For the CAP Instrumentation and Chemistry Resource Committees

Slides:



Advertisements
Similar presentations
Method Validation Studies It’s a Jungle out There!
Advertisements

CAP/POCT 2003 Carol Riley-Hunte, RRT Senior Education Specialist Bayer HealthCare.
Bay Area POCC Spring Meeting Stanford University Medical Center
Day to Day Management of Quality Control
User verification of performance Zahra Khatami/Robert Hill BHR University Hospitals/Sheffield Teaching Hospitals Focus 2013.
Automating Result Verification
PHARM 462 PART / /31 Good Manufacturing Practices (GMP) VALIDATION of ANALYTICAL TEST METHODS.
World Health Organization
Dr Samah Kotb Lecturer of Biochemistry 1 CLS 432 Dr. Samah Kotb Nasr El-deen Biochemistry Clinical practice CLS 432 Dr. Samah Kotb Nasr.
Laboratory Quality Control
This teaching material has been made freely available by the KEMRI-Wellcome Trust (Kilifi, Kenya). You can freely download,
MLAB 1335 Immunology/Serology Terry Kotrla
Kyiv, TRAINING WORKSHOP ON PHARMACEUTICAL QUALITY, GOOD MANUFACTURING PRACTICE & BIOEQUIVALENCE Validation of Analytical Methods Used For Bioequivalence.
Chemometrics Method comparison
PAT Validation Working Group Process and Analytical Validation Working Group Arthur H. Kibbe, Ph.D. Chair June 13, 2002.
Method Comparison A method comparison is done when: A lab is considering performing an assay they have not performed previously or Performing an assay.
Unit #7 - Basic Quality Control for the Clinical Laboratory
Quality Control In Measurements Tom Colella CLAS Goldwater Environmental Lab.
Validation of Analytical Method
Method Selection and Evaluation Method Selection and Evaluation D. Kefaya EL- Sayed Mohamed Prof. Of Clinical Pathology (Clinical Chemistry), Mansoura.
WWLC Standard Operating Procedures Presented by Frank Hall, Laboratory Certification Coordinator.
Introduction to ISO New and modified requirements.
THROUGH PROFICIENCY TESTING
VALIDATION METHODOLOGY
Analytical considerations
Method Validation and Verification: An Overview Patricia Hanson, Biological Administrator I Florida Department of Agriculture and Consumer Services, Food.
Quality WHAT IS QUALITY
1 / 9 ASTM D19 Method Validation Procedures William Lipps Analytical & Measuring Instrument Division July, 2015.
Quality Control Lecture 5
Quality Control and Patient Risk Curtis A. Parvin, Ph. D
Laboratory QA/QC An Overview.
Quality Assurance How do you know your results are correct? How confident are you?
QC THE MULTIRULE INTERPRETATION
Biochemistry Clinical practice CLS 432 Dr. Samah Kotb Lecturer of Biochemistry 2015 Introduction to Quality Control.
Validation Defination Establishing documentary evidence which provides a high degree of assurance that specification process will consistently produce.
Metrology for Chemical Analysis
Verification of precision and bias
All you wanted to know about laboratory tests, but were afraid to ask By Dietmar Stöckl STT Consulting The informed patient meeting the laboratory.
Industrial Technology Institute Test Method Validation & Verification H.P.P.S.Somasiri Principal Research Scientist / SDD-QAD /QM Industrial Technology.
Control Charts and Trend Analysis for ISO 17025
LECTURE 13 QUALITY ASSURANCE METHOD VALIDATION
 Remember Chemistry panel Quality Control:-  In a medical laboratory, it is a statistical process used to monitor and evaluate the analytical process.
 Routine viral diagnostics: indirect and direct detection of viruses. ◦ Indirect detection: serological tests; ◦ Direct detection:  Viral antigens;
Comparison Testing Demystified: Applications of Correlation Testing PAUL RICHARDSON. MSC, FIBMS. SENIOR HPTN QA/QC COORDINATOR
By: EID ALATAWI Serial Dilution. Introduction: Many of the laboratory procedures involve the use of dilutions. It is important to understand the concept.
SEMINAR ON PRESENTED BY BRAHMABHATT BANSARI K. M. PHARM PART DEPARTMENT OF PHARMACEUTICS AND PHARMACEUTICAL TECHNOLGY L. M. COLLEGE OF PHARMACY.
EQUIPMENT and METHOD VALIDATION
Diagnostic clinical chemistry
means to “TO CHECK OR PROVE THE VALIDITY OF” According to FDA – “ The goal of validation is to establish a documented evidence which provides a high degree.
Introduction to Quality Assurance. Quality assurance vs. Quality control.
Quality Assurance.
Method Validation-where do I start?
Quality Assurance Procedure Manuals.
Unit #6 - Basic Quality Control for the Clinical Laboratory
Introduction to Lab Quality
Author: Nurul Azyyati Sabri
Instrument Maintenance
1 2 3 INDIAN INSTITUTE OF TECHNOLOGY ROORKEE PROJECT REPORT 2016
EPA Method Equivalency
Practical clinical chemistry
Have Instrument, make method; How new methods are made and validated
This teaching material has been made freely available by the KEMRI-Wellcome Trust (Kilifi, Kenya). You can freely download,
EPA Method Equivalency
به نام خدا تضمين کيفيت در آزمايشگاه
What it Means, Why it Works, and How to Comply
Choice of Methods and Instruments
Quality Control Lecture 3
Quality Assessment The goal of laboratory analysis is to provide the accurate, reliable and timeliness result Quality assurance The overall program that.
08/ COMMUTABILiTY.
STANDARDISATION, HARMONISATION AND TRACEABILITY
Presentation transcript:

