TO ERR IS HUMAN ……. We detest it. We avoid even talking about it. ERRORS.

Slides:



Advertisements
Similar presentations
QUALITY CONTROL IN HEMATOLOGY
Advertisements

Day to Day Management of Quality Control
Module 6: QC Basic Rules and Charts
Errors in Chemical Analyses: Assessing the Quality of Results
Quality in Environmental Analysis Tuesday, June 27, 2000 Goldwater Center Room 487.
Control of Analytical Variables Dr. Roula Hamid MSc Clin Biochem Central Puplic Health Laboratory QC Chemistry.
World Health Organization
Dr Samah Kotb Lecturer of Biochemistry 1 CLS 432 Dr. Samah Kotb Nasr El-deen Biochemistry Clinical practice CLS 432 Dr. Samah Kotb Nasr.
Laboratory Quality Control
Quality Assurance.
How Science Works Glossary AS Level. Accuracy An accurate measurement is one which is close to the true value.
Internal Quality Control (QC) for Medical Laboratories: An introduction Dr. Otto Panagiotakis and Dr. Alexander Haliassos ESEAP – Greek Proficiency Testing.
Quality Assurance / Quality Control
LABORATORY QUALITY CONTROL
This teaching material has been made freely available by the KEMRI-Wellcome Trust (Kilifi, Kenya). You can freely download,
Control charts Pharos University. An internal system for quality control is needed to ensure that valid data continue to be produced. This implies that.
Quality Assurance in Haematology
Quality Assurance in the clinical laboratory
Quality Assurance / Quality Control
Chemometrics Method comparison
Method Comparison A method comparison is done when: A lab is considering performing an assay they have not performed previously or Performing an assay.
Unit #7 - Basic Quality Control for the Clinical Laboratory
Quality Assurance.
Quality Assessment 2 Quality Control.
Validation of Analytical Method
Method Selection and Evaluation Method Selection and Evaluation D. Kefaya EL- Sayed Mohamed Prof. Of Clinical Pathology (Clinical Chemistry), Mansoura.
ABC of Quality Control A problem based approach RT ERASMUS NHLS / FACULTY OF HEALTH SCIENCES, UNIVERSITY OF STELLENBOSCH 20th July, 2007, Bela Bela.
Analytical Goals. Valid data are essential in making medical decisions, the most important concepts used in judging analytical performance :- 1)Analytical.
1 Accreditation and Certification: Definition  Certification: Procedures by which a third party gives written assurance that a product, process or service.
Bias and Errors. Some Terms Used to Describe Analytical Methods Accuracy Precision LOD RDL LOQ Selectivity Sensitivity Linearity Ruggedness.
Chapter 5 Errors In Chemical Analyses Mean, arithmetic mean, and average (x) are synonyms for the quantity obtained by dividing the sum of replicate measurements.
Automated CBC Parameters
Quality Control Lecture 5
Laboratory QA/QC An Overview.
Quality Assurance How do you know your results are correct? How confident are you?
Why do we need to do it? What are the basic tools?
QC THE MULTIRULE INTERPRETATION
An Overview. Definitions (1)  Quality Control - QC refers to the measures that must be included during each assay run to verify that the test is working.
Biochemistry Clinical practice CLS 432 Dr. Samah Kotb Lecturer of Biochemistry 2015 Introduction to Quality Control.
Quality control & Statistics. Definition: it is the science of gathering, analyzing, interpreting and representing data. Example: introduction a new test.
Application of Westgard multi rules in medical laboratories
Quality Control Internal QC External QC. -Monitors a test's method precision and analytical bias. -Preparation of quality control samples and their interpretation.
LECTURE 13 QUALITY ASSURANCE METHOD VALIDATION
L ABORATORY Q UALITY C ONTROL. INTRODUCTION _A major role of the clinical laboratory is the measurement of substances in body fluids or tissues for the.
 Remember Chemistry panel Quality Control:-  In a medical laboratory, it is a statistical process used to monitor and evaluate the analytical process.
 Routine viral diagnostics: indirect and direct detection of viruses. ◦ Indirect detection: serological tests; ◦ Direct detection:  Viral antigens;
CLINICAL LABORATORIES THE ROLE: A science that uses sophisticated instruments and techniques with the application of theoretical knowledge to perform complex.
Chapter 5: Errors in Chemical Analysis. Errors are caused by faulty calibrations or standardizations or by random variations and uncertainties in results.
Diagnostic clinical chemistry
QC - THE OUT-OF-CONTROL PROBLEM 謝 昆 穎 Assistant Product Manager, CDG BIO-RAD Laboratories, Taiwan Branch.
Quality Assurance.
Quality Assessment.
Quality is a Lousy Idea-
Pakistan Society of Chemical Pathologists Zoom Series of Lectures ZT 25. Quality Management 2 Brig Aamir Ijaz MCPS, FCPS, FRCP (Edin), MCPS-HPE HOD.
Unit #6 - Basic Quality Control for the Clinical Laboratory
Pakistan Society Of Chemical Pathologists Zoom Series of Lectures ZT 24. Quality Managent 1 Brig Aamir Ijaz MCPS, FCPS, FRCP (Edin), MCPS-HPE HOD and.
Quality Assurance in the clinical laboratory
INTERNAL QUALITY CONTROL
Quality Assurance in the clinical laboratory
Laboratory Quality Control
Practical clinical chemistry
Quality is a Lousy Idea-
This teaching material has been made freely available by the KEMRI-Wellcome Trust (Kilifi, Kenya). You can freely download,
Choice of Methods and Instruments
Introduction To Medical Technology
Quality Control Lecture 3
▪Internal quality control:
Quality Assessment The goal of laboratory analysis is to provide the accurate, reliable and timeliness result Quality assurance The overall program that.
Westgard rules Dr Ahmed Aboamer MD in Clinical pathology.
Presentation transcript:

