Measurement of the neutron detection efficiency of the KLOE calorimeter P.Gauzzi (Universita’ La Sapienza e INFN – Roma) for the KLONE Group 10 th ICATPP.

Slides:



Advertisements
Similar presentations
Stefan Roesler SC-RP/CERN on behalf of the CERN-SLAC RP Collaboration
Advertisements

Preliminary studies for T2 primary target for the NA61 fragmentation beam run 11 th October 2010 – NA61 Collaboration Meeting M. Calviani on behalf of.
Advanced GAmma Tracking Array
Study of plastic scintillators for fast neutron measurements
Multi Pixel Photon Counters (MPPC) and Scintillating Fibers (Sci-Fi)
Salvatore Fiore University of Rome La Sapienza & INFN Roma1 for the KLOE collaboration LNF Spring School “Bruno Touscheck”, Frascati, May 2006 CP/CPT.
Luminosity Monitor Commissioning MICE Collaboration Meeting March 2010 Paul Soler, David Forrest Danielle MacLennan.
Design and test of a high-speed beam monitor for hardon therapy H. Pernegger on behalf of Erich Griesmayer Fachhochschule Wr. Neustadt/Fotec Austria (H.
Study of two pion channel from photoproduction on the deuteron Lewis Graham Proposal Phys 745 Class May 6, 2009.
The Time-of-Flight system of the PAMELA experiment: in-flight performances. Rita Carbone INFN and University of Napoli RICAP ’07, Rome,
Atmospheric Neutrino Oscillations in Soudan 2
The Transverse detector is made of an array of 256 scintillating fibers coupled to Avalanche PhotoDiodes (APD). The small size of the fibers (5X5mm) results.
The PEPPo e - & e + polarization measurements E. Fanchini On behalf of the PEPPo collaboration POSIPOL 2012 Zeuthen 4-6 September E. Fanchini -Posipol.
B. Sciascia 27 November 2006, XXXIII LNF Scientific Committee1 Neutron detection efficiency of the KLOE calorimeter B.Sciascia (LNF  INFN) for the KLONE.
Status of the Beamline Simulation A.Somov Jefferson Lab Collaboration Meeting, May 11, 2010.
A scintillation detector for neutrons below 1 MeV with gamma-ray rejection Scintillators are 3 mm BC408, 10 layers total Adjacent layers are optically.
Journées GDR Nucléon - Instrumentation April 8-9, 2008 – Saclay M.Anelli, G.Battistoni, S.Bertolucci, C.Bini, P.Branchini, C.Curceanu, G.De Zorzi, A.Di.
Possibilities of TOF measurements on NPI neutron generators Mitja Majerle Department of Nuclear Reactions Nuclear Physics Institute ASCR.
Neutron scattering systems for calibration of dark matter search and low-energy neutrino detectors A.Bondar, A.Buzulutskov, A.Burdakov, E.Grishnjaev, A.Dolgov,
V.L. Kashevarov. Crystal Collaboration Meeting, Mainz, September 2008 Photoproduction of    on protons ► Introduction ► Data analysis.
Luminosity Monitor UKNF Meeting 7 June 2010 Paul Soler, David Forrest Danielle MacLennan.
Calibration of the CMS Electromagnetic Calorimeter with first LHC data
PrimEx collaboration meeting Energy calibration of the Hall B bremsstrahlung tagging system using magnetic pair spectrometer S. Stepanyan (JLAB)
Lead Fluoride Calorimeter for Deeply Virtual Compton Scattering in Hall A Alexandre Camsonne Hall A Jefferson Laboratory October 31 st 2008.
Systematic studies of neutrons produced in the Pb/U assembly irradiated by relativistic protons and deuterons. Vladimír Wagner Nuclear physics institute.
09/10/2005NSTAR Graal collaboration1 The Graal collaboration results and prospects Presented by Carlo Schaerf Università di Roma “Tor Vergata” and.
Neutron detection in LHe ( HMI run 2004) R.Golub, E. Korobkina, J. Zou M. Hayden, G. Archibold J. Boissevain, W.S.Wilburn C. Gould.
Shashlyk FE-DAQ requirements Pavel Semenov IHEP, Protvino on behalf of the IHEP PANDA group PANDA FE-DAQ workshop, Bodenmais April 2009.
Anna Ferrari VCI 2007, Wien, February 22nd, Measurement and simulation of the neutron response and detection efficiency of a Pb – scintillating.
Apollo Go, NCU Taiwan BES III Luminosity Monitor Apollo Go National Central University, Taiwan September 16, 2002.
Experimental Nuclear Physics Some Recent Activities 1.Development of a detector for low-energy neutrons a. Hardware -- A Novel Design Idea b. Measure the.
