Frequency Scanned Interferometer for LC Tracker Alignment Hai-Jun Yang, Sven Nyberg, Keith Riles University of Michigan, Ann Arbor Department Of Energy(DOE)

Slides:



Advertisements
Similar presentations
Gravitational Wave Astronomy Dr. Giles Hammond Institute for Gravitational Research SUPA, University of Glasgow Universität Jena, August 2010.
Advertisements

Update of ILC Tracker Alignment Based on Frequency Scanned Interferometry University of Michigan ILC Group (Hai-Jun Yang, Alexander Harvey Nitz, Eui Min.
Frequency Scanned Interferometry for ILC Tracker Alignment University of Michigan ILC Group (Hai-Jun Yang, Eui Min Jung, Sven Nyberg, Keith Riles) Vancouver.
Direct Frequency Comb Spectroscopy for the Study of Molecular Dynamics in the Infrared Fingerprint Region Adam J. Fleisher, Bryce Bjork, Kevin C. Cossel,
Frequency Scanning Interferometry (FSI) measurements
Measuring the Speed of Light via an Open Cavity HeNe Laser James D. White, Daniel J. D’Orazio, Mark J. Pearson, Justin T. Schultz, Daniel Sidor, Michael.
R&D Towards a Solid-State Central Tracker in NA World-Wide Review of LC Tracking January 8, 2004 Bruce Schumm SCIPP & UC Santa Cruz.
Stephen Gibson, ATLAS Offline Alignment, 2 nd July Incorporating FSI with the Offline Alignment Overview ATLAS Group, University of Oxford Stephen.
S. M. Gibson, IWAA7 November ATLAS Group, University of Oxford, UK S. M. Gibson, P. A. Coe, A. Mitra, D. F. Howell, R. B. Nickerson Geodetic Grids.
Results The optical frequencies of the D 1 and D 2 components were measured using a single FLFC component. Typical spectra are shown in the Figure below.
Preliminary Results of Laser Ranging to Un-cooperative Targets at Shanghai SLR Station Yang FuMin, Zhang ZhongPing, Chen JuPing, Chen WanZhen, Wu ZhiBo,
MONALISA: Interferometric Position Monitor at the Nanometre Scale David Urner Paul Coe Matthew Warden Armin Reichold Oxford University.
Precision Displacement Measurement via a Distance Measuring Interferometer (DMI)
1 Laser Beam Coherence Purpose: To determine the frequency separation between the axial modes of a He-Ne Laser All sources of light, including lasers,
Currently: 3 year ( ) NSF-supported UF/IAP collaborative project "Methods and Instruments for High-Precision Characterization of LIGO Optical Components"
Uncertainty considerations for the calibration of transfer standard radiation thermometers Graham Machin, NPL.
Lecture 8: Measurement of Nanoscale forces II. What did we cover in the last lecture? The spring constant of an AFM cantilever is determined by its material.
The Linear Collider Alignment and Survey (LiCAS) Project Richard Bingham*, Edward Botcherby*, Paul Coe*, Grzegorz Grzelak*, Ankush Mitra*, Johannes Prenting.
LiCAS Project: FSI Overview Richard Bingham, Edward Botcherby, Paul Coe, John Green, Grzegorz Grzelak, Ankush Mitra, John Nixon, Armin Reichold University.
Initial Calibration and Stability Results from the LiCAS RTRS FSI System John Dale for the LiCAS Collaboration IWAA February 2008.
PERFORMANCE OF THE DELPHI REFRACTOMETER IN MONITORING THE RICH RADIATORS A. Filippas 1, E. Fokitis 1, S. Maltezos 1, K. Patrinos 1, and M. Davenport 2.
Use of a commercial laser tracker for optical alignment James H. Burge, Peng Su, Chunyu Zhao, Tom Zobrist College of Optical Sciences Steward Observatory.
R&D Towards a Solid-State Central Tracker in NA World-Wide Review of LC Tracking January 8, 2004 Bruce Schumm SCIPP & UC Santa Cruz.
Atacama Large Millimeter/submillimeter Array Expanded Very Large Array Robert C. Byrd Green Bank Telescope Very Long Baseline Array Green Bank Laser Rangefinders.
S. M. Gibson, P. A. Coe, Photon02, 5 th September Coordinate Measurement in 2-D and 3-D Geometries using FSI Overview ATLAS Group, University of.
LumiCal Alignment System Status report Leszek Zawiejski, Tomasz Wojtoń, Arkadiusz Moszczyński Institute of Nuclear Physics PAN, Cracow 25th FCAL Collaboration.
Study of optical properties of aerogel
§ 4 Optical Fiber Sensors
ISAT 303-Lab3-1  Measurement of Condition: Lab #3 (2005):  List of parameters of condition: –Linear distance, angular displacement, vibration, displacement,
NA62 Gigatracker Working Group Meeting 2 February 2010 Massimiliano Fiorini CERN.
1 Fabry-Perot Interferometer Fri. Nov. 15, Spectroscopy applications: Fabry- Perot Interferometer Assume we have a monochromatic light source.
GWADW, May 2012, Hawaii D. Friedrich ICRR, The University of Tokyo K. Agatsuma, S. Sakata, T. Mori, S. Kawamura QRPN Experiment with Suspended 20mg Mirrors.
