1 Radiometer Medical ApS, Åkandevej 21, DK-2700 Brønshøj, Tel: +45 38 27 38 27, www.radiometer.com RTC, December 2004 ABL800 FLEX Calibration.

Slides:



Advertisements
Similar presentations
The ISA for Physics What you need to revise.
Advertisements

SVCC Respiratory Care Programs Analysis of ABG Samples.
Instrumental Analysis
Nernst Equation Consider the half reaction:
Electrode measuring principle
CHEMISTRY ANALYTICAL CHEMISTRY Fall
ISE Ion Selective Electrodes Prepared By Michigan Department of Environmental Quality Operator Training and Certification Unit.
1 Calibration of pH meters, glass electrodes and conductometers at INPL Calibration of pH meters, glass electrodes and conductometers at INPL Dr. Elena.
BUILD-UP SIMULATIONS FOR DAFNE WIGGLER W/ ELECTRODES Theo Demma.
Stoichiometry: Quantitative Information about chemical reactions.
1 SSS II Lecture 1: Correlation and Regression Graduate School 2008/2009 Social Science Statistics II Gwilym Pryce
Announcements. Today Review membrane potential What establishes the ion distributions? What confers selective permeability? Ionic basis of membrane potential.
World Health Organization
Monroe L. Weber-Shirk S chool of Civil and Environmental Engineering pH measurements The probe, from chemistry to voltage to a number + + V.
Lecture 19: Tues., Nov. 11th R-squared (8.6.1) Review
Chapter 6 The Normal Distribution and Other Continuous Distributions
This work is licensed under a Creative Commons Attribution-Noncommercial-Share Alike 2.0 UK: England & Wales License Lab 6: Saliva Practical Beer-Lambert.
Monroe L. Weber-Shirk S chool of Civil and Environmental Engineering pH measurements The probe, from chemistry to voltage to a number + + V.
Union College Mechanical Engineering MER301 -Measurement System and Uncertainty Analysis LECTURE 16-17: Measurement System Analysis and Uncertainty Analysis-Homework.
Chemical calculations used in medicine part 1 Pavla Balínová.
BPT2423 – STATISTICAL PROCESS CONTROL.  Estimation of Population σ from Sample Data  Control Limits versus Specification Limits  The 6σ Spread versus.
The Siggaard-Andersen acid-base chart
Classification of Instruments :
Quality Assurance in the clinical laboratory
Chem. 31 – 2/18 Lecture. Announcements Turn in AP1.2 Quiz today Exam 1 coming up (1 week from next Monday) Today’s Lecture –Chapter 4 Material Calibration.
Measurement Concepts & Interpretation. Scores on tests can be interpreted: By comparing a client to a peer in the norm group to determine how different.
Lecture 2b. Electromagnetic Spectrum Visible range: = nm Ultraviolet: = nm Low energyHigh energy.
CALIBRATION METHODS.
Chemometrics Method comparison
Method Comparison A method comparison is done when: A lab is considering performing an assay they have not performed previously or Performing an assay.
Acids and Bases Biotechnology I. Life Chemistry  Based on water  Cells contain 80-90% water  Proper pH essential to ALL living systems Plants cannot.
1 Radiometer Medical ApS, Åkandevej 21, DK-2700 Brønshøj, Tel: , Transcutaneous measuring principles – tc and saturation.
Oximetry measuring principle
COLORIMETRYCOLORIMETRY Prepared By Michigan Department of Environmental Quality Operator Training and Certification Unit Note: A printed description of.
The following minimum specified ranges should be considered: Drug substance or a finished (drug) product 80 to 120 % of the test concentration Content.
Chem. 31 – 9/23 Lecture Guest Lecture Dr. Roy Dixon.
Accuracy and Precision
Biostatistics: Measures of Central Tendency and Variance in Medical Laboratory Settings Module 5 1.
Quality Control Lecture 5
Announcements:. Last lecture 1.Organization of the nervous system 2.Introduction to the neuron Today – electrical potential 1.Generating membrane potential.
Determining the Concentration of a Solution: Beer’s Law
Quality Assurance How do you know your results are correct? How confident are you?
1 Dec. 11, 1997 LEADS Quality Control Systems Robert Brewer (512) Monitoring Operations Division Network QA Manager.
Lesson declared p[H] After finished this lesson student will to able to # Say about p[H]. # Say about pH indicators. # Say about Buffer solution.
CALIBRATION METHODS. For many analytical techniques, we need to evaluate the response of the unknown sample against the responses of a set of standards.
Karst Chemistry II. Conductivity – Specific Conductance Conductance – the electrical conductivity of aqueous solution, and is directly related to the.
Chemical calculations I
Chapter Goals (Membrane Potentials) After studying this chapter, students should be able to 1. explain what is meant by active transport and describe how.
Basic Business Statistics
Applications of Intermolecular Potentials. Example 1. A gas chromatograph is to be used to analyze CH 4 -CO 2 mixtures. To calibrate the response of the.
CORRELATION ANALYSIS.
Uncertainty & Errors in Measurement. Waterfall by M.C. Escher.
POTENTIALS AND THERMODYNAMICS OF CELLS (1) POTENTIALS AND THERMODYNAMICS OF CELLS (1)
CORRELATION-REGULATION ANALYSIS Томский политехнический университет.
EQUIPMENT and METHOD VALIDATION
Instrumental Analysis Ion selective electrodes
L Berkley Davis Copyright 2009 MER301: Engineering Reliability Lecture 8 1 MER301: Engineering Reliability LECTURE 8: Chapter 4: Statistical Inference,
David Housman for Math 323 Probability and Statistics Class 05 Ion Sensitive Electrodes.
Chapter 13 Linear Regression and Correlation. Our Objectives  Draw a scatter diagram.  Understand and interpret the terms dependent and independent.
Experiment (6) : Kinetic Study of inversion of cane sugar catalyzed by an acid Theory In chemistry, specific rotation ([α]) is a property.
Quality Assurance in the clinical laboratory
Spectroscopy Techniques
Rapid Lab Systems measuring principle
Clinical Pharmacokinetics
Nernst Equation Consider the half reaction:
Spectrophotometry A method to determine concentration of a species exploiting the absorption of EMR.
Quality Assessment The goal of laboratory analysis is to provide the accurate, reliable and timeliness result Quality assurance The overall program that.
The Normal Distribution
Presumptions Subgroups (samples) of data are formed.
Presentation transcript:

1 Radiometer Medical ApS, Åkandevej 21, DK-2700 Brønshøj, Tel: , RTC, December 2004 ABL800 FLEX Calibration

2 Agenda What is calibration? Calibration material Why calibrate? Obtaining a calibration line Sensitivity Status Drift pO 2 electrode Metabolite electrodes Calibration results

3 What is calibration? Liquid or gas with a known parameter value is drawn into the measuring chamber The electrode measures an electrical signal corresponding to the known value of the liquid or gas This is repeated with liquid or gas of another known parameter value The analyzer establishes a calibration line for each parameter and adjusts its readings to measure the known value of each parameter “The calibration process determines and checks the accuracy with which the ABL measures its parameters” (quote from the Operator’s Manual)

4 Calibration material Two calibrating solutions, Cal 1 and Cal 2, are used to calibrate the pH, electrolyte and metabolite electrodes Two gas mixtures, Gas 1 and Gas 2, are used to calibrate the pCO 2 and pO 2 electrodes The tHb Calibration Solution is used to calibrate the optical system

5 Calibration solutions and gases Composition: Cal 1 contains: K +, Na +, Ca 2+, Cl -, Glu, Lac, pH = 7.4 Cal 2 contains: K +, Na +, Ca 2+, Cl –, pH = 6.8 Gas 1 contains: O 2 (19.7 %), CO 2 (5.6 %) Gas 2 contains: CO 2 (11.2 %) Nominal concentrations/percentages stated on container label Precise concentrations/percentages contained in the bar code

6 Why calibrate? The electrodes are active elements and must be calibrated regularly Signals from the electrodes change because of, e.g. ­ protein build-up ­ worn-out membranes ­ aging electrodes

7 Calibration line The calibration line represents the relationship between the electrical signal measured at the electrode and the concentration of the substrate specific to the electrode –106 –69 pH mV

8 Obtaining a calibration line Each electrode has a different calibration line ­ the pH electrode is used as an example The electrical signals from two calibration solutions of known pH are plotted ­ Cal 1 solution with pH = gives a reading of –106 mV ­ Cal 2 solution with pH = gives a reading of –69 mV –106 –69 mV pH

9 Obtaining a calibration line The relationship between measured potential and concentration is linear The calibration line forms the scale used to convert the measured potential to pH –106 –69 mV pH

10 A 1-point calibration line versus a 2-point Only Cal 1 is measured, giving one calibration point A calibration line with the same slope as the previous 2-point calibration is superimposed on this new calibration point –106 –69 mV pH P cal.

