1 Principle  LC-MS is interfacing HPLC system with mass spectrometer.  The difficulty in hyphenation is to transform the solute into gas phase ion. 

Slides:



Advertisements
Similar presentations
D e t e c t o r s f o r H P L C.
Advertisements

Liquid chromatography coupled with Mass spectroscopy
Background and Theory of Electrospray and Tandem Mass Spectrometry
Evolving Role of Mass Spectrometry in Bioanalytical Analysis Daniel Pentek February 1, UCONN Bioanalytical Chem 395.
FC-MS from Teledyne Isco CombiFlash ® a Name You Can Rely On.
Sanja Risticevic Chem 323 Poster Presentation Quadrupole Ion Trap Mass Spectrometry.
Andrew Rouff and Ben Berger
Mass spectroscopy. In a typical MS procedure:  1- a sample is loaded onto the MS instrument, and undergoes vaporization.  2- the components of the sample.
Mass Spectroscopy Skyline IB Chemistry HL Mass Spectroscopy.
Inductively Coupled Plasma Mass Spectrometry or ICP-MS
12-1 Molecular Mass Spectroscopy Molecular structure Composition of mixtures Molecular mass spectra Ion Source Mass Spectrometers Applications.
427 PHC.  Atomic emission spectroscopy (AES) is based upon emission of electromagnetic radiation by atoms.
Mass Spectrometry Inlet system Ion Source Mass Analyzer Ion Detector.
HPLC Coupled with Quadrupole Mass Spectrometry and Forensic Analysis of Cocaine.
Lecture 14 LC-MS Ionization. GC Computer MS GC-MS.
Mass Spectroscopy Mass Spectrometry ä Most useful tool for molecular structure determination if you can get it into gas phase ä Molecular weight of.
Lecture 8. GC/MS.
Chem. 133 – 4/23 Lecture.
Chem. 133 – 4/28 Lecture. Announcements Lab Report 2.3 due Today Pass back graded materials (lab reports 2.2, Q5, and AP3.1) Today’s Lecture Mass Spectrometry.
Chapter 20 Molecular Mass Spectrometry. Introduction... Mass spectroscopy is perhaps one of the most widely applicable of all the analytical tools available.
Mass Analyzers Double Focusing Magnetic Sector Quadrupole Mass Filter Quadrupole Ion Trap Linear Time-of-Flight (TOF) Reflectron TOF Fourier Transform.
Mass Analyzers Double Focusing Magnetic Sector Quadrupole Mass Filter Quadrupole Ion Trap Linear Time-of-Flight (TOF) Reflectron TOF Fourier Transform.
LC-MS Based Metabolomics. Analysing the METABOLOME 1.Metabolite Extraction 2.Metabolite detection (with or without separation) 3.Data analysis.
Molecular Mass Spectroscopy Chem Introduction In Mass Spectroscopy (MS), atomic and molecular weights are generally expressed in terms of atomic.
Molecular Mass Spectrometry
LC-MS Lecture 7.
Mass Spectroscopy Quantitative Chemical Analysis Harris, 6th Edition
LC/MS WORKSHOP IOWA STATE UNIVERSITY Kamel Harrata  Instrument Description  Data Acquisition  Data Processing.
Instant Notes Analytical Chemistry
Mass Spectroscopy 1Dr. Nikhat Siddiqi. Mass spectrometry is a powerful analytical technique that is used to identify unknown compounds, to quantify known.
6. Mass Spectrometry Adv. Inst. Techs. How does it work? a very small amount of sample is bombarded by a beam of high energy (usually electron beam) produces.
Russell Rouseff FOS 6355 Summer 2005 What is Mass Spectroscopy Analytical Chemistry Technique Used to identify and quantify unknown compounds Can also.
Gas Chromatography And Mass Spectrometry
Mass Spectrometer Mass spectrometere is a very powerful method to analyse the structure of organic compounds.
AUSTEN K. SCRUGGS CALIFORNIA STATE UNIVERSITY, FRESNO RIMI FACILITY INSTRUMENTATION.
Types of Chromatography Liquid chromatography versus gas chromatography? Applications? Volatile  GC Non-volatile  LC.
Chapter 8 – Mass Spectrometry. Mass Spectrometry The mass spectrometer can be used for: – Quantitative analysis – as a sophisticated and very sensitive.
Year 12 Chemistry Unit 3 – AOS 1 Chemical Analysis.
Molecular mass spectrometry Chapter 20 The study of “molecular ions” M + e -  M e -
1 Chemical Analysis by Mass Spectrometry. 2 All chemical substances are combinations of atoms. Atoms of different elements have different masses (H =
William Kou Jonathan Mertz. Introduction The field of Mass Spectrometry using Ambient Ionization techniques has grown exponentially since The direct.
Based on ionization of gas phase molecule followed by analysis of the masses of the ions produced. The Mass Spectrum: Graph of ion intensity versus mass-to-charge.
Mass Spectroscopy 1 Mass Spectroscopy (Mass Spec) Applying Atomic Structure Knowledge to Chemical Analysis.
Chemistry Topic: Atomic theory Subtopic : Mass Spectrometer.
Ionization energy?. Ionization energy? EI Ionization??
Combining and Choosing Analytical Techniques Chapter 8.
By- Bhavya, Harsh, Harshvardhan, Namrata, Ronit and Vidhatri
LIQUID CHROMATOGRAPHY-MASS SPECTROMETRY
Separates charged atoms or molecules according to their mass-to-charge ratio Mass Spectrometry Frequently.
Chromatography and Instrumentation. Chromatography Separate Analyze Identify Purify Quantify Components Mixture Chromatography is used by scientists to:
Mass Spectrometry Relative atomic masses and the mass of individual isotopes can be determined using a mass spectrometer. The principle behind mass spectrometry.
Chem. 133 – 4/26 Lecture. Announcements Return graded quiz and additional problem Lab – Lab report deadlines (2:4 – Thursday) Today’s Lecture – Mass Spectrometry.
Chapter 29 Mass Spectrometry. 29 A Principles of mass spectrometry In the mass spectrometer, analyte molecules are converted to ions by applying energy.
LIQUID CHROMATOGRAPHY – MASS SPECTROSCOPY (LC-MS) Presented by Md Akbar Siddiq Khan M.Pharm Nizam College Of Pharmacy Hyderabad - A.P.
LIQUID CHROMATOGRAPHY-MASS SPECTROMETRY (LC/MS)
Methodology of targeted mass spectrometry
Yonsei Proteome Research Center Peptide Mass Finger-Printing Part II. MALDI-TOF 2013 생화학 실험 (1) 6 주차 자료 임종선 조교 내선 6625.
SYED AHMAD RAZA ARSLAN SALEEM M. ABUBAKAR WASIF ZAFAR MANAN ARIF WALEED ALI M. USMAN INSTRUMENTATION, MASS SPECTRUM & APPLICATIONS.
CHEM133 Mass Spectrometry: Lecture 1
Components of Mass Spectrometer
Chem. 133 – 4/13 Lecture.
Mass Spectrometry Obaid M. Shaikh.
Mass Spectroscopy. Mass Spectroscopy Mass Spectrometry Most useful tool for molecular structure determination if you can get it into gas phase Molecular.
Lecture 14 LC-MS Ionization.
Lecture 14 LC-MS Ionization.
Molecular Mass Spectroscopy
Mass Spectrometry THE MAIN USE OF MS IN ORG CHEM IS:
M.Prasad Naidu MSc Medical Biochemistry, Ph.D,.
Suppose we make a new compound containing of C,H, and N
Presentation transcript:

1 Principle  LC-MS is interfacing HPLC system with mass spectrometer.  The difficulty in hyphenation is to transform the solute into gas phase ion.  This problem is solved by using interfaces.  Hence the general principle involves the vaporization of the effluent and then ionization of the analyte which is further analyzed through the MS.

PROBLEMS IN COMBINING HPLC AND MS HPLC Liquid phase operation deg. C No mass range limitations Inorganic buffers 1 ml/min eluent flow is equivalent to 500 ml/min of gas MS  Vacuum operation  deg. C  Up to 4000 Da for quadrupole MS  Requires volatile buffers  Accepts 10 ml/min gas flow

Components of LCMS: 3 INTERFACEDETECTOR Extraction of analyte from solvent. Ion evaporation and ionisation.

INSTRUMENTATION 1HPLC SYSTEM 2INTERFACES 3 MASS ANALYZERS 4 DETECTORS

1. INTERFACES 1. Moving Belt Interface. 2.Thermospary Interface. 3.ATMOSPHERIC PRESSURE IONIZATION: I. Electrospray Ionization. II.Atmospheric Pressure Chemical Ionization. III.Atmospheric Pressure Photoionization.

1. Moving Belt Interface 6

1. Moving belt interface:  STEPS: I. Sample deposited onto a moving belt or wire. II. Sample passes through multiple vaccum zones. III. Sample is desorbed into source using heat. IV. Belt cycles back.  ADVANTAGE: Used with wide range of HPLC conditions, Flow rates and mobile phase.  DISADVANTAGES: Appearance of intense chemical background. Not suitable for thermally labile compounds. Surface effect reduces detection limits. 7

2. Thermospray Interface: 8

 STEPS: I.Nebulisation of the eluent from a heated transfer tube. II.Ionisation of analytes.  ADVANTAGE: Easier to use at flow rate 2 ml\min.  DISADVANTAGES: Mobile phase should be volatile. Decomposition of thermally liable analytes. Not suitable for high molecular weight analyte.(>1000Da) Temperature optimization is critical. 9

I. Electrospray Ionisation: 10 High electrical potential

I. Electrospray ionisation:  STEPS: I.Analyte solution flow passes through electrospray needle having high potential difference. II.Formation of charged droplets. III.Gas phase ion formation.  ADVANTAGES: Applicable to volatile and non-volatile analytes. Study of higher molecular weight substances. Analyze multiply charged compounds. Practical mass range up to Da  DISADVANTAGES : Not applicable to non-polar and low polarity compounds. Salts and ion-pairing agents reduce sensibility. 11