For the CAP Instrumentation and Chemistry Resource Committees Calibration, Calibration Verification, Analytical Measurement Range, and Clinically Reportable Range For the CAP Instrumentation and Chemistry Resource Committees 1

The Truth, Please What’s your response when You receive the results of a CAP Linearity Survey? You come to Checklist Question CHM.13400 (one of several relating to Calibration Verification)? 2

The Truth, Please What’s your response when You receive the results of a CAP Linearity Survey? You come to Checklist Question CHM.13400 (one of several relating to Calibration Verification)? If you’re anything like me, it’s not “Wow, let me at it. I was really looking forward to taking care of this!” Who I Am: Gary L. Horowitz, MD, FCAP Director, Clinical Chemistry, Beth Israel Deaconess Medical Center Associate Professor of Pathology, Harvard Medical School Chair, CAP Chemistry Resource Committee 3

Terms Calibration 4

Terms that Instill Fear Calibration Calibration Verification Analytical Measurement Range (AMR) Clinically Reportable Range (CRR) 5

Terms that Instill Fear Calibration Calibration Verification Analytical Measurement Range (AMR) Clinically Reportable Range (CRR) CAP Linearity Surveys 6

Why We’re Here Today confusing, but important, concepts audioconference format: allows anyone to visit the web as often as needed objectives: explain the concepts reveal how they relate to the accreditation process show you in detail how the CAP Linearity Surveys enable you to address these issues for your laboratory ambitious goals, but we have two excellent speakers, experts in Laboratory Medicine in CAP Surveys 7

Introductions Greg Miller, PhD Professor of Pathology, Virginia Commonwealth University Consultant Member, CAP Chemistry Resource Committee 8

Introductions Greg Miller, PhD Professor of Pathology, Virginia Commonwealth University Consultant Member, CAP Chemistry Resource Committee Anthony Killeen, MD, PhD, FCAP Associate Professor of Pathology, University of Minnesota Member, CAP Instrumentation Resource Committee Former Member and Chair, CAP Chemistry Resource Committee Program developed as joint effort of these two CAP Resource Committees 9

Objectives Differentiate between the terms calibration, calibration verification, analytical measurement range (AMR) and clinically reportable range (CRR). Clarify the accreditation requirements regarding these items. Explain techniques to perform the appropriate verifications. Clarify CAP Linearity (LN) Survey reports. 10

CLIA 493.2 Reportable range: span of test result values over which the lab can establish or verify the accuracy of the measurement response. 11

CAP (two concepts) CLIA 493.2 Analytical measurement range: range of analyte values that a method can directly measure on the specimen without any dilution, or other pretreatment not part of the usual assay process. Reportable range: span of test result values over which the lab can establish or verify the accuracy of the measurement response. 12

