Uni Bergen:G. Eigen, J. Zalieckas, E. van der Kraaij FZU Prague: J. Cvach, J. Kvasnicka, I. Polák CALICE meeting, Argonne, 19/03/2014.

Slides:



Advertisements
Similar presentations
Optical Fiber Calibration System and Adaptive Power Supply Jiri Kvasnicka, Ivo Polak, Jaroslav Cvach Prague, Institute of Physics Gerald Eigen,
Advertisements

2 Introduction   MiniCal test-beam studies started at the beginning of March (till March 6 we only had 17 APD’s, then 33 APD’s)   A few days were.
CALICE WHCAL testbeam at SPS H8 27 sep – 03 oct: 6 days for energies up to 180 GeV (+ polarity)
Study of the MPPC Performance - contents - Introduction Fundamental properties microscopic laser scan –check variation within a sensor Summary and plans.
Performance of MPPC using laser system Photon sensor KEK Niigata university, ILC calorimeter group Sayaka IBA, Hiroaki ONO, Paul.
Mass test of MPPC with a prototype of MEG II liquid xenon detector
Cosmic Ray Test Stand with Scintillating Cells for Digital Hadron Calorimeter As of 9am 05/28/2003.
Calice, UT ArlingtonJiri Kvasnicka, FZU, Prague1 LED notched fiber system Jiri Kvasnicka 1.Introduction 2.Test setup and fiber layout on HBU0.
LED calibration systems for CALICE hadron calorimeter Jiri Kvasnicka, Ivo Polak Institute of Physics (FZU), Prague June 11, 2011TIPP 2011, Kvasnicka1 The.
Sep 13, 2007CALICE Collaboration Meeting Prague 1 Long term test of LED pulsing  Long LED system stability measurement  Aging of part of optical system.
DESY, DEC-2013Ivo Polák, FZU, Prague1 Activity on Adaptive power & Calibration at Prague Ivo Polák, on behalf of Prague’s group QMB1A update.
Feb 20, 2009 CALICE meeting, Daegu, Korea QRL in Magnetic Field 1 QRLed Driver in Magnetic Field Jaroslav Zalesak Institute of Physics of the ASCR, Prague.
Optical calibration system for EUDET AHCAL module Jaroslav Cvach IPASCR 1.Calibration system for SiPM photodetectors 2.QRLED driver photoelectron spectra.
LED calibration systems for CALICE hadron calorimeter Jiri Kvasnicka Institute of Physics, Prague June 11, 2011TIPP 2011, Chicago1 Under.
Reports from DESY Satoru Uozumi (Staying at DESY during Nov 11 – 25) Nov-21 GLDCAL Japan-Korea meeting.
2013 MAR 21Ivo Polák FZU Prague1 Tasks for AIDA SiPM Adaptive HV PSU AIDA subgroup Prague 2013 meeting Characterization of temperature influence to SiPM.
Development of Multi-pixel photon counters(2) M.Taguchi, T.Nakaya, M.Yokoyama, S.Gomi(kyoto) T.Nakadaira, K.Yoshimura(KEK) for KEKDTP photon sensor group.
CALICE Annecy SiPM Adaptive HV PSU Ivo Polák & Jiří Kvasnička, J.Cvach Prague, Institute of Physics, Prague Erik van der Kraaij, Gerald.
Calibration UV LED system for CALICE scintillator based Tile Hadron Calorimeter Ivo Polák on behalf of the CALICE Collaboration Institute of Physics of.
First Results from T3B Calice AHCAL Meeting – Desy, January 2011 Christian Soldner Max-Planck-Institute for Physics.
FLC Group Test-beam Studies of the Laser-Wire Detector 13 September 2006 Maximilian Micheler Supervisor: Freddy Poirier.
DESY July 2, 2003J. Cvach: APD & preamp1 Checks of APD matrix & Prague 16ch. preamplifier 1.Cross talk between pixels 2.Homogeneity of response from pixels.
The MPPC Study for the GLD Calorimeter Readout Introduction Measurement of basic characteristics –Gain, Noise Rate, Cross-talk Measurement of uniformity.
The Scintillator ECAL Beam Test at FNAL K. Kotera, Shinshu-u, 1st October 2009 CALICE Scintillator ECAL group; Kobe University, Kyungpook University, the.
LCWA, ALCPG 2009 Albuquerque, NM Ivo Polák, FZU, Prague1 Calibration system of CALICE AHCAL detector Ivo Polák 1.Two calibration light distribution.
Light Calibration System (LCS) Temperature & Voltage Dependence Option 2: Optical system Option 2: LED driver Calibration of the Hadronic Calorimeter Prototype.
1. DESY tests 2. Electronics developments 3. Optical developments 4. Outlook Casablanca J. Cvach, CALICE Coll. meeting.
Montpellier, November 15, 2003 J. Cvach, TileHCAL and APD readout1 TileHCAL- fibre readout by APD APDs and preamplifiers Energy scan with DESY beam –Energy.
Uni Bergen:G. Eigen, E. van der Kraaij, J. Zalieckas FZU Prague: J. Cvach, J. Kvasnička, I. Polák Linear Collider Workshop, Tokyo 13/11/2013.
R&D of MPPC for T2K experiment PD07 : Photosensor Workshop /6/28 (Thu) S.Gomi T.Nakaya M.Yokoyama H.Kawamuko ( Kyoto University ) T.Nakadaira.
Scintillator tile-SiPM system development for CALICE Engineering AHCAL Prototype Michael Danilov, ITEP & CALICE LCWS10 Beijing March 2010 Outline New scintillator.