D EDICATED S PECTROPHOTOMETER F OR L OCALIZED T RANSMITTANCE A ND R EFLECTANCE M EASUREMENTS Laetitia ABEL-TIBERINI, Frédéric LEMARQUIS, Michel LEQUIME.
IPBSM status and plan ATF project meeting M.Oroku.
David Urner, Oxford University, RHUL – June StaFF Stabilization of Final Focus Motion Stabilization with Nano-Meter Precision David Urner Paul Coe.
LECTURER PROF.Dr. DEMIR BAYKA AUTOMOTIVE ENGINEERING LABORATORY I.
September 14, Monday 4. Tools for Solar Observations-II Spectrographs. Measurements of the line shift.
ATF2 meeting LAPP; David Urner 1 David Urner Paul Coe Matthew Warden Armin Reichold Geoffrey Rayner MPhys alumnus MONALISA Monitoring, Alignment.
18 October Observational Astronomy SPECTROSCOPY and spectrometers Kitchin, pp
Modern Optics Lab Lab 6 Part 2: Interference Experiments  Observe interference by plane-parallel plates: Measure the thickness of the plates based on.
Frequency Scanned Interferometer for ILC Tracker Alignment Hai-Jun Yang, Sven Nyberg, Keith Riles University of Michigan, Ann Arbor SLAC LCD Tele-Conference.
ALCPG Tracking Session 1/11/03 K. Riles - U. Michigan1 Status Report on Michigan Group Tracker Alignment R&D Keith Riles, Jin Yamamoto, Haijun Yang University.
Tunable, resonant heterodyne interferometer for neutral hydrogen measurements in tokamak plasmas * J.J. Moschella, R.C. Hazelton, M.D. Keitz, and C.C.
Frequency Scanned Interferometer Demonstration System Tim Blass, Jason Deibel, Sven Nyberg, Keith Riles, Haijun Yang University of Michigan, Ann Arbor.
Frequency Scanned Interferometer Demonstration System Jason Deibel, Sven Nyberg, Keith Riles, Haijun Yang University of Michigan, Ann Arbor American Linear.
Multiple interference Optics, Eugene Hecht, Chpt. 9.
Haifeng Huang and Kevin K. Lehmann
1/10 Tatsuya KUME Mechanical Engineering Center, High Energy Accelerator Research Organization (KEK) ATF2-IN2P3-KEK kick-off meeting (Oct. 10, 2006) Phase.
Effects of long term annealing in p-type strip detectors irradiated with neutrons to Φ eq =1·10 16 cm -2, investigated by Edge-TCT V. Cindro 1, G. Kramberger.
A new method for first-principles calibration
Frequency Scanned Interferometry for ILC Tracker and Final Focus Magnet Alignment Haijun Yang, Tianxiang Chen, Keith Riles University of Michigan, Ann.
Charles University Prague Charles University Prague Institute of Particle and Nuclear Physics Absolute charge measurements using laser setup Pavel Bažant,
Frequency Scanned Interferometer for ILC Tracker Alignment Hai-Jun Yang, Sven Nyberg, Keith Riles University of Michigan, Ann Arbor International Linear.
MONALISA Update David Urner ATF2 Meeting Dec
IPBSM Operation 11th ATF2 Project Meeting Jan. 14, 2011 SLAC National Accelerator Laboratory Menlo Park, California Y. Yamaguchi, M.Oroku, Jacqueline Yan.
Laser Alignment System for LumiCal and BeamCal Arkadiusz Moszczyński, Wojciech Wierba, Leszek Zawiejski Institute of Nuclear Physics PAN, Cracow 22nd FCAL.
1 Electromagnetic waves: Multiple beam Interference Wed. Nov. 13, 2002.
ST7 Interferometer LIGO-G Z1 The ST7 Interferometer Andreas Kuhnert Robert Spero Jet Propulsion Laboratory California Institute of Technology.
10-meter Interferometer Results M. Woods (special thanks to Steve Myers and Tim Slaton) Jan. 31, 2000 Commissioning Setup System Noise Monte Carlo simulation.
Quantum optical experiment on measurement of the gravitational force acting on the neutron Session of the Programme Advisory Committee for Nuclear Physics.
SPM Users Basic Training August 2010 Lecture VIII – AC Imaging Modes: ACAFM and MAC Imaging methods using oscillating cantilevers.
Progress of Shintake Monitor (ATF2 IP-BSM) ATF2 weekly meeting 2008/9/3 Takashi Yamanaka The University of Tokyo.
Accuracy of the orbit measurement by KEKB BPM system for the study of ILC damping ring H. Fukuma (KEK) Requirement for the accuracy of BPM data.
Concentration Dependence of Line Shapes in the Band of Acetylene Matthew Cich, Damien Forthomme, Greg Hall, Chris McRaven, Trevor Sears, Sylvestre.
Date of download: 6/24/2016 Copyright © 2016 SPIE. All rights reserved. FSI setup. While the laser sweeps the frequency, the interferometer detector acquires.
Continuous Frequency Scanning Interferometry
Frequency Scanned Interferometry for ILC Tracker Alignment
Frequency Scanned Interferometer for ILC Tracker Alignment
Fabry-Perot Resonator
R&D Towards a Solid-State Central Tracker in NA
Presentation transcript:

Frequency Scanned Interferometer for LC Tracker Alignment Hai-Jun Yang, Sven Nyberg, Keith Riles University of Michigan, Ann Arbor Department Of Energy(DOE) Review July 19, 2004

Hai-Jun Yang - FSI for LC Tracker Alignment2 LC - Silicon Detector Barrel – 5 layers, inner/outer radii – 20/125 cm, Silicon drift detector or microstrips  r  = 10  m,  rz = 20  m Forward – 5 disks, double-sided silicon microstrips  r  = 7  m,  rz = 7  m Coverage - |cos(  )|=0.99 Boundary between barrel and forward disks - |cos(  )|=0.80 Wafer size – 10cm x 10cm Wafer thickness – 150  m Ref: SLAC-R-570 (2001) hep-ex/

July 19, 2004Hai-Jun Yang - FSI for LC Tracker Alignment3 Physics Goals and Background  To Carry out R&D toward a direct, quasi real time and remote way of measuring positions of critical tracker detector elements during operation.  The 1-Dimension accuracy of absolute distance is on the order of 1 micron. Basic idea: To measure hundreds of absolute point-to-point distances of tracker elements in 3 dimensions by using an array of optical beams split from a central laser. Absolute distances are determined by scanning the laser frequency and counting interference fringes. Assumption: Thermal drifts in tracker detector on time scales too short to collect adequate data samples to make precise alignment. Background – some optical alignment systems RASNIK system: used in L3, CHORUS and CDF, will be used in ATLAS and CMS Frequency Scanned Interferometer(FSI): will be used in ATLAS SCT [A.F. Fox-Murphy et al., NIM A383, 229(1996)] Focusing here on FSI system for LC tracker detector

July 19, 2004Hai-Jun Yang - FSI for LC Tracker Alignment4 Principle of Distance Measurement The measured distance can be expressed by + constant end corrections c - speed of light,  N – No. of fringes,  - scanned frequency n g – average refractive index of ambient atmosphere Assuming the error of refractive index is small, the measured precision is given by: (  R / R) 2 = (   N /  N) 2 + (   v /  ) 2 Example: R = 1.0 m,  = 6.6 THz,  N ~ 2R  /c = To obtain  R  1.0  m, Requirements:   N ~ 0.02,   v ~ 3 MHz

July 19, 2004Hai-Jun Yang - FSI for LC Tracker Alignment5 FSI Demonstration System (I)  Tunable Laser: New Focus Velocity 6308, 3-4 mW, nm.  Retroreflector: Edmund, D=2”, angle tolerance:  3 arc seconds.  Photodiode: Thorlabs PDA55, DC-10MHz, Amplified Si Detector, 5 Gain Settings.  Thorlabs Fabry-Perot Interferometer SA200, high finesse(>200) to determine the relative frequency precisely, Free Spectral Range (FSR) is 1.5 GHz, with peak FWHM of 7.5 MHz.  Thermistors and hygrometer are used to monitor temperature and humidity respectively.  PCI Card: NI-PCI-6110, 5 MS/s/ch, 12-bit simultaneous sampling DAQ.  PCI-GPIB Card: NI-488.2, served as remote controller of laser.  Computers: 1 for DAQ and laser control, 3 for analysis.