CAP (two concepts) CLIA 493.2 Clinically reportable range: range of analyte values that are reported as a quantitative result, allowing for specimen dilution or other pretreatment used to extend the actual AMR. Reportable range: span of test result values over which the lab can establish or verify the accuracy of the measurement response. 13

CAP CLIA Calibration: set of operations that establish the quantitative relationship between the reagent system / instrument response and the concentration or activity values of an analyte. Calibration: process of testing and adjusting to establish a correlation between the measurement response and the concentration. 14

CAP (two concepts) CLIA 493.2 Calibration verification: assaying materials of known concentration to substantiate the system’s calibration throughout the reportable range. 15

CAP (two concepts) CLIA 493.2 Calibration verification: process of confirming that the current calibration settings remain valid for a method. Calibration verification: assaying materials of known concentration to substantiate the system’s calibration throughout the reportable range. 16

CAP (two concepts) CLIA 493.2 Calibration verification: assaying materials of known concentration to substantiate the system’s calibration throughout the reportable range. Analytical measurement range verification: process of confirming that the assay system will correctly recover the concentration or activity of the analyte over the AMR. 17

CLIA 493.1255: Calibration and calibration verification procedures (a) Perform and document calibration procedures. Follow manufacturer’s instructions. Use criteria established by lab during validation of method when put in service. Whenever calibration verification fails to meet the lab’s specifications. 18

CLIA 493.1255: Calibration and calibration verification procedures (b) Perform and document calibration verification procedures. Follow manufacturer’s instructions Use criteria established by lab during validation of the method when put in service Include minimum, mid-point and maximum value to verify the laboratory’s reportable range (this is AMR) At least every 6 months or when the following occur: Change of reagents unless it does not affect the range to report patient results (this is AMR), or QC values (this is CAL) Major preventive maintenance (this is CAL) QC trend, shift or out of limits (this is CAL) Schedule for verifying reportable range (this is AMR) requires more frequent calibration verification 19

Calibration (two point) Analytical Measurement Range extends beyond the high calibrator = Reportable Results SIGNAL CONCENTRATION CALIBRATOR TARGET VALUES May be measurement system specific 20

Calibration (two point) Example: typical serum glucose method Low calibrator = reagent blank (0 mg/dL) High calibrator = 356 mg/dL AMR = 10-600 mg/dL The AMR is based on the expectation that the signal is linearly related to glucose concentration over the 10-600 mg/dL range. Verification of the AMR expectation is required because the calibrators do not define the complete measurement range. 21 21

Calibration (multi-point) Analytical Measurement Range = Reportable Results Note that when calibrators span the AMR, the calibration is fully defined over the AMR SIGNAL CONCENTRATION CALIBRATOR TARGET VALUES May be measurement system specific 22

Accuracy (Trueness) Traceability Manufacturer provides calibration traceability to the highest order reference method or reference material available Manufacturer’s product calibrator has method (and sometimes reagent lot) specific target values Cannot mix reagents and calibrators from different manufacturers 23

Terminology Trueness: the mean bias between a routine method and a reference method determined from replicate measurements of multiple patients’ samples Imprecision is removed (or minimized) by replication Accuracy: the difference between a routine method and a reference method for a single measurement on a single patient’s sample Difference includes influence of both bias and imprecision for a single measurement Refer to the CLSI harmonized terminology database for more detailed definitions at www.clsi.org/resources 24

Calibration Verification Verify that measurement system calibration has not changed since the last calibration event And, where applicable (many routine methods), Verify that method calibration is in conformance to the manufacturer’s specifications (i.e. is consistent with the manufacturer’s calibration traceability) Special case (uncommon) when a lab can verify calibration traceability to an accuracy based reference system 25

Commutable Reference/ PT/ QC Material Has equivalent mathematical relationships among the results of different measurement procedures for a reference material and for native clinical samples (International Standards Organization, Clinical and Laboratory Standards Institute derived). Produces a numeric result from a routine method that is the same as would be obtained for a native clinical sample that has the same quantity of an analyte (measurand) present. 26