Measurement and simulation of neutron detection efficiency in lead-scintillating fiber calorimeter Active material:  1.0 mm diameter scintillating fiber.
Hadronic interaction studies with the ARGO-YBJ experiment (5,800 m 2 ) 10 Pads (56 x 62 cm 2 ) for each RPC 8 Strips (6.5 x 62 cm 2 ) for each Pad ( 
1 NaI calibrationneutron observation NaI calibration and neutron observation during the charge exchange experiment 1.Improving the NaI energy resolution.
A search for strange tribaryonic states in the reaction Heejoong Yim Seoul National University For KEK-PS E549 collaboration.
Catania 11 ICATPP october, 2009 Como 1/12 Catania Comparative measurements of the performances of four super bialkali large.
Nucleon Decay Search in the Detector on the Earth’s Surface. Background Estimation. J.Stepaniak Institute for Nuclear Studies Warsaw, Poland FLARE Workshop.
Test beam preliminary results D. Di Filippo, P. Massarotti, T. Spadaro.
Progress on F  with the KLOE experiment (untagged) Federico Nguyen Università Roma TRE February 27 th 2006.
09/10/2005NSTAR Graal collaboration1 The Graal collaboration results and prospects Presented by Carlo Schaerf Università di Roma “Tor Vergata” and.
N_TOF commissioning INTC-P-249 Spokespersons: J.L.Tain, V.Vlachoudis Contactperson: V.Vlachoudis th Nov 2009 for the n_TOF collaboration.
5-9 June 2006Erika Garutti - CALOR CALICE scintillator HCAL commissioning experience and test beam program Erika Garutti On behalf of the CALICE.
00 Cooler CSB Direct or Extra Photons in d+d  0 Andrew Bacher for the CSB Cooler Collaboration ECT Trento, June 2005.
Search for the  + in photoproduction experiments at CLAS APS spring meeting (Dallas) April 22, 2006 Ken Hicks (Ohio University) for the CLAS Collaboration.
M.Anelli, G.Battistoni, S.Bertolucci, C.Bini, P.Branchini, C.Curceanu, G.De Zorzi, A.Di Domenico, B.Di Micco, A.Ferrari, S.Fiore, P.Gauzzi, S.Giovannella,F.Happacher,
Luminosity Monitor Design MICE Collaboration Meeting 31 May 2009 Paul Soler.
 0 life time analysis updates, preliminary results from Primex experiment 08/13/2007 I.Larin, Hall-B meeting.
MEG 実験 2009 液体キセノン検出器の性能 II 西村康宏, 他 MEG コラボレーション 東京大学素粒子物理国際研究セン ター 第 65 回年次大会 岡山大学.
Hall C Summer Workshop August 6, 2009 W. Luo Lanzhou University, China Analysis of GEp-III&2γ Inelastic Data --on behalf of the Jefferson Lab Hall C GEp-III.
Energy Reconstruction in the CALICE Fe-AHCal in Analog and Digital Mode Fe-AHCal testbeam CERN 2007 Coralie Neubüser CALICE Collaboration meeting Argonne,
Rita Carbone, RICAP 11, Roma 3 26/05/2011 Stand-alone low energy measurements of light nuclei from PAMELA Time-of-Flight system. Rita Carbone INFN Napoli.
1 Cosmic Ray Physics with IceTop and IceCube Serap Tilav University of Delaware for The IceCube Collaboration ISVHECRI2010 June 28 - July 2, 2010 Fermilab.
Three years of cross-section measurements of (n,xn) threshold reactions at TSL Uppsala and NPI Řež O. Svoboda, A. Krása, A. Kugler, M. Majerle, J. Vrzalová,
Measurement of the CR light component primary spectrum B. Panico on behalf of ARGO-YBJ collaboration University Rome Tor Vergata INFN, Rome Tor Vergata.
ICARUS T600: low energy electrons
COMPASS calorimeters S. Platchkov IRFU, CEA-Saclay GDR, 8-9 avril 2008
The Transition Radiation Detector for the PAMELA Experiment
Study of Hypernuclei with Heavy Ion Beams (HypHI) at GSI Shizu Minami GSI, Germany on behalf of HypHI collaboration Introduction Phase 0 experiment R.
Stanislav Pospíšil et al, Brightness Vidyo meeting
at TSL high energy neutron facility
Cross-section Measurements of (n,xn) Threshold Reactions
Filimon Roukoutakis, MC-PAD ITN ER
an ecofriendly possibility
Neutron detection efficiency of the KLOE calorimeter
Very high efficiency, about 4 times
for collaboration “Energy plus transmutation”
Jin Huang Los Alamos National Lab
Measurement and simulation of neutron detection
O. Svoboda, A. Krása, A. Kugler, M. Majerle, J. Vrzalová, V. Wagner
A Neutron Counter for the CLAS12 Central Detector
Presentation transcript:

Measurement of the neutron detection efficiency of the KLOE calorimeter P.Gauzzi (Universita’ La Sapienza e INFN – Roma) for the KLONE Group 10 th ICATPP Conference October 8-12, 2007 – Como M.Anelli, G.Battistoni, S.Bertolucci, C.Bini, P.Branchini, C.Curceanu, G.De Zorzi, A.Di Domenico, B.Di Micco, A.Ferrari, S.Fiore, P.G., S.Giovannella,F.Happacher, M.Iliescu, M.Martini, S.Miscetti, F.Nguyen, A.Passeri, A.Prokofiev, P.Sala, B.Sciascia, F.Sirghi

P.Gauzzi10 th ICATPP Conference - October 9 th 2007 – Como 2 The KLOE calorimeter Pb - scintillating fiber sampling calorimeter of the KLOE experiment at DA  NE (LNF): 1 mm diameter sci.-fi. (Kuraray SCSF-81 and Pol.Hi.Tech 0046) –Core: polystyrene,  =1.050 g/cm 3, n=1.6, peak ~ 460 nm 0.5 mm groved lead foils Lead:Fiber:Glue volume ratio = 42:48:10 X 0 = 1.6 cm  =5.3 g/cm 3 Calorimeter thickness = 23 cm Total scintillator thickness ~ 10 cm 1.2 mm 1.35 mm1.0 mm Lead

P.Gauzzi10 th ICATPP Conference - October 9 th 2007 – Como 3 The KLOE calorimeter Operated from 1999 to 2006 with good performance and high efficiency for electron and photon detection, and also good capability of π/  /e separation Energy resolution:  E /E=5.7%/  E(GeV)  t =54 ps/  E(GeV)  147 ps Time resolution (see KLOE Collaboration, NIM A482 (2002),364) (  K S K L ; K S  π + π ─ ; K L  2π 0 ) 44

P.Gauzzi10 th ICATPP Conference - October 9 th 2007 – Como 4 Why neutrons at KLOE ? Detection of n of few to few hundreds of MeV is traditionally performed with organic scintillators (elastic scattering of n on H atoms produces protons detected by the scintillator itself)  efficiency scales with thickness  1%/cm Use of high-Z material improves the neutron efficiency (see C.Birattari et al., NIM A297 (1990) and NIM A338 (1994) and also T.Baumann et al., NIM B192 (2002)) Preliminary estimate with KLOE data (n produced by K  interactions in the apparatus) showed a high efficiency (  40%) for neutrons with E n < 20 MeV, confirmed by the KLOE MC n detection is relevant for the DA  NE-2 program at LNF; two proposals: – search for deeply bounded kaonic nuclei (AMADEUS) – measurement of the neutron time-like form factors (DANTE) A test has been performed with the neutron beam of the The Svedberg Laboratory of Uppsala (October 2006 and June 2007)

P.Gauzzi10 th ICATPP Conference - October 9 th 2007 – Como 5 The neutron TSL E KIN (MeV) 5.31 m KLOE calorimeter module (  2 cm) Neutrons produced in the reaction 7 Li(p,n) 7 Be Proton beam energy from 180 MeV to  20 MeV Neutron energy spectrum peaked at max energy (at 180 MeV  f p =42% of n in the peak) Tail down to termal neutrons

P.Gauzzi10 th ICATPP Conference - October 9 th 2007 – Como 6 Experimental setup 1. Old KLOE prototype: total length  60 cm 3  5 cells (4.2 cm  4.2 cm) read out at both ends by Hamamatsu/Burle PMTs 2. Beam Position Monitor: array of 7 scintillating counters 1 cm thickness 3.Reference counter : NE110; 5 cm thick; 10  20 cm 2 area (in June 2007  two other NE110 counters 2.5 cm thick) All mounted on a rotating frame allowing for vertical (data taking with n beam) and horizontal (for calibration with cosmic rays) positions (1) (3) (2)

