Preliminary Characterization Results: Fiber-Coupled, Multi-channel, Hyperspectral Spectrographs Carol Johnson, NIST MODIS-VIIRS Team Meeting January 26-28,

Slides:



Advertisements
Similar presentations
Good Practices of Fluorescence Spectroscopy
Advertisements

HIRES Technical concept and design E. Oliva, HIRES meeting, Brera (Milan, Italy)1.
OL 750 Diffuse Spectral Reflectance Measurement System
Brown Bag Lunch Lecture ABI Calibration
Spectral Resolution and Spectrometers
DESpec spectrographs Jennifer Marshall Darren DePoy Texas A&M University.
Optical Astronomy Imaging Chain: Telescopes & CCDs.
7. Beer’s Law and It’s Implications for Instrument Construction.
ABRF meeting 09 Light Microscopy Research Group. Why are there no standards? Imaging was largely an ultrastructure tool Digital imaging only common in.
Echelle spectra reduction with IRAF*
Reflected Solar Calibration Demonstration System Update J. McCorkel K. Thome D. Jennings B. McAndrew J. Hair D. Rabin A. Daw National Aeronautics and Space.
1 Lites FPP-SP Performance SOT #17 Meeting, NAOJ, April Solar-B FPP As-Built Performance of the FPP Spectro- Polarimeter October, 2004 FPP Team Bruce.
Calibration and Status of MOBY Dennis Clark, NOAA/NESDIS Carol Johnson, NIST Steve Brown, NIST Mark Yarbrough, MLML Stephanie Flora, MLML Mike Feinholz,
NOAA Research and Operations Marine Optical Buoy Design Review July 18-19, 2006 Summary Carol Johnson NIST.
NOAA Research and Operations Marine Optical Buoy Design Review July 18-19, 2006 Plan for calibration and maintenance of AHAB Uncertainty Budget: Laboratory.
CCD testing Enver Alagoz 12 April CCD testing goals CCD testing is to learn how to – do dark noise characterization – do gain measurements – determine.
Rachel Klima (on behalf of the MASCS team) JHU/APL MASCS/VIRS Data Users’ Workshop LPSC 2014, The Woodlands, TX March 17,2014 MASCS Instrument & VIRS Calibration.
IFTSUV: an Imaging Fourier Transform Spectrometer in the UV Anne Millard, 3 rd year PhD student, directed by P. Lemaire and J.-C. Vial.
Blue: Histogram of normalised deviation from “true” value; Red: Gaussian fit to histogram Presented at ESA Hyperspectral Workshop 2010, March 16-19, Frascati,
”MOBY-Net, an ocean color vicarious calibration system” PI: Kenneth Voss, Physics Dept., Univ of Miami Co-I’s: Carol Johnson (NIST), Mark Yarbrough (MLML),
14 October Observational Astronomy SPECTROSCOPY and spectrometers Kitchin, pp
First results of the tests campaign in VISIBLE in VISIBLE for the demonstrator 12 October 2007 SNAP Collaboration Meeting Paris Marie-Hélène Aumeunier.
18 October Observational Astronomy SPECTROSCOPY and spectrometers Kitchin, pp
Optical Aeronomy Calibration Facility CEDAR WORKSHOP JUNE, 2007 Jeff Baumgardner, Center for Space Physics Boston University.
Optronic Laboratories Inc. OL Series 750 Automated Spectroradiometric Measurement System Optronic Laboratories Inc. OL Series 750 Automated Spectroradiometric.
Eusoballoon optics test Baptiste Mot, Gilles Roudil, Camille Catalano, Peter von Ballmoos Test configuration Calibration of the light beam Exploration.
Figure 8.Color map of the geometric correction along the dispersion axis for segment A. Figure 4. Measured distortions for all PSA positions for segment.
DECam Daily Flatfield Calibration DECam calibration workshop, TAMU April 20 th, 2009 Jean-Philippe Rheault, Texas A&M University.
Overview of the “Geostationary Earth Radiation Budget (GERB)” Experience. Nicolas Clerbaux Royal Meteorological Institute of Belgium (RMIB) In collaboration.
1 Leonardo Pinheiro da Silva Corot-Brazil Workshop – October 31, 2004 Corot Instrument Characterization based on in-flight collected data Leonardo Pinheiro.
September 16, 2008LSST Camera F2F1 Camera Calibration Optical Configurations and Calculations Keith Bechtol Andy Scacco Allesandro Sonnenfeld.
Headwall Instrument Overview Laboratory Characterizations Geo-Location Field Characterization Data Product Description References.
GALFA Software Group GALFA-HI –Narrow-band spectral line data cubes of HI emission GALFA-RRL –Narrow-bad spectral line data cubes of recombination lines.
1 Reflected Solar Calibration Demonstration System - SOLARIS K. Thome, D. Jennings, B. McAndrew, J. McCorkel, P. Thompson NASA/GSFC.
OL 750 Measurement Systems OL 750 Measurement Systems Optronic Laboratories, Inc.
Jan 15, 2004WFC3 TIPS – John W. MacKenty1 WFC3 TIPS Presentation January 15, 2004 Optical Stimulus.
CALCON September B. Carol Johnson, Steven W. Brown and Howard W. Yoon National Institute of Standards and Technology Gaithersburg, MD Radiometric.
Science with Giant Telescopes - Jun 15-18, Instrument Concepts InstrumentFunction range (microns) ResolutionFOV GMACSOptical Multi-Object Spectrometer.
On the Evaluation of Optical Performace of Observing Instruments Y. Suematsu (National Astronomical Observatory of Japan) ABSTRACT: It is useful to represent.
Recent Solar Irradiance Data From SBUV/2 and OMI Matthew DeLand and Sergey Marchenko Science Systems and Applications, Inc. (SSAI) SOLID WP-2 Workshop.
Astronomical Observational Techniques and Instrumentation
Radiometry and Uncertainties from SORTIE (Spectral Ocean Radiance Transfer Investigation and Experiment) Kenneth Voss and Howard Gordon, Univ. of Miami.
Status of MOBY and Future Plans Dennis Clark Carol Johnson, NIST Steve Brown, NIST Mark Yarbrough, MLML Stephanie Flora, MLML Mike Feinholz, MLML.
Visible Spectro-polarimeter (ViSP) Conceptual Design David Elmore HAO/NCAR
Stray Light or Spectral Out-of-Band in MOS/MOBY Steve Brown & Carol Johnson (NIST) Michael Feinholz & Stephanie Flora (MLML) Jim Mueller (CHORS)
In conclusion the intensity level of the CCD is linear up to the saturation limit, but there is a spilling of charges well before the saturation if.
Multi-channel, Hyperspectral Spectrograph System using a Volume Phase Holographic Transmission Grating for In-Water Radiometry B. Carol Johnson, 1 Steven.
Building an Integrated Ocean Color Sensor Web at the Land-Sea Interface UCSC UARC/ARC Ames GSFC BSI SARP.
Satlantic HyperPro II Freefalling hyperspectral radiometer Can be deployed in “floater” mode 16 bit ADC at 12 Hz (maximum) 0 - ~100 m depth rating, descent.
# x pixels Geometry # Detector elements Detector Element Sizes Array Size Detector Element Sizes # Detector elements Pictorial diagram showing detector.
Optical Non-Invasive Approaches to Diagnosis of Skin Diseases
TEMPO Instrument Update
Relative Spectral Response and Flat Fields with Internal Calibration Lamps Luisa M. Lara IAA-CSIC Granada (SPAIN)
Single Object & Time Series Spectroscopy with JWST NIRCam
R&O Buoy Spectrograph System
TEMPO Instrument Update
VIRTIS Operations at Lutetia
AIRS (Atmospheric Infrared Sounder) Instrument Characteristics
An IFU slicer spectrometer for SNAP
Instrument Characterization: Status
What Is Spectral Imaging? An Introduction
Volume 84, Issue 6, Pages (June 2003)
Detective Quantum Efficiency Preliminary Design Review
Optical Non-Invasive Approaches to Diagnosis of Skin Diseases
Status of Equatorial CXRS System Development
Sensor calibration.
Astronomical Observational Techniques and Instrumentation
OL 750 Measurement Systems
Hyperspectral Terminology
Volume 106, Issue 5, Pages (March 2014)
Presentation transcript:

Preliminary Characterization Results: Fiber-Coupled, Multi-channel, Hyperspectral Spectrographs Carol Johnson, NIST MODIS-VIIRS Team Meeting January 26-28, 2010 Washington, DC

Marine Optical System & Data Stream In MOS, light is input sequentially into the dual spectrographs using optical fibers and a rotating mirror assembly (in place of the cosine collector). The full slit is imaged onto CCD detectors. On a typical day, it took 27 min to acquire a full data set, with integration times of between 1 and 30 sec (Es vs Lu collectors) for the CCDs. A dark scan, three light scans, and a dark scan are taken at each channel. The MOS optical system In-Water Diver Cals MOBY

New Sensor – Simultaneous Acquisition Resonon Volume Phase Holographic (VPH) in line spectrograph Romack fiber optic input (currently 14 channels) Princeton Instrument PIXIS CCD Blue In Line Spectrograph (BILS) Red In Line Spectrograph (RILS) The inputs to the optical fibers are at the desired locations. The fiber outputs are aligned vertically at the entrance slit. The prism-grating-prism in-line optical system (Resonon, Inc.) images the different input channels at the same time on the CCD camera), spaced along the slit direction. RILS image of diffuse solar flux Average net ADU/sec for each channel

Project History & Status Breadboard two systems, SIRCUS characterization, field tests –JY CP140 spectrograph, Andor camera, 4 inputs –Kaiser Holospec, Apogee camera, 6 inputs Custom optical designs –Resonon spectrographs (blue and red), Romack input fiber bundles, and Princeton Instruments cameras are procured and operational –Characterizations well underway Testing of packaged field system Aug 2010

