Energy calibration of the threshold of Medipix for ATLAS Céline Lebel Université de Montréal presenting for the Institut of Experimental.

Slides:



Advertisements
Similar presentations
NUCP 2371 Radiation Measurements II
Advertisements

IEAP CTU Prague 8 th December, 2004Tomáš Holý Neutron tomography with Medipix-2 Tomáš Holý, Jan Jakůbek, Stanislav Pospíšil, Josef Uher.
Dante Nakazawa with Prof. Juan Collar
Gamma-Ray Spectra _ + The photomultiplier records the (UV) light emitted during electronic recombination in the scintillator. Therefore, the spectrum collected.
Study of plastic scintillators for fast neutron measurements
Radiation Detectors / Particle Detectors
Beam tests of Fast Neutron Imaging in China L. An 2, D. Attié 1, Y. Chen 2, P. Colas 1, M. Riallot 1, H. Shen 2, W. Wang 1,2, X. Wang 2, C. Zhang 2, X.
Detection of Gamma-Rays and Energetic Particles
1 Pixel Readout of GEMs Motivation The Setup Results Next Steps A. Bamberger, K. Desch, J. Ludwig, M. Titov, U. Renz, N. Vlasov, P. Wienemann, A. Zwerger.
Test of Pixel Sensors for the CMS experiment Amitava Roy Purdue University.
Third International Conference on Frontier Science Villa Mondragone – Monteporzio Catone Physics and Astrophysics in Space June 14, 2004 Mercedes Paniccia.
Particle Interactions
Stopping Power The linear stopping power S for charged particles in a given absorber is simply defined as the differential energy loss for that particle.
Institute of Experimental and Applied Physics Czech Technical University in Prague 11th December 2007 Michal Platkevič RUIN Rapid Universal INterface for.
Medipix2-based radiation camera and it's application in RELAXd and ATLAS projects Zdenek Vykydal Institute of Experimental and Applied Physics, Czech Technical.
Why silicon detectors? Main characteristics of silicon detectors: Small band gap (E g = 1.12 V)  good resolution in the deposited energy  3.6 eV of deposited.
Applications of a Pixellated Detection System to Digital Mammography
Future usage of quasi-infinite depleted uranium target (BURAN) for benchmark studies Pavel Tichý Future usage of quasi-infinite depleted uranium target.
GERMANIUM GAMMA -RAY DETECTORS BY BAYAN YOUSEF JARADAT Phys.641 Nuclear Physics 1 First Semester 2010/2011 PROF. NIDAL ERSHAIDAT.
1 G. Pellegrini The 9th LC-Spain meeting 8th "Trento" Workshop on Advanced Silicon Radiation Detectors 3D Double-Sided sensors for the CMS phase-2 vertex.
Surface events suppression in the germanium bolometers EDELWEISS experiment Xavier-François Navick (CEA Dapnia) TAUP Sendai September 07.
Fully depleted MAPS: Pegasus and MIMOSA 33 Maciej Kachel, Wojciech Duliński PICSEL group, IPHC Strasbourg 1 For low energy X-ray applications.
Digital analysis of scintillator pulses generated by high-energy neutrons. Jan Novák, Mitja Majerle, Pavel Bém, Z. Matěj 1, František Cvachovec 2, 1 Faculty.
Performance limits of a 55  m pixel CdTe detector G.Pellegrini, M. Lozano, R. Martinez, M. Ullan Centro Nacional de Microelectronica, Barcelona, 08193,
Advanced semiconductor detectors of neutrons
© Jimoid.com 2005 Ionising Radiation There are two types of radiation; ionising and non-ionising. Radiation Ionising Non-ionising Indirectly ionising (neutral.
MX-10: Pixel Particle Detector V.Viswanathan, S.Martinek, M.Honig, P.Hubner Dr.Vijayaragavan VISWANATHAN 27/10/2013 IEEE-NSS: New Detector.
Alpha Particle Scintillation Analysis in High Pressure Argon Daniel Saenz, Rice University Advisor: Dr. James White, Texas A&M University.
NA62 Gigatracker Working Group 28 July 2009 Massimiliano Fiorini CERN.
Experimental Nuclear Physics Some Recent Activities 1.Development of a detector for low-energy neutrons a. Hardware -- A Novel Design Idea b. Measure the.
Chapter 5 Interactions of Ionizing Radiation. Ionization The process by which a neutral atom acquires a positive or a negative charge Directly ionizing.
Institute of Experimental and Applied Physics Czech Technical University in Prague IEAP – CTU Prague 1 Current status and future development of neutron.
Review of synthesis of super heavy elements: reactions, decays and characterization. Experimental Setup of MASHA. Results of first experiments. study.