July 19, 2004Hai-Jun Yang - FSI for LC Tracker Alignment6 FSI Demonstration System (I) Laser Mirror Beamsplitters Photodetector Fabry-Perot Interferometer Retroreflector

July 19, 2004Hai-Jun Yang - FSI for LC Tracker Alignment7 FSI with Optical Fibers (II)

July 19, 2004Hai-Jun Yang - FSI for LC Tracker Alignment8 FSI with Optical Fibers (II) ♦ A key issue for the optical fiber FSI is that the intensity of the return beams received by the optical fiber is very weak. e.g. the core of the single mode optical fiber has diameter of 5  m. Geometrical Efficiency: ~ 6.25  10 –10 for a distance of 0.5 m  A novelty in our design is the use of a gradient index lens (GRIN lens – 0.25 pitch lens with D=1mm, L=2.58mm) to collimate the output beam from the optical fiber. The density of the outgoing beam is increased by a factor of ~1000 by using the GRIN lens. This makes it possible to split the laser beam into many beams to serve a set of interferometers simultaneously.

July 19, 2004Hai-Jun Yang - FSI for LC Tracker Alignment9 Multiple-Measurement Techniques If drift error(  ) occurs during the laser scanning, it will be magnified by a factor of Ω(Ω  /  ~ 67 for full scan of our tunable laser), OPD measured = OPD true + Ω   Plastic box and PVC pipes are constructed to reduce thermal drift. Assuming a vibration with one frequency: x vib (t) = a vib  cos(2  f vib t +  vib ) Fringe phase at time t:  (t) = 2   [OPD true + 2x vib (t)]/ (t)  N = [  (t)  (t0)]/2  = OPD true  /c + [2x vib (t)/ (t)- 2x vib (t0)/ (t0)] If we assume (t) ~ (t0) =, measured OPD can be written as, OPD meas = OPD true + Ω  [2x vib (t)- 2x vib (t0)] (1) OPD meas = OPD true  Ω  4a vib sin[  f vib (t-t0)]  sin[  f vib (t+t0)+  vib ] (2)  Two new multiple-distance measurement techniques are presented to extract vibration and to improve the distance measurement precision based on Eq.1 and Eq.2, respectively.

July 19, 2004Hai-Jun Yang - FSI for LC Tracker Alignment10 Two Multiple-Measurement Techniques  Fringes  F-P Peaks FSR=1.5 GHz  Fix the measurement window size (t-t0) and shift the window one F-P peak forward each time to make a set of distance measurements. The average value of all measurements is taken to be the final measured distance of the scan.  If t0 is fixed, the measurement window size is enlarged one F-P peak for each shift. An oscillation of a set of measured OPD reflects the amplitude and frequency of vibration.

July 19, 2004Hai-Jun Yang - FSI for LC Tracker Alignment11 Vibration Test  A PZT transducer was employed to produce controlled vibration of the retroreflector, f vib = 1.01  0.01 Hz, a vib = 0.14  0.02  m  Since the vibration is magnified for FSI during the scan, the expected reconstructed vibration amplitude is: A vib = 10.0  1.4  m.  The extracted vibration f vib =  Hz, A vib = 9.82  0.06  m  Measurable range f vib = 0.1 ~ 100 Hz, a vib = few nm ~ 1  m

July 19, 2004Hai-Jun Yang - FSI for LC Tracker Alignment12 Absolute Distance Measurements The scanning rate was 0.5 nm/s and the sampling rate was 125 KS/s. The measurement residual versus the No. of measurements/scan shown in Fig., (a) for one typical scan, (b) for 10 sequential scans.  It can be seen that the distance errors decrease with increasing N meas. N meas =1, precision=1.1  m (RMS) N meas =1200, precision=41 nm (RMS)  Multiple-distance measurement technique is well suited for reducing vibration effects and uncertainties from fringe & frequency determination, BUT not good for drift errors such as thermal drift. (a) (b)

July 19, 2004Hai-Jun Yang - FSI for LC Tracker Alignment13 Dual-Laser FSI (III) Two Lasers Two Choppers  A dual-laser FSI intended to reduce the drift errors is under study currently. Two lasers are operating simultaneously, but the laser beams are isolated by using two choppers. Laser #1: D 1 = D true + Ω 1  1 Laser #2: D 2 = D true + Ω 2  2 Drift errors:  1   2 =  D true = (D 2 -  D 1 ) / (1 -  ), Where  = Ω 2 / Ω 1

