MPPC status M.Taguchi(kyoto) T2K ND280 2006/7/7.

Slides:



Advertisements
Similar presentations
Scintillator Tile Hadronic Calorimeter Prototype (analog or semidigital) M.Danilov ITEP(Moscow) CALICE Collaboration LCWS04, Paris.
Advertisements

New test result of MPPC M.Taguchi(kyoto).
Comparison of 3 types of scintillator strips and 2 types of reflector films Miho NISHIYAMA Shinshu-U Light yield of 3 scintillator strips have been measured.
The Multi-Pixel Photon Counter for the GLD Calorimeter Readout Jul Satoru Uozumi University of Tsukuba, Japan 1.Introduction 2.Recent.
Study of the MPPC Performance - contents - Introduction Fundamental properties microscopic laser scan –check variation within a sensor Summary and plans.
Performance of MPPC using laser system Photon sensor KEK Niigata university, ILC calorimeter group Sayaka IBA, Hiroaki ONO, Paul.
Study of Photon Sensors using the Laser System 05/7/12 Niigata University, Japan Sayaka Iba, Editha P. Jacosalem, Hiroaki Ono, Noriko.
1 A proposal of new simple system for testing a large number of MPPC for the R&D phase of GLD calorimeter 2007/Feb/6 ACFA ILC Workshop ICEPP, University.
Scintillator based muon upgrade / BELLE Super B Factory Workshop In Hawaii Jan 2004, Honolulu, Hawaii 1.Scintillator strip option 2.Geiger photodiodes.
Mass test of MPPC with a prototype of MEG II liquid xenon detector
CFT Calibration Calibration Workshop Calibration Requirements Calibration Scheme Online Calibration databases.
6mm 【 Development of Readout Electronics for MPPC 】 We report the read out electronics of MPPC( Multi-Pixel Photon Counter ). MPPC is a new photodetector.
The Transverse detector is made of an array of 256 scintillating fibers coupled to Avalanche PhotoDiodes (APD). The small size of the fibers (5X5mm) results.
MPPC R&D status Kobe Univ. CALICE collaboration meeting Yuji SUDO Univ. of Tsukuba ~ contents ~ Introduction Linearity curve Recovery time.
GLD Calorimeter Status Oct 学術創成会議 S. Uozumi Shinshu University FJPPL meeting held at the end of September. Preparation underway toward the ECAL.
Selection of Silicon Photomultipliers for ILC Analogue Hadron Calorimeter Prototype Lay-out * ILC Hadron Calorimeter prototype with SiPM readout * Selection.
CALICE Meeting DESY ITEP&MEPhI status report on tile production and R&D activities Michael Danilov ITEP.
MPPC Radiation Hardness (gamma-ray & neutron) Satoru Uozumi, Kobe University for Toshinori Ikuno, Hideki Yamazaki, and all the ScECAL group Knowing radiation.
Development of Multi-pixel photon counters(2) M.Taguchi, T.Nakaya, M.Yokoyama, S.Gomi(kyoto) T.Nakadaira, K.Yoshimura(KEK) for KEKDTP photon sensor group.
SiPM: Development and Applications
The MPPC Study for the GLD Calorimeter Readout Introduction Measurement of basic characteristics –Gain, Noise Rate, Cross-talk Measurement of uniformity.
The Scintillator ECAL Beam Test at FNAL Adil Khan International Linear Collider Workshop 2010 LCWS10 & ILC10, Beijing, China CALICE Scintillator ECAL group.
The Scintillator ECAL Beam Test at FNAL K. Kotera, Shinshu-u, 1st October 2009 CALICE Scintillator ECAL group; Kobe University, Kyungpook University, the.
MPPC update including plastic connector T2K experiment collaboration meeting 2007/4/18 (Wed) S.Gomi T.Nakaya M.Yokoyama ( Kyoto University ) T.Nakadaira.
R&D of MPPC for T2K experiment PD07 : Photosensor Workshop /6/28 (Thu) S.Gomi T.Nakaya M.Yokoyama H.Kawamuko ( Kyoto University ) T.Nakadaira.
Study of the MPPC performance - R&D status for the GLD calorimeter readout – Nov 6-10.
Study of the Multi-Pixel Photon Counter for ILC calorimeter Satoru Uozumi (Kobe University) Atami Introduction of ILC and MPPC The MPPC performance.
MEG II 実験 液体キセノンガンマ線検出器に用いる 光検出器 MPPC の 実装に向けた最終試験 家城 佳 他 MEG II collaboration + 九大の方々.
Development of Multi-pixel photon counters(2) M.Taguchi, T.Nakaya, M.Yokoyama, S.Gomi(kyoto) T.Nakadaira, K.Yoshimura(KEK)
1 SiPM Gain Test SiPM Multiple APD pixels operating at Geiger mode. Output is the sum of the outputs from all APD pixels. Advantages Compact size High.
1 NIU Integrated Readout Layer Kurt Francis NICADD Northern Illinois University March 19, 2014.
