Are The Less Invasive Techniques For Monitoring Cardiac Output As Accurate As The Pulmonary Artery Catheter? Dr Andrew Rhodes St George’s Hospital London.

Slides:



Advertisements
Similar presentations
Paul Nolan, Galway University Hospitals
Advertisements

Measurement of CV variables. In vitro Total control of confounding variables –Vasomotion, temperature changes, autoregulation, mean BP Most accurate because.
Monitoring The Emergency Patient ABDULRAHMAN SINDI ( Emergency Medicine Resident)
Preoperative Assessment in Private Practical Pointers for Private Practitioners Dr Adam Molnar MBBS FANZCA Victorian Anaesthetic Group.
Haemodynamic Monitoring
Arterial Catheters Systemic arterial blood pressure is most accurately measured by placing a catheter directly into a peripheral artery. Peripheral arterial.
Cardiac Output And Hemodynamic Measurements Iskander Al-Githmi, MD, FRCSC, FCCP Asst. Professor of Surgery King Abdulaziz University.
Serum Lactate Post-Antibiotic antibiotic † † * * Kumar et al. J Infect Dis 2006; 193:
Sepsis.
liver function monitoring
A REVIEW OF FUNCTIONAL HAEMODYNAMIC MONITORING AJ van den Berg.
Haemodynamic Monitoring
Swan-Ganz Catheter- Balloon flotation Pulmonary Artery catheter Use for monitoring critically ill patients (mostly in the ICU) Catheterization only possible.
Pre-Operation Evaluation of Thoracic Surgery Patient: Spirometry and Pulmonary Exercise test (PXT) 吳惠東.
Set your intravascular volume right Jost Mullenheim James Cook University Hospital, Middlesbrough.
Chapter 15 Assessment of Cardiac Output
Hemodynamic monitoring Prof. Jean-Louis TEBOUL University Paris-South
Pressure, Flow, and Resistance Understanding the relationship among pressure, flow and resistance can help you understand how cardiac output and vascular.
Hemodynamics Is defined as the study of the forces involved in blood circulation. Hemodynamic monitoring is used to assess cardiovascular function in the.
Simple approaches to difficult topics
H e a r t S T A T blood flow and disease monitoring intervention Cover.
Protective Lung Strategy Mazen Kherallah, MD, FCCP
Circulatory Failure 6 th November Physiology and pathophysiology of the heart and circulation Pathophysiological effects of altered intravascular.
Wet, Dry, or Even? Some Ways of Looking at Volume Status without a Pulmonary-Arterial Catheter Anne K. Sutherland, MD Critical Care Medicine St. Barnabas.
Richard Beale Guy’s and St Thomas’ Hospital Trust London, UK
Cardiac Output Monitoring in Cardiac Surgery with Cardiopulmonary Bypass K. Lebedinski, A. Vetchinkin St. Petersburg.
Haemodynamic Monitoring Theory and Practice. 2 Haemodynamic Monitoring A.Physiological Background B.Monitoring C.Optimizing the Cardiac Output D.Measuring.
Haemodynamic Monitoring Theory and Practice. 2 Haemodynamic Monitoring A.Physiological Background B.Monitoring C.Optimising the Cardiac Output D.Measuring.
How Can I Measure Cardiac Output In A Patient With Shock? Jon Sevransky MD International Consensus Conference Paris France April 27, 2006.
Can pharmacists improve outcomes in hypertensive patients? Sookaneknun P (1), Richards RME (2), Sanguansermsri J(1), Teerasut C (3) : (1)Faculty of Pharmacy,
Use the right tool for the right job!
Diagnosing cardiac disease states using a minimal cardiovascular model Christina Starfinger Wednesday, 21 March 2007 HRSC Scientific Meeting.
Health Time Smart-Skin sensing for Real-Time Lifestyle guidance.
PWV & PWA Parameter Extraction / Transfer Function.
