Quality Control Lecture 5

Slides:



Advertisements
Similar presentations
Why do Q.A.? Physician’s Expectations Diagnosis Therapy Technology Physician’s Expectations Diagnosis Therapy Technology.
Advertisements

Module 6: QC Basic Rules and Charts
Internal quality control
Dr Samah Kotb Lecturer of Biochemistry 1 CLS 432 Dr. Samah Kotb Nasr El-deen Biochemistry Clinical practice CLS 432 Dr. Samah Kotb Nasr.
Laboratory Quality Control
Summary 1 l The Analytical Problem l Data Handling.
Limitations of Analytical Methods l The function of the analyst is to obtain a result as near to the true value as possible by the correct application.
Control Charts for Variables
Types of Errors Difference between measured result and true value. u Illegitimate errors u Blunders resulting from mistakes in procedure. You must be careful.
Quality Assurance.
Lecture 2 Data Processing, Errors, Propagation of Uncertainty.
Quality Control Barbara Weberman MT(ASCP) Oakland County Health Division Laboratory Supervisor.
Internal Quality Control (QC) for Medical Laboratories: An introduction Dr. Otto Panagiotakis and Dr. Alexander Haliassos ESEAP – Greek Proficiency Testing.
This teaching material has been made freely available by the KEMRI-Wellcome Trust (Kilifi, Kenya). You can freely download,
InterQC Delivering Simplicity in the Quality Control Management for your Laboratory Developed by Vitro.
Quality Assurance in Haematology
Quality Assurance in the clinical laboratory
Quality Assurance / Quality Control
Chemometrics Method comparison
Method Comparison A method comparison is done when: A lab is considering performing an assay they have not performed previously or Performing an assay.
Unit #7 - Basic Quality Control for the Clinical Laboratory
Quality Assurance.
Quality Assessment 2 Quality Control.
V. Rouillard  Introduction to measurement and statistical analysis ASSESSING EXPERIMENTAL DATA : ERRORS Remember: no measurement is perfect – errors.
ABC of Quality Control A problem based approach RT ERASMUS NHLS / FACULTY OF HEALTH SCIENCES, UNIVERSITY OF STELLENBOSCH 20th July, 2007, Bela Bela.
The following minimum specified ranges should be considered: Drug substance or a finished (drug) product 80 to 120 % of the test concentration Content.
PARAMETRIC STATISTICAL INFERENCE
Biostatistics: Measures of Central Tendency and Variance in Medical Laboratory Settings Module 5 1.
Prof. of Clinical Chemistry, Mansoura University.
Chapter 5 Errors In Chemical Analyses Mean, arithmetic mean, and average (x) are synonyms for the quantity obtained by dividing the sum of replicate measurements.
Automated CBC Parameters
Lecture 4 Basic Statistics Dr. A.K.M. Shafiqul Islam School of Bioprocess Engineering University Malaysia Perlis
Laboratory QA/QC An Overview.
Quality Assurance How do you know your results are correct? How confident are you?
Data Collection and Processing (DCP) 1. Key Aspects (1) DCPRecording Raw Data Processing Raw Data Presenting Processed Data CompleteRecords appropriate.
5. Quality Assurance and Calibration Quality assurance is We do to get the right answer for our purpose. Have Sufficient accuracy and precision to support.
Why do we need to do it? What are the basic tools?
Biochemistry Clinical practice CLS 432 Dr. Samah Kotb Lecturer of Biochemistry 2015 Introduction to Quality Control.
Quality Control – Introduction
Quality control & Statistics. Definition: it is the science of gathering, analyzing, interpreting and representing data. Example: introduction a new test.
Application of Westgard multi rules in medical laboratories
Quality Control: Analysis Of Data Pawan Angra MS Division of Laboratory Systems Public Health Practice Program Office Centers for Disease Control and.
ERT 207 Analytical Chemistry ERT 207 ANALYTICAL CHEMISTRY Dr. Saleha Shamsudin.
Quality Control Internal QC External QC. -Monitors a test's method precision and analytical bias. -Preparation of quality control samples and their interpretation.
LECTURE 13 QUALITY ASSURANCE METHOD VALIDATION
 Remember Chemistry panel Quality Control:-  In a medical laboratory, it is a statistical process used to monitor and evaluate the analytical process.
10 March 2016Materi ke-3 Lecture 3 Statistical Process Control Using Control Charts.
Levey Jennings Chart Activity Staff Meeting Topic.
Diagnostic clinical chemistry
QC - THE OUT-OF-CONTROL PROBLEM 謝 昆 穎 Assistant Product Manager, CDG BIO-RAD Laboratories, Taiwan Branch.
Quality Assessment.
Lesson 1-9 Quality Assessment.
ABG Instrumentation Analytical Errors
Unit #6 - Basic Quality Control for the Clinical Laboratory
Quality Assurance in the clinical laboratory
Westgard rules.
INTERNAL QUALITY CONTROL
Laboratory Quality Control
Practical clinical chemistry
Quality Control (2) Lecture 6
Process control in Hematology section
9 Tests of Hypotheses for a Single Sample CHAPTER OUTLINE
Choice of Methods and Instruments
Process Capability.
Introduction To Medical Technology
Quality Control Lecture 3
▪Internal quality control:
Quality Assessment The goal of laboratory analysis is to provide the accurate, reliable and timeliness result Quality assurance The overall program that.
Westgard rules Dr Ahmed Aboamer MD in Clinical pathology.
Presentation transcript:

Quality Control Lecture 5 Quality Assurance Quality Control Lecture 5

What is Quality Control? Quality Control in the clinical laboratory is a system designed to increase the probability that: each result reported by the laboratory is valid and can be used with confidence by the physician to make a diagnostic or therapeutic decision

Internal Quality control Internal quality control involves the analysis of control samples with patient specimens, then evaluating the results statistically to determine the acceptability of the analytical run They are used to analyze the accuracy and precision performance of the assay or analyzer If the control is in range, it is assumed that the reagents and analyzer are performing correctly and patient testing can begin When control samples do not produce accurate and precise results, it can be assumed that any patient results obtained at the same time are also erroneous

Control A solution that contains the same constituents as those being analyzed in the patient sample Most are commercially produced from pooled sera The manufacturer has analyzed each lot of serum for a variety of test components and the expected range of assay values for each component is provided to the laboratory when shipped

Internal Quality Control For the Internal Quality Control, the preparation of the control samples and their interpretation are handled within the laboratory In contrast, External quality control involves the estimation of a test method's accuracy by the analysis of unknown samples sent to the laboratory from outside sources The samples are sent to the laboratory where they are analyzed and the results returned to the agency that supplied the control

Calibrators Controls and calibrators or standards differ; and are not interchangeable Calibrators and standards are used to adjust instrumentation or to define a "standard curve" for analysis The value assigned for each constituent are to be determined by a reference method Definitions The set of operations that establish, under specified conditions, the relationship between values indicated by a measuring instrument and the corresponding values of the analyte to be measured Calibrators are solutions with known values that establish the relationship between the amount of signal produced in the assay and analyte concentration

Characteristics of A Good Control The composition of the control material should be as similar to the patient sample as possible, reacting in the same manner The analyte concentration should be at medically significant levels The constituents should be stable under storage for a long period of time prior to preparation Material should have low vial-to-vial variability The material should be ready to use or require a minimum of preparation and be readily available for emergency use After a vial has been opened and the material prepared, it should remain stable for the period of use The material should be available in large quantities The material should be reasonably priced (but cost should not be the primary factor)

Qualitative "bipolar" and Semi-Quantitative Procedures No entirely satisfactory statistical evaluation Control levels should consist of a minimum of a negative and weak positive control A strong positive control is useful in monitoring the sensitivity of the method in the upper range but is not always necessary The control is used as a point of reference from which the technologist makes a judgment for comparison Semi quantitative tests should have control concentrations at each of the graded levels, that is, trace, 1 +, 2+, and so on As with the qualitative methods, each control level can be used as a comparison for grading the test results

Quantitative Procedures In quantitative procedures, commercially prepared quality control sera are used with patient samples to detect systematic analytical errors and monitor precision Prepare and test the material daily for a minimum of 20 consecutive working days, paying careful attention to instrument function At the end of 20- day period all of the data is collected to calculate a mean, standard deviation, and coefficient of variation excluding the data known to be the result of mistakes and explained errors coefficient of variation: It is defined as the ratio of the standard deviation to the mean .

Quantitative Procedures The distribution of data set should be a bell-shaped Gaussian curve If the data distribution is skewed, some sort of large systematic shift has occurred during the test period and the data should not be used to calculate the control's acceptable limits Investigate possible problems and restart data collection Control sample placement should be random within the analytical run to estimate more accurately the amount of imprecision Sometimes controls are analyzed at fixed intervals e.g. every 20 samples

Interpretation Interpretation of the control result can take one of several forms: Graphical interpretation using levey-Jennings or shewhart charts Statistical and graphical interpretation by: multi-rules, cumulative summaries and trend analysis

Shewhart or Levey- Jennings control charts The levey-Jennings control chart is derived from the Gaussian distribution indicating the mean and the one, two, and three standard deviation ranges on both sides of the mean The chart illustrates the relationship between the levey-Jennings control chart and the Gaussian distribution from which it is derived This figure also shows the expected percentage of results that should fall within each of the standard deviation ranges

Shewhart or Levey- Jennings control charts On the control chart used to evaluate the result of the control runs the dates or number of analyses are listed along the X-axis and the values of the control along the Y- axis The mean and the 1,2 and 3 standard deviation (s) limits of the control analysis to date are marked As data is obtained it is plotted one point at a time along the chart

Shewhart or Levey- Jennings control charts In a random distribution and in a correctly operating test system approximately: 65% of the values will be between the ± Is ranges and will be evenly distributed on either side of the mean 95% of the values should fall between the ± 2s ranges and 99% between the ± 3s limits

Shewhart or Levey- Jennings control charts The ± 2s limits are considered as warning limits A value between the 2s and 3s limit indicates the analysis should be repeated The ± 3s limits are rejection limits. Analysis should stop, patient results held and the test system investigated

Example of a shift in calibration can be failure in recalibration when lot numbers of reagents during analytical run

A trend can start on one side of the mean and move across it or it can occur entirely on one side of the mean Trends can be caused by the deterioration of reagents, pump tubing or light sources in instruments Shifts and trends can occur together or independently