Scintillator Tiles and Test Beam Plans Calice Collaboration Meeting – Casablanca 2010 Christian Soldner Max-Planck-Institute for Physics.
Development of Multi-pixel photon counters(2) M.Taguchi, T.Nakaya, M.Yokoyama, S.Gomi(kyoto) T.Nakadaira, K.Yoshimura(KEK)
MPPC Measurements at LSU Brandon Hartfiel LSU Hardware Group Thomas Kutter, Jessica Brinson, Jason Goon, Jinmeng Liu, Jaroslaw Nowak Sam Reid January 2009.
5 July 2010Ivo Polák, FZU, Prague LED notched fibre distributing system Calibration system for SiPM Ivo Polák, Ji ř í Kvasni č ka
AIDA Plans for WP9 in Bergen In an HCAL for ILD we have ~10 6 channels  corrections due to temperature changes ( Δ T) need to be done automatically online.
MPPC status M.Taguchi(kyoto) T2K ND /7/7.
Development of Multi-Pixel Photon Counters(MPPC) Makoto Taguchi Kyoto University.
AHCAL Electronics. Status Commissioning Mathias Reinecke for the AHCAL developers HCAL main meeting Hamburg, Dec. 10th, 2009.
Status of photon sensor study at Niigata University -- SiPM and MPPC -- Photon sensor mini workshop 05/9/16 University Niigata University.
AHCAL Electronics. SPIROC2 and HBU measurement results Mathias Reinecke CALICE main meeting Univ. HASSAN II, Casablanca, Morocco Sept. 23rd, 2010.
Timing Studies of Hamamatsu MPPCs and MEPhI SiPM Samples Bob Wagner, Gary Drake, Patrick DeLurgio Argonne National Laboratory Qingguo Xie Department of.
Gain stability and the LYSO beam radiation monitor measurements
Linearity Tests The laser induced high noise pulses on the APD rendering it useless. We tried several methods to shield and filter the noise. With a stabilized.
Multipixel Geiger mode photo-sensors (MRS APD’s) Yury Kudenko ISS meeting, KEK, 25 January 2006 INR, Moscow.
CALICE Tungsten HCAL Prototype status Erika Garutti Wolfgang Klempt Erik van der Kraaij CERN LCD International Workshop on Linear Colliders 2010, October.
Scintillator tile – SiPM development at ITEP Michael Danilov, ITEP CALICE Meeting, Casablanca, 23 Sep 2010.
1 Calorimeters LED control LHCb CALO meeting Anatoli Konoplyannikov /ITEP/ Status of the calorimeters LV power supply and ECS control Status of.
DESY BT analysis - MPPC Saturation Correction - S. Uozumi Feb Sci-ECAL meeting 1.MPPC Gain Measurement (with LED data) 2.Inter-calibration of readout.
5-9 June 2006Erika Garutti - CALOR CALICE scintillator HCAL commissioning experience and test beam program Erika Garutti On behalf of the CALICE.
(s)T3B Update – Calibration and Temperature Corrections AHCAL meeting– December 13 th 2011 – Hamburg Christian Soldner Max-Planck-Institute for Physics.
Performance of new MPPC Nov. 21 Korea-Japan joint meeting Takashi Maeda Hideki Yamazaki Yuji Sudo (University of Tsukuba) --- Contents ---
G. Eigen, Paris, Introduction The SiPM response is non-linear and depends on operating voltage (V-V bd ) and temperature  SiPMs need monitoring.
CALICE, CERN June 29, 2004J. Zálešák, APDs for tileHCAL1 APDs for tileHCAL MiniCal studies with APDs in e-test beam J. Zálešák, Prague with different preamplifiers.
Energy Reconstruction in the CALICE Fe-AHCal in Analog and Digital Mode Fe-AHCal testbeam CERN 2007 Coralie Neubüser CALICE Collaboration meeting Argonne,
Upgrade of the MEG liquid xenon calorimeter with VUV-light sensitive large area SiPMs Kei Ieki for the MEG-II collaboration 1 II.
AHCAL Electronics. Status of Integration Mathias Reinecke for the DESY AHCAL developers AHCAL main and analysis meeting Hamburg, July 16th and 17th, 2009.
Study of the MPPC for the GLD Calorimeter Readout Satoru Uozumi (Shinshu University) for the GLD Calorimeter Group Kobe Introduction Performance.
Development of Multi-pixel photon counters(2) M.Taguchi, T.Nakaya, M.Yokoyama, S.Gomi(kyoto) T.Nakadaira, K.Yoshimura(KEK) for KEKDTP photon sensor group.
M.Taguchi and T.Nobuhara(Kyoto) HPK MPPC(Multi Pixel Photon Counter) status T2K280m meeting.
Performance of 1600-pixel MPPC for the GLD Calorimeter Readout Jan. 30(Tue.) Korea-Japan Joint Shinshu Univ. Takashi Maeda ( Univ. of Tsukuba)
LED notched fibre distributing system Calibration system for SiPM
Light mixer proposal Arjan Heering, Anton Karneyeu
LED notched fibre system short HBU0 party with QMB6
Igor Mandić1, Vladimir Cindro1, Gregor Kramberger1 and Marko Mikuž1,2
ScECAL+AHCAL+TCMT Combined Beam FNAL
BESIII EMC electronics
R&D of MPPC for T2K experiment
Comparison of CPTA and Hamamatsu SiPMs
The MPPC Study for the GLD Calorimeter Readout
Presentation transcript:

Uni Bergen:G. Eigen, J. Zalieckas, E. van der Kraaij FZU Prague: J. Cvach, J. Kvasnicka, I. Polák CALICE meeting, Argonne, 19/03/2014

Introduction The gain of SiPMs depends both on bias voltage and on temperature Gain decreases with temperature Gain increases with bias voltage For stable operations, we need to keep gain constant In an analogue hadron calorimeter with millions of channels, this is a difficult task due to temperature variations Thus, it is desirable to adjust V bias to compensate for T changes to keep the gain constant Goal: build a V bias regulator to keep gain constant (<1%) First, we need to measure SiPM gain vs temperature and vs V bias  determine dV/dT to obtain constant gain  determine dV/dT to obtain constant gain   apply the compensation and demonstrate proof of principle This is work conducted in the framework of the EU project AIDA 2 J.Kvasnicka, Argonne, 19/03/2014

SiPM Test Setup We work in a climate chamber at CERN that is stable within 0.2˚C Use digital oscilloscope read out by PC, low voltage & bias voltage power supplies Use pulse generator for LED signal Blue LED shines on SiPM 3 J.Kvasnicka, Argonne, 19/03/2014 SiPM + preamp +T sensor + LED

Temperature Measurement Use 3 pt 1000 sensors Near SiPM, inside and outside black box Use LM35 sensor for compensation feedback Vary T from 5˚C to 40˚C in 5˚C steps except in 20˚– 30˚C range (2˚C steps) except in 20˚– 30˚C range (2˚C steps)  T SiPM ~T SET + 0.4˚C,  T SiPM ~T SET + 0.4˚C,  offset is same over entire range 4 J.Kvasnicka, Argonne, 19/03/2014 pt 1000 near SiPM pt 1000 outside black box LM35 sensor SiPM