Calorimeter in front of MUCh Mikhail Prokudin. Overview ► Geometry and acceptance ► Reconstruction procedure  Cluster finder algorithms  Preliminary.
Introductory presentation Benedikt Bergmann Institute of Experimental and Applied Physics, Czech Technical University in Prague, Czech Republic Vienna.
Geant4 User Workshop, Catania, Italy, 15th October, 2009
Radiation hardness of Monolithic Active Pixel Sensors (MAPS)
Present and future activities in LSM (2 nd LSM Extension WORKSHOP, Oct.16, 2009) (1) TGV experiment – measurement of 2 EC/EC decay of 106 Cd (2) R&D of.
1 Cost Room Availability Passive Shielding Detector spheres for accelerators Radiation Detection and Measurement, JU, First Semester, (Saed Dababneh).
Goddard February 2003 R.Bellazzini - INFN Pisa A new X-Ray Polarimeter based on the photoelectric effect for Black Holes and Neutron Stars Astrophysics.
Activities within the ARDENT- project: November 2012 – June 2013 Benedikt Bergmann Institute of Experimental and Applied Physics, Czech Technical University.
Studies on Activation in the ATLAS cavern with MPX Detectors Benedikt Bergmann Institute of Experimental and Applied Physics, Czech Technical University.
We are investigating the dependence of efficiency of diamond detector samples on accumulated radiation dose. We have Sr 90  -source of known activity.
A Prototype Diamond Detector for the Compton Polarimeter in Jefferson lab, Hall C Medium Energy Physics Group Amrendra.
Progress with GaAs Pixel Detectors K.M.Smith University of Glasgow Acknowledgements: RD8 & RD19 (CERN Detector R.&D. collaboration) XIMAGE (Aixtron, I.M.C.,
Ray Bunker (UCSB) – APS – April 17 th, 2005 CDMS SUF Run 21 Low-Mass WIMP Search Ray Bunker Jan 17 th -DOE UCSB Review.
Spectra distortion by the interstrip gap in spectrometric silicon strip detectors Vladimir Eremin and.
ATLAS Transition Radiation Tracker End-cap Quality Control and the Characterization of Straw Deformations Michael Kagan University of Michigan Supervisor:
Characterization of 3D and planar Si diodes with different neutron converter materials R. Mendicino, M. Dalla Palma, M. Boscardin, S. Ronchin, G.-F. Dalla.
Equalization of Medipix2 imaging detector energy thresholds using measurement of polychromatic X-ray beam attenuation Josef Uher a,b, Jan Jakubek c a CSIRO.
Medipix3 chip, downscaled feature sizes, noise and timing resolution of the front-end Rafael Ballabriga 17 June 2010.
Institute of Experimental and Applied Physics Czech Technical University in Prague IEAP – CTU Prague 1 BrightnESS T4.1 Meeting Activity at IEAP CTU in.
F. Murtas CERN-INFN Frascati December 2 th 2013 – He3 free detectors Italy Workshop 1 Front End Electronics for GEM detector  Neutron detectors and GEM.
Neutrinoless double beta decay (0  ) CdTe Semico nductor Band gap (eV) Electron mobility (cm 2 /V/s) Hole mobility (cm 2 /V/s) Density (g/cm 3.
F.Murtas G.Claps, E.Baracchini 1 EDIT FrascatiOctober 22 h 2015 GEMPIX detectors : GEM with higly pixelated readout Introduction GEMPIX Detectors GEMPIX.
Report (2) on JPARC/MLF-12B025 Gd(n,  ) experiment TIT, Jan.13, 2014 For MLF-12B025 Collaboration (Okayama and JAEA): Outline 1.Motivation.
TPC for ILC and application to Fast Neutron Imaging L. An 2, D. Attié 1, Y. Chen 2, P. Colas 1, M. Riallot 1, H. Shen 2, W. Wang 1,2, X. Wang 2, C. Zhang.
DEVELOPMENT OF PIXELLATED SEMICONDUCTOR DETECTORS FOR NEUTRON DETECTION Prof. Christer Fröjdh Mid Sweden University.
Progress with GaAs Pixel Detectors
Particle tracking with semiconductor pixel detector Timepix
C. Leroy Montreal – Prague Collaboration
GLAST LAT tracker signal simulation and trigger timing study
Performance of the Medipix and Timepix devices for the recognition of electron-gamma radiation fields C. Teyssier1,3, J. Bouchami1, F. Dallaire1, J. Idarraga1,
The response function of semiconductor pixel detectors for highly ionizing particles Jan Jakubek, P. Soukup Institute of Experimental and Applied Physics,
Scintillation Counter
High Rate Photon Irradiation Test with an 8-Plane TRT Sector Prototype
HE instrument and in-orbit performance
PERFORMANCE OF THE METAL RADIATION MONITORING SYSTEMS
Why silicon detectors? Main characteristics of silicon detectors:
Gain measurements of Chromium GEM foils
Presentation transcript:

