should be used –For exposure times < 20ms, are limited to 90 kVp Doses –As low as 50% reduction possible ACCEPTED FOR PUBLICATION IN BJR

Optimisation & Standardisation in Hull? Agreed with Consultant Radiologists: –60 kVp & 10 mAs After a ‘settling in’ period: –60 kVp & 8 mAs –Want to go lower than 8mAs eventually!!! Implication on patient dose? –Using PCXMC effective dose calculation software: –80 kVp/5 mAs = mSv –60 kVp/8 mAs = mSv –Approx 45% drop in effective dose