Radiological Physics Center David Followill, Ph.D. and RPC Staff.

Slides:



Advertisements
Similar presentations
بسم الله الرحمن الرحيم ﴿و قل رب زدنى علماً﴾ صدق الله العظيم.
Advertisements

IAEA Quality Audits in Radiotherapy
RapidArc plan verification using ArcCHECK™
Commissioning an Anthropomorphic Spine and Lung Phantom for Remote Dose Verification of Institutions Participating in RTOG 0631 Douglas Caruthers, M.S.;
Guideline for Monitoring of Clinical Trials for Cooperative Groups, CCOP Research Bases, and the Clinical Trials Support Unit (CTSU) Lore Lagrone Program.
RADIOTHERAPY ACCIDENT IN COSTA RICA - CAUSE AND PREVENTION OF RADIATION ACCIDENTS IN HOSPITALS Module XIX.
Slice Thickness Interpolation: The effect of interpolated slice thickness on the 2D vs 3D gamma results are shown in Table 3 for the QA data set only.
FDA-QA-DAS/2010 FDA’s Public Meeting: Device Improvements to Reduce the Number of Under-doses, Over-doses, and Misaligned Exposures from Therapeutic Radiation.
Challenges in Credentialing Institutions and Participants in Advanced Technology Clinical Trials Geoffrey Ibbott, David Followill, Andrea Molineu, Jessica.
Medical Physics Support of Linear Accelerators. Overview of Physics Support Accelerator safety issues Accelerator safety issues Task Group Report #35.
High-Energy Photon Standard Dosimetry Data: A Quality Assurance Tool Jessica R. Lowenstein, Stephen F. Kry, Andrea Molineu, Paola Alvarez, J. Francisco.
Photon Beam Monitor-Unit Calculations
Algorithms used in heterogeneous dose calculations show systematic error as measured with the Radiological Physics Center’s anthropomorphic thorax phantom.
Maria Skłodowska-Curie Memorial Cancer Centre and Institute of Oncology, Warsaw, Poland 8th ECMP, Athens, Dosimetry audits in radiotherapy.
Evaluation of the characteristics of TLD LiF:Mg.Ti-100 Powder: A Measure of Consistency Between Multiple Batches of Powder Paola Alvarez,Jose Francisco.
Results of IAEA supported TPS audit in Europe
Quality Assurance: Manufacturer & Clinical Aspects  Alan Cohen, M.S. DABR  Paul Naine, MSc. MIPEM  Jim Schewe, PhD, DABMP Accuray Incorporated Elekta.
IAEA International Atomic Energy Agency IAEA quality audits in radiotherapy Joanna Izewska Head, Dosimetry Laboratory Dosimetry & Medical Radiation Physics.
IAEA International Atomic Energy Agency PREVENTION OF ACCIDENTAL EXPOSURE IN RADIOTHERAPY Part 3: Analysis of causes and contributing factors IAEA Training.
Innovation/Impact: By designing a simulated human shaped (anthropomorphic) plastic phantom with targets, organs at risk (OAR) and heterogeneities, the.
Food and Drug Administration Public Meeting June 9-10: Quality Assurance of Therapeutic Medical Devices In Radiation Oncology Nabil Adnani, Ph.D., DABR.
IAEA International Atomic Energy Agency International Reporting Systems Ola Holmberg, PhD Head, Radiation Protection of Patients Unit Radiation Safety.
COMPUTERS AND RADIATION THERAPY/ONCOLOGY Lydia Eichler CIS 109.
An introduction to the National Voluntary Laboratory Accreditation Program Sally Bruce, Chief for the National Voluntary Laboratory Accreditation Program.
Results The measured-to-predicted dose ratio criteria used by the RPC to credential institutions is , however for this work, a criteria of
11. – , Athens 8th European Conference on Medical Physics DOSIMETRY AUDITS IN RADIOTHERAPY IN THE CZECH REPUBLIC Irena Koniarová Daniela Ekendahl.
به نام خداوند بخشایندۀ بخشایشگر
 Quality & Safety Considerationsin Stereotactic Radiosurgery and Stereotactic Body Radiation Therapy.
TWO FIELD BREAST PLAN VS. OPTIMIZED CONFORMAL BREAST PLAN: COMPARISON OF PLAN PARAMETERS Authors: Borko Basarić, Ozren Čudić, Milan Teodorović, Borislava.
CALGB Informational Session June 22, 2007 David Hurd, MD Interim Chair Data Audit Committee.
Evaluation of New Pre-Treatment In-Air Patient Specific QA Software for TomoTherapy Treatments Lydia L. Handsfield¹, Quan Chen¹, Kai Ding¹, Wendel Renner²,
Quality Control in Radiation Therapy, A New Concept: Dosimetry Check