Non-commutable materials may have a matrix bias Bias in a result that is attributable to modification of the usual clinical sample matrix caused by preparation of a reference material, PT material or QC material CAP has investigated matrix bias in PT materials using fresh frozen off-the-clot samples (that are expected to be commutable with clinical samples) CAP Home (www.cap.org) > CAP Reference Resources and Publications > Chemistry CAP has developed accuracy based PT Surveys from this research 27

CLSI EP14 to validate commutability 10 95% prediction interval 8 6 Test Method 4 Clinical Specimen 2 2 4 6 8 10 Comparison Method (1 RMP if available) 28

CLSI EP14 to validate commutability 10 95% prediction interval 8 6 Test Method 4 Clinical Specimen RM Commutable 2 2 4 6 8 10 Comparison Method (1 RMP if available) 29 29

CLSI EP14 to validate commutability 10 95% prediction interval 8 6 Test Method 4 Clinical Specimen RM Commutable RM Not-Commutable 2 2 4 6 8 10 Comparison Method (1 RMP if available) 30 30

Effect of non-Commutable PT materials 1 5 2 3 d u P o n t D i m e s C h l r ( g / L ) A b - K a S v y p f : N , . c Patient bias = 0.2% PT bias = -9.5% Reference Method L a b M e d . 1 9 9 3 ; 1 1 7 : 3 4 5 - 5 1

CAP Accuracy Based Surveys Glycohemoglobin (GH2) – uses fresh whole blood with National Glycohemoglobin Standardization Program target values Creatinine Accuracy Calibration Verification and Linearity (LN24) – uses off-the-clot serum with NIST target values Accuracy Based Lipids (ABL) – uses off-the-clot serum with CDC reference method target values Additional accuracy based Surveys are in development for analytes with clinical practice guidelines 32

Calibration Verification Run assay materials with measurement system and compare results to measurement system appropriate target values Result = Target ? Calibration is OK Re-calibrate YES NO 33

Calibration Verification Materials Method product calibrator(s) Method vendor provided materials for calibration verification Previously tested clinical specimens Reference materials with matrix and target values appropriate for the method PT (or PT validated) materials with matrix and target values appropriate for the method, including materials designed for calibration verification 34

Verification of Calibration Trueness (or accuracy for individual samples) Requires a reference material that is commutable with native clinical samples (not commonly available) Most PT materials are not commutable due to the cost to prepare commutable materials and the practical need to have many analytes in the same material 35

LN Surveys for Calibration Verification Lab verifies conformance to the manufacturer’s calibration, using: Commutable reference materials that have been validated to give results equivalent to those for native clinical samples (e.g. CAP LN 24 Creatinine accuracy based Survey with NIST target values) Non-commutable materials with target values appropriate for a specific method (e.g. most CAP LN Survey materials with peer group evaluation) 36

Calibration Verification Concentration Levels Approximately the same as the calibrators used for method calibration i.e. to confirm correct calibration Can also verify the AMR if concentrations, or activities, of the calibrators cover the full measurement range. LN Survey provides a series of defined concentrations to verify calibration over the AMR 37

Frequency of Calibration Verification Manufacturer’s instructions Troubleshoot a QC problem Calibration OR calibration verification at least every 6 months (CLIA), or when: Change in critical reagents, component, or maintenance If cal <6 months do not need calibration verification unless part of manufacturer’s calibration process 38

Analytical Measurement Range Verification Non-linear region Specimen concentration relationship is linear over the AMR CONCENTRATION RECOVERED CONCENTRATION EXPECTED Based on: (a) admixture ratio (b) assigned value 39

Method Calibration and AMR Verification can be a single process ANALYTICAL MEASUREMENT RANGE Calibrator must: 1. Have at least 3 values that span the AMR, and 2. Have target values specific for the method (and reagent lot, if necessary). = reportable values Analytical signal No separate verification of AMR is necessary Method Calibrator 40

AMR Verification 1. Specimens (can use patients) with low and high values 2. Mix 1:1 to get midpoint value 3. Mix low:mid and mid:high to get additional values 4. Expected values are: Sample Proportion Concentration expected Low 0% 20 mg/dL 0.75 L + 0.25 H 25% 160 mg/dL 0.50 L + 0.50 H 50% 299 mg/dL 0.25 L + 0.75 H 75% 438 mg/dL High 100% 578 mg/dL Criteria for acceptability = recovery of concentration in proportion to the amount present (linear relationship) 41