SiPM Tested We measured the dG/dT & dG/dV dependence for 15 SiPMs from 3 manufacturers We tested the V bias adjustment on 4 SiPMs: CPTA 857 CPTA 1677 KETEK W 12 MPPC 50um MPPC 20um “B” Note that CPTA sensors were sensors were attached to attached to 3x3 cm 2 scintillator tiles while the other sensors were directly 3x3 cm 2 scintillator tiles while the other sensors were directly illuminated by blue LED illuminated by blue LED 5 J.Kvasnicka, Argonne, 19/03/2014    

Layout of the Adaptive Power Regulator Ivo designed V bias regulator that operates in 10 V to 80 V region plus T correction effects Temperature slope: <1 to100 mV/K, both positive and negative First design of test board done summer 2013 Tested in CERN Sep 2013, results shown in this presentation 6 J.Kvasnicka, Argonne, 19/03/2014 Analogue voltage regulator Voltage reference

Gain Determination (being updated) Determine gain by fitting Gaussian functions to peaks Gaussian functions to peaks of single pe spectra of single pe spectra Define gain as Distance between 1 pe & 2 pe peaks (MPPCs) Distance between pedestal & 1 pe peak (CPTA, KETEK) Define the error on the gain as the errors of the two fitted Gaussian mean values added in quadrature Quality cut: Require gain extracted from Gaussian peaks and that from multikernel peaks not to differ by more than 20% (a sign of a bad fit) 7 J.Kvasnicka, Argonne, 19/03/2014 RooFit multikernel function Individual Gaussian fits

Gain vs Voltage for CPTA 857 Each point is the gain extracted from 50k waveforms of 80 ns from 50k waveforms of 80 ns Linear fit to these distributions yields break-down voltage & dG/dV yields break-down voltage & dG/dV Break-down voltage drops with T dG/dV~ capacitance is not constant  it increases linearly with temperature  it increases linearly with temperature 8 J.Kvasnicka, Argonne, 19/03/2014

Gain vs Temperature for CPTA 857 At V bias =33.4 V & T=25˚C: G=6.4×10 5 Extract V breakdown =31.9 V 0 th order polynomial fit yields: dG/dV=(4.32±0.01 stat )×10 5 /K 1 st order polynomial fit yields: dG/dV=(4.41±0.11 sys )×10 5 /K dG/dV=(4.41±0.11 sys )×10 5 /K 9 J.Kvasnicka, Argonne, 19/03/ th order polynomial fit yields: 0 th order polynomial fit yields: dG/dT=(-5.76±0.02 stat )×10 3 /K dG/dT=(-5.76±0.02 stat )×10 3 /K Determine: Determine: dV/dT= / dV/dT= /  = 13.33±0.35 mV/K Gain stability: σ(dV/dT)  (dG/dV)  (ΔT/G) = 0.31%

Gain after V bias Adjustment for CPTA 857 Adjust voltage using V bias regulator test board (Note: color code gives the range of voltage applied (Note: color code gives the range of voltage applied e.g. blue: <V bias <33.45) At each temperature, take 20 samples with 50k waveform each Linear fit to all data points offset: (6.71±0.02)x10 5 slope: 15.38±75.87 Gain is uniform in range of interest 5˚- 40˚C interest 5˚- 40˚C  measured non-uniformity is < ±0.04% is < ±0.04% 10 J.Kvasnicka, Argonne, 19/03/2014 CPTA 857 Gain stability: 0.31%

Gain after V bias Adjustment for other SiPMs Perform similar study with 3 other SiPMs from CPTA, KETEK other SiPMs from CPTA, KETEK and Hamamatsu and Hamamatsu CPTA 1677: non-uniformity <± 0.3% KETEK W12: non-uniformity <±1.2% Hamamatsu 11759: non-uniformity <±1.7% 11 J.Kvasnicka, Argonne, 19/03/2014 KETEK W12 CPTA 1677 Hamamatsu Gain stability: 0.28% Gain stability: 0.05% Gain stability: 0.9%