Simultaneous Systems & Sample Number Band averaged results with a prototype 6-channel hyperspectral system tested in Case 1 waters off Oahu. Five to 100 scans were acquired with 4 sec integration times. Yarbrough, M.A., S. Flora, M.E. Feinholz, T. Houlihan, Y.S. Kim, S.W. Brown, B.C. Johnson, K. Voss and D.K Clark (2007). “Simultaneous measurement of up-welling spectral radiance using a fiber-coupled CCD spectrograph.” Proc. SPIE 6680, 66800J-1 to 66800J-11. The simultaneous design places no restriction on the number of samples averaged. This is an advantage compared to MOBY, where the sequential measurements of Es, LuMid, Es, LuTop, Es, LuBot, Es, LuMOS, Es means an increase in number of samples would increase the time between these data collections, impacting the determination of Lw.

Simultaneous Systems & Correlated Noise The effect of correlations in the light field was investigated by deriving L w ( ) from four simultaneous L u ( )s (Kaiser/Apogee), and by randomly sampling the L u ( ) scans in time to simulate the current MOBY sampling statistics. The measurement uncertainty was reduced between 20% to 60% for the ocean color bands.

In-Line Spectrograph Parameters from Optical Design ParameterBlueRed Size, cm13.7 x x 43.2 Spectral coverage, nm Spectral resolution, nm Image at focal plane, mm13 x 13 Slit dimensions, mm13 x Thermal effect, pixel/deg C< 0.05 pixel 38 line pr / mm76 at 545 nm61 at 700 nm Throughput, %74.8 at 430 nm72.5 at 700 nm Ghosting / Stray Light< 0.5% at 420 nm< 0.6% at 520 nm “Report on Blue and Red Imaging Spectrometers for MOBY,” Michael Kehoe and Casey Dodge, Resonon, Inc.

Characterization Results To Date Two Spectrographs: –Blue system: Romack fiber bundle (14 fibers, 800  m core); Blue In-Line Spectrograph (BILS); Princeton Pixis 1024B back illuminated CCD (13  m pixels) –Red system: Romack fiber bundle (14 fibers, 800  m core); Red In-Line Spectrograph (RILS); Princeton Pixis 1024BR back illuminated, deep depletion CCD (13  m pixels) Tests to date: –noise, wavelength calibration, spectral stray light, imaging behavior, system response

Noise Tests showed –low dark count rate,  =0.0028,  = ADU/sec (track 7) –digitization at 2MHz vs 100kHz did not compromise dark noise CCD Camera System Features –Fast & Slow digitization rate; three gain settings (number of electrons to get one ADU) –four stage TEC holds CCD at -75  C –~ 600 ADU built-in offset

Wavelength Calibration w/ Hg Lamp Coverage 372 to 734 nm, step nm No dependence with track – if smile were an issue we would see this here Preliminary fits to polynomials

Laser Characterization on SIRCUS BILS initial testing on SIRCUS; focus was on Track 7. Results are very encouraging, both in spectral and spatial dimension: ~ The spectral stray light is the best we’ve seen so far for a single grating system. BILS is 20x better than MOS in terms of integrated area.

Imaging Behavior There is measureable keystone, which affects our planned on-chip hardware binning (a method to reduce noise and increases dynamic range). This can be corrected by optimizing the optical design in the final systems. Track 8Track 1 There is an obvious non-uniformity in the spatial (slit) direction, the “saddle.” This effect appears stable and is under investigation: partial coherence at the slit or spatial mode effects in the short fibers used in the test? All results shown here were averaged over the 41 rows between the vertical lines indicated.

Inter-reflections in BILS Track 2 illuminated, artifact on Track 13: Note scale change in this split image (100 ADU vs 15,000 ADU) We discovered that there is an artifact that appears in the same wavelength region (~410 nm) independent of the wavelength of the input flux and paired with the track illuminated: light on 1 shows on 14, light on 2 shows on 13,.... light on 7 shows on 7, etc. The explanation by Resonon’s modeling is dispersed light reflecting off the CCD is recombining upon reentry into the prism/grating/prism assembly (e.g, as in a double subtractive system), and then being imaged in zero order. This inter- reflection is caused by a protective coating of SiO 2 that will not be used in a final production system.

System Response & Preliminary Stray Light Correction Algorithms A preliminary SLC matrix was determined from the SIRCUS data. Validation data were acquired using various sources (filtered lamp; LED, laser). No SLC corrections are applied to the data shown here. (A)the known L( ) for this validation source; (B)the net ADU/sec for BILS; (C)the derived system response from B/A Note: A and B are not that different for PER! (A) (B)(C) The responses disagree in regions where stray light (or the artifact) is contributing strongly. This will be removed with the SLC and algorithm.

Conclusion We’re on track; no hardware show stoppers for a superb sensor for field radiometry Team members: –NIST (Johnson, Saunders, Li, Clark, Parr) –MLML (Yarbrough, Feinholz, Flora, Houlihan) –Resonon (Kehoe, Dodge, Swanson) –UM (Voss) Implementation for MOBY-C depends on funding profile