ITU Journal Club: Dr. Clinton Jones. ST4 Anaesthetics.
SEVERAL MAJOR CAPABILITIES OF CRITICAL CARE INVIRONMENT 1.Process store and intergrate physiologic and diagnostic information from various sources. 2.Accept.
COMBINED USE OF TRANSPULMONARY THERMODILUTION (TPTD) TECHNIQUE IN FLUID MANAGEMENT FOR SEPSIS PATIENTS 1 St. Marianna University School of Medicine, Kanagawa,
Pierre SQUARA, MD Clinique Ambroise Paré, Neuilly Should we (can we) measure and optimize VO 2 in shock.
What is the future of haemodynamic monitoring? Steven M. Hollenberg, MD Professor of Medicine Robert Wood Johnson Medical School/UMDNJ Director, Coronary.
Cardiac Output Monitoring
CVS Monitoring in Shock
Haemodynamic Monitoring Theory and Practice. 2 Haemodynamic Monitoring A.Physiological Background B.Monitoring C.Optimising the Cardiac Output D.Measuring.
Arterial Line. Outline Definition. Indication Contraindication. EQUIPMENT Arterial Sites Nursing Skills Standard.
PulseCO Monitoring System. Estimates of Preload Clinical: BP, HR, capillary refill, urine Postural changes CVP PAC Echo.
Hemodynamic Monitoring John Nation RN, MSN Thanks to Nancy Jenkins.
Assessment of the Cardiovascular System p
Functional Hemodynamic Monitoring NEANA Spring Meeting April 2016 Donna Adkisson, R.N., M.S.N. Clinical Educator LiDCO, Limited.
FloTrac Sensor & Vigileo Monitor
Impact of Heart Rate on Central Aortic Pressures and Hemodynamics Analysis From the CAFE (Conduit Artery Function Evaluation) Study : CAFE-Heart Rate Bryan.
SECONDARY PREVENTION IN HEART DISEASE CATHY QUICK AUBURN UNIVERSITY/AUBURN MONTGOMERY EBP III.
© 2016 Global Market Insights, Inc. USA. All Rights Reserved U.S. Hemodynamic Monitoring Systems Market will the drive Regional Share in 2024.
Technologic Principles for CO Monitoring
The use of pulmonary artery catheters and cardiac output monitoring devices on a busy combined cardiac and general intensive care unit Dr U. Puar1, Dr.
Cardiac Output And Hemodynamic Measurements
B – The Cardiovascular System
Objectives Describe the cardiovascular monitoring techniques used in the care of critically ill patients and how to interpret the results of hemodynamic.
Copyright © 2018 American Academy of Pediatrics.
Cardiac Output Monitoring FCCNC 2018
Flow Monitoring Approaches
The Pulmonary Artery Catheter
The Pulmonary Artery Catheter
Comparison of respiration-induced variations in arterial blood pressure and multi-site photoplethysmographic signals during surgery S Sun1,2, WH Peeters1,
Estimation of changes in cardiac output from the arterial blood pressure waveform in the upper limb†,††  N.W.F. Linton, R.A.F. Linton  British Journal.
Pulseco: a less-invasive method to monitor cardiac output from arterial pressure after cardiac surgery  Timothy T Hamilton, MD, Lynne M Huber, RN, Michael.
Uncalibrated pulse contour-derived stroke volume variation predicts fluid responsiveness in mechanically ventilated patients undergoing liver transplantation 
BY: OTEWA, MESHACK OPONDO, EDDAH
Continuous and minimally invasive cardiac output monitoring by long time interval analysis of a radial arterial pressure waveform: assessment using a.
Less Invasive Methods of Advanced Hemodynamic Monitoring AHMED ALSADDIQUE PROFESSOR & CONSULTANT CARDIAC SURGEON.
INVASIVE PRESSURE MONITORING
Pulmonary artery pressure in a) healthy subjects and b) pulmonary hypertension (PH) subjects. Pulmonary artery pressure in a) healthy subjects and b) pulmonary.
Presentation transcript:

Are The Less Invasive Techniques For Monitoring Cardiac Output As Accurate As The Pulmonary Artery Catheter? Dr Andrew Rhodes St George’s Hospital London Haemodynamic monitoring in shock and implications for management

Conflicts of Interest Consultant for Edwards Lifesciences. Perform research on LiDCO, Vigilance and Vigileo haemodynamic monitors. Member of DSMB for Orion pharma.

Grade Process for Diagnostic Technologies. 1.Assess the evidence of accuracy for the diagnostic test. 2.Assess the evidence that using the technology influences outcome. R. Jaeschke

Summary Description of Technology Assessment of Evidence for accuracy of technology. Assessment of evidence that technology can influence outcome. Recommendations.

How accurate is the PAC? 12-15% for triplicate injection of ice saline –Stetz, 1982 Am Rev Respir Dis –Nilsson 2004 Acta Anesth But worse if –Old –Obese –Less injections –Continuous technologies

Arterial Pulse Analysis for the Measurement of Cardiac Output Arterial pulse pressure analysis is a technique of measuring and monitoring stroke volume on a beat to beat basis from the arterial pulse pressure waveform. Advantages Minimally invasive No incremental risk to the patient Beat to beat information

Pulse Pressure Relationship to Stroke Volume The fluctuations of blood pressure around a mean value are caused by the volume of blood forced into the arterial conduit by each systole. The magnitude of the change in pressure – known, as the pulse pressure, is a function of the magnitude of the change in stroke volume One factor, however, that is of particular importance is the compliance of the arterial wall.

The relationship between Pulse Pressure and Stroke Volume is difficult however….. Problems with measurement of Pulse Pressure –Wave reflection less in an aortic waveform than in a radial waveform –Damping – in the measuring apparatus –Morphology. Problems with assessing aortic compliance –Non linear relationship, prevents any simple approach to estimate volume from the pressure change

Compliance Is altered by –Age –Sex –Disease process –Pharmacology –Etc etc.

Pressure Compliance Female Male Change of Arterial Compliance with Sex Langouwouters J Biomechan 1984

Pressure Compliance AGE Change of Arterial Compliance with Age Langouwouters J Biomechan 1984

Most companies have concluded that this non-linear compliance relationship implies that some sort of calibration technique is necessary for accurate determination of stroke volume from arterial pressure traces.

Assessment of Accuracy Calibration –Very little in shocked or haemodynamically unstable patients. –Study quality limited by design Stable patients No gold standard (mostly against PAC) Pulse contour analysis –Mostly in anaesthetic or post surgical patients.

Int Care Med : Comparison of pulmonary artery and arterial thermodilution cardiac output in critically ill patients. Sakka SG, Reinhart K, Meier-Hellmann A. Bias 0.7 L/min Limits of agreement 1.9 to -0.5 L/min 37 patients with sepsis / septic shock (33) or SAH (4)

AuthorValidationMean CO Bias2 x SD of bias % Error KuritaPAC, EMF MasonPAC LintonPAC CorleyPAC Garcia-RodriguezPAC LintonTPTD LintonPAC Validation Studies Of Lithium Dilution Calibration.

Validation Studies of PCO from PiCCO No. PtsPopulationBiasLimits of agreement Weissman11Neurosurgery0.1+/- 1.9 Jansen54Cardiac surgery0.1+/- 1.0 Gratz27Cardiac surgery0.0+/- 1.1 Rodig26Cardiac surgery0.2+/- 2.0 Goedje24ICU0.1+/- 1.4 Zollner19ICU0.3+/- 2.5 Buhre12Cardiac surgery0.0+/- 1.3

Crit Care Med ;52-8. Reliability of a new algorithm for continuous cardiac output determination by pulse contour analysis during hemodynamic instability. Godje O, Hoke K, Goetz A et al. 24 cardiac surgical patients with change in cardiac output >20% during study period. Bias -0.2 L/min Limits of agreement 2.1 to -2.5 L/min

No. PtsPopulationBiasLimits of agreement Hamilton20Cardiac surgery0.1+/- 1.2 Pittman22Cardiac surgery0.1+/- 1.3 Validation Studies of PCO from PulseCO

Evidence that the Use of Pulse Analysis monitoring Improves Outcome. In post surgical patients –Pearse, Crit care 2005 In shocked patients –???

Yes, (level of evidence, C) but it depends…. –Not all monitors are the same. –In stable patients they perform to a clinically acceptable level and have other advantages. Continuous data Less invasive Offer other variables. –In shocked patients the evidence is less clear. Are the pulse analysis techniques as accurate as the PAC for monitoring CO?

No. –But we should recommend more studies evaluating the performance and efficacy of these monitors in shocked patients. Can we recommend this new technology as an alternative to the PAC in shock?