Study of the MPPC for the GLD Calorimeter readout Satoru Uozumi (Shinshu University) for the GLD Calorimeter Group May 29 – Jun 4 DESY Introduction.
1 Development of Multi-Pixel Photon Counters (1) S.Gomi, T.Nakaya, M.Yokoyama, M.Taguchi, (Kyoto University) T.Nakadaira, K.Yoshimura, (KEK) Oct
1 MPPC update S.Gomi, T.Nakaya, M.Yokoyama, M.Taguchi, (Kyoto University) T.Nakadaira (KEK) Nov KEK.
Development and Study of the Multi Pixel Photon Counter
Study of the MPPC for the GLD Calorimeter readout Satoru Uozumi (Shinshu University) Feb Beijing Introduction Basic performances Future.
Development of Multi-Pixel Photon Counters(MPPC) Makoto Taguchi Kyoto University.
MPPC Saturation Corection Idea -> how to do. MPPC Gain measurement To apply saturation correction to the beam data, all MPPC are needed to be calibrated.
Status of photon sensor study at Niigata University -- SiPM and MPPC -- Photon sensor mini workshop 05/9/16 University Niigata University.
Development of Multi-Pixel Photon Counters and readout electronics Makoto Taguchi High Energy Group.
Study of the MPPC for the GLD Calorimeter Readout Satoru Uozumi (Shinshu University) for the GLD Calorimeter Group (KNU, Kobe, Niigata, Shinshu, ICEPP.
Linearity Tests The laser induced high noise pulses on the APD rendering it useless. We tried several methods to shield and filter the noise. With a stabilized.
Multipixel Geiger mode photo-sensors (MRS APD’s) Yury Kudenko ISS meeting, KEK, 25 January 2006 INR, Moscow.
R&D of Calorimeter using Strip/Block Scintillators with SiPM
Scintillator tile – SiPM development at ITEP Michael Danilov, ITEP CALICE Meeting, Casablanca, 23 Sep 2010.
SiPM for CBM Michael Danilov ITEP(Moscow) Muon Detector and/or Preshower CBM Meeting ITEP
DESY BT analysis - MPPC Saturation Correction - S. Uozumi Feb Sci-ECAL meeting 1.MPPC Gain Measurement (with LED data) 2.Inter-calibration of readout.
December Status of MRS photodiodes ND280 Convener’s Meeting, 9 June 2006 Yury Kudenko INR, Moscow.
Systematics on ScECAL 1 st prototype DESY Apr , CALICE Satoru Uozumi for the CALICE collaboration NIM A, 763 (2014) 278.
F Don Lincoln, Fermilab f Fermilab/Boeing Test Results for HiSTE-VI Don Lincoln Fermi National Accelerator Laboratory.
Performance of new MPPC Nov. 21 Korea-Japan joint meeting Takashi Maeda Hideki Yamazaki Yuji Sudo (University of Tsukuba) --- Contents ---
Beta-ray test for strip scintillator readout MPPC GLD Cal group KEK 06/2/28 (Tue) Niigata university Sayaka IBA.
The Multi-Pixel Photon Counter for the GLD Calorimeter Readout Jul Satoru Uozumi University of Tsukuba, Japan for the GLD Calorimeter.
Study and Development of the Multi-Pixel Photon Counter for the GLD Calorimeter Satoru Uozumi (Shinshu, Japan) on behalf of the GLD Calorimeter Group Oct-9.
Status of NEWCHOD E.Guschin (INR), S.Kholodenko (IHEP), Yu.Kudenko (INR), I.Mannelli (Pisa), O.Mineev (INR), V.Obraztsov (IHEP), V.Semenov(IHEP), V.Sugonyaev.
Development of UV-sensitive MPPC for upgrade of liquid xenon detector in MEG experiment Daisuke Kaneko, on behalf of the MEG Collaboration µ γ Liquid xenon.
Silicon Photomultiplier Development at GRAPES-3 K.C.Ravindran T.I.F.R, OOTY WAPP 2010 Worshop On behalf of GRAPES-3 Collaboration.
Study of the MPPC for the GLD Calorimeter Readout Satoru Uozumi (Shinshu University) for the GLD Calorimeter Group Kobe Introduction Performance.
Development of Multi-pixel photon counters(2) M.Taguchi, T.Nakaya, M.Yokoyama, S.Gomi(kyoto) T.Nakadaira, K.Yoshimura(KEK) for KEKDTP photon sensor group.
M.Taguchi and T.Nobuhara(Kyoto) HPK MPPC(Multi Pixel Photon Counter) status T2K280m meeting.
Performance of 1600-pixel MPPC for the GLD Calorimeter Readout Jan. 30(Tue.) Korea-Japan Joint Shinshu Univ. Takashi Maeda ( Univ. of Tsukuba)
Development of Multi-Pixel Photon Counters (1)
Progress report on SiPM development and its applications
Characteristics of S12045(X) photon sensors for GlueX
Upgrade of LXe gamma-ray detector in MEG experiment
Upgrade of LXe gamma-ray detector in MEG experiment
R&D of MPPC for T2K experiment
T2Kロシアグループに向けたMPPC(SiPM)の性能評価
R&D of MPPC in kyoto M.taguchi.
The MPPC Study for the GLD Calorimeter Readout
Presentation transcript:

MPPC status M.Taguchi(kyoto) T2K ND /7/7

Contents Test sample Device-by-device variation Calibration test Conclusion

Test sample latest samples which was delivered in Jan.2006 We have three samples for each type device-by-device variation is tested at the point bias V=69.0V (at this point gain~10 6, noise threshold <1MHz) Number of pixels Pixel pitch(μm) AreaOperating voltage Geometrical efficiency x1.0mm V46% x1.0mm V55% HPK now develops 100pixel sample with larger geometrical efficiency

Gain ~device-by-device variation~ 100pixel gain400pixel gain variation ~7% variation ~10% 69.0V 3×10 6 1× ℃

Noise rate ~device-by-device variation~ 69.0V 100pixel Noise threshold 69.0V vairation~30% vairation~13% 400pixel Noise threshold 700kHz 800kHz 15 ℃

PDE ~device-by-device variation~ 2 100pixel 69.0V 100pixel PDE(MPPC)/QE(PMT) 400pixel PDE(MPPC)/QE(PMT) variation ~10% variation ~20 % measure the PDE of MPPC relative to that of a PMT 15 ℃ 2

Summary of device-by-device variation 100pixel400pixel gain2.6x10 6 ~ 3.0x x10 5 ~ 9.8x10 5 Noise th 480~620 KHz750~850kHz PDE(MPPC)/ QE(PMT) 1.8~ ~2.0 T=15 ℃, bias V=69.0V Device-by-device variation is small, but it is necessary to test much larger number of samples

Calibration test Motivation Gain, PDE, crosstalk of MPPC are all sensitive to the temperature and bias voltage It is necessary to calibrate the variation of Gain, PDE, crosstalk if temperature of bias voltage change MPPC Signal ∝ Gain(T,V) x PDE(T,V) x 1-crosstalk(T,V) 1 T: temperature V: bias voltage evaluate the stability of device response by two calibration methods(explain each method later)

Set up for calibration test 1inch PMT cosmic-ray 1mm φfiber MPPC2(100) MPPC1(100) MPPC3(400) MPPC4(400) scintillator blue LED put scintillators in four layers inserted fibers are viewed by four MPPCs(two are 400 pixel and two are 100pixel) change temperature intentionally like 20 ℃  25 ℃ The same bias voltage (69.0V) is applied to four MPPCS temperature chamber

Method 1 100pixel gain100pixel crosstalk 100pixel PDE(MPPC)/QE(PMT) gain, PDE, crosstalk are all functions of V-V bd V-V bd V bd :breakdown voltage (derived by linearly extrapolating the gain-voltage curve to the point where gain becomes zero) V Gain V bd

Method 1 corrected light yield = gainxPDEx MIP ADC counts 1-crosstalk 1 20 ℃ 25 ℃ measure the variation of gain(V-V bd ) estimate variation of V-V bd estimate the variation of PDE(V-V bd ),crosstalk(V-V bd ) variation of gain (100pixel)

Stability of device response(method1) +3% -3% 20 ℃ 25 ℃ corrected light yield(100pixel) device response is stable within ~3% after calibration of method 1 response of other three samples is also well calibrated

Method2 MIP ADC count ∝ gain(T,V)×PDE(T,V)× LED ADC count ∝ gain(T,V)×PDE(T,V)× 1 1- crosstalk(T,V) 1 MIP ADC count LED ADC count we can calibrate the variation of gain, PDE, crosstalk by taking the ratio of MIP ADC count to LED ADC count corrected light yield = MIP ADC count LED ADC count

Stability of device response(method2) +3% -3% +3% 20 ℃ 25 ℃ corrected light yield(100pixel) device response is stable within ~3% after calibration of method 2 response of other three samples are also well calibrated

MPPC1(100)MPPC2(100)MPPC3(400)MPPC4(400) Method12.5%2.3%3.8%3.1% Method22.5%1.3%2.4%1.4% We need to guarantee the stability of light from LED for method 2 measure the PDE, cross-talk rate as a function of V-V bd before installation for method 1 Summary and discussion about calibration test Stability of each MPPC response after calibration in RMS/mean calibration is possible by two methods (the precision of calibration is better for method 2)