TO ERR IS HUMAN ……

We detest it. We avoid even talking about it. ERRORS

Do we really understand them? Not taught properly during UG and PG Medical errors Medical Negligence: Not doing something, which a reasonable person would do or doing something which a reasonable person would not do.

Zero Error Paradigm  Doctors And Health Sector Are Expected To Deliver Every Time  Doctors Themselves & Lay Public Thinks That Doctors Can Achieve This Impossible Ideal Every Time

Objectives  Understand Errors  Factors that lead to errors  Error management

Significance of Error Potential Leads To:  Misdiagnosis  Delays In Treatment  Increased Costs Avoidable Retests Cost US 200million USD Per Year Even A Small Calibration Bias Can Affect Treatment Rates:  1% +Ve Bias In Cholesterol Result  5% Increase In Patients Exceeding The Treatment Cut-off  3% +ve Bias  15% Increase In Patient Treatment

Comparative Perspective Error Rates Could Be Reported In Cases Per Million To Allow Easy Comparison:  2.5 Deaths / Million Anaesthesia Cases  164 Deaths / Million Car Crashes  5000 Cases Of Lost Baggage / Million Air Passengers  4566 Laboratory Errors (Results > 6SD) / Million Results Kazmierczak & Catrou. Arch Pathol Lab Med 1993;117:714-8

Consequences of the Error  Loss of lives  Increase in total cost  Loss of trust in System  Diminished satisfaction – Patients/Docs Institute of Medicine Report Nov-1999Institute of Medicine Report Nov-1999

Understanding Human Behaviour and Error David Embrey Human Reliability Associates 1, School House, Higher Lane, Dalton, Wigan, Lancashire. WN8 7RP

Basic Concept  True value - this is an ideal concept which cannot be achieved.  Accepted true value - the value approximating the true value, the difference between the two values is negligible.  Error - the discrepancy between the result of a measurement and the true (or accepted true value).