II. Atmospheric Pressure Chemical Ionization. 12

II. Atmospheric Pressure Chemical Ionisation:  STEPS: I.HPLC effluent is passed through a pneumatic nebulizer where the droplets are both generated and desolvated. II.The spray so formed then passes through a heated region where the vapour is dried. III.Solvent molecules are ionised by corona discharge to generate stable reaction ions. IV.Liquid flow rate ml/min.  ADVANTAGES: It is best applied to compounds with low to moderately high polarities.  DISADVANTAGES: Thermal degradation can occur. Limited structural information as multiple charging do not take place. Not appropriate for higher MW (e.g.,>1000 Da) and analyte that are charged in solution. 13

III.Atmospheric Pressure Photoionization 14

III. Atmospheric Pressure Photoionisation: STEPS: I.A vaporizer converts the LC eluent to the gas phase. II.A discharge lamp generates photons in a narrow range of ionization energies. ADVANTAGE OVER APCI: Applicable to highly non polar compounds and low flow rates (<100 μl/min). 15

2. MASS ANALYSERS: 1.Magnetic Sector System. 2.Quadrapole Mass Analyser. 3.Ion Trap. 4.Time Of Flight. 16

1. Magnetic Sector System 17 Flight Tube

1. Magnetic Sector System: I.Ion optics in the ion-source chamber extract and accelerate ions. II.The ions enter the flight tube between the poles of a magnet and are deflected by the magnetic field. III.By varying the acceleration voltage or the magnetic field strength, a complete spectrum (an m/z range of e.g. 50 to 2000) can be obtained. IV.The accelerated ions of different masses were detected at different impact points on the detector plate and mass ratios are measured. 18

2. Quadrapole Mass Analyser: F 19

2. Quadrapole Mass Analyser:  It allows ions of any m/z ratio to pass through the metallic rods where a potential difference is applied and the instrument acts as a mass filter.  Quadrupole mass analyzers can operate in two modes: I. Scanning (scan) mode. II. Selected ion monitoring (SIM) mode.  A triple quadrupole tandem mass unit(ms-ms) has three quadrupoles arranged in a series. It can be set either for the analysis of intact peptides or their fragment ions.  Q1 is used to scan across a preset m/z range and select an ion of interest.  Q2 focuses and transmits the ions while introducing a collision gas into the flight path of the selected ion.  Q3 serves to analyze ion fragments. 20

3. Ion Trap Mass Analyser: 21

3.Ion Trap Mass Analyser: I.It consists of a circular ring electrode & two end caps that together form a chamber. II.Ions entering the chamber are trapped by electromagnetic field. Another field can be added to selectively eject ions from the trap. Mechanisam:  Destabilise the ions and eject them progressively from the trap.  Keep only one ion of a given m/z value in the trap and then eject it to observe it specifically.  Keep only one ion in the trap fragment it by inducing vibrations and observe the fragments. This is MS/MS.  The advantages of the ion-trap mass spectrometer include compact size and the ability to trap and accumulate specific ions. It carry out MS n i.e. multiple levels of fragmentation with out addition mass analyser. 22

4. Time Of Flight. 23

4. Time Of Flight: I. Uniform electromagnetic force is applied to all ions. II. Lighter ions travel faster & arrive at detector first.  After the initial acceleration phase, the velocity reached by an ion is inversely proportional to its mass.  These are commonly used in studies of proteins and protein fragments because this type of detector can handle and analyze very large molecular and fragmentation ions. 24

3. Detectors: 1.Electron Multiplier Detector. 2.Dynolyte Photomultiplier Detector. 25

1. Electron Multiplier Detector. 26

1. Electron multiplier Detector:  A conversion dynode is used to convert either negative or positive ions into electrons.  These electrons are amplified by a cascade effect to produce a current. 27

2. Dynolyte Photomultiplier Detector. 28 Photocathode

2. Dynolyte Photomultiplier Detector:  Ions exiting the quadrupole are converted to electrons by a conversion dynode.  These electrons strike a phosphor which when excited,emit photons.  The photons strike a photocathode at the front of the photomultiplier to produce electrons and the signal is amplified by photomultiplier. 29

Applications Molecular Weight Determination Differentiation of similar octapeptides Determining the molecular weight of green fluorescent protein Structural Determination Structural determination of ginsenosides using MSn analysis Pharmaceutical Applications Rapid chromatography of benzodiazepines Detection of degradation products for salbutamol Identification of bile acid metabolites Biochemical Applications Rapid protein identification using capillary LC/MS/MS and database searching

Clinical Applications High-sensitivity detection of trimipramine and thioridazine Food Applications Identification of aflatoxins in food Determination of vitamin D3 in poultry feed supplements using MS3 Environmental Applications Detection of phenylurea herbicides Detection of low levels of carbaryl in food