Energy calibration of the threshold of Medipix for ATLAS Céline Lebel Université de Montréal presenting for the Institut of Experimental and Applied Physics of the Czech Technical University in Prague

/ 52C. Lebel2 People involved in this project Institut of Experimental and Applied Physics of the Czech Technical University in Prague –Stanislav Pospíšil –Jan Jakůbek –Josef Uher –Vlastimil Král –Michal Platkevič –Vladimír Tichý Charles University –Michal Suk –… Université de Montréal –Claude Leroy –Céline Lebel

/ 52C. Lebel3 Outline Medipix Medipix in ATLAS: challenges foreseen Energy calibration of the low threshold –Equalization –Photons: Decreasing Flux –Alphas –Electrons –Neutrons Summary and Outlook

/ 52C. Lebel4 Medipix2 device - a single X-ray photon counting pixel detector Medipix2 Pixels: 256 x 256 Pixel size: 55 x 55 μm 2 Area: 1.5 x 1.5 cm 2 Medipix2 Quad Pixels: 512 x 512 Pixel size: 55 x 55 μm 2 Area: 3 x 3 cm 2 Planar pixellated detector (Si, GaAs, CdTe, thickness: 300/700/1000mm) Bump-bonded to Medipix readout chip containing in each pixel cell: - amplifier, - double discriminator - and counter Detector chip Medipix chip Bump-bonding 

ATLAS

/ 52C. Lebel6 LHC

/ 52C. Lebel7 ATLAS Height: 22m Width: 44 m Weight: 7000 tons

/ 52C. Lebel8 Medipix in ATLAS: the challenge ATLAS environment has high radiation fluxes Problem: is the device radiation hard? –We don’t know. maximum information in the shortest timeThe data acquisition must be optimized to obtain the maximum information in the shortest time. What kind of information are we looking for? –Composition and spectroscopic characteristics of the radiation field inside ATLAS Number of particles, energy and type

/ 52C. Lebel9 Approach for ATLAS: Layers All events are accepted Particles depositing less than X keV are rejected Minimum ionizing particles are rejected Counting mode: all particles deposit more than X MeV in a single pixel Low threshold

/ 52C. Lebel10 Approach for ATLAS: multiple area Polyethylene: Fast neutron detection 6 LiF: Thermal neutron detection Comet-type track from protons Cluster-type track from α and triton

/ 52C. Lebel11 Adaptation of the Medipix2 device for position sensitive detection of neutrons Silicon pixel detector can not detect neutrons directly. Conversion of neutrons to detectable radiation in a converter layer (via nuclear reactions or recoiled protons) deposited on the detector surface is needed. Converter materials for thermal neutrons: 6 Li: 6 Li + n  a (2.05 MeV) + 3 H (2.72 MeV) 10 B: 10 B + n  a (1.47 MeV) + 7 Li (0.84 MeV) + g (0.48MeV)(93.7%) 10 B + n  a (1.78 MeV) + 7 Li (1.01 MeV) (6.3%) 113 Cd: 113 Cd + n  114 Cd + g (0.56MeV) + conversion electrons 155 Gd: 155 Gd + n  156 Gd + g (0.09, 0.20, 0.30 MeV) + conversion electrons 157 Gd: 157 Gd + n  158 Gd + g (0.08, 0.18, 0.28 MeV) + conversion electrons Converter for fast neutrons: polyethylene foil Detector: 150 – 700 mm thick silicon pixel detector (pixel size 55 mm) bump bonded to Medipix-2 readout chip.

Equalization and parameters

/ 52C. Lebel13 Equalization Medipix is composed of pixel channels. All are not created equal! –Pixelman allows for the equalization of the low threshold (THL) This is a small but necessary adjustment Two methods: –Use noise edge Useful to remove the noisiest pixels –Use noise center Centers the noise → This puts the necessary offsets so that all pixels read the same energy

/ 52C. Lebel14 Parameters to adjust THL FBK IKrum

Particle detection

/ 52C. Lebel16 Photons Photons can be detected in silicon only by transferring their energy to charged particles. Photoelectric effect All energy transferred to the material Compton diffusion Energy partially transferred to the material Pair production All energy transferred to the material

/ 52C. Lebel17 Photons

/ 52C. Lebel18 Setup for sources

/ 52C. Lebel19 Photons – 55 Fe Production of X-rays with hν = 6 keV –Primary interaction in silicon: Photoelectric effect Energy of the photoelectrons: E = hν – BE for silicon, 1.8 keV is the binding energy of the K-shell (predominant at that photon energy) Energy deposited by the photoelectron→ 4.2 keV Range of such electrons < 1 μm (dE/dx = 7.33 keV/μm) Deposition is always in a single pixel!