Sources of Error  Analytical – Random & Systematic  Non-analytical

Random Error  An error which varies in an unpredictable manner, in magnitude and sign, when a large number of measurements of the same quantity are made under effectively identical conditions.  Random errors create a characteristic spread of results for any test method and cannot be accounted for by applying corrections. Random errors are difficult to eliminate but repetition reduces the influences of random errors.

Random Errors

Causes of R.E. (Imprecision)  Reagent  Variable Reagent Mixing  Splashing  Calibrators  Too Frequent calibrations  Variable Age  QC material  Variable Preparation  Variable Age  Sample  Carry-over  Interference Common to all these materials Evaporation / Condensation  Instrument  Worn out spares  Electrical Noise ( E.M.F.)  Inconsistent wavelength  Inconsistent Temperature  Volumetric / Timing Error  Contamination of Tubings/Valves  Partial blockages in Sample/ Rgt probes  variable cuvette thickness  Intermittent failure

Systematic Error  An error which, in the course of a number of measurements of the same value of a given quantity, remains constant when measurements are made under the same conditions, or varies according to a definite law when conditions change.  Systematic errors create a characteristic bias in the test results and can be accounted for by applying a correction.

Systematic Errors

Common Causes of S.E. (Bias)  Manual Error in Reconstitution of Control/ Calibrator/ Reagent / Water quality  Over a Period of Time Changes in  Volumetric / Timing  Light Source  Instrument Spares

Accuracy…. …The agreement between your value and the 'true' value, that is how correct your result is.

Precision…. The reproducibility of your results or the agreement between replicate measurements.

The Closer Your Results, Are To Each Other, For The Same Analyte In The Same Serum, The Better Your Precision. For Method evaluation, precision should be assessed in terms of:- - Within Run Performance (Intra-assay precision) - Between Run Performance (Inter-assay precision)

Accurate and Precise Imprecise But Will Be Accurate If You Average The Results Precise but Inaccurate

Specificity…. …is the ability of a method to measure solely the component of interest. A lack of specificity will affect accuracy − Falsely elevated or lowered values may be obtained

…is the ability to detect small quantities of a measured component. Sensitivity… − will affect both precision and accuracy at the bottom end of the assay range.

AIMS OF QC & QA Quality control simply ensures that results generated by the system are correct. Quality Assurance is concerned with much more: Right test is carried out on the right specimen, and that the right result and right interpretation is delivered to the right person at the right time

 Quality Assessment - (Proficiency testing) means to determine the quality of the results. & is a challenge to the effectiveness of the QA and QC programs.  Quality Assessment - RANDOX RIQAS, BIORAD EQAS.

Objectives Of Internal QC  To Improve & Maintain Performance (Analytical, Medical & Other Aspects)  To Alert Staff Towards Erroneous Results & To Take Remedial Actions  To Detect Undue Variations In Results & Identify Its Source

LEVY JENNING’S CHART

Levy Jennings Chart +3 sd -3 sd +2 sd -2 sd -1 sd +1 sd Analyte conc. Target value Daily Assay Run

Westgard Rules  Formulation Based On Statistical Methods  To Analyse Data.  To Define Specific Performance Limits  To Detect Random And Systematic Errors.  Six Commonly Used Westgard Rules –  Two - Warning  Four - Mandatory  Violation of Warning rules should Trigger A Review Of Test Procedures, Reagent Performance And Equipment Calibration.  Violation of Mandatory Rules Should Result In Rejection of The Results

Warning Rules  Warning 1 2SD : When the IQC value exceeds the mean by  2SD. It is an event likely to occur normally in less than 5% of cases. – Random Error  Warning 4 1SD : If four consecutive IQC values exceed the same limit (mean  1SD) and this may indicate the need to perform instrument maintenance or reagent calibration. – Systematic Error/ Bias