P.Gauzzi10 th ICATPP Conference - October 9 th 2007 – Como 7 Trigger & DAQ Trigger: No beam extraction signal available Scintillator trigger: Side 1 – Side 2 overlap coincidence Calorimeter trigger: analog sum of the signals of the first 12 cells (4 planes out of 5)   A  B overlap coincidence Trigger signal is phase locked with the RF signal ( ns) DAQ: Simplified version of the KLOE experiment DAQ system (VME standard) Max DAQ rate : 1.7 kHz - Typical run: 10 6 events For each configuration/energy: scans with different trigger thresholds Three data-sets: –E peak = 180 MeV -- October two weeks –E peak = 46.5 MeV -- June 2007 –E peak = 21.8 MeV -- “ 4 days n Y X Z

P.Gauzzi10 th ICATPP Conference - October 9 th 2007 – Como 8 Efficiency measurement Global efficiency measurement: integrated over all the energy spectrum f live = fraction of DAQ live time  = acceptance (assuming the beam fully contained in the calorimeter surface    1) Sizeable beam halo at low peak energy  Rate(trigger) must be corrected E n = 180 MeV f live

P.Gauzzi10 th ICATPP Conference - October 9 th 2007 – Como 9 Neutron rate Absolute flux of neutrons measured after the collimator; 2 monitors of beam intensity (see A.Prokofiev et al., PoS (FNDA2006) 016) : –Ionization Chamber Monitor (7 cm  ): online monitor, not position sensitive –Thin-Film Breakdown Counter (1 cm  ): offline monitor; used to calibrate the ICM by measuring the neutron flux at the collimator exit Rate(n) = Rate(ICM)  K  πr 2 / f p r = collimator radius (1 cm) K = calibration factor (TFBC to ICM) f p = fraction of neutrons in the peak  accuracy: 10% at higher peak energy (180 MeV) 20% at lower peak energy (20 – 50 MeV)

P.Gauzzi10 th ICATPP Conference - October 9 th 2007 – Como 10 Scintillator calibration Trigger threshold calibration in MeV eq.el.en.: ADC counts  source to set the energy scale in MeV: 90 Sr  ─ endpoint = 0.56 MeV 90 Y  ─ endpoint = 2.28 MeV 25 keV/ADC count ADC counts Thr. [mV] ADC counts  6 counts/mV Events

P.Gauzzi10 th ICATPP Conference - October 9 th 2007 – Como 11  (%) E peak = 180 MeV Scintillator efficiency Agrees with the “thumb rule” (1%/cm) at thresholds above 2.5 MeV el.eq.en. Check of the method and of the beam monitor accuracy EnEn  (%)/ cm of scintillator Agrees with previous measurements in the same energy range after rescaling for the thickness

P.Gauzzi10 th ICATPP Conference - October 9 th 2007 – Como 12 Calorimeter calibration Cell response equalized: MIP peak at  550 ADC counts Trigger threshold calibration ADC counts Thr. [mV] Energy scale calibration with the MIP/MeV conversion factor from KLOE ( 1 MIP in one calorimeter cell  35 MeV eq. en.) ( see KLOE Collaboration, NIM A354 (1995),352 )  6 counts/mV Events

P.Gauzzi10 th ICATPP Conference - October 9 th 2007 – Como 13  (%) Comparison with our scintillator normalized to the same active material thickness Calorimeter efficiency Stable for different run conditions Very high efficiency w.r.t. the naive expectation ( ~ 2 MeV thr.) E peak = 180 MeV

P.Gauzzi10 th ICATPP Conference - October 9 th 2007 – Como 14 Energy spectrum from TOF n Rephasing is needed, since the trigger is phase locked with the RF (45 ns period) From TOF   spectrum of the neutrons Assuming the neutron mass  kinetic energy spectrum Energy spectrum can be reconstructed from TOF

P.Gauzzi10 th ICATPP Conference - October 9 th 2007 – Como 15 Efficiency vs energy Fast MC to test the sensitivity of the time distribution to the shape of the efficiency curve Better agreement with an efficiency decreasing with energy Some discrepancy in the low energy part of the spectrum T cell (ns) MC ● Data E kin (MeV) ε(%) T cell (ns) MC ● Data E kin (MeV) ε(%)