AMR Verification AMR verification can be based on demonstrating linearity either by measuring recoveries or by demonstrating expected proportions 42

AMR Verification Measured mg/dL Expected Proportion 200 400 600 50 100 Expected Proportion Measured mg/dL Linear relationship verifies recovery of correct concentration relationship over AMR. 43

Materials for AMR Verification Linearity material of appropriate matrix PT (or PT validated) materials Previously tested clinical specimens Previously tested clinical specimens, altered by admixture, dilution, spiking Reference materials with matrix and target values appropriate for the method Method product calibrator(s) 44

AMR: how close to low and high limits Goal is to verify acceptable clinical performance For low limit: use clinical judgment, e.g. creatinine, 0.2 or 0.3 mg/dL For high limit: within 10-15% of upper limit of AMR Challenging to find residual samples for some analytes; need to use clinical judgment, e.g. blood gas (use commercial materials) cardiac markers, e.g. Troponin (use LN Survey) 45

CAP LN Surveys Most evaluate performance vs. peer group due to commutability limitations (matrix effect bias) LN24 offers matrix-free creatinine challenges with reference method values assigned Verify that calibration conforms to manufacturer’s specifications Verify that calibration is correct over the analytical measurement range When accuracy based (e.g., LN24), verify that calibration is traceable to a reference system 46

Example of LN2 Report (Calcium) 47

Example of LN2 Report (Calcium) 48

Example of LN24 Report (Creatinine) 49

Example of LN24 Report (Creatinine) 50

Example of a Non-Linear Result Creatinine 51

Calibration Verification and Linearity Users Guide 52

Clinically Reportable Range Range of analyte values that will be reported as a quantitative result, allowing for specimen dilution or other pretreatment to extend the AMR Policy based on the laboratory director’s medical judgment regarding the clinical use requirements for a laboratory test 53

Clinically Reportable Range Truncation or extension of the AMR based on clinical requirements. Established once when method is put in service Does not need to be verified on a recurring schedule Values outside the CRR are reported as < or > a numeric value. Checklist question is: Are dilution protocols and diluents (or concentration protocols) specified for all methods for which the CRR exceeds the AMR? 54

Clinically Reportable Range Example: hCG may have a CRR with both low and high limits. AMR is 3-1,000 mIU/mL (method specification). CRR is 5-1,000,000 mIU/mL (clinical usefulness). Results of <3, 3, or 4 are reported as “<5 mIU/mL”. Results >1,000 are diluted and rerun to obtain quantitative values up to 1,000,000 mIU/mL. Results >1,000,000 are reported as “>1,000,000 mIU/mL”. NOTE: CRR limits are the judgment of the lab director and may be different for different labs 55

Clinically Reportable Range Example: AST may have a CRR with only a low limit. AMR is 4-900 IU/L (method specification). CRR is 4 IU/L to any value (clinical usefulness). Results of <4 are reported as “<4 IU/L”. Results >900 are diluted and re-assayed. The upper CRR is not specified because specimens are diluted until a quantitative value is obtained. 56

Summary Calibration or Calibration Verification at least every 6 months. Calibration verification is not necessary if calibration is performed at least every 6 months. 57

Summary Analytical Measurement Range verification at least every 6 months. May be required when change a critical reagent lot or other analytical parameter that can influence the AMR. Not required when a recalibration is performed. Not required when method calibrators include at least 3 concentrations that span the full AMR. REQUIRED for initial method validation. 58

Summary Clinically Reportable Range is established once when a method is introduced based on clinical requirements as defined by the lab director. SOP for dilution or other specimen pre-treatment must be defined. Not required to verify CRR on a recurring schedule 59

Summary For additional information on the content of the Audioconference or questions about the CAP Calibration Verification/Linearity Surveys, please contact the CAP: By telephone: Customer Contact Center at 800-323-4040, option 1# By email: contactcenter@cap.org By mail: Surveys Program 325 Waukegan Road Northfield IL 60093-2750