Stability Indicating Assays for Dose Formulation Stability Testing Teresa Fuller GSK.

Slides:



Advertisements
Similar presentations
Training Brochure 2012 Industrial Training in Advance Pharmaceutical Science.
Advertisements

Dr. Birgit Schmauser, BfArM, Bonn
D e t e c t o r s f o r H P L C.
Analytical Method Development and Validation
ANALYTIC METHODS II. PART Jana Švarcová.  Chromatography  Electrophoresis  Potentiometry  Titration  Spectrophotometry.
Caffeine as a Surrogate for Cocaine in an HPLC Forensic Experiment Ray A. Gross Jr, Indravadan Shah and Muhamed Jasarevic Department of Physical Sciences.
KROMATOGRAFİ Sedat Türe. HPLC Liquid Chromatography.
High Performance Liquid Chromatography. HPLC originally refered to: High Pressure Liquid Chromatography currently refers to: High Precision Liquid Chromatography.
World Health Organization
Results Initial chromatographic conditions [Figure 2a caption] for the separation of the degradation products of aspirin were problematic due to the long.
Chromatography.
In carbon-13 NMR, what do the number of peaks represent?
HPLC Systems. Column Chromatography HPLC Modes HPLC – System Components.
Purity Identificaion of Crack-Coccaine Through the use of HPLC Rony Anderson 1, Eduardo de Jesus Oliveira 1, Kyle Wojciechowski 2 1- Laboratorio de Tecnologia.
Presented by Meng Shi The C holesterol in plasma was a important factor to lead Thrombus.
Vinh Tran Chem Fall 2011 December 9, 2011.
Chromatographic detectors for Liquid Chromatography.
High Performance Liquid Chromatography. HPLC originally refered to: High Pressure Liquid Chromatography currently refers to: High Precision Liquid Chromatography.
Lecture 21 HPLC-2 Instrumentation. Column Injector Detector Chromatograph Data Processing.
Instant Notes Analytical Chemistry
QUALITY CONTROL OF PHYSICO-Chemical METHODS Introduction :Validation توثيق المصدوقية.
B IOCHEMICAL INSTRUMENTAL ANALYSIS -11 Dr. Maha Al-Sedik.
Pn. Suryati Bt. Syafri 2009 High Performance Liquid Chromatography (HPLC)
CHAPTER 29 Supercritical Fluid Chromatography The mobile phase is a supercritical fluid (a fluid above its critical T and critical pressure) Supercritical.
Chromatography Year 12.
Validation of Analytical Method
1. 2 Vitamin Losses During Food Storage and Processing 16-June-2008.
Created with MindGenius Business 2005® Performance parameters (Advanced) Theoretical Plate Model Performance parameters (Advanced) Theoretical Plate Model.
The following minimum specified ranges should be considered: Drug substance or a finished (drug) product 80 to 120 % of the test concentration Content.
By: Thilag.k & Stephen. What is Hpcl??? Hplc or high performance liquid chromatography is the most widely used analytical separation technique. The difference.
Year 12 Chemistry Unit 3 – AOS 1 Chemical Analysis.
GC Advantages 1. Very Large N (Very Long Columns) 2. No Packing Material (A=0) 3. Simple Mobile Phase (Compressed Gas) 4. Universal Detectors (FID) 5.
Peak-purity by LC-MS and LC-DAD Knut Dyrstad Erlend Hvattum Sharon Jara Arnvid Lie.
Simple Scale-up on a 940-LC Analytical to Preparative HPLC
Analysis of Ibuprofen in Tablets According to the US Pharmacopoiea using a Varian 920-LC Analytical HPLC and Pursuit ® XRs column By Phuong Truong Varian.
Combining and Choosing Analytical Techniques Chapter 8.
HPLC – High Performance Liquid Chromatography
Predictive safety study of ticagrelor's degradation combining LC-MS and in silico approach H. SADOU YAYE*, B. DO, N. YAGOUBI * Pharmacist at Pitié-Salpêtrière.
Diversity of molecules and the resulting analytical challenges CHAOTIC 24 April 2013.
Created with MindGenius Business 2005® Instrumentation (1) Instrumentation (1) Origins Preparative column chromatography (Glass columns) Internal diameter:
Chromatography High Performance Liquid Chromatography HPLC Chapter Dr Gihan Gawish.
Biochemical instrumental analysis - 11 Dr. Maha Al-Sedik 2015 CLS 332.
Lecture 10 ANALYTICAL METHOD DEVELOPMENT AND VALIDATION IN HPLC AND GC. Lecture 10 – Chromatography, Dr. Rasha Hanafi 1© Dr. Rasha Hanafi, GUC.
Determination of metformin in urine (by Liquid Chromatography LC)
Instrumental Analysis
Instrumental Analysis (I) HPLC Tutorial 8. Graded presentation Students in groups of 4-5 individuals are asked to prepare a presentation (weight=5% of.
3M Drug Delivery Systems 3 Introduction A family of hydrofluoroalkane-compatible excipients based on oligomeric lactic acid (OLA) has been proposed for.
EQUIPMENT and METHOD VALIDATION
DETERMINATION OF  9 -TETRAHYDROCANNABINOL BY HPLC/DAD IN FOOD SUPPLEMENT SAMPLES OF HEMP SEED OIL Maja Shishovska 1, Dragica Doneva 1, Katerina Starkovska.
Table 2. Summary of chromatographic methods of Terazosin in different matrices Alankar Shrivastava et al. Various Analytical Methods for the Determination.
Multi-Analyte LC-MS/MS Methods – Best Practice.
HIGH-PERFORMANCE LIQUID CHROMATOGRAPHIC EXPERIMENT
Chromatography- TLC & HPLC
High Performance Liquid Chromatography (HPLC)
High Performance Liquid Chromatography HPLC
Determination of Vismodegib by
Size Exclusion Chromatography (SEC)
HPLC detectors Pharmawiki 11/18/2018.
Satish Pradhan Dnyanasadhana College, Thane (W)
BASICS OF SUPERCRITICAL FLUID CHROMATOGRAPHY
ANALYTICAL METHOD VALIDATION
Chromatography Daheeya Alenazi.
High Performance Liquid Chromatography (H. P. L. C. ) dr. Sham b
Liquid Chromatography - Method Development and Validation
Levofloxacin structural formula
Selvadurai Muralidharan, Jayaraja kumar, Venugopal Vijayan
Transferring LC-UV to LC-MS.
CHROMATOGRAPHY.
Fig. 6. HPLC analysis of RP key components
Damiana Gentili Qualified Person/QU Director 09/05/2019
Presentation transcript:

Stability Indicating Assays for Dose Formulation Stability Testing Teresa Fuller GSK

Stability Indicating Assays for Dose Formulation Stability Testing –Stability Studies for Dose Formulations –What the white paper says –What is a stability indicating assay? –So do we need a stability-indicating assay? –Closing thoughts

Stability Studies for Formulations –Performed for every formulation to be used on a GLP study –Test the formulation for concentration on day 0 –Test the formulation for concentration on further timepoints to assess stability, eg. day 1, day 7 and day 14 –Based on concentration results over time, assess stability

What the white paper says “The analytical method should be suitable for the intended purpose. It is conceivable to have a rapid analytical method for test article only, which would not be suitable as a stability-indicating assay where degradents are separated from the test article or a more elaborate method which may have the additional advantage of identifying degradents. Both types of assays are acceptable and would require the appropriate level of validation for their intended purposes.”

What is a Stability Indicating Assay? –According to the US-FDA stability guideline of 1998 ‘validated quantitative analytical methods that can detect the changes with time in the chemical, physical, or microbiological properties of the drug substance and drug product, and that are specific so that the contents of active ingredient, degradation products, and other components of interest can be accurately measured without interference.’ –But this applies to drug substance and drug product not dose formulations! –For dose formulations we do not necessarily need to quantify the degradents –So what should the formulation analytical method look like?

Isocratic UPLC/UV –Chromatography column, %organic solvent and wavelength chosen for analyte of interest –K’ of at least 3 [K’ = (V t – V 0 )/V 0 ] –Eg. HPLC System Waters Acquity UPLC with TUV Detector using High Sensitivity Flow cell Typical Injection Volume1 µL Flow Rate1 mL/min Wavelength245 nm Analytical Column50 x 2.1 mm i.d. Waters BEH C µm Column Temperature60 ºC Run Time1 min Typical Retention Time0.8 min Mobile Phase A0.05% Trifluoroacetic Acid Mobile Phase BMethanol Isocratic Composition A:B69:31

Isocratic UPLC/UV Reference standard Stability sample day 7, ambient PFL 0.02 mg/mL in water

Generic UPLC-UV Gradient HPLC SystemWaters Acquity UPLC Typical Injection Volume0.5 – 5 uL Flow Rate0.6 mL/min WavelengthAnalyte specific Analytical Column100 x 2.1 mm i.d. Waters BEH C µm Column Temperature50 ºC Run Time12 min Mobile Phase A0.1% Trifluoroacetic Acid in water Mobile Phase B0.1% Trifluoroacetic Acid in acetonitrile Gradient profile Time%B

Generic UPLC-UV Gradient

Isocratic and Gradient Compared IsocraticGradient Run Time1-2 mins12 mins Method development1 day to several weeksMinimal Visible degradentsSignificantly different degradents probably not visible Most degradents and impurities likely to be visible Stability-indicatingpossibly

How can we assess how stability indicating the assay is? –Forced degradation –Store sample at 30°C, 65% RH or 40°C, 75% RH or 50°C, ambient RH –UV light exposure –Oxidation, eg. By addition of hydrogen peroxide –Acid/alkali hydrolysis –Compare forced samples to a fresh sample –If any additional peaks are seen in the forced samples but these are well resolved from the analyte then assume assay is adequate

How can we assess how stability indicating the assay is? –Peak purity analysis –Acquire UV spectrum at several points (at least 3) across the chromatographic peak –Compare spectra to look for differences which may indicate underlying peaks co-eluting with the analyte

How can we assess how stability indicating the assay is? –Mass spectral peak purity –Acquire mass spectrum at several points across the peak, similarly to UV peak purity

The ideal stability indicating method? –UPLC-MS/MS –Use specific parent and daughter ions to monitor the analyte Analyte Precursor ion (m/z) Product Ion (m/z) GSK Product 290

Pros and Cons TechniqueProsCons Forced Degradation No additional equipment reqd Time Degradation may not be the same in real samples UV Peak purityQuickDegradation products may have similar spectrum Needs PDA detector Mass specSpecificExpensive equipment Mass spectral comparison difficult

So do we need a stability-indicating assay? –White paper says we don’t! –Unstable formulation considered to be greater than 10% drop from t0 –Complex oral and iv formulations are likely to have had previous development performed by a Product Development group –storage stability already tested to some degree –We need some confidence in our assay

Closing Thoughts –Several factors to consider –Balancing act between having confidence in the method and minimising method development time and effort –Is a gradient better than isocratic? –Is MS detection better than UV as it becomes more affordable?

Acknowledgements –Emma Simmonds, GSK PD –Tammy Clegg, GSK DMPK –Louise Kay, GSK DMPK

Useful Reference –Review of Development of Validated Stability Indicating Assay Methods –Journal of Pharmaceutical and Biomedical Analysis 28 (2002) 1011–1040