P.Gauzzi10 th ICATPP Conference - October 9 th 2007 – Como 16  (%)/ cm of scintillator  (%) - scint. In June 2007 we took data at lower peak energies: 21.8 and 46.5 MeV Large errors: big uncertainty in the beam halo evaluation worse accuracy of the beam monitors Correction factor for beam halo  0.9  0.1 Low energy data

P.Gauzzi10 th ICATPP Conference - October 9 th 2007 – Como 17 The ratio is almost independent from the halo (Ratio normalized to the same active material thickness)  (%) - calorimeter Calorimeter efficiency Very high efficiency at low threshold Agreement with the high energy measurements Correct. factor for beam halo  0.8  0.1

P.Gauzzi10 th ICATPP Conference - October 9 th 2007 – Como 18 MC simulation A detailed simulation of the calorimeter structure and of the beamline (source, collimator and concrete shielding) has been carried out with the FLUKA Code Neutron fluence along the beamline (n/cm 2 /29000n) collimator concrete shielding KLOECAL 7 Li target GLUE FIBERS base module replicas 200 layers LEAD

P.Gauzzi10 th ICATPP Conference - October 9 th 2007 – Como 19 Preliminary results of MC n Some discrepancy in the low energy part of the spectrum No cut in released energy No trigger simulation  Upper limit on   (%) E n (MeV) Data - Low thresh.

P.Gauzzi10 th ICATPP Conference - October 9 th 2007 – Como 20 Conclusions The first measurement of the detection efficiency for neutrons of MeV of a high sampling Pb-sci.fi. calorimeter has been performed at the The Svedberg Laboratory in Uppsala. Measurement of the n efficiency of a NE110 scintillator agrees with published results in the same energy range. The calorimeter efficiency, integrated over the whole neutron energy spectrum, ranges between % at the lowest trigger threshold. Study of the efficiency as a function of n energy is in progress. Full simulation with FLUKA is in progress. Further test foreseen for beginning 2008 at Louvain-la-Neuve (E n = 10 – 70 MeV; larger interbunch time)

P.Gauzzi10 th ICATPP Conference - October 9 th 2007 – Como 21 Spares

P.Gauzzi10 th ICATPP Conference - October 9 th 2007 – Como 22 Time structure 4.2 ms 2.4 ms 40 ns 41 ns  5 ns FWHM

P.Gauzzi10 th ICATPP Conference - October 9 th 2007 – Como 23 Preliminary results of MC Preliminary results of MC Simulated neutron beam: E kin = 180 MeV Each primary neutron has a high probability to have elastic/inelastic scattering in Pb In average 5.4 secondaries per primary neutron are generated, counting only neutrons above 19.6 MeV. target P el (%) P inel (%) Pb fibers glue neutrons above 19.6 MeV 62.2% photons 26.9% protons6.8% He-43.2% deuteron 0.4% triton 0.2% He-3 0.2% neutrons 94.2% protons 4.7% photons 1.1% Secondaries created in interactions of low energy neutrons (below 19.6 MeV) are - in average particles per primary neutron.

P.Gauzzi10 th ICATPP Conference - October 9 th 2007 – Como 24 Preliminary results of MC The enhancement of the efficiency appears to be due to the large inelastic production of neutrons in Pb. These secondary neutrons: - are produced isotropically; - are associated with a non negligible fraction of e.m. energy and of protons, which can be detected in the nearby fibers; - have low energy and then have a large probability to do new interactions in the calorimeter with neutron/proton/γ production. n Z(cm) p n1n1 n2n2 n3n3 n4n4 X(cm)

P.Gauzzi10 th ICATPP Conference - October 9 th 2007 – Como 25 Beam halo TSL beam experts measured a sizeable beam halo at low peak energy (21.8 and 46.5 MeV): –TFBC scan of the area near the collimator  integrated flux over the ICM area  5% of the “core” flux (with large uncertainty)  halo shape also measured Confirmed by our background counters Our calorimeter is larger than the projection of ICM area By integrating over the calorimeter we obtain an estimate of the halo contribution to the trigger rate of (20  10)% Only 10% on the reference scintillator due to the smaller area

P.Gauzzi10 th ICATPP Conference - October 9 th 2007 – Como 26  (%) KLOE vs others Comparison with other calorimeters: –KLOE: 23 cm thick –Crystall Ball: NaI 40.7 cm thick (NIMA462(2001),463) –GRAAL: BGO 24 cm thick (NIMA562(2006),85)