Conclusion device-by-device variation among three samples is small,  but it is necessary to test much larger number of samples calibration is possible by two methods  further test of much larger number of samples

back up

MPPC(Multi pixel photon coutner) 100~1000 APD pixel in 1mm 2 Each pixel operates as Geiger mode (independent of input light) The output is a sum of all the APD signals Compact Low-cost Insensitive to the magnetic field Low bias voltage :40~75V High gain:10 5 ~10 7 MPPC characters:

Basic performance ~Gain~ MPPC gain = output charge from a single pixel 0p.e. 1p.e. 2p.e. 3p.e. ADC counts calculate gain from the number of ADC counts between 1p.e. and pedestal peak

Basic performance ~gain~ 3× pixel gain bias V 68.4V69.6V 1× V69.4V bias V 400pixel gain 3×10 5

Basic performance ~Noise rate~ count the rate above 0.5 and 1.5p.e. threshold without external light Noire rate measured at the 1.5p.e. threshold increases as the temperature becomes lower 68.2V 69.8V 1MHz ■ ■ ■ : 0.5p.e threshold ▲ ▲ ▲ : 1.5p.e threshold 100kHz because the cross-talk rate increases as the temperature becomes lower 100pixel Noise threshold bias V

Basic performance ~crosstalk~ ・ Assuming 2p.e. noise is caused by crosstalk of 1p.e noise(accidental coincidence of 1p.e noise is subtracted) Cross-talk rate = Data taken by random trigger 0.5p.e.1.5p.e.

Basic performance ~PDE~ MPPC(total area 1mm 2 ) ½ inch PMT 1mmφslit MPPC x Y slit PMT The view from this side ・ only the light going through 1mmφslit is detected ・ Scan the MPPC and PMT with moving stage and search the position with maximum light yield ・ The ratio of MPPC p.e to PMT p.e is taken as relative PDE of MPPC to that of PMT WLS fiber

Basic performance ~linearity(100pixel)~ 3080 injected p.e injected p.e. fired pixel number (Data-Exp)/Exp(%) 6 expected curve calculated from number of pixels and measured cross-talk rate

Basic performance ~linearity(400pixel)~ injected p.e injected p.e. fired pixel number (Data-Exp)/Exp(%) 6 expected curve calculated from number of pixels and measured cross-talk rate

Photoelectrons for MIP photoelectrons at 20 ℃ are about 20% smaller than that at 25 ℃  need to correct for the variation of PDE, cross-talk rate photoelectrons for MPPC3 are much smaller than that for other MPPCs  due to misalignment of a fiber to the MPPC MPPC1(100)MPPC2(100) MPPC3(400)MPPC4(400) ℃ 25 ℃

Uniformity within 1pixel RMS/mean=2% efficiency Gain RMS/mean=2% Cross-talk rate pixel

efficiency 100pixel Gain RMS/mean=3% Cross-talk rate 0.22 Uniformity of each pixel

efficiency RMS/mean=1.7% Gain RMS/mean=1.6% Cross-talk rate Uniformity within 1pixel 400pixel 0.25

RMS/mean=2.9% RMS/mean=3.4% Gain efficiency Cross-talk rate Uniformity of each pixel 400pixel 0.18

Laser test laser 1pixel spot size~ a few μm wavelength ~532nm light intensity ~a few photons 100pixel efficiency measured the geometrical efficiency( ε geom ) by scanning the laser with 0.1μm pitch efficiency is calculated from the fraction of events with more than 0.5p.e. to the total events 68μm 100μm ε geom =46% for 100 pixel ε geom =55% for 400 pixel HPK now develops new 100pixel sample with larger geometrical efficiency 100μm

Variation of 1p.e. ADC count 1p.e. ADC count changes about 20% when temperature changes from 20 ℃ to 25 ℃ MPPC1(100) MPPC2(100) MPPC3(400) MPPC4(400) 20 ℃ 25 ℃ 40hours

MPPC1(100)MPPC2(100) MPPC3(100)MPPC4(100) -3% +3% corrected light yield = MIP ADC count 1p.e. ADC count x PDE x 1 (1-cross- talk rate) Stability of light yield ~method 1~ 20 ℃ 25 ℃

Stability of light yield ~method2~ MPPC1(100)MPPC2(100) MPPC3(400)MPPC4(400) +3% -3% 20 ℃ 25 ℃ LED ADC count = corrected light yield MIP ADC count

light yield distribution for MIP is fitted with Landau distribution temperature coefficient of scintillator and fiber is very small MPPC1(100)MPPC2(100) MPPC3(400) MPPC4(400) MIP ADC count 20 ℃ 25 ℃

400pixel gain 400pixel PDE(MPPC)/QE(PMT) 400pixel crosstalk V-V bd Gain,PDE,crosstalk are functions of V-V bd