Summary of the 2013 measurements Results from the CERN 2013: 4 SiPM measured and compensated T range 5˚ – 40˚C Large statistics taken, dG/dV, dG/dT calculated dV/dT extrapolated  used as a compensation constant Measurement taken with the Adaptive V bias adjustment test board Various dV/dT: mV/K for CPTA & KETEK sensors 55 mV/K for Hamamatsu sensors Hamamatsu sensors operate at higher V bias than CPTA/KETEK sensors Adjustment range ΔV=0.7 (1.9) V for CPTA & KETEK (Hamamatsu) for The gain stabilization with the V bias regulator test board works well  for all four tested SiPMs, gain stability is < 1% as required 12 J.Kvasnicka, Argonne, 19/03/2014

Adaptive Power Regulator 2 nd Prototype Prototype finished in Jan 2014 Extended V range: SiPM <130 V APD <450V Has a front panel Delivered to CERN in Feb SiPMs measured, same setup T range extended: 1~50˚C Data analysis ongoing 13 J.Kvasnicka, Argonne, 19/03/2014

Analysis (being updated) Some bad fits observed in the data  loosing statistics  we want to improve that Problem observed in the histogram Cause: Numerical rounding issue? Charge extraction procedure rewritten Ongoing evaluation of best fit option Individual peak fits RootKeysPDF Sum of Gaussians (deals better with noisy histograms from high T) 14 J.Kvasnicka, Argonne, 19/03/2014

Summary (2014) and Next Steps ADApower prototype finished Jan 2014 New data was CERN Feb 2014 and will be analyzed  Aim for better gain extraction Check the previous (2013) data AIDA note almost finished (delayed), will be released soon Q assemble and finalize few other units of ADApower box Integrate the V bias regulator PCB into the FE readout electronics of the AHCAL  summer 2014 Write final report for AIDA project  fall J.Kvasnicka, Argonne, 19/03/2014

Backup Slides 16 J.Kvasnicka, Argonne, 19/03/2014

Voltage Temperature Relation dV/dT is obtained from a ratio of 2 first-order differential equations dV/dT is obtained from a ratio of 2 first-order differential equations dG/dT=a-b*V+O(V 2 ) dG/dT=a-b*V+O(V 2 ) dG/dV=c+d*T+O(T 2 ) dG/dV=c+d*T+O(T 2 ) Thus, an analytical solution exists Thus, an analytical solution exists For the above linear case, we get: For the above linear case, we get: For a linear model in dG/dT For a linear model in dG/dT and dG/dV find following and dG/dV find following V versus T dependence V versus T dependence 17 J.Kvasnicka, Argonne, 19/03/2014 V T [˚C] Linear approximation for CPTA 857 (d≠0)

Summary of dV/dT Measurements V bias for Hamamatsu is ~70 V while V bias for CPTA is ~33 V while V bias for CPTA is ~33 V & V bias for KETEK is ~28 V & V bias for KETEK is ~28 V For KETEK and CPTA, dV/dT is ~15-20 mV/K dV/dT is ~15-20 mV/K for Hamamatsu, for Hamamatsu, dV/dT is ~55 mV/K dV/dT is ~55 mV/K Thus, compensation will be Thus, compensation will be simpler for CPTA and KETEK simpler for CPTA and KETEK SiPMs SiPMs We tested the compensation on: 4 SiPMs in 2013 The same 4 and 1 additional SiPMs in J.Kvasnicka, Argonne, 19/03/2014

Results for CPTA 1677 Extract dV/dT=18.51±0.31 mV/K and gain stability of 0.1% Extract dV/dT=18.51±0.31 mV/K and gain stability of 0.1% 19 J.Kvasnicka, Argonne, 19/03/2014

Results for KETEK W12 Extract dV/dT=21.29±0.08 mV/K and gain stability of 0.01% Extract dV/dT=21.29±0.08 mV/K and gain stability of 0.01% 20 J.Kvasnicka, Argonne, 19/03/2014

Results for Hamamatsu Extract dV/dT=54.3±0.75 mV/K and gain stability of 0.28% Extract dV/dT=54.3±0.75 mV/K and gain stability of 0.28% 21 J.Kvasnicka, Argonne, 19/03/2014

22 J.Kvasnicka, Argonne, 19/03/2014