/ 52C. Lebel20 Photons – 6 keV X-Rays (1 sec) THL-FBK = THL-FBK = THL-FBK =

/ 52C. Lebel21 Photons – 6 keV X-Rays

/ 52C. Lebel22 Photons – 241 Am Decays to 237 Np, emitting alphas and gammas. –Main gamma lines: 59.5 keV at 35.9 % and 26.3 keV at 2.4 % At 59.5 keV: –Photoelectric effect with the K-shell Energy of the electron: 57.7 keV Range using CSDA approximation: ~30 μm (Source:

/ 52C. Lebel23 Photons – 60 keV THL-FBK = THL-FBK = THL-FBK =

/ 52C. Lebel24 Photons – 60 keV gammas THL-FBK = → E = 58 keV !

Alphas

/ 52C. Lebel26 Alphas Bethe-Bloch dE/dx x

/ 52C. Lebel27 THL-FBK = THL-FBK = THL-FBK = THL-FBK =

/ 52C. Lebel28 Changing Alpha Energy E α = 4.20 MeV E α = 1.41 MeV E α = 0.69 MeV The threshold is too low. cannot These energies cannot calibrate the threshold with this FBK. The threshold is too low. cannot These energies cannot calibrate the threshold with this FBK.

/ 52C. Lebel29 Raising FBK E α = 4.20 MeV E α = 0.74 MeV Tentatively: THL-FBK = → 735 keV Tentatively: THL-FBK = → 735 keV

/ 52C. Lebel30 About full depletion voltage … Flux (part/sec) Average Cluster Size MeV deposit → Range of ~21 μm V fd = 46 ± 1 V Collection starts at 40 V → V fd = 46 ± 1 V

/ 52C. Lebel31 Cluster Size as function of U bias E E

/ 52C. Lebel32 Does the threshold depend on the USB or the Medipix used? NO! USB1-Medipix1 USB2-Medipix2 USB1-Medipix2 (with USB2 equalization)

Electrons

/ 52C. Lebel34 THL-FBK = Electrons Source 90 Sr-Y Average electron energy: 935 keV → mip

/ 52C. Lebel35 THL-FBK = THL-FBK= Electrons

/ 52C. Lebel36 THL-FBK = THL-FBK= Electrons

/ 52C. Lebel37 THL-FBK = THL-FBK= Electrons

/ 52C. Lebel38 THL-FBK = THL-FBK= Electrons

/ 52C. Lebel39 THL-FBK = THL-FBK= Electrons

/ 52C. Lebel40 THL-FBK = THL-FBK= Electrons

/ 52C. Lebel41 THL-FBK = THL-FBK= Electrons

/ 52C. Lebel42 Some – very preliminary – results E α = 4.20 MeV E α = 0.69 MeV Electrons

/ 52C. Lebel43 Some – very preliminary – results 6 keV X-Rays 60 keV Gammas Electrons

Neutrons

/ 52C. Lebel45 Neutrons detection through reaction with Polyethylene Using the reaction n + 1 H → p + n Fast neutrons from reactor Thank you, Radek Skoda, for making this measurement possible!

/ 52C. Lebel46 Position of the polyethylene layer Polyethylene

/ 52C. Lebel47 Results obtained at the reactor Sparrow

/ 52C. Lebel48 Importance of threshold THL-FBK = FBK = 128 THL-FBK = FBK = 200 THL-FBK = FBK = 200

/ 52C. Lebel49 Examples of response of MEDIPIX-USB device to fast monochromatic neutrons: 17MeV neutrons, flux about 10 4 n/(s.cm 2 ) The direction of the neutrons with respect to the image was upstream (from bottom to top). The huge background is due to gamma rays which accompany neutrons. Half of the sensor (the right-hand side) was covered with a CH2 foil about 1.3 mm thickness. One can clearly recognize long and rather thick tracks of recoiled protons (up to 2 mm, vertically oriented) and big tracks and clusters generated via 28Si(n,α)25Mg, 28Si(n,p)28Al nuclear reactions in the body of the silicon detector. These events are displayed on the dense background caused by tracks and traces of electrons from interactions of gamma rays. One can even recognize that proton tracks shapes follows a Bragg law.

/ 52C. Lebel50 Conclusions ONEONE energy (58 keV) has been definitively attributed to a threshold value heavy charged particle from electronsA threshold has been identified to isolate the contribution of heavy charged particle from electrons All USB and Medipix give the same variation as a function of effective threshold PhotonsHeavy charged particles ElectronsPhotons, Heavy charged particles and Electrons give specific tracks which can be identified. Adding a Polyethylene layer allows the detection of fast neutrons

/ 52C. Lebel51 Outlook More energy lines must be attributed to specific values of THL-FBK –Different X-Rays! More on neutrons Analysis of heavy charged particle tracks: influence of plasma effect in silicon

Děkuji vám! Thank you!