11 Monitoring the calibration line The calibration line is updated at every calibration and is monitored with respect to: Sensitivity Status/zero point Describes the slope and position of the calibration line Drift Describes the variation in the calibration line between consecutive calibrations Each electrode has its own status and sensitivity limits

12 Sensitivity (slope) The sensitivity is an evaluation of the slope of a 2-point calibration line in relation to the slope of the theoretical calibration line ­ the theoretical calibration line is derived from the Nernst equation The theoretical pH calibration line has a slope = –61.5 mV/pH –106 –69 mV pH Actual Theoretical

13 Sensitivity, an example During a 2-point calibration the following is measured: –106 –69 mV pH –101 –65 EE  pH Theo. 2P-cal.  E –101.0 – (–65.0)  E –101.0 – (–65.0)  pH × (–61.5) (7.4–6.8)×(–61.5)  pH × (–61.5) (7.4–6.8)×(–61.5) Sensitivity = × 100 % Example: Sensitivity = × 100 % = 97.6 % Cal 1 solution with pH = gives a reading of –101.0 mV - Cal 2 solution with pH = gives a reading of –65.0 mV

14 ABL800 FLEX sensitivity limits EE  pH Theo. 2P-cal. EE  pH Theo. 2P-cal. % ParameterSensitivity limits pH % pCO % pO2pO pA/mmHg ( pA/kPa) cK+cK % cNa % cCl % cCa % cGlu pA/mM cLac pA/mM For a calibration to be accepted, the sensitivity must fall within the following limits:

15 Status The status describes the deviation of a 1-point calibration point from the theoretical calibration line ­ the status is evaluated using only one calibration point; i.e. we assume that the sensitivity (slope) is unchanged  E –100 – (–106) –61.5–61.5 Status = Example: Status = = pH –69 Theo. 1P-cal. mV –106 –100 EE 7.300

16 ABL800 FLEX status limits For a calibration to be accepted, the status and zero point must fall within the following limits Theo. cal. Theo. cal. % ParameterStatus limits pH pCO mmHg cK+cK mmol/L cNa mmol/L cCl mmol/L cCa mmol/L Zero point pO2pO2 < 6 mmHg (0.8 kPa) cGlu< pA cLac< pA

17 Sensitivity and Status - during measurements The sample pH is calculated as follows pH sample = pH in sample E sample = measured potential in sample E cal1 = measured potential in Cal 1 pH cal1 = pH in Cal 1 –61.5= sensitivity of theoretical calibration line

18 Drift The drift of an electrode is a description of how much a calibration differs from the last calibration Drift is a check of whether the adjustment performed during calibration is acceptable This is defined by the Drift Tolerances A drift value of –0.001 pH units indicates that during calibration the analyzer has been adjusted to measure 1 mpH unit higher

19 Calculation of drift Drift 1 reflects the change in position between two calibrations (given at both 1-point and 2-point calibrations) Drift 2 reflects the change in slope between two 2-point calibrations (given only at 2-point calibrations) mV pH Slope of first 2P-cal. First 2P-cal. Second 2P-cal. Drift 1 Drift 2

20 Drift tolerances Drift tolerances express the extent to which drift values for an electrode can fluctuate Drift values lying outside the tolerance limits indicate that the electrode is unstable and thus incapable of providing reliable calibrations Drift tolerances can be user- defined in the Setup programs, but should be done so with care Access from Calibration Setup

21 Sensitivity of pO 2 electrode Obtained from a 2-point calibration using Gas 1 and pA kPa 2P-cal  I 2800 – 60  pO – 0.0 Sensitivity = Example: Sensitivity = = 147 pA/kPa

22 pO 2 zero point Calculated from the sensitivity and the current measured at pO 2 = kPa First 2P- cal. Second 2P-cal I 0 85 I 0 85 Sensitivity, previous cal. 147 Sensitivity, previous cal. 147 Zero point = Example: Zero point = = 0.58 kPa pA

23 Metabolite sensitivity Base line formed by extrapolation of consecutive zero-point currents measured on rinse solution Time I cal I0I0 I cal – I – 400 C Met 10.0 Sensitivity = Example, sensitivity, Glu = = 510 pA/mmol/L pA

24 Calibration of optical system Zero point measured during each calibration Cuvette path length, L, measured during tHb calibration A =  × c × L = absorbance = extinction coefficient = concentration of optically active substance in tHb calibration solution = cuvette path length

25 tHb calibration It is recommended that a tHb calibration is performed manually every three months

26 Summary pH, pCO 2, K, Na, Ca, Cl SensitivityStatusDrift pO 2 Sensitivity Zero point Drift Glu, Lac Base line SensitivityDriftOXI Zero point Cuvette path length

27 Calibration status The Analyzer Status – Calibration screen displays the results of the most recently performed calibration of each type available

28 Calibration Log Results from previous calibrations are viewed via the Calibration Log

29 2-Point calibration result

30 1-Point calibration result

31 Calibration printout

32 Radiometer Medical ApS, Åkandevej 21, DK-2700 Brønshøj, Tel: , Radiometer Training Center, December 2004