Mandatory Rules  Mandatory 1 3SD : When the IQC value exceeds the mean by  3SD. Regarded as out of control. – Systematic Error  Mandatory 2 2SD : When two consecutive IQC values exceed the mean on the same side of the mean by  2SD.– Systematic Error  Mandatory R 4SD : Applied when the IQC is tested in duplicate or N=2. when the difference in SD between the duplicates exceeds 4SD.  Mandatory 10x : When the last 10 consecutive IQC values are on the same side of the mean or target value. – Systematic Error

Westgard Rules: 1 3SD +3 sd -3 sd +2 sd -2 sd -1 sd +1 sd Analyte Conc. Target value Daily Assay Run

Westgard Rules: 10X +3 sd -3 sd +2 sd -2 sd -1 sd +1 sd Analyte Conc. Target value Daily Assay Run

Westgard Flowchart Control data 1 point outside 2 SD 1 point outside 3 SD 2 consecutive values outside the same 2 SD Difference between 2 controls within a run exceeds 4 SD 4 consecutive control values on one side of the mean and further than 1 SD from the mean 10 consecutive values on one side of the mean In control – report data Out of control – reject analytical run Yes No Yes

Follow-up Action in the Event of a Violation  Accept the test run in its entirety - only when a warning rule is violated. Initiate a review.  Reject the whole test run - only when a mandatory rule is violated.  Enlarge the grey zone and thus re-test range for that particular assay run - this option can be considered in the event of a violation of either a warning or mandatory rule.

HOW TO INVESTIGATE THE CAUSE….. …… ELIMINATE THE GUESSWORK LAY DOWN A PROTOCOL

Generic Protocol for Remedial Actions 1.Reanalyze the same control immediately. 2.Repeat the test using a fresh vial of control. 3.Repeat the test using a new control from a different lot. 4.Check integrity of Reagent system & make necessary changes. 5.Perform maintenance and rerun the control. 6.Recalibrate (Only if necessary) and rerun the control. 7.Call an expert for help.

“Knowing is Not Enough; We Must Apply. Willing is Not Enough; We Must Do.” —Goethe

Planning QC Strategy  Define quality specifications - TEa  Measure – Assess bias & imprecision - TE  Analyse with Normalised OPSpecs Charts, Power Function graphs, Critical-error graphs, QC selection grids etc.  Implement & Improve  Control  Monitor with SIGMA MATRIX

WHAT IS ‘BEYOND QC’ ?

OTHER ISSUES TO BE CONSIDERED  Maintenance & Monitoring of - Materials, Machine & Man  Method Validation  QC/ Calibrator Material  S.O.P.s  Comparison Of Data

Maintenance & Monitoring  Temperature maintenance – For Material stability - Refrigerator, Analysers  Water Quality – NCCLS Grade II  Electrical Stability– Earthing, E.M.F.  Instruments – P.M., Cuvettes, Bulbs, Tubings.  Calibration of equipments - Pipettes  Maintenance of Procedure Manuals  Inventory managements & Storage - Materials

MONITORING - Reagents  Reagent characteristics – Vital data  pH  O.D. of Blanks  Factor Range  Conductivity for diluents etc.  Calibration factors – Delta checks  Reaction Curves – For Ap & As  Errors & Action Taken Logs

METHOD VALIDATION Should be Performed Before Introducing Test / New Method / Systems in Routine Practice by Thoroughly Testing Reference Material or by Comparison of Test Results Performed by an Alternative Method

Criteria & Information Sought In Method Validation  Accuracy by running known value samples of M.D.Levels & comparing test results – For Constant & Proportional errors, Regressional analysis  Precision by running replicates of QC samples with within- the-run, Across-the-run & Between-the-run precision checks & comparison  Linearity Information

QC / CALIBRATOR MATERIAL  Third Party Control Materials  Controls with values for your method/ instruments with wider database  Stability & Type of Material (Matrix Effect)  Running Two levels of controls – Provides better error checks across a range

Man  Training of Staff  Basic  Continuous education  Laying down S.O.P.

COMPARISON OF TEST RESULTS  Peer Group Comparison (Randox 24x7)  Delta Checks (Previous values)  Clinical correlation

THANK YOU !!