Patient Plan Results: Table 3 shows the ratio of the Pinnacle TPS calculation to the DPM recalculation for the mean dose from selected regions of interest.
David Followill, Ph.D. Radiological Physics Center
In vivo dosimetry Eirik Malinen Eva Stabell Bergstrand Dag Rune Olsen.
Comparison of Clinical Parameters for Proton Therapy in the United States Paige Summers, MS.
IMRT QA Plan Site 5%/3mm3%/3mm2%/2mm 0% noise1% noise2% noise0% noise1% noise2% noise0% noise1% noise2% noise HN
The RPC Proton Therapy Approval Process
FDA Public Meeting: Device Improvements to Reduce the Number of Under-doses and Misaligned Exposures from Therapeutic Radiation.
Institute for Advanced Radiation Oncology
Commissioning of an Optically-Stimulated Luminescence (OSL) system for remote dosimetry audits J.F. Aguirre, P. Alvarez, C. Amador, A. Tailor, D. Followill,
Medical Accelerator F. Foppiano, M.G. Pia, M. Piergentili
Integrating the Health Care Enterprise- Radiation Oncology Use Case: In Vivo Patient Dosimetry Editor: Juan Carlos Celi - IBA Reviewer: Zheng Chang – Duke.
The Radiological Physics Center’s Anthropomorphic Quality Assurance Phantom Program Carrie F. Amador, Nadia Hernandez, Andrea Molineu, Paola Alvarez, and.
1 A Comprehensive Study on the Heterogeneity Dose Calculation Accuracy in IMRT using an Anthropomorphic Thorax Phantom S Davidson 1, R Popple 2, G Ibbott.
Laurie Cuttino MD, Dorin Todor PhD, Douglas Arthur MD, Rohini George, Lynn Pacyna CMD Medical College of Virginia Campus Department of Radiation Oncology.
Investigation of 3D Dosimetry for an Anthropomorphic Spine Phantom R. Grant 1,2, G. Ibbott 1, J. Yang 1, J. Adamovics 3, D Followill 1 (1)M.D. Anderson.
Development and Implementation of a Remote Audit Tool for High Dose Rate (HDR) 192 Ir Brachytherapy Using Optically Stimulated Luminescence Dosimetry Kevin.
K. Pulliam, MS 1,2., D Followill, PhD 2., L Court, PhD 2., L Dong, PhD 3., M Gillin, PhD 2., K Prado, PhD 3., S Kry, PhD 2 1 The University of Texas Graduate.
Introduction The Radiological Physics Center (RPC) anthropomorphic quality assurance (QA) phantom program is one tool the RPC uses to remotely audit institutions.
A Comparison Between Two Leading Stereotactic Platforms in the Treatment of Multiple Metastases Sandra Vermeulen MD, James Raisis.
TLD POSTAL DOSE QUALITY AUDIT FOR 6MV AND 15MV PHOTON BEAMS IN RADIOTHERAPY CLINICAL PRACTICE Sonja Petkovska 1, Margarita Ginovska 2, Hristina Spasevska.
Radiation Therapy Trials - Quality Assurance:  patient safety  adherence to protocol constraints  uniformity of patient treatments  efficient review.
COUNTRY REPORT: PHILIPPINES Dr. Miriam Joy C. Calaguas.
Commissioning of a commercial treatment planning system for IMAT and Dose Painting treatment delivery. G. Pittomvils 1,,L. Paelinck 1, F. Crop 2, W. De.
Rapid Arc Treatment Verification: post evaluation on Delta-4 and proposal of a new verification protocol G. Pittomvils 1,,L. Paelinck 1, T. Boterberg 1,
Adapting A Clinical Medical Accelerator For Primary Standard Dosimetry
E. Mezzenga 1, E. Cagni 1, A. Botti 1, M. Orlandi 1, W.D. Renner 2, M. Iori 1 1. Medical Physics Unit, ASMN-IRCCS of Reggio Emilia, Italy 2. MathResolution.
Treatment Chart Record of patients radiation therapy history. Must contain: History and diagnosis Rationale for treatment Treatment plan Consent Documentation.
SIMULATION CHECK LIST IMAGING PROTOCOL.
IAEA E-learning Program
CHAPTER 3 DOSE DETERMINATION FOR EXTERNAL BEAMS
The magnitude of H&N IMRT dose delivery errors from three possible failure modes: beam quality, symmetry, and MLC position Jackie Tonigan, M.S. Advisor:
Developing a Comprehensive Site Selection Process for a Cancer Network in a Resource-Limited Settings in Sub-Saharan Africa Meg Wirth AMC Operations &
IAEA E-learning Program
Very High Energy Electron for Radiotherapy Studies
Left Posterior Superior Right Anterior Inferior
Fig. 4. Percentage of passing rate between clinical and 544 plans.
GHG meeting at ESTRO36 May, 2017
SBRT trial QA IROC GHG meeting at ESTRO May, 2017
Presentation transcript:

Radiological Physics Center David Followill, Ph.D. and RPC Staff

Radiological Physics Center Formed in 1968 and located at MD Anderson Cancer Center (1 of 12 longest running grants). Our Mission is to assure NCI and cooperative groups that institutions participating in clinical trials deliver prescribed radiation doses that are comparable and consistent, (minimize dose uncertainty), make corrections and report findings to the community. Funded continuously for 44 years as cooperative clinical trial groups have changed and expanded internationally Use of remote and onsite dosimetry audits

RPC Scope of Monitoring Monitoring 1888 inst. participating in clinical trials - includes 210 non-North American sites 41 countries (since %  ) - ~23,000 beams - ~3500 machines

Components of RPC QA Program 1. Remote audits of machine output 1,888 institutions, ~14,000 beams measured with TLD and OSLD in North America and Internationally 2. Patient Treatment record reviews 474 charts reviewed for GOG, NSABP, NCCTG, RTOG (brachytherapy) 3. On-site dosimetry reviews 41 institutions visited in 2011 (~150 accelerators/450 beams measured) 4. Credentialing - Phantoms ~500 irradiations in 2011

Reference calibration (NIST traceable) Correction Factors: Field size & shape Depth of target Transmission factors Treatment time Evaluated by RPC Dosimeters Evaluated by RPC visits and chart review RPC Verification of Institutions’ Delivery of Tumor Dose Tumor Dose Evaluated by RPC phantoms

So, how are we doing?

OSLD/TLD Beam Output Checks 3-4% of the beams require a repeat

Comprehensive On-Site Audits Reference Beam Calibration Percent of Inst. with ≥ 1 beam out of Criteria (since 2002) PhotonsElectrons OSLD/TLD (±5%) 7-11% 6-12% Visits (±3%) ~13% ~15%

Discrepancies Regarding: Number of Institutions Receiving rec. (n = 156) Review QA Program115 (74%) Photon Field Size Dependence (small FSD)62 (40%) Wedge Factor (WF)50 (32%) Off-axis Factors (OAF)/Beam symmetry46 (29%) Electron Calibration27 (17%) Photon Depth Dose25 (16%) Electron Depth Dose18 (12%) Photon Calibration13 (8%) Review Temp/Press Correction11 (7%) Change to TG-519 (6%) Electron Cone Ratios8 (5%) Using Multiple Sets of Data8 (5%) Discrepancies Discovered (Jan. ’05 – Mar. ’11) On-Site Dosimetry Review Audit

Treatment record reviews RPC performs independent retrospective review and recalculation of doses for RTOG, NCCTG and GOG brachy. patients Errors in dose calculation and doses reported to study groups are discovered and corrected The RPC review has resulted in changing the reported dose on 546 (27%) of the1993 protocol patients reviewed since % are EBRT dose errors - 87% are brachytherapy dose errors We revise the dose data in 1 of every 3 charts

RPC Phantoms Pelvis (10) Thorax (10) Liver (6) H&N (30) SRS Head (10) Spine (8)

Independent “end to end” audit Imaging Planning/dose calculation Setup delivery Uniform phantoms and dosimeters Standardized analysis Uniform pass/fail criteria Allows inst. to inst. comparison Established infrastructure PhantomPatient Benefits of RPC Phantoms PhantomPatient

Phantom Results PhantomH&NProstateSpineLung Irradiations Pass 686 (79%) 162 (82%) 22 (63%) 178 (75%) Fail Criteria7%/4mm 5%/3m m 5%/5m m RTOG Inst. Acceptable 557 (54%) 206 (20%) 83 (8%) 289 (28%) Pass 928 (81%) 265 (85%) 78 (65%) 361 (79%) Fail Comparison between institution’s plan and delivered dose.

Phantom Results Comparison between institution’s plan and delivered dose. H&NProstateSpineLung Irradiations (all years) Pass (all years)928 (81%)265 (85%)78 (65%)361 (79%) Fail (all years) Irradiations (2011) Pass (2011)101 (93%)45 (80%)31 (78%)68 (85%) Fail (2011) Criteria7%/4mm 5%/3mm5%/5mm Failure rate doubles going to ±5%/3mm criteria

Why do we continue to find errors? 1.Too busy 2.Advanced technology/ Don’t understand process 3.Communication/Fear of punishment 4.Training/Failure to ask for help 5.Can’t accept the fact that an error could be mad WHO report on “Radiotherapy Risk Profile” states that 60% of all radiotherapy incidents are attributable to human error. Human Errors!

Let’s get past these hurdles! Questions?