July 19, 2004Hai-Jun Yang - FSI for LC Tracker Alignment14 Comparison of FSI performances  National Institute of Standards and Technology (NIST): Air transport FSI, Distance: 30 cm – 5 m, Precision: ~ 250 nm by averaging measurements of 80 independent scans. [J.A. Stone et.al, Appl. Opt. V38. No. 28, 5981(1999)]  University of Oxford – ATLAS Group Optical fiber FSI, Distance: 20 cm – 1.2 m, Precision: ~215 nm by using dual-laser technique to reduce drift errors [P.A. Coe, Doctoral Thesis, U. of Oxford, 2001]  University of Michigan – NLC Group Optical fiber FSI, Distance: 10 cm – 0.6 m (measurable distance limited by bandwidth of our femtowatt photodetector, Hz) Precision: ~50 nm by using new multiple-distance measurement technique Vibration: Hz, > few nanometers, can be extracted precisely using new vibration extraction technique. [Submitted to Applied Optics, 2004]

July 19, 2004Hai-Jun Yang - FSI for LC Tracker Alignment15 Summary and Outlook  Two FSI demonstration systems, with or without optical fibers, were constructed to make high-precision absolute distance measurements.  Two new multi-distance-measurement analysis techniques were presented to improve absolute distance measurement and to extract the amplitude and frequency of vibration.  A high accuracy of ~50 nm for distances up to 60 cm under laboratory conditions was achieved.  Major error sources were estimated, and the expected error was in good agreement with measured error from real data.  We are investigating dual-laser scanning technique used by Oxford ATLAS group currently.  Michigan group rapidly coming up to speed on technology, but much work lies ahead.

July 19, 2004Hai-Jun Yang - FSI for LC Tracker Alignment16 BACKUP SLIDES

July 19, 2004Hai-Jun Yang - FSI for LC Tracker Alignment17 Absolute Distance Measurements Measured Distance(cm) Precision for Open box(  m) Precision for Close box(nm) Scan rate (nm/s) FSI(Air or Optical Fiber) ( ) 3.5 ~ 5.047, 43, 51 30, Air FSI ( ) 1.1, , Optical Fiber FSI ( ) N/A36, 32 45, Optical Fiber FSI ( ) 5.7, 4.456, 530.4Optical Fiber FSI ( ) N/A510.25Optical Fiber FSI Each precision listed is for standard deviation (RMS) of 10 scans.

July 19, 2004Hai-Jun Yang - FSI for LC Tracker Alignment18 Temperature Measurements Outside of Box Inside of Box

July 19, 2004Hai-Jun Yang - FSI for LC Tracker Alignment19 Error Estimations Error from uncertainties of fringe and frequency determination, dR/R ~ 1.9 ppm; if N meas = 1200, dR/R ~ 77 ppb Error from vibration. dR/R ~ 0.4 ppm; if N meas = 1200, dR/R ~ 11 ppb Error from thermal drift. Temperature fluctuations are well controlled down to 0.5 mK(RMS) in Lab by plastic box on optical table and PVC pipes shielding the volume of air near the laser beam. An air temperature change of 1 0 C will result in a 0.9 ppm change of refractive index at room temperature. The drift will be magnified during scanning. if N meas = 1200, dR/R ~ 0.9 ppm/K  0.5mK   (94) ~ 42 ppb. Error from air humidity and pressure, dR/R ~ 10 ppb. The total error from the above sources is ~ 89 ppb which agrees well with the measured residual spread of 90 ppb.

July 19, 2004Hai-Jun Yang - FSI for LC Tracker Alignment20 Systematic Error Estimations The major systematic bias comes from uncertainty of the Free Spectral Range (FSR) of the Fabry Perot interferometer used to determine scanned frequency range precisely, the relative error would be dR/R ~ 50 ppb if the FSR was calibrated by an wavemeter with a precision of 50 ppb. A wavemeter of this precision was not available for the measurement described here. Systematic bias from uncertainties of temperature, air humidity and barometric pressure scales should have negligible effect. The total systematic error of above sources is ~ 50 ppb.

July 19, 2004Hai-Jun Yang - FSI for LC Tracker Alignment21 RASNIK Demonstration System RASNIK provides alignment monitoring with submicron precision, developed at NIKHEF.

July 19, 2004Hai-Jun Yang - FSI for LC Tracker Alignment22 Fringe & F-P peaks for One Laser Fringe is intensity oscillation of two interference laser beams while scanning. FSR = 1.5 GHz

July 19, 2004Hai-Jun Yang - FSI for LC Tracker Alignment23 Fringes & F-P Peaks for Dual-Laser Laser-1 Laser-2

July 19, 2004Hai-Jun Yang - FSI for LC Tracker Alignment24 Components of FSI System